J Thorac Cardiovasc Surg 2006;131:447-454
© 2006 The American Association for Thoracic Surgery
Cardiothoracic Transplantation |
a Department of Biomedical Engineering, Lerner Research Institute, The Cleveland Clinic Foundation, Cleveland, Ohio
b Department of Thoracic and Cardiovascular Surgery, The Cleveland Clinic Foundation, Cleveland, Ohio
c George M. and Linda H. Kaufman Center for Heart Failure, The Cleveland Clinic Foundation, Cleveland, Ohio
d Transplant Center, The Cleveland Clinic Foundation, Cleveland, Ohio
e Department of Quantitative Health Sciences, The Cleveland Clinic Foundation, Cleveland, Ohio
f Department of Cardiovascular Medicine, The Cleveland Clinic Foundation, Cleveland, Ohio
Received for publication April 8, 2005; revisions received August 31, 2005; accepted for publication September 19, 2005. * Address for reprints: Eugene H. Blackstone, MD, Department of Thoracic and Cardiovascular Surgery/JJ40, The Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195 (Email: blackse{at}ccf.org).
| Abstract |
|---|
|
|
|---|
METHODS: Between 1991 and 2002, left ventricular assist devices were implanted in 207 patients, exclusive of those receiving preoperative mechanical circulatory support, which precluded measuring right ventricular stroke work. Duration of inotropic support was analyzed as a continuous variable, truncated by death or transplantation, and in turn as a risk factor for these 2 events.
RESULTS: Inotropic support decreased from 100% on the day of implantation to 57%, 33%, and 22% by days 7, 14, and 21. Its duration was strongly associated with lower preimplantation right ventricular stroke work index, older age, and nonischemic cardiomyopathy and was associated (P < .04) with higher mortality before transplantation but not with transition to transplantation. We identified no preimplantation risk factors for right ventricular assist device use because of its relatively infrequent use in this population (18 patients, only 4 of whom survived to transplantation).
CONCLUSION: Duration of inotropic support after left ventricular assist device insertion is strongly correlated with low preimplantation right ventricular stroke work index. In turn, it was associated with reduced survival to transplantation. Thus, right ventricular stroke work measured before implantation might be useful in decision making for biventricular support, destination therapy, or total artificial heart.
| Introduction |
|---|
|
|
|---|
For example, right ventricular assist device (RVAD) support is a surrogate for severe RVD after LVAD implantation.
1-6
Not only is this not known before implantation, it identifies only the extreme of the RVD spectrum. Our clinical impression has been that RVD is an underappreciated problem with serious consequences that is more prevalent than reflected by contemporary use of RVAD support.
Prolonged inotropic support, arbitrarily defined as 14 days or longer, has been used as another surrogate for less severe forms of RVD.
5,6
It also is a postimplantation variable; in addition, it has not been explicitly recognized that duration of support is a continuous variable that might be truncated by transplantation and death. Therefore, to better manage critically ill patients with clinically evident RVD and to assist in preimplantation decision making for biventricular support or cardiac replacement, we reexamined the duration of inotropic support as a surrogate for RVD by (1) identifying preimplantation factors associated with it, with particular emphasis on preinsertion right ventricular (RV) stroke work, and (2) assessing its impact on clinical outcomes.
| Patients and Methods |
|---|
|
|
|---|
Patients
From December 1991 through July 2002, 259 patients underwent LVAD implantation at the Cleveland Clinic as a bridge to transplantation. Fifty-two patients on pre-LVAD extracorporeal membrane oxygenation or other temporary mechanical support devices (eg, Abiomed BVS 5000) were excluded because using these devices precludes accurately measuring hemodynamic components of RV stroke work. Thus, 207 patients were included in this analysis. During support, patients were maintained on the transplantation list. Demographic, medical history, hemodynamic, and laboratory values, and outcomes were extracted from our Unified Transplantation Database, which has been approved for use in research by the institutional review board (Table 1).
|
3 µg · kg1 · min1) was administered. In 18 patients an RVAD was placed for severe clinically evident RVD with persistently low LVAD flow despite high-dose inotropic support.
Data Analysis
Data are summarized as percentages, means ± standard deviation, medians with 15th and 85th percentiles, or 68% confidence limits equivalent to ±1 standard deviation, as appropriate.
Duration of inotropic support
Because duration of inotropic support might be truncated (censored) by death or transplantation, it was analyzed from the time of LVAD implantation until the earliest occurrence of either of these competing events
8,9
by using nonparametric Kaplan-Meier and parametric hazard function methods.
10
Risk factors for inotropic support and RVAD use
Preoperative risk factors for longer duration of inotropic support were identified by hazard function regression using variables listed in the Appendix,
10
and for RVAD use by means of logistic regression. Bootstrap bagging (1000 cycles) was used for risk factor selection,
11
retaining variables in each bootstrap model with a P value of less than .05. Factors appearing in at least 50% of model replications were considered reliably identified with a P value of less than .05.
Of particular importance to this study was pre-LVAD right ventricular stroke work index (RVSWI). It was calculated from preimplantation Swan-Ganz catheter data as follows:
|
| (1) |
Impact of inotropic support on clinical outcomes
Association of duration of inotropic support with death before transplantation and with transplantation itself was assessed in the context of a cross-over phenomenon (technically, a modulated renewal process).
12
Patients were initially considered as being "on inotropes"; at the time inotropes were discontinued (cross-over), patients were restarted at a new time zero ("not on inotropes"). A variable for duration of previous inotropic support was generated whose value was 0 before inotropic support was discontinued and, thereafter, number of days of support. This allowed us to quantify the association with outcome of inotropic support and its duration.
To depict the influence of inotropic support on survival, we used the cross-over model to simulate 6-month survival for a continuous set of durations of inotropic support. Specifically,
|
| (2) |
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
RVAD Support
An RVAD was placed in 18 (8.7%) patients, 16 within 2 days of LVAD implantation. The other 2 were placed on days 3 and 12 because of respiratory failure followed by severe RVD. In 17 of the 18 patients, high-dose inotropic support for severe RVD preceded RVAD placement. Need for RVAD support was not related reliably to any preoperative risk factor investigated. Survival before transplantation of these 18 patients was 47%, 29%, and 22% at 10, 20, and 30 days, respectively (Figure 5).
|
| Discussion |
|---|
|
|
|---|
Clinical Correlates of RVD after LVAD Implantation
Several studies have used clinical surrogates of RVD after LVAD implantation, particularly need for RVAD or inotropic support.
3,5,6,13
Need for RVAD support identifies patients with severe RVD, which is often fatal. We have previously shown an association between poor preoperative RV contractility and subsequent RVAD implantation. In this study we did not find an association between RVAD use and preoperative factors. However, this apparent discrepancy is explained by our excluding 52 patients with pre-LVAD temporary mechanical circulatory assistance, which was a risk factor for RVD in prior studies.
2,14,15
We excluded these patients because pre-LVAD hemodynamic measurements (and specifically RVSWI) with transiently turned off mechanical support were not available in many cases.
We believe that use of inotropic support as a surrogate for less severe RVD requires a more sophisticated approach than simply dichotomizing its duration into short versus long support, as has been done by others.
5,6,16
Its duration is a continuous variable without a sharp cut-off point with respect to outcomes. Consider the limitations of any arbitrary cut-off duration. With a cut-off duration of 14 days, a patient on inotropic support for 13 days would be classified as not having RVD, whereas one supported for 14 days would be classified as having RVD. This is not a clinically helpful distinction.
Clinical Implications
Factors associated with prolonged inotropic support identify patients at risk for RVD, delayed rehabilitation, and other complications after LVAD implantation. Although high preoperative pulmonary vascular resistance is a well-recognized risk factor for RVD after heart transplantation
17
and thus serves as an exclusion criterion, a preoperative predictor of RVD after LVAD implantation useful for decision making has yet to be established. Our results suggest that preoperative RV contractility as reflected in RVSWI, but not preload or afterload, might be one such factor. We now routinely measure RVSWI to assess RV contractility. Perhaps RVSWI will be measured and manipulated as we do pulmonary vascular resistance to assess transplant candidacy. Ideally, a more load-independent measure than RVSWI will be identified to quantify RV contractility. We speculate that such a measure could allow us to better anticipate need for biventricular support versus LVAD support alone.
Current clinical strategies for treating RVD focus on increasing RV contractility with inotropes, providing circulatory support only in cases of severe RVD. For example, we use milrinone, a phosphodiesterase III inhibitor with positive inotropic and vasodilatory effects.
18
Nevertheless, as found in this study, pharmacologic treatment of RVD is associated with poor survival and thus requires alternative strategies. Although the need for an RVAD is associated with high mortality, this likely reflects current indications. If RVADs were used for less severe forms of RVD, their benefit might become more apparent. Unfortunately, current RVAD use is associated with high risks of bleeding and thromboembolic events, which preclude broader clinical application. Thus, developing blood-compatible implantable RVADs is needed. Cardiac replacement with total artificial hearts might become an alternative for biventricular failure.
19,20
Results presented in this report should be helpful in selecting candidates for these devices.
The most practical use of the knowledge from this study might be in decision making for destination therapy with LVADs alone. If patients remain inotrope dependent, their prognosis is predicted to be poor in the absence of a biocompatible RVAD or other treatment option.
15
As demonstrated from this study, nonischemic cardiomyopathy, advanced age (likely a large group of these will not be transplantation candidates), and poor preoperative RV function with low RVSWI might prove to be relative clinical contraindications to destination therapy with LVADs.
| Limitations |
|---|
|
|
|---|
In identifying risk factors for duration of inotropic support, we considered only preimplantation variables. Certainly, postoperative complications, such as pump malfunction, could also result in need for inotropic support. In addition, many surgeons wean inotropes after LVAD implantation according to device flow and central venous pressure. These postoperative variables were deliberately not considered because our objective was to identify preimplantation risk factors that might help guide device selection. We also did not examine time-varying dosage of inotropic support as a longitudinal end point, only its duration.
Finally, patients receiving pre-LVAD temporary mechanical circulatory support were excluded from this study to ensure accurate assessment of hemodynamics. For this reason, we were limited in some aspects of our analyses that have been covered adequately in our previous studies.
3,7
Thus, we recognize that pre-LVAD circulatory support is associated with occurrence of severe post-LVAD RVD (need for RVAD), even though we could not reanalyze this in the present study. In our experience, these patients also have preoperative RVD and likely have a low RVSWI, if it could be measured accurately.
| Conclusion |
|---|
|
|
|---|
| Appendix |
|---|
|
|
|---|
Demography: Age (y), sex, body surface area (m2), body mass index (kg · m2), weight (kg), height (cm).
Medical history: Diagnosis of heart disease, prior thoracic surgery, presence of implantable cardioverter-defibrillator.
Temporary support (within 3 days of implantation): Mechanical ventilation, intra-aortic balloon pump use, inotropic support, duration of pre-LVAD hospital stay (d).
Comorbidities (within 3 days of implantation): Acute myocardial infarction, infection, ventricular tachycardia or fibrillation.
Hemodynamics (immediately before implantation): Pulmonary artery systolic, diastolic, and mean pressures (mm Hg), cardiac index (L · min1 · m2), pulmonary vascular resistance (Wood units), central venous pressure (mm Hg), right ventricular stroke work index (mm Hg · mL · m2) calculated from pulmonary artery (Swan-Ganz) data as follows:
|
|
Laboratory values (1 day before implantation): Creatinine (mg · dL1), total bilirubin (mg · dL1), aspartate transaminase (U · L1), total protein (g · dL1), albumin (g · dL1).
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. M. Adamson, M. Stahovich, S. Chillcott, S. Baradarian, J. Chammas, B. Jaski, P. Hoagland, and W. Dembitsky Clinical Strategies and Outcomes in Advanced Heart Failure Patients Older Than 70 Years of Age Receiving the HeartMate II Left Ventricular Assist Device: A Community Hospital Experience J. Am. Coll. Cardiol., June 21, 2011; 57(25): 2487 - 2495. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. R. Ensor, C. A. Paciullo, W. D. Cahoon Jr, and P. E. Nolan Jr Pharmacotherapy for Mechanical Circulatory Support: A Comprehensive Review Ann. Pharmacother., January 1, 2011; 45(1): 60 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Matthews, T. M. Koelling, F. D. Pagani, and K. D. Aaronson The Right Ventricular Failure Risk Score: A Pre-Operative Tool for Assessing the Risk of Right Ventricular Failure in Left Ventricular Assist Device Candidates J. Am. Coll. Cardiol., June 3, 2008; 51(22): 2163 - 2172. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Stone Current Status of Mechanical Circulatory Assistance Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2007; 11(3): 185 - 204. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||