J Thorac Cardiovasc Surg 2005;130:1645-1652
© 2005 The American Association for Thoracic Surgery
Evolving Technology |
a Department of Cardiac Surgery, University of Technology Dresden, Dresden, Germany
b Department of Cardiac Surgery, Heart Center, University of Leipzig, Leipzig, Germany
c Department of Cardiac Surgery, Cardiocentro Ticino, Lugano, Switzerland
d Department of Cardiovascular and Thoracic Surgery, Klinikum Braunschweig, Braunschweig, Germany
e Department of Cardiovascular Medicine, Stanford University, Stanford, Calif
f Department of Cardiac Surgery, Heart Center Hamburg, University of Hamburg, Hamburg, Germany
Read at the Eighty-fifth Annual Meeting of The American Association for Thoracic Surgery, San Francisco, Calif, April 10-13, 2005.
Received for publication March 31, 2005; revisions received August 10, 2005; accepted for publication August 17, 2005. * Address for reprints: Klaus E. Matschke, MD, Department of Cardiac Surgery, Heart Center Dresden University Hospital, Fetscherstr 76, D-01307 Dresden, Germany (Email: k.matschke{at}herzzentrum-dresden.com).
| Abstract |
|---|
|
|
|---|
METHODS: Five centers enrolled 133 patients awaiting elective coronary artery bypass grafting surgery. Outcome variables were intraoperative device performance, incidence of device-related adverse events, predischarge and 6-month angiographic graft patency, and 12-month clinical outcome. Independent core laboratories performed qualitative and quantitative angiographic and computed tomographic assessments.
RESULTS: The C-Port was used to perform a vein-to-coronary anastomosis in 130 patients. Intraoperative conversion to a hand-sewn anastomosis was necessary in 11 patients because of inadequate target site preparation, inappropriate target vessel selection, or both. Inadequate blood flow related to poor runoff required conversion in 3 additional patients. Three patients died before discharge of causes unrelated to the device. At discharge, 113 patients had a C-Port implant in place, and 104 C-Port anastomoses were studied by means of angiography, resulting in 100 FitzGibbon A, 3 FitzGibbon B, and 1 FitzGibbon 0 classifications. At 6 months, one additional patient died of a device-unrelated cause, and 98 patients were evaluated by means of angiography (n = 89). Overall patency (FitzGibbon A) was 92.1%. Three C-Port anastomoses were rated FitzGibbon B, and 4 were rated FitzGibbon 0. At 12 months, 107 (98.2%) of 109 alive patients were followed up, without any reports of device-related major adverse cardiac events.
CONCLUSIONS: The C-Port System allows for a rapid, reliable, and compliant distal anastomosis and yields favorable 6-month angiographic and 12-month clinical results when compared with published studies.
| Introduction |
|---|
|
|
|---|
Recently, the Cardica C-Port Anastomotic System (Cardica, Inc, Redwood City, Calif) has been introduced to facilitate saphenous venous grafttocoronary artery anastomoses. The C-Port System allows rapid automated distal coronary anastomoses by integrating all the functions necessary for completing an anastomosis. A compliant, angled, end-to-side anastomosis is performed by automatically placing 8 individual clips and creating an arteriotomy with the push of a button.
This multicenter clinical trial investigated the safety and efficacy of the C-Port device in the construction of distal coronary anastomoses.
| Materials and Methods |
|---|
|
|
|---|
Study inclusion criteria were age greater than 18 years and less than 85 years, ejection fraction of greater than 30%, creatinine value of less than 200 mmol/L, and the requirement for nonemergency bypass of at least one coronary artery. Exclusion criteria were previous CABG surgery, the preoperative need for an intra-aortic balloon counterpulsation, bleeding diathesis, a recent cerebrovascular accident, and patients in New York Heart Association (NYHA) class IV.
The study design included a perioperative assessment with predischarge angiography, intermediate patency evaluation at 6 months, and longer-term (12-month) clinical follow-up with resting and stress electrocardiography (ECG) and assessment of major adverse cardiac events (MACEs). A MACE was defined as the presence of death, myocardial infarction, or the need for target vessel revascularization.
