J Thorac Cardiovasc Surg 2006;131:621-624
© 2006 The American Association for Thoracic Surgery
Surgery for Congenital Heart Disease |
Cardiothoracic Centre, All India Institute of Medical Sciences, New Delhi, India.
Received for publication August 23, 2005; revisions received September 29, 2005; accepted for publication October 10, 2005. * Address for reprints: Shiv Kumar Choudhary, MCh, Department of Cardiothoracic and Vascular Surgery, All India Institute of Medical Sciences, New Delhi, 110029, India. (Email: shivchoudhary{at}hotmail.com).
| Abstract |
|---|
|
|
|---|
METHODS: Between December 2003 and May 2005, 32 patients aged 18 days to 2 years (mean age, 2.5 ± 0.5 months) and weighing 2.1 to 6.3 kg (mean, 3.6 ± 1.3 kg) underwent adjustable pulmonary artery banding.
RESULTS: All patients survived the operation. There were 2 deaths, one caused by meningitis and another caused by aspiration pneumonitis. Satisfactory band gradients were achieved between 3 and 10 days (7.2 ± 2.6 days) in 3 to 6 sittings. Mean follow-up was 7.5 ± 3.8 months (1-16 months). One patient required reoperation for unsatisfactory band gradient 2 weeks after discharge. There were no late deaths. Follow-up computed tomographic angiograms (n = 4) demonstrated proper band placement and ruled out distortion of the pulmonary arteries. Four patients underwent uneventful definitive operations after an interval of 7 to 13 months.
CONCLUSION: This technique of percutaneously adjustable pulmonary artery banding is simple and inexpensive and allows easy band adjustments without the need for multiple reoperations.
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Standard anesthesia techniques with endotracheal intubation were used. A small cannula was placed in the radial artery for arterial pressure monitoring, and 2 peripheral veins were cannulated for venous access.
Informed consent was obtained from all patients. The preferred approach was through a median sternotomy. Lately we have started performing this operation through an upper partial sternotomy in which only the manubrium is divided. After sternotomy, the thymus was split open but not excised. The pericardium was opened just enough to expose the great vessels. The aorta and the main pulmonary artery (PA) were separated from each other proximally. The ductus was dissected out and clipped in all cases. A right-angle forceps was passed between the aorta and the PA, and the PA was looped with a No. 2 Ethibond (Johnson and Johnson Inc, Somerville, NJ) suture. The right-angle forceps was again passed in a similar fashion, and the suture was grasped again by the tip of the right-angle forceps so that the PA was doubly looped. Now both the ends of the thread were passed through a 0.5-cm x 0.5-cm polytetrafluoroethylene (PTFE) pledget (IMPRA, Inc, Tempe, Ariz), which was anchored to the adventitia of the PA with interrupted 6-0 polypropylene sutures to prevent migration of the band. Both the arms of the Ethibond suture were clipped to each other with a ligaclip (LT200; Ethicon Endo-Surgery, Inc, Cincinnati, Ohio) flush with the PTFE pledget. These sutures were then brought out through the pericardium, which was closed after placing one temporary epicardial pacing wire on the right ventricle. After pericardial closure, the 2 ends of the suture were brought out through the lateral edge of the sternum and then through the subcutaneous tissue and skin. The sternum was then closed. The 2 ends of the suture were then passed through a 1-cm x 2-cm PTFE pledget and clipped together with a big ligaclip (LT400, Ethicon Endo-Surgery, Inc). The ends of the suture were then tied together to form a loop. The patient was then shifted to the intensive care unit. The two ends of the suture were then tied together to form a loop (Figure 1).
|
For single-ventricle candidates, the band was tightened as much as possible. However, we ensured that the saturation did not decrease to less than 75%. If there was hemodynamic compromise during tightening or there was bradycardia or a decrease in the saturation, the band was loosened by simply removing the lowermost clip and allowing the threads to retract down.
On achieving satisfactory gradients, the band was internalized by simply opening up the subcutaneous plane for about 1 cm and clipping the 2 threads together. Above this, the sutures were cut to release the portion of the band outside the skin. The skin was then closed with 2 interrupted sutures. We observed that the band gradient increased by 5 to 10 mm Hg at the time of internalization.
After discharge from the hospital, these patients were followed up after 1 month and then at every 3-month interval. At each follow-up visit, they underwent clinical examination, measurement of saturation and echocardiography for band gradients, and estimation of PA pressure and ventricular function.
| Results |
|---|
|
|
|---|
There were no late deaths. Follow-up computed tomographic angiograms (n = 4) demonstrated proper band placement and ruled out distortion of the PAs. The band gradients increased by 2 to 12 mm Hg without hemodynamic changes. Four patients underwent definitive operations after an interval of 7 to 13 months. In 2 patients undergoing a bidirectional Glenn shunt at reoperation, the main PA was divided and repaired with autologous glutaraldehyde-fixed pericardium. In the other 2 patients undergoing closure of ventricular septal defect, a longitudinal pulmonary arteriotomy was made across the band, and the PA was repaired with autologous glutaraldehyde-fixed pericardium. The remaining patients are awaiting reoperation.
| Discussion |
|---|
|
|
|---|
For these critically ill patients, there is a need to develop a method of banding that is delicately adjustable not only during the operation but also postoperatively because the patient's hemodynamic and respiratory condition might change postoperatively, thereby altering the gradient created at the time of the operation. Several investigators have developed a variety of adjustable bands.
3-8
Bonnet and colleagues
8
have recently described their early results with a telemetric adjustable FloWatch pneumatic device in which tightening and loosening are regulated by an external control unit. Although effective, this device and the other described methods are expensive, require special equipment, and have not found general acceptance.
Gradual and delicate tightening and loosening of the band are possible with our method, making it particularly useful. Two-dimensional echocardiography and continuous-wave Doppler study were used to ascertain the best-tolerated degree of tightening along with the dynamic performance of right and left ventricular size and function. This is important because gradient alone is not an adequate means of assessing tolerable outflow obstruction. The most important advantage is that the tightening is carried out under physiologic conditions when the patient is off ventilator and off inotropes. Also, because the band is tightened at multiple sittings, the increase in afterload to the ventricle is gradual and can be adjusted to that tolerated by the patient without hemodynamic instability under normal physiologic conditions. It can be loosened immediately in the case of poor tolerance. Furthermore, this technique is simple, inexpensive, does not require specialized equipment, and is minimally invasive in those instances in which it is performed through a ministernotomy.
An initial concern about this type of band was that the PA could get tented toward the sternum and could lead to PA distortion, as well as making sternal re-entry hazardous at reoperation. To address these concerns, we performed computed tomographic angiography in 4 randomly selected patients at follow-up and found that these concerns were misplaced. Also, the sternal re-entry at reoperation was not difficult, and the PA was not adherent to the back of the sternum. Another concern was that the suture used for the band might cut through a dilated and tense PA. However, we have not observed this. Careful dissection and placement of the band, avoiding trauma to the adventitia of the PA, appear to have prevented this complication. Also, we initially feared pressure necrosis of the skin underlying the PTFE pledget; to prevent this, we ensured that the size of the pledget was large enough (1 cm x 2 cm) to distribute the pressure in a wider area.
A limitation of our study is that it is an early experience, and we do not have direct measurements of the PA pressures. However, the echocardiographically achieved gradients have been satisfactory, and the band has been an adequate initial palliation in patients who have undergone reoperation. The findings at reoperation suggest that there were not many adhesions around the band, and there was no distortion of the PA. In addition, there was no band migration, and it was well positioned without compressing the left or right PAs.
None of our patients have required loosening of the band at follow-up. We believe that the purpose of the band is to provide short-term palliation, and any loosening that might be required in the initial postoperative phase can be easily achieved before internalization of the band, which is usually performed after 10 to 20 days. However, if long-term palliation is desired, the potential for somatic growth is a concern.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
V Chaturvedi, S Talwar, B Airan, and B Bhargava Interventional cardiology and cardiac surgery in India Heart, March 1, 2008; 94(3): 268 - 274. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Talwar, S. K. Choudhary, A. Mathur, B. Airan, R. Singh, R. Juneja, S. S. Kothari, and A. Saxena Changing Outcomes of Pulmonary Artery Banding With the Percutaneously Adjustable Pulmonary Artery Band Ann. Thorac. Surg., February 1, 2008; 85(2): 593 - 598. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno Invited Commentary Ann. Thorac. Surg., February 1, 2008; 85(2): 598 - 598. [Full Text] [PDF] |
||||
![]() |