Tomaso Bottio
Giulio Rizzoli
Gianluigi Bisleri
Gino Gerosa
Gaetano Thiene
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bottio, T.
Right arrow Articles by Thiene, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bottio, T.
Right arrow Articles by Thiene, G.
Related Collections
Right arrow Valve disease

J Thorac Cardiovasc Surg 2006;132:1017-1022
© 2006 The American Association for Thoracic Surgery


Surgery for Acquired Cardiovascular Disease

Commissural dehiscence: A rare and peculiar cause of porcine valve structural deterioration

Tomaso Bottio, MD, PhDa,*, Marialuisa Valente, MDb, Giulio Rizzoli, MDa, Vincenzo Tarzia, MDa, Gianluigi Bisleri, MDa, Elena Pettenazzo, BSc, PhDb, Gino Gerosa, MDa, Gaetano Thiene, MD, FRCPb

a Department of Cardiovascular Surgery, University of Padua Medical School, Padua, Italy
b Department of Pathology, University of Padua Medical School, Padua, Italy.

Received for publication February 2, 2006; revisions received May 23, 2006; accepted for publication June 7, 2006.

* Address for reprints: Tomaso Bottio, MD, PhD, Department of Cardiovascular Surgery, Piazza Spedali Civili 1, 25100, Brescia, Italy. (Email: bottio{at}med.unibs.it).


    Abstract
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
OBJECTIVE: Calcification is the main cause of structural valve deterioration; however, other causes of failure have been identified, and among them, dehiscence of a commissure from the stent has been reported in several models of porcine valves. The aim of this study was to analyze the rate and mode of occurrence of this complication in first- and second-generation porcine bioprosthetic explants.

METHODS: Among 586 porcine xenografts explanted and analyzed at the Institute of Pathological Anatomy of the University of Padua, 17 (2.9%) have been replaced for incompetence because of commissural dehiscence. All these explants were in the mitral position, with the exception of a Carpentier-Edwards supra-annular aortic valve prosthesis.

RESULTS: Dehiscence was observed in 9 (1.9%) of 455 Hancock standard explants, in 1 (3.2%) of 31 Hancock II, in 3 (8.6%, 2 standard and 1 supra-annular) of 35 Carpentier-Edwards, in 1 (2.4%) of 42 Bioimplants, and in 3 (50%) of 6 Xenotech after a mean time function of 157 ± 50, 156, 96 ± 29, 143, and 130 ± 8 months, respectively. Dehiscence was the sole cause of incompetence in 6 cases. An impending commissural dehiscence caused by blood creeping was observed in one case. This might be an explanation for the dehiscence other than excessive trimming of the aortic wall.

CONCLUSIONS: Commissural dehiscence is an uncommon and peculiar mode of failure of porcine valves implanted in the mitral position and was observed earlier and more frequently with Carpentier-Edwards porcine explants (P < .05). We speculate that pericardial strip protection of the suture between the Dacron fabric and porcine aortic wall, as used in the Biocor porcine valve, might prevent this complication.



Abbreviations and Acronyms SVD = structural valve deterioration



    Introduction
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Tissue valve xenografts have been implanted for more than 40 years, and different generations of porcine bioprostheses have been available. The first generation failed mostly because of calcification and suboptimal valve design. Then several improvements, such as supra-annular configuration, stent flexibility, low-pressure glutaraldehyde fixation, strut height reduction, new methods of tissue preservation, and calcium retardant agents, have been introduced,1-3Go thus enhancing their function and retarding structural valve deterioration (SVD). However, manufacturing factors have seldom been implicated as a direct cause of SVD. The dehiscence of a commissure, namely the commissural detachment of the porcine aortic wall from the stent post, leading to a cusp prolapse and regurgitation, has been reported in first-generation (Carpentier-Edwards standard and Hancock standard)4,5Go and second-generation (Carpentier-Edwards supra-annular, Medtronic Intact, and Hancock II) porcine valves.1,6,7Go

By using an implant retrieval analysis of all types of bioprostheses implanted at the University of Padua, we investigated the occurrence of commissural dehiscence as a cause of valve incompetence in various valve explants.


