|
|
||||||||
J Thorac Cardiovasc Surg 2006;131:403-411
© 2006 The American Association for Thoracic Surgery
Evolving Technology |
a New York Presbyterian HospitalWeill Cornell Medical Center, Department of Cardiothoracic Surgery, New York, NY
b New York Presbyterian HospitalWeill Cornell Medical Center, Department of Anesthesiology, New York, NY
c New York Presbyterian HospitalWeill Cornell Medical Center, Department of Medicine, New York, NY
Read at the Thirty-first Annual Meeting of The Western Thoracic Surgical Association, Victoria, BC, Canada, June 22-25, 2005.
Received for publication June 16, 2005; revisions received October 18, 2005; accepted for publication October 28, 2005. * Address for reprints: Charles A. Mack, MD, Department of Cardiothoracic Surgery New York-Presbyterian Hospital, Weill Cornell Medical College, M404, 525 East 68th St, New York, NY 10021 (Email: cmack{at}med.cornell.edu).
| Abstract |
|---|
|
|
|---|
METHODS: Six mongrel dogs underwent sternotomy. A clamp employing a 10-cm argon-based linear cryoablation device was used epicardially to isolate the pulmonary veins and left atrial appendage. After clamping of lesions, the probe was removed from the cryoclamp device, and the remaining linear lesions, analogous to the Cox maze III, were performed. Pulmonary vein stenosis was evaluated with the use of magnetic resonance imaging. Left atrial function and pulmonary venous flow velocities were assessed with transesophageal echocardiography. Transmurality was confirmed both electrically and histologically. Animals were then put to death at 30 days.
RESULTS: All acute and chronic cryoclamp lesions produced conduction block. There was no change in right (RPV) or left pulmonary vein (LPV) diameter on the basis of magnetic resonance imaging at baseline and at planned death (RPV-1, 19.6 ± 2.9 mm vs 16.9 ± 2.8 mm, P = .22; RPV-2, 13.2 ± 2.0 mm vs 11.8 ± 1.6 mm, P = .22; and LPV, 12.2 ± 2.4 mm vs 11.2 ± 1.9 mm, P = .30). Left atrial function and pulmonary venous flow velocities were unchanged. Tissue sections determined transmurality in 93% of cryoclamp lesions and 84% of linear ablations performed with the 10-cm malleable probe.
CONCLUSIONS: Epicardial application of this cryoclamp device on the beating heart produced transmural lesions, which persisted 30 days. Linear epicardial cryoablation was not as effective as the cryoclamp device at producing consistent transmural lesions. This novel, versatile device may be useful in treating patients with atrial fibrillation on the beating heart without cardiopulmonary bypass.
| Introduction |
|---|
|
|
|---|
|
Atrial fibrillation (AF) is present in 1% of the general population, affecting over 2 million people in the United States.
1
The prevalence of this arrhythmia has been increasing as the population ages, afflicting nearly 10% of patients older than 80 years.
2
AF is associated with significant morbidity through its compromise of cardiac hemodynamics and increased risk of thromboembolic events. The Cox maze III procedure was described by Cox and associates
3
in 1991 and is considered the gold standard for the surgical treatment of AF. The procedure involves a "cut-and-sew" technique, which is thought to direct reentrant circuits into dead ends while allowing the normal propagation of sinoatrial stimuli. The long-term success of the Cox maze III is well established with 96.6% freedom from AF at a mean follow-up of 5.4 years.
4
Despite the success of the Cox maze III with regard to restoring sinus rhythm and preventing strokes and anticoagulant-related bleeding, it failed to gain widespread use, largely because of its complexity, prolonged cardiopulmonary bypass time, increased risk of bleeding from multiple suture lines, and overall increased morbidity.
5
As a means of avoiding the morbidity of the cut-and-sew technique, new ablative devices that use cryo, radiofrequency, microwave, ultrasound, and laser energy to block these reentrant circuits have been established. Cryoablation employs freezing to ablate tissue by forming intracellular ice crystals. These ice crystals irreversibly damage organelles leading to cell death once the tissue has thawed.
