Median sternotomy is a common approach to expose the heart for cardiac surgery. The sternum is usually closed with stainless wires or thick-braided sutures after surgery. We have used Polydioxane (PDS) cord sutures 1.0 mm in diameter for closing the sternum on 350 patients for the last one and a half year. But recently, three patients suffered broken PDS sutures and instability of the sternum several weeks after surgery. The sternum had to be closed immediately in two of them after the broken sutures because of wound dehiscence. One patient was observed closely without surgical intervention because of renal failure and MRSE carrier on skin culture.
Cases
All three patients had some risk factors for wound healing. One was a 73 -year-old man, who suffered an old myocardial infarction with poor left ventricle function, chronic atrial fibrillation and diabetes mellitus. He underwent on-pump 4-vessel aorto-coronary bypass grafting concomitant with Maze procedure. He suffered a minor stroke and required long mechanical ventilation therapy and tracheostomy on the 10th postoperative day. Sternum instability was observed on the 12th postoperative day after bronchoscopy. He revealed wound dehiscence on the 14th day and sternum re-closure was performed on the same day. All 6 PDS cord sutures were broken beside the knot without any loose ligation or broken sternum.
One was a 75-year-old man with aortic arch aneurysm and angina pectoris. He had chronic obstructive pulmonary dysfunction and severe obesity. Total arch replacement was performed under hypothermic circulatory arrest concomitant with aorto-coronary bypass grafting. He was extubated on the second postoperative day and discharged from the intensive care unit on the third postoperative day. Sternal instability was revealed on the 7th postoperative day and wound dehiscence on the 9th postoperative day. Reclosure of the sternum was performed on the 10th postoperative day. Only the lower part of the sternum was broken, but all 6 PDS code sutures were broken without loose knot.
The last was a 68-year-old woman with angina pectoris. She had a history of atrial septal defect closure via right thoracotomy at 31 years of age. The case was complicated with asthma, diabetes mellitus and renal failure. Off pump 3 vessels coronary bypass grafting was performed. She was extubated on the next day and recovered well. But sternal instability occurred during an asthma attack on the 10th postoperative day with a cracking sound. Her chest CT revealed a separation of several cm throughout the sternum. She was followed without sternal reclosure because she had no wound dehiscence, suffered from renal failure and MRSE carrier on her skin.
Macroscopic Finding
The broken PDS cord sutures of case 1 were carefully examined. All six PDS cords had completely snapped. Two of them were placed in the sternum manubrium and the other 4 were located just around the sternum body at the intercostal space. No loose knots were observed. Broken sites were 3 to 15 mm apart from the knot. The cord stumps gradually became thin, kinked with curled filaments and possibly had been torn off (Figures 1 and 2).
Twenty mm away from the broken end on 2 of them, what appeared to be abrasion injures were observed. On the sternum manubrium where the cord protruded, the cord was thought to have been abraded on the edge of the bone.
|
|
PDS cord suture is designed to close the sternum. It has enough strength for traction force, but it may be damaged by shear stress with a bone edge. The damaged part of the suture loses strength and may break under severe traction due to coughing or bronchial suction. Macroscopic findings of broken sutures indicated abrasion injury due to friction with the bone edge or that they were torn off under severe traction.
PDS cord is a synthetic absorbable suture. PDS cord of 1.0 mm in diameter has sufficient tensile strength (180 N) for closing the sternum before use. But it decreases gradually after use in the body. Manufacturer data show the tensile strength of 50-65% at 3 weeks and 30-40% at 6 weeks after implantation. This is another reason for broken sutures.
Broken sutures made of PDS cord may occur several weeks after surgery when patients have a persistent cough. PDS cord sutures do not have sufficient reliability to close the sternum, especially in patients with chronic obstructive pulmonary dysfunction and obesity.
1
References
This article has been cited by other articles:
![]() |
X. N. Yang and U. Pastorino Are absorbable sutures inadequate to close the sternum? J. Thorac. Cardiovasc. Surg., December 1, 2006; 132(6): 1503 - 1503. [Full Text] [PDF] |
||||
![]() |
N. Luciani, A. Anselmi, and G. Possati Adjusting the indication to polydioxane suture for elective sternal closure. J. Thorac. Cardiovasc. Surg., November 1, 2006; 132(5): 1243 - 1244. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||