    Materials and Methods
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Between January 1970 and January 2004, 2212 patients underwent heart valve replacement with valve xenografts. The majority of the tissue valves used were porcine bioprostheses (1676 porcine vs 536 pericardial). Concerning the porcine valves in the aortic position, the distribution was as follows: 39% Hancock valves (1970-2000), 47% Biocor valves (1990-2005), and 4% Carpentier-Edwards standard and supra-annular valves (1980-1990), and the remaining were other models, such as Labcor, SJM X-cell, Bioimplant, and Xenotech valves. On the contrary, in the mitral position the distribution was as follows: 68% Hancock valves (1970-2000), 18% Biocor (1990-2005), and 8% Carpentier-Edwards standard and supra-annular (1980-1990), and the remaining were other models, such as Bioimplant, Xenotech, Labcor, and Angell-Shiley valves. The mean age of the patients at the time of the operation was 65 ± 10 years, and 54% were female. Five hundred eighty-six porcine bioprostheses have been explanted at reoperation and retained for pathologic study. The protocol of the investigation consisted of gross examination, photography of the specimens with a stereoscope, and radiography for a semiquantitative analysis of Ca++ deposits using a 0- to 4-point scale, as previously reported.8Go All the valves were examined in each case at the time of explantation by the same pathologist (GT) with long-term experience in this field, who was skilled enough to rule out surgical manipulations. Each valve was sent to the pathologist with an attached surgical report. There were 538 first-generation porcine bioprostheses (455 Hancock standard, 35 Carpentier-Edwards [26 standard and 9 supra-annular], 42 Bioimplant, and 6 Xenotech) and 48 second-generation porcine bioprostheses (31 Hancock II and 17 Biocor). Details are summarized in Table 1. Four hundred thirty-six were mitral, 136 were aortic, and 14 were tricuspid bioprostheses. Four hundred eighty prostheses were in place for less than 15 years, and the remaining 106 were in place for more than 15 years.


View this table:
[in this window]
[in a new window]
 
TABLE 1. Explanted bioprostheses
 
In the entire series of explants, 17 (2.9%) bioprostheses have been replaced for incompetence because of a peculiar type of SVD, namely commissural dehiscence.

The incidence of commissural dehiscence in the aortic and mitral position and among different valve models was compared by using the Fisher exact test.


    Results
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Sixteen bioprostheses with a dehiscent commissure were explanted from the mitral position (28-33 mm range of label size), and only 1 (Carpentier-Edwards supra-annular) was explanted from the aortic position (23 mm; P < .005). Twelve were implanted in female subjects (mean age, 65.5 ± 10 years), and 5 were implanted in male subjects (mean age, 63.8 ± 6 years).

Macroscopic Examination
The commissural dehiscence rate is reported in Table 1, and it was significantly more frequent in Carpentier-Edwards and Xenotech prostheses (P < .05). Commissural dehiscence, with (4 cases) or without (2 cases) calcification, was the sole cause of dysfunction in 6 cases after a mean period of function of 187 ± 62 months (Figures 1 and 2). Go In the remaining 11 explants, tears and wears were concomitantly present (mean time of function, 159 ± 35 months), with (8 cases) or without (3 cases) calcification. Dehiscence occurred at the anterior commissure (ie, between the right and left cusps) in 13 explants (Table 2).


Figure 1
View larger version (42K):
[in this window]
[in a new window]
 
Figure 1. Dehiscence at the anterior commissure in a mitral Carpentier-Edwards standard prosthesis in place for 102 months in the absence of calcification.

 

Figure 2
View larger version (66K):
[in this window]
[in a new window]
 
Figure 2. Dehiscence at the anterior commissure in a mitral Hancock standard prosthesis in place for 225 months. Mild calcific deposits were seen at radiography.

 

View this table:
[in this window]
[in a new window]
 
TABLE 2. Porcine bioprostheses with commissural dehiscence
 
In 2 prostheses (1 Hancock standard and 1 Xenotech) dehiscence was concomitantly present in 2 different commissural posts; in the Xenotech valve the dehiscence of the anterior commissure was associated with a hematoma of the left posterior commissure, with evidence of incipient detachment (Figure 3).


Figure 3
View larger version (72K):
[in this window]
[in a new window]
 
Figure 3. Double commissural dehiscence in a Xenotech mitral valve: one overt and one impending with a hematoma (arrow).

