J Thorac Cardiovasc Surg 2006;132:20-26
© 2006 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
Department of Health Sciences, University of York, York, United Kingdom.
Received for publication October 12, 2005; accepted for publication January 13, 2006. * Address for reprints: Ole Lund, MD, DSc (PhD), 30 Gresham Close, Darlington, County Durham DL1 2YT, United Kingdom. (Email: olelund{at}btinternet.com).
| Abstract |
|---|
|
|
|---|
METHODS: Following a formal study protocol, we performed a dedicated literature search of publications during 1989 to 2004 and included articles on adult aortic valve replacement with a mechanical or stented bioprosthetic valve if age, mortality statistics, and prevalences of well-known risk factors could be extracted. We used standard and robust regression analyses of the case series data with valve type as a fixed variable.
RESULTS: We could include 32 articles with 15 mechanical and 23 biologic valve series totaling 17,439 patients and 101,819 patient-years. The mechanical and biologic valve series differed in regard to mean age (58 vs 69 years), mean follow-up (6.4 vs 5.3 years), coronary artery bypass grafting (16% vs 34%), endocarditis (7% vs 2%), and overall death rate (3.99 vs 6.33 %/patient-year). Mean age of the valve series was directly related to death rate with no interaction with valve type. Death rate corrected for age, New York Heart Association classes III and IV, aortic regurgitation, and coronary artery bypass grafting left valve type with no effect. Included articles that abided by current guidelines and compared a mechanical and biologic valve found no differences in rates of thromboembolism.
CONCLUSION: There was no difference in risk factor-corrected overall death rate between mechanical or bioprosthetic aortic valves irrespective of age. Choice of prosthetic valve should therefore not be rigorously based on age alone. Risk of bioprosthetic valve degeneration in young and middle-aged patients and in the elderly and old with a long life expectancy would be an important factor because risk of stroke may primarily be related to patient factors.
| Introduction |
|---|
|
|
|---|
Recent years have nonetheless witnessed a shift in prosthetic heart valve preference from mechanical to biologic valves, which is not just explained by the increasing age of patients. Bioprosthetic valves are increasingly being implanted in younger patients
7
spurred by the belief that third-generation stented xenografts have superior durability compared with previous models.E1 The risk of SVD accelerates with time (ie, requires long follow-up to become apparent) and is inversely related to age.
6
High SVD rates of earlier models implanted in younger patients through the 1970s and 1980s resulted in a high proportion of redo AVR causing the preference to swing back toward mechanical valves. Our knowledge base on SVD of third-generation xenografts is consequently based predominantly on valve series with too few young patients who underwent operation too late in the series to have an impact at the present time. Extrapolation from seemingly low rates of SVD in current relatively short "old age series" to younger patients may thus be in error.
Thromboembolism and anticoagulant-related bleeding remain the dominant complications of mechanical valves. The incidence rates of these complications seem to be fairly constant after AVR, but reported rates vary considerably especially for bioprosthetic valves including rates of zero or below the age-specific stroke rates of the background population.
6,8
Meta-analysis of published data taking both death and the main prosthesis-related complications into account to assess the benefits of mechanical versus biologic valves
9-11
may thus be misleading. Previous meta-analyses have, furthermore, included series with obsolete mechanical valves
9,10
and bioprosthetic valve series with follow-up of less than 10 years with related low SVD rates.
10,11
Finally, meta-analyses have been coupled with microsimulation, which is difficult to comprehend and for which the chosen modeling parameters may introduce bias.
9,11
A death is an extremely well-defined event, and it would make common sense to postulate that a prosthetic valve complication entity that does not impact mortality of the patients is merely trivial, and none of them are. We therefore had as our primary aim to perform a meta-analysis on pertinent publications from 1989 to 2004 to determine the age-corrected impact of currently available mechanical (bileaflet and single disc) versus stented bioprosthetic (porcine and bovine pericardial) valves on crude mortality after AVR and to correct for the impact of other well-known risk factors using standard and robust regression analysis techniques. To minimize bias related to mortality of various complications with peculiar time effects, which may differ between mechanical and biologic valves, we chose to include only publications with a maximum follow-up of at least 10 years.
| Materials and Methods |
|---|
|
|
|---|
Inclusion and Exclusion Criteria and Essential Data Extraction
Articles were included if essential data (defined below) could be extracted from a pure mechanical or stented bioprosthetic AVR series with a longest follow-up of at least 10 years, with at least 75 patients and a maximum of 5% aged less than 15 years, and with currently available prosthetic valves. We accepted up to 10% obsolete valve types in mixed mechanical or biologic valve series and excluded articles in which the operative period started before 1975 to avoid time bias. We included only articles that claimed conformance to either the original
13
or current
14
guidelines for reporting (or that detailed comparable definitions and method of follow-up), had at most 5% of patients lost to follow-up,
13,14
and originated in the industrialized world (Appendix E1). To avoid confounding between case series by factors not related to AVR per se, we excluded series with prevalences of redo operations, concomitant surgery other than coronary artery bypass grafting (CABG), and operation for active endocarditis of more than 20%. It was essential that we could extract the following data: (1) Death statistics: total number of deaths (early and late) of any cause and total follow-up (accumulated number of patient-years). (2) Demographic and operative data: mean age of the patients with standard deviation or range; prevalence of women, concomitant CABG, concomitant other procedures, redo operation, and operation for active endocarditis. (3) A measure of disease severity: We chose New York Heart Association (NYHA) functional classes because they are usually reported. Other "disease severity" variables including left ventricular function and comorbidities could only be extracted from a fraction of the included articles (Appendix E1). (4) Valve disease: Prevalence of aortic stenosis, regurgitation, and mixed lesion could be recorded in all but 1 of the included articles. Specific exclusion criteria are given per inference from the above. We also excluded pure specialty articles, such as those on the small aortic root/small-sized prosthetic valves and so forth. In case of repeat publications from the same institution, we chose the last published or the one from which the essential data could be extracted. Further details are given in Appendix E1. In total we were able to include 32 articles
E1-E32
with 38 valve series, which are detailed in Table E1
relative to valve type.
|
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The main result of the present analysis is simple and clear: There was no significant difference in death rate between mechanical and bioprosthetic valves when corrected for age and well-known risk factors with no interaction between age and valve type, that is, we could find no difference irrespective of age. The strong impact of age, advanced NYHA class, and concomitant CABG on mortality after AVR needs no explanation; these risk factors have been identified in countless articles. Aortic regurgitation had a negative regression coefficient in the least squares regression model, indicating a reduction in death rate. The explanation probably is that age and NYHA classes III and IV had statistically far stronger impact in the model, thus leaving "young patient in good NYHA class" for aortic regurgitation. This is strengthened by the fact that we were testing case series data and not individual patient data: Patients with aortic regurgitation are younger on average than those with aortic stenosis
E2
and will then dominate in series with a low mean age. The fact that the regression models made perfect clinical sense is an important strength of our results. It also indicates that correction for important risk factors is crucial for a meta-analysis to make sense.
We could not analyze the influence of small valve (eg, prevalence of valve size
21 mm) because such data were available in only a fraction of the included valve series (Appendix E1). An impact of prosthetic valve to patient size mismatch seems to have been noted primarily in studies dominated by stented porcine valves.
17
Small stented porcine valves have always been problematic, and many surgeons avoid using stented bioprostheses smaller than 23 mm, implanting a mechanical valve instead irrespective of age.
18
Blackstone and coworkers,
19
analyzing 13,258 patients with AVR from several centers, noted that small bioprosthetic valves that potentially could cause size mismatch were rarely implanted. Several of the presently included bioprosthetic valve series did not include 19-mm valves.
E19,E25,E30,E32
One may thus assume that mechanical valves often serve as "bail-out" valves for bioprostheses, which may compromise some mechanical valve series.
There were only 2 biologic valve series
E21,E22
with subjects with a mean age less than 60 years in whom the majority of the mechanical valve series were located, which is a weakness of the present study. The group with a young mean age is the group for which one would especially expect an impact of bioprosthetic valve SVD. One of the two "young" bioprosthetic valve series with a decent follow-up was small,
E21
whereas the other had a maximum follow-up of only 12 years.
E22
In general, the included bioprosthetic valve series had a mean follow-up of only 5.3 years, which may be too short to involve an impact of SVD-related mortality and thus may have significantly skewed the present results.
There are 2 more problems associated with gauging the risk of SVD and its impact. Many of the presently included articles report only freedom from explant because of SVD
E1,E17,E18,E29,E32
or indicate that more than 80% with SVD actually underwent a redo operation.
E23,E27,E28
All of those studies were retrospective and did not report regular echocardiographic follow-up, and many patients with SVD may thus have been overlooked. In a large 25-year study of first-time AVR with allografts, involving annual echocardiography, from a center renowned for accepting patients for redo AVR who had been turned down elsewhere, 258 patients with significant SVD were identified but only 53% underwent a redo AVR.
20
Furthermore, the 5-year survival after the diagnosis of SVD was 78% (± a standard error of 3%) with redo AVR but only 33% (±5%) without; patients who were not offered a redo AVR were on average 12 years older and had a 3-fold higher prevalence of concomitant CABG at the primary AVR than those who were (previously unpublished analysis by OL). The patients who were not offered a redo operation had obviously become too old or sick with a risk associated with reoperation that was deemed unacceptably high. Remarkably, the 5-year survival after the first appearance of congestive heart failure in the general population is almost exactly similar to the above poor 5-year survival after unoperated SVD.
21
This means that freedom from explant (redo) because of SVR tells only half the tale, and that SVD is a very serious complication. Without a redo operation the prognosis is extremely poor, but the risk of reoperation is high with death and serious complications in more than 80% of elderly patients
22
who are thus less likely to be given the option. A recent analysis of 12-year freedom from SVD in first-, second-, and third-generation stented bioprosthetic valves indicated that differences in mean age of the valve series rather than valve design was the decisive factor.
6
One may thus suspect that it is not in the best interest of many patients, young in general and old with long life expectancy, to be offered a bioprosthesis.
The main problem with mechanical valves is the need for lifelong oral anticoagulant treatment. Bleeding and thromboembolism generally account for some 80% of valve-related complications associated with bileaflet disc valves.
E2
However, the rate of thromboembolism has in general been halved in aortic bileaflet disc valves compared with caged ball valves on similar "old high-intensity" anticoagulant treatment.
23
A decrease of the international normalized ratio (INR) to the currently recommended level (2.0-3.0) for bileaflet disc valves resulted in a further reduction in overall complication rates.
24
It is well established that patient self-management of oral anticoagulation (AC) treatment manages to keep INR within the desired interval significantly more often than with conventional management, which has resulted in a further decrease of complication rates.
25
Two large studies pooling data from 34
26
and 46
27
AC clinics indicated that the risk of bleeding was related to the primary indication for oral AC treatment, with the lowest risk in the group with a prosthetic heart valve and an INR target of 2.0 to 3.0 and a death rate close to that of the background population.
27
Several pertinent factors may indicte that it would be extremely difficult to engage bleeding and thromboembolic complications in a meta-analysis of mechanical versus biologic valve performance. The reported complication rates vary within the same mechanical valve type, but the variability is especially high with biologic valves.
6
In the present bioprosthetic valve series with a mean age of 71 to 74 years and claiming conformance to current guidelines,
14
the rates of thromboembolism varied from 1.40% to 6.45% per patient-year.
E18,E20,E26,E30
The bleeding rates ranged from nearly nothing to 1.18%/patient-year even though the guidelines clearly specify that all bleeding events must be recorded also in patients who are not on oral AC or antiplatelet treatment.
14
Limited emphasis has traditionally been put on thromboembolism and bleeding in bioprosthetic valve series partly because of the dogma that biologic valves are not thrombogenic and do not require AC treatment, which possibly explains reporting standards. A bioprosthetic valve does not protect a patient from the "normal occurrence" of, for instance, gastrointestinal, urogenital, and cerebral bleeding or from the "background rate" of stroke in general,
3,8
which "trickles over into mortality" whether the authors reported thromboembolism and bleeding or not. Neither does a biologic prosthetic valve protect the patient from requiring oral AC treatment for the usual (nonprosthetic valve) indications, and more than 20% of patients have been reported to be taking oral warfarin at a mean of 2.6 to 5.8 years after AVR.
E13,E18,E30
It has been shown that patients with a beneficial preoperative risk score obtained an age- and gender-specific normal rate of stroke (thrombotic, embolic, bleeding) and normal survival in the long term after AVR despite an obsolete mechanical valve and high-intensity AC treatment.
3
This is in accordance with the discovery by Butchart and coworkers
28
that nearly all patients with an aortic disc valve and a cerebrovascular event had at least 1 of the main risk factors for stroke that are also operational in the general population. This would indicate that the composition of a patient material has as much or higher impact on the rates of cardiovascular events as has the valve type. All 3 of the presently included articles that abided by the current guidelines
14
and compared a mechanical and biologic AVR series with a mean age greater than 70 years reported similar rates of thromboembolism
E11,E13,E31
and bleeding.
E11
Meta-analysis of randomized studies has become the gold standard in evidence-based medicine, but heart valve research has not lent itself easily to randomization because of the extended time required to get meaningful results. Only 2 randomized trials between mechanical and bioprosthetic valves have been published since 1989: the Edinburgh
29
and Veterans Affairs trials.
30
Both trials used valves that are now obsolete and pooled patients with isolated AVR, isolated mitral valve replacement, and multiple valve replacement, thus making interpretations difficult.
| Conclusion |
|---|
|
|
|---|
| Appendix E1 |
|---|
|
|
|---|
Table E1 details the included valve series relative to the specific prosthetic valve. In a few instances data were extracted from several articles detailing the same patient series. We contacted the corresponding author of articles published between 2000 and 2004 that potentially could be included but in which essential data were missing and received the requested information for 9 articles. Certain valve types have been modified several times since their introduction (eg, the Carpentier-Edwards standard porcine valve and Perimount bovine pericardial valve, Edwards Lifesciences, Irvine, Calif; and the variations of the St Jude bileaflet disc valve, St Jude Medical Inc, St Paul, Minn) but have retained their original name, whereas others (eg, Sorin's bileaflet and single disc valves, Sorin Biomedica, Modena, Italy) have been renamed with each change. We chose to accept such valve types as current. We included articles originating in the industrialized world defined as North America, Western Europe (Sweden, Finland, Norway, Denmark, Germany, Holland, Belgium, Luxembourg, United Kingdom, Ireland, France, Austria, Switzerland, Italy, Spain, Portugal, and Greece), Australia, New Zealand, and Japan. The following variables could only be extracted from a fraction of the publications and could therefore not be used in the statistical tests: prevalence of small prosthetic valves (
21 mm; 65%), prior myocardial infarct (50%), diabetes (20%), kidney failure (18%), hypertension (15%), poor left ventricular function (13%), and so forth.
| Appendix E2 |
|---|
|
|
|---|
For least squares linear regression, the outcome variable, death rate, should follow an approximately normal distribution with uniform variance, which was shown to be the case and substantiated by an approximately linear normal plot. Least squares regression gives the estimated difference between valve types as deaths per 100 patient-years. The case series were weighted by size using the total patient-years of follow-up. Negative binomial regression is a development of Poisson regression, which allows for more variation between the series than the Poisson distribution would allow. This is an approximation because deaths should happen randomly in time, but in AVR series the death rate is high at the beginning, decreases after the immediate postoperative period, and increases over the more than 10 years of follow-up as subjects age. Thus, we would not expect a uniform death rate over follow-up, and deaths would not be generated by a Poisson process.
We cannot treat study (or article) as a fixed factor, because most studies include only 1 series. We therefore adopted a random effects approach, treating each study as a cluster and using robust standard errors using the Huber-White-sandwich estimator of variance;
15
the effect is to increase the standard error and reduce power and precision when there is more than 1 series of the same valve type (eg, 2 different mechanical valves) in a cluster, and to reduce the standard error and increase power and precision when there are series of different valve types (mechanical vs biologic) in a cluster.
The difference between valve types was adjusted for by variables that may influence survival. There are 2 reasons for doing this. A variable that is related to survival and that differs between patients given different valve types may either produce a spurious difference between them or mask one that is otherwise present. For example, patients given mechanical valves tend to be younger with a smaller annual risk of death than patients given bioprosthetic valves. The second reason for adjustment is that we may have variables that predict survival, even though they do not differ between patients given the 2 valve types. Adjusting for such variables will not change the estimate of the difference between valve types, but will reduce the random variation and improve the precision of the estimate. We cannot adjust for all the variables recorded because we have data only at the level of the case series, that is, 38, and a good guideline is that we should have at least 10 observations per variable; a regression equation with more than 4 predicting variables may be unreliable. Because each series has only 1 type of valve (mechanical or bioprosthetic), the analysis should be efficient for estimating the difference between valves. It will not be efficient for estimating the effect of other factors. A variable that may greatly influence survival at the individual level may not produce differences between the case series because too few patients had the adverse characteristic or the series did not differ much in the proportion of patients with the variable.
We had data on NYHA classes in all articles but in only 5 articles as mean and standard deviation (SD). We could extract the proportion of patients in NYHA classes III or IV in all other series. We calculated the mean and SD from 22 series in which all 4 classes were given, then used proportion in classes III and IV as the dichotomous outcome variable and mean NYHA class and SD as the predictors in a logistic regression. This gave the log odds of being in class III or IV, which we converted to a proportion. The observed and predicted proportions in NYHA classes III or IV were closely and linearly related (r = 0.98). We then used the same equation to calculate the proportion of patients in NYHA classes III or IV in the 5 series in which we only had mean NYHA class and SD.
|
|
| Footnotes |
|---|
Appendices E1 and E2, Table E1, and References E1 to E32
are available online.
| References |
|---|
|
|
|---|
| Online References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. Stassano, L. Di Tommaso, M. Monaco, F. Iorio, P. Pepino, N. Spampinato, and C. Vosa Aortic valve replacement: a prospective randomized evaluation of mechanical versus biological valves in patients ages 55 to 70 years. J. Am. Coll. Cardiol., November 10, 2009; 54(20): 1862 - 1868. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. H. Stone Current selection of optimal prosthetic aortic valve replacement in middle-aged patients: still dealer's choice. J. Am. Coll. Cardiol., November 10, 2009; 54(20): 1869 - 1871. [Full Text] [PDF] |
||||
![]() |
P. S.U. Myken and O. Bech-Hansen A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis. J. Thorac. Cardiovasc. Surg., January 1, 2009; 137(1): 76 - 81. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Vahanian, B. Iung, L. Piérard, R. Dion, and J. Pepper CHAPTER 21 Valvular Heart Disease ESC Textbook of Cardiovascular Medicine, January 1, 2009; 2(1): med-9780199566990-chapter - med-9780199566990-chapter. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Dal-Bianco, B. K. Khandheria, F. Mookadam, F. Gentile, and P. P. Sengupta Management of Asymptomatic Severe Aortic Stenosis J. Am. Coll. Cardiol., October 14, 2008; 52(16): 1279 - 1292. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Borer and L. N. Girardi Repair of the congenitally bicuspid regurgitant aortic valve a strategic advance. J. Am. Coll. Cardiol., July 1, 2008; 52(1): 50 - 51. [Full Text] [PDF] |
||||
![]() |
T. Sakamoto, G. Matsumura, Y. Kosaka, Y. Iwata, N. Yamamoto, S. Saito, K. Ishihara, and H. Kurosawa Long-term results of Konno procedure for complex left ventricular outflow tract obstruction. Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 37 - 41. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Eitz, S. Schenk, D. Fritzsche, A. Bairaktaris, O. Wagner, H. Koertke, and R. Koerfer International Normalized Ratio Self-Management Lowers the Risk of Thromboembolic Events After Prosthetic Heart Valve Replacement Ann. Thorac. Surg., March 1, 2008; 85(3): 949 - 955. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. El Oakley, P. Kleine, and D. S. Bach Choice of Prosthetic Heart Valve in Today's Practice Circulation, January 15, 2008; 117(2): 253 - 256. [Full Text] [PDF] |
||||
![]() |
R. H. Jones The Year in Cardiovascular Surgery J. Am. Coll. Cardiol., May 8, 2007; 49(18): 1887 - 1898. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, A. Vahanian, H. Baumgartner, J. Bax, E. Butchart, R. Dion, G. Filippatos, F. Flachskampf, R. Hall, B. Iung, et al. Guidelines on the management of valvular heart disease: The Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology Eur. Heart J., January 26, 2007; (2007) ehl428v1. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||