Anastomotic System
The C-Port device (Figure 1) integrates in one tool all functions necessary to enable rapid automated distal coronary anastomoses. When using the Cardica C-Port System, a vein is harvested from the patient's leg by conventional means and is attached to the cartridge of the C-Port deployment device (Figure 2). The surgeon creates a small incision in the target vessel. The graft is then inserted between the 2 cartridge arms and attached by 4 small spikes. This system is suitable for coronary arteries as small as 1 mm in internal diameter. A small incision is created in the target vessel, and the anvil of the C-Port is inserted. The connection between the vein graft and the coronary artery can then be performed rapidly, without the need for temporary occlusion of the target vessel. Deployment is accomplished by depressing a trigger button on the deployment handle, which first approximates the cartridge onto the coronary target and then deploys 8 individual stainless-steel anastomotic clips that penetrate the graft and target vessels and form against the anvil inside the coronary artery. Each clip has 4 tines that form in pairs for a total of 32 points of tissue approximation. A knife located within the anvil is deployed simultaneously and cuts the coronary artery from the inside out, thereby creating the arteriotomy. The cartridge then unclamps, and the device is removed from the anastomosis site through the previously created incision. A stitch is necessary in most instances to occlude the anvil insertion hole, and additional stitches might be necessary to provide hemostasis at the toe and heel (Figure 3).
|
|
|
Core Laboratory Analysis
The patency of the index grafts was evaluated by means of either angiography or computed tomography (CT). The digital-film images were assessed by respective core laboratories. Assessment of angiographies and CT scans were performed at Stanford University Medical Center under the direction of Edwin Alderman (angiography) and Geoffrey Rubin (CT). For angiographic images, distal anastomosis patency was classified according to the FitzGibbon scoring system.
3
C-Port anastomoses evaluated by means of CT were classified as patent if there was positive contrastation in the graft, including the distal anastomosis and target vessel. Because of the known limitations of CT, these data were excluded in the patency statistics.
The Core Angiographic Laboratory at Stanford used both qualitative and quantitative methods to evaluate each anastomosis. Qualitative methods included estimates of thrombolysis in myocardial infarction flow grade and general assessments of patency. The quantitative coronary angiography laboratory included measurement of the internal lumen dimensions of the graft proximal to the distal anastomosis (A), the coronary target distal to the anastomosis (C), and the internal diameter of the anastomosis itself (B). This was performed by the core laboratory using standard angiographic computer software (PlusPlus; Sanders Data Systems, Stanford, Calif). These dimensions made it possible to calculate the average percentage diameter stenosis, as described by Caskey and associates
4
:
![]()
| Results |
|---|
|
|
|---|
|
|
| Discharge Patency |
|---|
|
|
|---|
| Six-month Follow-up |
|---|
|
|
|---|
Of these patients, 98 (87.5% [98/112]) were evaluated by means of angiography (n = 89) or CT (n = 9). The results indicated that 85 C-Port index grafts evaluated by means of angiography and all 9 C-Port index grafts evaluated by means of CT were patent, for a rate of 95.9% (94/98). Three C-Port index grafts evaluated by means of angiography were rated FitzGibbon B, and 4 were rated FitzGibbon 0. Excluding the 4 occluded anastomoses, the average distal anastomosis stenosis was 10.7% ± 30.3%, with a minimum of 116% and a maximum of 71%. The overall freedom of stenosis greater than 50% (FitzGibbon A) in grafts evaluated by means of angiography at 6 months was therefore 92.1% (82/89). In the core laboratory analysis there was no evidence of any narrowing of the coronary artery at the site of the anvil insertion hole.
| Twelve-month Follow-up |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
|
Because of the design of the product, the distance between clips is marginally larger at the toe and heel. In many instances during this trial, additional conventional stitches were placed to achieve hemostasis. However, this limitation is offset by the ability to perform a consistent, reproducible, and compliant anastomosis on target vessels as small as 1 mm in diameter. Future product iterations will address this shortcoming.
In patients in whom the C-Port deployment was clinically unsuccessful, a safe bailout with conversion to a hand-sewn anastomosis was possible in all cases, without any compromise to patient long-term outcome. At the end of each procedure, the patient was informed of the success of the deployment, and many patients in whom the deployment was unsuccessful elected to refuse angiographic follow-up.
Any alternative technique used to construct a distal anastomosis in coronary surgery should provide patency similar to that obtained with conventional suturing techniques using the same type of graft. Mack and associates presented a meta-analysis at the Techno College, EACTS 2004 (European Association of Cardio-Thoracic Surgeons), comprising more than 28,000 grafts. According to these researchers, the patency of hand-sewn saphenous vein grafts can be expected to be 87.9% after 30 days, 84.1% between 3 and 6 months, 82.7% after 12 months, and 74.3% between 2 and 5 years. In this initial C-Port trial, graft patency (FitzGibbon A and B) exceeded the meta-analysis benchmark, with discharge patency of 99% and 6-month patency of 95.9%. Further evaluation of the C-Port grafts with significant stenosis or occlusion revealed that 5 of these 7 grafts were anastomosed to coronary arteries with a native stenosis of less than or equal to 60%. According to Manninen and coworkers,
6
reduced native artery stenosis is a predictor of graft occlusion caused by competitive flow.
The 12-month evaluation, including assessment by means of stress ECG and determination of MACEs, demonstrated a favorable outcome in patients after implantation of the C-Port System. None of the patients experienced a myocardial infarction or required a target vessel revascularization since discharge. The overall mortality since discharge (3.6%) compares with published data by Cheng and colleagues
7
and Unger and associates,
8
who describe a 1-year mortality of 2.6% and 3.6%, respectively. In addition, Unger and associates found a myocardial infarction rate of 4.1% in more than 600 patients studied 1 year postoperatively.
The C-Port anastomosis system has some important characteristics that might help explain this favorable outcome. The attachment of the graft to the target vessel is accomplished with individual clips, which results in a compliant anastomosis. The vascular anastomosis originally developed by Alexis Carrel used a modified interrupted technique. Cardiac surgeons adopted this technique, and some of the pivotal studies describing long-term patency outcome in CABG surgery by Loop and coworkers
9
and Lytle and colleagues
10
are based on the interrupted technique. Wolf and associates
11
have recently published results from a clinical trial in which multiple clips were used to perform an interrupted anastomosis in the left internal thoracic arterytoleft anterior descending coronary artery bypass grafts. According to these authors, this technique avoided a stenosis of the anastomosis, which can amount to 24% in the hand-sewn continuous stitch technique. In addition to providing a compliant anastomosis, the C-Port System clips have a small amount of blood-exposed nonintimal surface (Figure 3). Scheltes and coworkers
12
studied various anastomotic devices, compared them with sutured anastomoses, and postulated that a reduction in blood-exposed nonintimal surface would be beneficial for long-term outcome. The automated creation of the anastomosis ensures consistency from one anastomosis to another. Finally, the C-Port System allows placement of the distal anastomosis without interrupting native artery blood flow, thereby avoiding the need for intracoronary shunts or temporary myocardial ischemia.
Limitations of the study include a nonrandomized patient population selected for conformance with prespecified inclusion and exclusion criteria.
In summary, the 6-month patency rate and 12-month clinical outcome of the C-Port device is very promising. On the basis of these successful results, we have continued evaluation of this product in successive studies with longer-term follow-up. The integrated design of this product allows rapid and safe deployment and might prove to be a valuable asset in coronary surgery, especially in beating-heart procedures (off-pump coronary artery bypass).
| Discussion |
|---|
|
|
|---|
With this study, the authors demonstrate good safety and efficacy of this novel technology and have to be congratulated for their excellent results. Efficacy in this context involves both successful application of the device and creation of the patent anastomosis. Because this is a feasibility trial with an intention-to-treat model, the success rate of delivering the system was a fair 85% because it was successfully applied in 116 of 133 patients initially enrolled. In 11 patients conversion to a hand-sewn anastomosis was necessary; in 3 patients poor blood flow across the graft also made the redo of the anastomosis by hand necessary; 2 other patients were deemed unsuitable for the use of the device; and finally, in another patient the device failed because of a technical problem.
What about efficacy in terms of creation of a patent anastomosis? Although not all patients were followed up by means of angiography or CT, patency rates were excellent, with a FitzGibbon A rating at discharge of 96% and 93% after 6 months. Given the fact that all these grafts were veins and the target vessels were mainly branches of the circumflex, the diagonal, or the right coronary arteries, this is a very good result. However, more detailed information on how these anastomoses looked should be provided. I believe that classifying these anastomoses into FitzGibbon scores is not precise enough, and the anastomoses should be better characterized. Can you comment on this?
Dr Demertzis. Yes. I would like to address the issue of efficacy first. You mentioned the 11 unsuccessful deployments. The unsuccessful deployments were due to the fact that the cartridge has to align with the anvil and the tissue around the coronary artery has to be dissected to a sufficient amount. If this is not the case, the clips could miss the anvil and not deform, leading to a clinically unsuccessful anastomosis. This is not a true failure. It is a technique failure and not a technical failure.
The other reason to have a clinically unsuccessful deployment is if you do not respect the specs on the wall thickness of the target vessel. Therefore in a target vessel that is heavily diseased and thickened, you can have an unsuccessful deployment. This was the case in the initial phase of the study. As we understood the reasons for that and applied these criteria to our target vessels, this problem dissolved. At the end of this study and later on during the C-Port's commercial use, this problem has actually disappeared.
With regard to the FitzGibbon classification, we do have a complete quantitative graft and anastomosis analysis for every graft in every single patient, very precise and exact. Of course, with the FitzGibbon's classification, he have a rough way to group the grafts, and this is why it is widely used in the literature. Therefore for the sake of comparison, I think it is a rather adequate solution.
Dr Berdat. The use of this device seems rather safe, but there are some technical particularities. The extensive preparation of the vessel with dissection of 50% of the circumference and the need for creation of a separate incision and insertion of this device is not what could be called a no-touch technique. Furthermore, additional stitches precisely at the most sensitive points of the anastomoses, at the toe and heel, were usually necessary in this series. Can you comment on this?
Dr Demertzis. It was not intended to have a coronary anastomosis by not touching the coronary artery, of course.
The point on the stitches, yes, I have to mention that the tacking stitches were really superficial adventitial stitches and not transmural, which were absolutely sufficient in all the cases. You have to deal with the anvil insertion hole, which is somewhere around 1 mm behind the heel of the anastomosis. Quantitative analysis on the postoperative angiographs at discharge and at 6 months postoperatively did not reveal any problem in respect to that issue. Therefore it is a theoretic concern, but in real-world practice, it did not have any effect.
Dr Berdat. Again, the authors have to be congratulated to put their efforts in a field that has turned out to be controversial most recently.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
W. J.L. Suyker and C. Borst Coronary Connector Devices: Analysis of 1,469 Anastomoses in 1,216 Patients Ann. Thorac. Surg., May 1, 2008; 85(5): 1828 - 1836. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Kappert, S.-M. Tugtekin, R. Cichon, M. Braun, and K. Matschke Robotic totally endoscopic coronary artery bypass: A word of caution implicated by a five-year follow-up. J. Thorac. Cardiovasc. Surg., April 1, 2008; 135(4): 857 - 862. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Jacobs, D. Holzhey, H. Stein, F. W. Mohr, and V. Falk Catheter-Based Endoscopic Bypass Grafting: An Experimental Feasibility Study Ann. Thorac. Surg., November 1, 2007; 84(5): 1724 - 1727. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Vicol, S. Eifert, M. Oberhoffer, P. Boekstegers, and B. Reichart Mid-term patency after magnetic coupling for distal bypass anastomosis in coronary surgery. Ann. Thorac. Surg., October 1, 2006; 82(4): 1452 - 1456. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Martens, M. Argenziano, and M. C. Oz New Technology for Surgical Coronary Revascularization Circulation, August 8, 2006; 114(6): 606 - 614. [Full Text] [PDF] |
||||
![]() |
J. Wippermann, C. Konstas, M. Breuer, H. Kosmehl, T. Wahlers, and J. M. Albes Long-term effects in distal coronary anastomoses using different adhesives in a porcine off-pump model. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 325 - 331. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||