 
Time of Function of Different Prosthesis Models
Commissural detachment of the aortic wall occurred after a mean time of 157 ± 50 months in Hancock standard explants, 156 months in Hancock II explants, 96 ± 29 months in Carpentier-Edwards (standard and supra-annular) explants, 143 months in the Bioimplant, and 130 ± 8 months in the Xenotech. Therefore commissural dehiscence occurred earlier in the Carpentier-Edwards prostheses than in others (P < .05). Of the 3 Carpentier-Edwards prostheses, 2 were standard models implanted in the mitral position in 1980 and 1984, and one was a supra-annular model implanted in the aortic position in 1986.

Radiographic Findings
Two prostheses showed no signs of calcification (score 0 at radiography) after 102 months (Carpentier-Edwards standard, Figure 1) and 160 months (Hancock standard), respectively. Eleven prostheses showed initial signs of commissural mineralization (score, 1-2) at a mean time of 163 ± 50 months (range, 106-260 months). Four tissue valves showed heavy mineralization at a mean time of 151 ± 71 months (score, 3-4; Figure 4).


Figure 4
View larger version (55K):
[in this window]
[in a new window]
 
Figure 4. Commissural dehiscence and heavy mineralization with tears in a mitral Carpentier-Edwards standard prosthesis implanted in 1984 and in place for 64 months. Score was 4 at radiography.

 

    Discussion
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Among 586 porcine bioprostheses explanted at reoperation at our institute, 496 (85%) have been replaced as a result of SVD, mostly because of dystrophic calcification.9Go Seventeen porcine valve xenografts failed because of a detachment of the porcine aortic wall from the stent at the top of the commissural post, so-called commissural dehiscence.4Go

SVD has been extensively documented as the major complication of porcine bioprostheses.7,8Go However, it has seldom been attributed to a specific manufacturing cause. Nistal and coworkers4Go first ascribed the cause of commissural dehiscence of Carpentier-Edwards porcine valve models to a manufacturing flaw. According to their results, the commissural dehiscence occurred frequently in those models because of an excessive trimming of the outer layers of the porcine aortic wall accomplished to favor the hemodynamic performance. This manufacturing improvement was introduced in 1986 in the large mitral prostheses, and in 1987, this practice was extended to all sizes of mitral and aortic prostheses.10Go Moreover, the major stress sustained in the mitral position was an additional risk factor explaining the higher incidence of commissural dehiscence in mitral bioprostheses.4,11Go Also, in the experience of Jamieson and colleagues,1Go commissural dehiscence occurred mostly in mitral prostheses, namely 29 (14%) of 205 explants and only 1 case in the aortic position (1/81 [1.2%] explants). The average time of function of the affected mitral prostheses was 102.9 months (range, 44.7-162.7 months), whereas commissural dehiscence occurred as late as 172 months in the single affected aortic bioprosthesis. In their series 28 of 29 mitral prostheses were second-generation Carpentier-Edwards porcine valves implanted between 1982 and 1986. Concerning position and timing, our results are very similar to those reported by Jamieson and colleagues.1Go These authors, as well as Kimura and associates,12Go concluded that the modifications introduced in the second-generation Carpentier-Edwards supra-annular model did not improve the durability over the first-generation standard valve. The Vancouver group recently reported follow-up to 18 years of the Carpentier-Edwards supra-annular porcine valve in the mitral position and found that commissural dehiscence accounted for 19% of incompetence (39 of 204 explants with incompetence).13Go

However, this complication was not unique to the Carpentier-Edwards porcine valve because it has been observed also in other second-generation porcine bioprostheses, such as the Hancock II7Go and the Medtronic Intact.6,14Go In these valves aortic wall trimming was not performed, casting doubt on the offered explanation. The Hancock II porcine bioprosthesis was introduced in 1982, with some modifications aimed to improve durability over the Hancock standard bioprosthesis, and such improvements have been confirmed in clinical and pathologic studies.2,3,7,15,16Go Among 31 Hancock II bioprostheses explanted (our series), a commissural dehiscence not related to calcification or infection was the cause of incompetence in a mitral prosthesis at 156 months after implantation.

Our data confirm the higher incidence of cusp dehiscence in the mitral prostheses, and in agreement with the echocardiographic study of Naqvi and coworkers,17Go the commissure mostly affected was the anterior commissure, the one that probably supports the higher stress. Only twice have 2 commissures appeared detached, which means that a single dehiscent commissure with double-cusp prolapse might cause a significant valve incompetence requiring valve replacement. As previously described,4Go at the time of gross examination, the attachment of porcine cusps to the aortic wall at the commissure appeared intact, whereas it was the aortic wall that was completely detached from the stent post. At radiography, mineral deposits in the valve specimens were minimal or absent in 6 patients, thus proving that this commissural complication can occur solely and lead to valve dysfunction per se.

We share the current opinion that an excessive trimming of the aortic wall leads to weakening of the commissure and to an increased risk of dehiscence of the aortic wall from the stent post caused by mechanical stress. Nevertheless, commissural dehiscence, at a lower rate, occurred also in the Hancock II and Medtronic Intact bioprostheses, which do not undergo aortic wall trimming. Therefore other reasons might be applicable. Most of the porcine bioprostheses have a similar commissural suspension based on the suturing of the aortic wall to the fabric-covered stent. An infiltration of blood behind the commissure might create a hematoma, which, by means of progressive swelling, might detach the commissure, as in the case of the Xenotech prosthesis (Figure 3). Cuspal hematoma has been reported as a rare complication causing valve stenosis as a result of blood infiltration at the basal cusp suturing, accounting for hematoma dissection.18Go

The Biocor prosthesis has a pericardial shield tailored to cover the superior profile of the junction between stent–fabric–aortic wall and sutured between the inner and the outer side of the stent,19Go thus preventing this complication (Figure 5). Number and time of function of Biocor explants in our series of both aortic19Go and mitral valve replacement,20Go in which no commissural dehiscence occurred, was too short to confirm this hypothesis, and therefore pathologic studies on long-term explants are mandatory. However, it is fundamental to add to these considerations that (1) in both our studies19,20Go freedoms from SVD were calculated, including also the echocardiographic results; (2) only 3 SVD cases were observed for the aortic valves and none for the mitral valves at 8-year follow-up; and (3) in no patient was a commissural dehiscence observed at 2-dimensional echocardiography.


Figure 5
View larger version (71K):
[in this window]
[in a new window]
 
Figure 5. Drawing depicting a Biocor valve. Note that a pericardial sheet covers the superior profile of the stent to which the aortic wall and the fabric are sutured.

 

    Conclusions
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Commissural dehiscence from the stent post in the porcine valves is a peculiar and uncommon cause of dysfunction occurring in 2.9% of explants. It was observed mainly in the mitral position, and it was responsible for valve incompetence severe enough to require a reoperation. In our experience, among various models, commissural dehiscence was observed earlier and more frequently in Carpentier-Edwards porcine valve explants. Manufacturing solutions, such as those introduced in the second-generation Biocor porcine prostheses, might prevent or reduce this complication.


    Acknowledgments
 
We thank Mr Nicola Paccagnella for his assistance in preparing the figures and Mrs Chiara Carturan for editing the manuscript.


    Footnotes
 
Supported by grants from Istituto Superiore di Sanità and Ministero Istruzione Università e Ricerca, Rome, Italy, and by Regione Veneto, Venice, Italy.


    References
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 

  1. Jamieson WR, Burr LH, Janusz MT, Munro AI, Hayden RI, Miyagishima RT, et al. Carpentier-Edwards standard and supraannular porcine bioprostheses: comparison of technology. Ann Thorac Surg 1999;67:10-17.[Abstract/Free Full Text]
  2. David TE, Ivanov J, Armstrong S, Feindel CM, Cohen G. Late results of heart valve replacement with the Hancock II bioprosthesis. J Thorac Cardiovasc Surg 2001;121:268-277.
  3. Rizzoli G, Bottio T, Thiene G, Toscano G, Casarotto D. Fifteen years durability of Hancock II porcine bioprostheses. J Thorac Cardiovasc Surg 2003;126:66-74.[Abstract/Free Full Text]
  4. Nistal JF, Hurle A, Gutierrez JA, Mazorra F, Revuelta JM. Commissural dehiscence of Carpentier-Edwards mitral bioprostheses. Explant analysis and pathogenesis. J Thorac Cardiovasc Surg 1995;110:688-696.[Abstract/Free Full Text]
  5. Gallucci V, Bortolotti U, Milano A, Valfre C, Mazzucco A, Thiene G. Isolated mitral valve replacement with the Hancock bioprosthesis: a 13-year appraisal. Ann Thorac Surg 1984;38:571-578.[Abstract]
  6. Jamieson WR, Lemieux, MD, Sullivan JA, Munro AI, Metras J, Cartier PC. Medtronic intact porcine bioprosthesis: 10 years' experience. Ann Thorac Surg 1998;66(suppl):S118-S121.[Medline]
  7. Bottio T, Thiene G, Pettenazzo E, Ius P, Bortolotti U, Rizzoli G, et al. Hancock II bioprosthesis: a glance at the microscope in mid-long term explants. J Thorac Cardiovasc Surg 2003;126:99-105.[Abstract/Free Full Text]
  8. Milano A, Bortolotti U, Talenti E, Valfrè C, Arbustini E, Valente M, et al. Calcific degeneration as the main cause of porcine bioprosthetic valve failure. Am J Cardiol 1984;53:1066-1070.[Medline]
  9. Bottio T, Rizzoli G, Valente M, Casarotto D, Thiene G. Cause of structural valve deterioration in porcine bioprostheses implanted for more than 15 years 2001In: Proceedings of the 1st Biennial Meeting of the Society for Heart Valve Disease; London, United Kingdom. p. 104.
  10. Jamieson EWR, Marchand MA, Pelletier CL, Norton R, Pellerin M, Dubiel TW, et al. Structural valve deterioration in mitral replacement surgery: comparison of Carpentier-Edwards supra-annular porcine and Perimount pericardial bioprostheses. J Thorac Cardiovasc Surg 1999;118:297-304.[Abstract/Free Full Text]
  11. Allard MF, Thompson CR, Baldelli RJ, McNab JS, Babul SA, Betts JM, et al. Commissural region dehiscence from the stent post of Carpentier-Edwards bioprosthetic cardiac valves. Cardiovasc Pathol 1995;4:155-162.
  12. Kimura M, Iwakuma A, Morishige N, Nakamura K, Tachikawa Y, Shibano R, et al. Commissural dehiscence and pannus formation of porcine heart valve bioprostheses. Artif Organs 2003;27:706-713.[Medline]
  13. Gudas VM, Jamieson WRE, Burr H, Janusz MT, Fradet GJ, Lichtenstein SV, et al. Carpentier-Edwards Supra-annular mitral porcine bioprosthesis: reassessment to 18 years 2005In: Proceedings of the 3rd Biennial Meeting of The Society for Heart Valve Disease; Vancouver, British Columbia, Canada; June 17-20, p. 97.
  14. Barratt-Boyes BG, Jaffe WM, Whitlock RM. The Medtronic Intact porcine valve: ten-year clinical review. J Thorac Cardiovasc Surg 1998;116:1005-1014.[Abstract/Free Full Text]
  15. Ruel M, Kulikn A, Rubens FD, Bédard P, Masters RG, Pipe AL, et al. Late incidence and determinants of reoperation in patients with prosthetic heart valves. Eur J Cardiothorac Surg 2004;25:364-370.[Abstract/Free Full Text]
  16. Rizzoli G, Mirone S, Ius P, Polesel E, Salvador L, Zussa C, et al. Fifteen year results of the Hancock II valve: a multicenter experience. J Thorac Cardiovasc Surg 2006;132:602-609.[Abstract/Free Full Text]
  17. Naqvi TZ, Siegel RJ, Buchinder NA, Miroshnik S, Saedi G, Trento A, et al. Echocardiographic and pathologic features of explanted Hancock and Carpentier-Edwards bioprosthetic valves in the mitral position. Am J Cardiol 1999;84:1422-1427.[Medline]
  18. Thiene G, Bortolotti U, Talenti E, Guerra F, Valente M, Milano A, et al. Dissecting cuspal hematomas: a rare form of porcine bioprosthetic valve dysfunction. Arch Pathol Lab Med 1987;111:964-967.[Medline]
  19. Bottio T, Rizzoli G, Thiene G, Nesseris G, Casarotto D, Gerosa G. Hemodynamic and clinical outcomes with the Biocor valve in the aortic position: an 8-year experience. J Thorac Cardiovasc Surg 2004;127:1616-1623.[Abstract/Free Full Text]
  20. Rizzoli G, Bottio T, Vida V, Nesseris G, Caprili L, Thiene G, et al. Intermediate results of isolated mitral valve replacement with a Biocor porcine valve. J Thorac Cardiovasc Surg 2005;129:322-329.[Abstract/Free Full Text]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Tomaso Bottio
Giulio Rizzoli
Gianluigi Bisleri
Gino Gerosa
Gaetano Thiene
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bottio, T.
Right arrow Articles by Thiene, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bottio, T.
Right arrow Articles by Thiene, G.
Related Collections
Right arrow Valve disease


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS