J Thorac Cardiovasc Surg 2005;130:1639-1644
© 2005 The American Association for Thoracic Surgery
Evolving Technology |
, MD
b
a Biomedical Engineering
b Lerner Research Institute; Cardiovascular Medicine
c Thoracic and Cardiovascular Surgery, The Cleveland Clinic Foundation, Cleveland, Ohio
Received for publication May 18, 2005; revisions received August 11, 2005; accepted for publication August 15, 2005. * Address for reprints: Kiyotaka Fukamachi, MD, PhD, Department of Biomedical Engineering/ND20, The Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195 (Email: fukamak{at}ccf.org).
| Abstract |
|---|
|
|
|---|
METHODS: Eight mongrel dogs (mean weight 29.1 ± 4.0 kg) were used in this study. The occlusion device, constructed from 2 stainless steel strips covered with a knit braided polyester fabric, was implanted at the base of the left atrial appendage through a left thoracotomy on a beating heart. Dogs were evaluated at 7 days (n = 2), 30 days (n = 2), and 90 days (n = 4) by epicardial echocardiography, left atrial angiography, histologic inspection, and gross pathology.
RESULTS: Device implantation was performed without complications in all animals. Complete exclusion of the left atrial appendage from the circulation was confirmed acutely and chronically by echocardiographic and angiographic evaluations. There was no device migration or damage to adjacent structures.
CONCLUSION: This novel device enables rapid, reliable, and safe exclusion of the left atrial appendage. The device provides a new therapeutic option for reducing the risk of stroke in patients with atrial fibrillation.
| Introduction |
|---|
|
|
|---|
Accordingly, there is a need for a simple, minimally invasive technique that provides complete exclusion of the LAA. Recently, several novel techniques for LAA exclusion using thoracoscopic or percutaneous approaches have been reported.
6-10
Although an epicardial approach to LAA exclusion has the potential advantages of simplicity (no transseptal puncture) and low procedural risk of embolism (no left heart endocardial catheters), there is currently no instrumentation available to facilitate rapid, reliable, and safe epicardial exclusion of the LAA.
A novel device, the atrial exclusion device (AED) (AtriCure, Inc, Cincinnati, Ohio), has been specifically designed for the exclusion of structures such as the LAA. This study was conducted to evaluate the ability of the AED to exclude the LAA during short and intermediate follow-up periods in a canine model.
| Material and Methods |
|---|
|
|
|---|
|
In Vivo Study
Eight mongrel dogs (mean weight 29.1 ± 4.0 kg) were used in this study. This study was approved by the Cleveland Clinic's Institutional Animal Care and Use Committee, and all animals received humane care in compliance with the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources, National Research Council, and published by the National Academy Press, revised 1996.
Implant Study
All surgical procedures were performed using sterile techniques. The animals were anesthetized with intravenous thiopental (20 mg/kg). Anesthesia was maintained with isoflurane (0.5%-2.5%). The animals were placed in the right lateral position. Electrocardiogram leads were attached to the extremities. The left carotid artery was used for continuous monitoring of arterial pressure.
A left fourth intercostal thoracotomy was performed, and the pericardium was opened to expose the LAA. A 14-gauge angiocatheter was inserted into the left upper pulmonary vein to monitor LA pressure and to inject contrast material for LA angiography in the 60-degree left anterior oblique and 30-degree right anterior oblique planes. Two-dimensional epicardial echocardiography (2D EE) was performed to confirm blood flow into the LAA before AED application. Hemodynamic data, including systemic arterial and LA pressures and heart rate, were also collected at baseline. Ventilatory support was transiently stopped during data-acquisition periods. Hemodynamic parameters were digitized in real time at a sampling rate of 200 Hz with a data-acquisition system (PowerLab, AD Instruments, Inc, Mountain View, Calif) and stored on a hard disk for subsequent analyses by a custom-made visual basic program on Excel software (Excel 2000, Microsoft Corporation, Redmond, Wash).
After all baseline data were obtained, the AED in the open position was placed over the end of the LAA and aligned with the base of the LAA. Once in the desired position, the AED was closed by gently squeezing the flexible and rigid members together. If AED deployment and position were not satisfactory, the AED was simply repositioned and reapplied in the proper location. After implantation, echocardiographic, angiographic, and hemodynamic data were obtained as before.
The chest was closed with a chest drain tube in place. Animals were followed for 7 days (2 dogs), 30 days (2 dogs), or 90 days (4 dogs).
Explant Study
On the day of the explant study, animals were placed under general anesthesia as for the implant study. The right femoral artery was used for continuous monitoring of arterial pressure. The chest was reentered through the original incision to expose the AED. Hemodynamic assessment, LA angiography, and 2D EE were performed in the same manner as during the implant study.
In one 7-day and two 90-day dogs, electrical stimulation was applied to the LAA remnant. Standard epicardial pacing wires were placed on the LA body and LAA remnant. LA pacing was performed in asynchronous mode to confirm that the heart could be entrained. Then, stimulation (20 mA) was applied to 2 different locations on the LAA to attempt to pace the heart at a rate greater than the native sinus rate.
Before sacrifice, 500 U/kg of heparin was administered. Animals were then put to death by rapid intravenous injection of sodium pentobarbital (50 mg/kg) and potassium chloride (120 mEq). The heart was excised with the AED in place, and the external and internal surfaces of the LA were photographed and examined. The heart was thoroughly investigated to evaluate the completeness of the LAA exclusion, and surrounding tissue and structures were examined for signs of damage and to evaluate tissue response.
Histologic Evaluation
Cardiac tissue specimens were excised from both sides of the AED for evaluation of tissue ingrowth into the polyester fabric coating. Tissue from the LAA distal to the implanted AED was also evaluated. The tissues were fixed in formalin for 48 hours and prepared appropriately for evaluation. The sections were stained with hematoxylin-eosin.
Data Analysis
All values were expressed as mean ± standard deviation. Repeated-measures analysis of variance was used to assess differences between data at preimplantation, postimplantation, and follow-up.
| Results |
|---|
|
|
|---|
|
Gross and Histologic Examinations
The heart was excised with the AED in place in all animals. Examination of the inside of the LA revealed a smooth endocardial surface at the origin of the LAA; there was no cul-de-sac, and there was no thrombus associated with the AED. In one 7-day animal, a 1-cm, firmly adherent LA thrombus was identified arising from the body of the LA directly opposite to the LAA orifice; this likely represented a site of endocardial damage caused by the catheter tip. There was no communication between the orifice of the LAA and the thrombus, indicating that there was no relationship between the thrombus formation and the AED application.
At 7 days, the distal LAA was intact. The polyester fabric was covered with a thin, pink tissue. Figure 3 shows a representative gross anatomic view of the external LAA and the internal surface of the LAA orifice at 90 days. Tissue ingrowth completely covered the AED and the LAA was atrophied in 3 of 4 dogs (Figure 3, A). In addition, the occlusion line was smooth and linear without any trabeculated portions (Figure 3, B). Hematoxylin-eosinstained sections of the distal LAA tissue from non-atrophied LAAs showed myocardium with fibrotic remodeling and fatty infiltration, whereas the tissue from atrophied LAAs was completely replaced by fibrous tissue. In all cases, the polyester fabric was covered by fibrous connective tissue.
|
| Discussion |
|---|
|
|
|---|
AF, Stroke, and the LAA
AF is responsible for 10% to 20% of strokes. Recent randomized trials have demonstrated that anticoagulation substantially reduces, but does not eliminate, the risk of stroke in patients with AF.
1,11,12
However, anticoagulant therapy with warfarin is associated with both major (1%-2%/year) and minor (5%-10%/year) hemorrhagic complications.
13,14
Because of the increased risk of anticoagulant-related complications with advancing age, many physicians are reluctant to prescribe anticoagulant therapy for elderly patients with AF.
13,14
In addition, many patients are not compliant with anticoagulation. Because of these issues, recommended guidelines for anticoagulation are followed in less than half of outpatients with AF.
4,5
A potential alternative or adjunctive approach to stroke prevention in patients with AF is surgical management of the LAA. The LAA is the primary site of thrombus formation in patients with AF. Approximately 60% of LA thrombi in rheumatic AF form in the LAA, but this figure is 91% in patients with nonrheumatic AF.
2
These reports led to the recommendation that the LAA be excised or excluded at the time of mitral valve surgery.
15
This strategy has been associated with a reduction in the risk of late stroke in patients undergoing mitral valve surgery.
16
In addition, early data suggest that LAA excision or exclusion reduces the risk of stroke in selected patients with nonvalvular AF.
2,17
Techniques for LAA Management
Cardiac surgeons have used a variety of techniques to manage the LAA. Suture ligation from the endocardial surface of the LA requires the use of cardiopulmonary bypass and may be associated with bleeding or injury to the circumflex coronary artery. In addition, endocardial suture ligation is incomplete in 10% to 30% of patients, and incomplete exclusion may predispose the patient to thromboembolic events.
18-20
In contrast, external ligation or excision can be performed without opening the LA and without cardiopulmonary bypass. Such approaches may include suture ligation or use of cutting or noncutting surgical staplers.
10,21
These techniques can cause troublesome bleeding if friable LA tissue tears.
22
Epicardial application of a purse-string suture or loop may fail to occlude the LAA or may cause bunching of LA tissue; these might predispose to thrombus formation on the endocardial surface. In addition, these epicardial approaches, including this AED application, incur the potential risk of perioperative thromboembolism in patients with LAA thrombi that extend to the orifice of the LAA, because of the required manipulation of the LAA. Therefore, these approaches should be avoided in this patient population.
More recently, percutaneous approaches to LAA exclusion have been introduced. Several reports of preclinical data and human feasibility trials document successful endocardial application and the possibility of stroke prevention.
6-9
However, residual flow around the occlusion device immediately after implant and at follow-up was observed in every patient, and the importance of this finding is uncertain.
8
The AED
Unlike most surgical staplers, the AED has been developed specifically for exclusion of structures such as the LAA. It is configured for open surgical or thoracoscopic placement. It is easily placed on the beating heart and may be atraumatically repositioned if initial placement is unsatisfactory. When compared with epicardial stapling devices, the AED offers 2 advantages: (1) easy reapplication that ensures proper placement onto the base of the LAA and (2) elimination of the risk of bleeding that may originate along the occlusion (staple) line. The promotion of distal LAA atrophy by the AED, as indicated in our 30- and 90-day experiments, may eliminate the possibility of late recanalization of the LAA lumen, which is one of the potential problems of noncutting stapling devices. Angiography and echocardiography confirmed complete exclusion of the LAA at 7, 30, and 90 days without damage to adjacent cardiac structures. The physical characteristics of the device and tissue ingrowth around the polyester covering prevented device migration. By 90 days, the LAA generally atrophied, leaving a smooth endocardial surface where its origin had been. Although there was no thrombus in the vicinity of the LAA, 1 animal did have a LA thrombus that was likely attributable to catheter-induced trauma at angiography. At present, the AED is further evolving to obtain confidence for clinical trial.
Limitations
These data include early follow-up in a relatively small number of healthy animals. None of the animals had AF, so the possibility of LA thrombus formation after AED application in the setting of AF cannot be excluded. However, the smooth endocardial surface at the origin of the LAA and the absence of a cul-de-sac in this region are very encouraging. This study does not test the hypothesis that exclusion of the LAA reduces the risk of stroke or other thromboembolism in patients or animals with AF.
Further Clinical Implications
This study suggested that the AED can occlude and isolate the LAA completely, safely, and rapidly without cardiopulmonary bypass and provide favorable biocompatibility over a mid-term period (90 days). The results may stimulate the development of novel device that allows for the AED deployment under a thoracoscopic approach.
To date, no report shows clear evidence of the role of LAA exclusion in the prevention of strokes for patients with nonvalvular AF. A randomized trial of LAA occlusion during routine coronary artery bypass surgery (Left Atrial Appendage Occlusion Study) is ongoing and will provide definitive evidence about the LAA exclusion in patients without mitral valve disease.
23
Meanwhile, several reports have suggested that the LAA may be one of the sources of thrombus in patients with significant left ventricular dysfunction (eg, dilated cardiomyopathy), even in those who maintain sinus rhythm.
24,25
If LAA exclusion is proven to reduce stroke risk in patients with nonvalvular AF and to be free from other complications, this novel device, the AED, may be suitable not only for patients with nonvalvular AF but also for the aforementioned high-risk patients in sinus rhythm for prevention of embolic events resulting from LAA thrombi.
| Conclusion |
|---|
|
|
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. Fumoto, A. M. Gillinov, Y. Ootaki, M. Akiyama, D. Saeed, T. Horai, C. Ootaki, D. G. Vince, Z. B. Popovic, R. Dessoffy, et al. A novel device for left atrial appendage exclusion: the third-generation atrial exclusion device. J. Thorac. Cardiovasc. Surg., October 1, 2008; 136(4): 1019 - 1027. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Su, K P McCarthy, and S Y Ho Occluding the left atrial appendage: anatomical considerations Heart, September 1, 2008; 94(9): 1166 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Lacomis, O. Goitein, C. Deible, P. L. Moran, G. Mamone, S. Madan, and D. Schwartzman Dynamic multidimensional imaging of the human left atrial appendage Europace, December 1, 2007; 9(12): 1134 - 1140. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov Advances in Surgical Treatment of Atrial Fibrillation Stroke, February 1, 2007; 38(2): 618 - 623. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Onalan and E. Crystal Left Atrial Appendage Exclusion for Stroke Prevention in Patients With Nonrheumatic Atrial Fibrillation Stroke, February 1, 2007; 38(2): 624 - 630. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Kamohara, Z. B. Popovic, M. Daimon, M. Martin, Y. Ootaki, M. Akiyama, F. Zahr, F. Cingoz, C. Ootaki, M. W. Kopcak Jr, et al. Impact of left atrial appendage exclusion on left atrial function J. Thorac. Cardiovasc. Surg., January 1, 2007; 133(1): 174 - 181. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Kamohara, K. Fukamachi, Y. Ootaki, M. Akiyama, F. Cingoz, C. Ootaki, D. G. Vince, Z. B. Popovic, M. W. Kopcak Jr, R. Dessoffy, et al. Evaluation of a novel device for left atrial appendage exclusion: The second-generation atrial exclusion device. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 340 - 346. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stollberger, B. Schneider, and J. Finsterer Stroke prevention by means of epicardial occlusion of the left atrial appendage J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 207 - 208. [Full Text] [PDF] |
||||
![]() |
K. Kamohara, K. Fukamachi, and A. M. Gillinov Reply to the Editor J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 208 - 208. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||