6
Argon-based cryothermy has been used to create endocardial lesions on the arrested heart in patients with AF during concomitant cardiac procedures.
7,8
However, the ability to create transmural epicardial cryolesions on the beating heart may be limited by the heat from intracardiac blood flow. In this study, a novel, 10-cm, argon-based cryoprobe and cryoclamp device, which can reach temperatures of 160°C, was evaluated in an off-pump beating heart canine model, to determine whether both the clamp device and the linear probe are able to produce transmural lesions without causing pulmonary vein stenosis or left atrial (LA) dysfunction.
| Materials and Methods |
|---|
|
|
|---|
Clamp Cryoablation Lesions
A cryoclamp, which houses a 10-cm linear cryoprobe, was used epicardially on the beating heart (Figure 1). The lesion-set consisted of an atrial cuff adjacent to the ostia of the left pulmonary veins (LPVs; Figure 2). Next, the superior and inferior right pulmonary veins (RPVs) and the base of the left atrial appendage (LAA) were cryoablated individually (Figure 3). The clamp lesions were tested for conduction block with an epicardial pacing device.
|
|
|
Postoperative Care
Postoperative care consisted of pain management with a 75-µg fentanyl patch for 72 hours, oral carprofen for 5 days, and intermittent doses of subcutaneous hydromorphone as needed. Oral antibiotics were continued until postoperative day 2, when the chest tube was removed. The dogs survived for 30 days. Before the dogs were returned to the operating room for planned death and tissue harvest, postoperative TEE and MRI were performed to compare with baseline studies.
Planned Death and Tissue Harvest
The chest was reopened through the previous sternotomy and the pericardial space was reentered. The PVs and LAA were dissected free and again test-paced to assess for conduction block. A lethal dose of pentobarbital was then administered. The heart and lungs were removed en bloc. Gross inspection of the atria and PVs was made to identify lesions, inspect for mural thrombi, and to assess for PV stenosis. The entire sample was then immersed in 10% formalin for 24 hours before sectioning for histopathology.
Data Acquisition
Cryoablation
The cryoclamp consists of a 10-cm, argon-based, malleable probe (CryoCath Technology, Montreal, Quebec, Canada), which fits into a clamp device (Figure 1). Both the probe and the clamp have a thermostat, which displays real-time temperatures on the console. For the clamp lesions, once the cryoclamp was placed into position, a 20-second dose was applied as soon as the tissue temperature reached 40°C. The temperatures recorded are those at the end of the 20-second dosing period (Table 1). The temperatures described for the linear lesions are the final temperatures at the end of the 3-minute ablation (Table 2).
|
|
Transesophageal echocardiography
A 5-MHz TEE (Acuson, Mountain View, Calif) was used with an XP Acuson echocardiographic system. The mitral valve was examined for anatomic abnormalities and function with 2-dimensional and color flow Doppler. The LV inflow velocities were recorded at the tips of the mitral valve leaflets with the use of a sample volume of 5 mm. Color flow Doppler was used to align the ultrasonography beam with the direction of blood flow. The following diastolic LV inflow velocities were identified: early (E), after the beginning of diastole, and late (A), during atrial contraction. The left and right PVs were then identified with the use of 2-dimensional echocardiography and color flow Doppler. The systolic (S) and diastolic (D) velocities were recorded. When systolic velocities consisted of more than 1 peak, the highest was recorded.
Magnetic resonance imaging
MRI of the PVs was obtained with the use of a 3-Tesla scanner and V.H.3.0 software (General Electric, Milwaukee, Wis). All dogs were sedated with a single dose of pentobarbital (30 mg/kg) and intubated for airway control. Cine cardiac-gated MR images were obtained with an image matrix of 256 x 192 or 256 x 256 pixels, echo time (TE) 3.4, repetition time (TR) of 8.8, flip angle of 30°, and slice thickness of 6 mm. The images were acquired in the transverse, coronal, and sagittal planes to visualize the PVs. Cross-sectional images of the PVs were then used to measure vessel diameters (Figure 4).
|
|
| Results |
|---|
|
|
|---|
For the linear lesions, mean probe length used for the RA and LA lesions was 6.0 ± 2.5 cm and 2.8 ± 1.1 cm, respectively. Mean final temperatures for these lesions were 124.5°C ± 20.4°C and 139.2°C ± 16.2°C, respectively. Longer segments of probe exposed to the tissue lead to warmer temperatures at the end of the dosing period (Table 2).
All dogs survived the 30-day study period.
MRI
Cine MRI of the PVs was obtained in coronal, axial, and sagittal views (Figure 4). Preoperative and postoperative images were then compared to evaluate for PV stenosis. None of the postoperative images identified a stenotic lesion. In addition, measurements of the preoperative and postoperative images did not demonstrate significant changes in PV diameters (Table 3).
|
Postmortem Analysis
Visual inspection of the clamp lesions to the PV and LAA did not demonstrate stenosis or endoluminal thrombi (Figure 5). Lesion width ranged from 0.4 to 0.7cm. All PV tissue sections demonstrated 100% transmurality, whereas 70% of LAA sections were fully transmural (Table 1). All linear lesions were readily identified on gross inspection. Linear lesion width ranged from 0.5 to 1.6 cm (Table 2). Final pathologic examination demonstrated full transmurality in 84% of the tissue sections (Figure 6). Figure 6 demonstrates examples of nontransmural linear cryolesions. Linear lesions performed from the inferior LPV across the coronary sinus (CS) to the mitral annulus demonstrated segments of full transmurality in 4 of 6 dogs (Figure 7). The lesions to the CS were sectioned parallel to the lesion so as to clearly define the depth of penetration below the CS. With this type of sectioning, it is conceivable that the slide may not represent an equal plane along the lesion; therefore, it may not demonstrate full-thickness transmurality, which may be a limitation to the CS lesion analysis. The tricuspid annulus lesions were cut into cross sections, which may be a more sensitive way of determining lesion transmurality.
|
|
| Discussion |
|---|
|
|
|---|
The histologic results of the clamp lesions are comparable with other animal studies in which bipolar radiofrequency was used.
11,12
Histologic assessment of clamp cryolesions demonstrated nearly 100% transmurality in PV clamp lesions, with the exception of the LAA lesions, which had 5 of 17 (29.4%) tissue sections with nontransmural lesions. These tissue sections demonstrated thin regions of subendocardial sparing, particularly in the thicker trabeculated regions of the LAA. Despite this finding, the LAA demonstrated complete conduction block on test pacing (Table 1). However, even though test pacing was performed, we do acknowledge that we cannot prove for certain that we had local capture on test pacing, as we did not attempt to record electrograms in the isolated LAA or PV. One of the important principles in performing ablative procedures for AF is the creation of lesions that are transmural.
13
While the Cox maze III assures 100% transmurality with cut-and-sew lesions, only bipolar radiofrequency is reliable in ensuring 100% transmurality.
12
Endocardial cryoablation on the arrested heart provides the most optimum condition for transmural lesion with a smooth endocardial surface, and without heat dissipation from the intracardiac blood flow. Epicardial ablation on the beating heart, however, introduces the heat sink of intracardiac blood flow; in addition to epicardial fat, which prevents good tissue contact with the probe.
14
Although these thin subendocardial regions of cardiac tissue were spared, the relevance of this is unknown, as these lesions continued to demonstrate complete conduction block.
A similar phenomenon was noted in a dog study with the use of microwave energy to create a single circular lesion to isolate the PVs.
10
The investigators were able to achieve nearly 100% electrical isolation of the PVs in both acute and 3-week survival dog models. On histologic analysis, however, a mean of 33% transmurality was seen in the acute lesions, while the 3-week survival lesions demonstrate a mean 66% transmurality. The investigators concluded that lesions leading to electrical isolation are an effective means of blocking atrial conduction, despite being histologically incomplete. Linear lesions to both right and left atria also did not lead to 100% transmurality in all tissue sections. Areas of thickened tissue such as the RAA lead to the highest failure rates, and thinner tissue had more transmural tissue sections (SVC to IVC). While tissue thickness likely played a factor in these results, the length of the probe exposed to the tissue also led to warmer temperatures, and therefore decreased its efficacy, leading to failures (Table 2). Short lesions with thin tissue such as the LAA to LPV lesions led to 100% transmurality. Furthermore, although 4 of 6 LPV to mitral annulus lesions demonstrated areas of complete transmurality, none of these lesions demonstrated full transmurality throughout the entire length of the lesion. Cardiac tissue underneath the CS moving toward the annulus remained viable in all instances. Reasons for this failure include the presence of thick regions of epicardial fat in the interatrial groove, close proximity of the circumflex coronary artery preventing adequate probe alignment, and technical limitations of the prototype probe. To avoid injury to the circumflex artery, we were careful to avoid any contact with the vessel, which prevented adequate position to ablate the region of cardiac tissue underneath the CS. In addition, the tip of this prototype probe had nearly a half-centimeter region that did not cool as well as the rest of the probe. We believe this played a factor in preventing the creation of fully transmural lesions in this anatomic region. Modifications have since been made to avoid this limitation.
As with most animal studies, data from this study cannot be directly extrapolated to the human heart, whose tissue thickness and fat content are different. Although it appears that this ablation device was able to produce transmural lesions in this dog model, we cannot extrapolate this information to predict the degree of transmurality achievable in a human heart, which may be affected by chronic conditions such as rheumatic heart disease and long-standing mitral regurgitation. Most, but not all, lesions of the Cox maze III procedure were performed in this study. The LPV to RPV connecting lesion was not performed because of the anatomic proximity of the inferior pulmonary veins. Such close proximity may have led to an inability to differentiate the clamp lesions from the linear connecting lesion.
In conclusion, recent advances in technology have led to the creation of ablative devices that can replace the cut-and-sew technique used in the Cox maze III procedure. Cryoclamp ablation of the PVs and LAA in this canine model resulted in lesions with complete conduction block. Linear epicardial cryoablation using the 10-cm linear probe demonstrated transmurality in 85% of the tissue sections. Thus, compared with the CC device, linear epicardial cryoablation was not as effective at producing consistent transmural lesions.
This novel cryoablation probe and clamp device may offer yet another option for cardiac surgeons performing ablative procedures to treat AF. With this prototype, the clamp and probe easily disassemble and reassemble, allowing the surgeon to perform both clamp ablations and linear ablations with a single instrument. Future prototypes of this cryothermy technology may increase further the ability to perform transmural lesions via an epicardial approach on a beating heart, which may then continue to improve minimally invasive approaches using cryothermy.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Reyes, A. Benedicto, J. Bustamante, A. Sarraj, J. Manuel Nuche, P. Alvarez, and J. Duarte Restoration of atrial contractility after surgical cryoablation: clinical, electrical and mechanical results Interactive CardioVascular and Thoracic Surgery, October 1, 2009; 9(4): 609 - 612. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Groh, O. A. Binns, H. G. Burton III, S. W. Ely, and A. M. Johnson Ultrasonic Cardiac Ablation for Atrial Fibrillation During Concomitant Cardiac Surgery: Long-Term Clinical Outcomes Ann. Thorac. Surg., December 1, 2007; 84(6): 1978 - 1983. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Blomstrom-Lundqvist, B. Johansson, E. Berglin, L. Nilsson, S. M. Jensen, S. Thelin, A. Holmgren, N. Edvardsson, G. Kallner, and P. Blomstrom A randomized double-blind study of epicardial left atrial cryoablation for permanent atrial fibrillation in patients undergoing mitral valve surgery: the SWEDish Multicentre Atrial Fibrillation study (SWEDMAF) Eur. Heart J., December 1, 2007; 28(23): 2902 - 2908. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Saltman Must we cross the wall to get to the other side? J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 224 - 225. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |