J Thorac Cardiovasc Surg 2006;132:1150-1155
© 2006 The American Association for Thoracic Surgery
Evolving Technology |
a Cardiovascular Surgery Service, Texas Heart Institute at St Luke's Episcopal Hospital, Houston, Tex
b Department of Cardiovascular Pathology, Texas Heart Institute at St Luke's Episcopal Hospital, Houston, Tex
c Cullen Cardiovascular Surgical Research Laboratories, Texas Heart Institute at St Luke's Episcopal Hospital, Houston, Tex
d Divisions of Cardiothoracic Surgery, Michael E. DeBakey Department of Surgery, Baylor College of Medicine, Houston, Tex
e Congenital Heart Surgery, Michael E. DeBakey Department of Surgery, Baylor College of Medicine, Houston, Tex
f Somanetics Corporation, Troy, Mich.
Received for publication March 8, 2006; revisions received April 27, 2006; accepted for publication May 8, 2006. * Address for reprints: Scott A. LeMaire, MD, One Baylor Plaza, BCM 390, Houston, TX 77030 (Email: slemaire{at}bcm.tmc.edu).
| Abstract |
|---|
|
|
|---|
METHODS: Transcutaneous near-infrared spectroscopic sensors were placed in the midline over the upper and lower thoracic vertebrae of 4 progressively larger pigs (weight range 21-70 kg). After the entire aorta was exposed, segmental arteries from T6 through L1 were sequentially ligated while regional oxygen saturation was monitored. Decreases in regional oxygen saturation were calculated as percentage changes from baseline. The degrees of ischemia in the upper and lower spinal cord were compared histopathologically.
RESULTS: Baseline regional oxygen saturations were similar in the upper (68.8% ± 9.0%) and lower (68.0% ± 11.5%, P = .82) cord. After ligation, however, regional oxygen saturation levels were significantly lower in the lower cord (41.3% ± 10.1%) than in the upper cord (64.8% ± 9.3%, P = .037). The regional oxygen saturation had decreased by 39.0% ± 11.5% in the lower cord but only by 6.3% ± 7.6% in the upper cord (P = .026). This difference was confirmed microscopically: upper-cord sections had fewer ischemic neurons (8.8 ± 9.4) than did lower-cord sections (21.3 ± 13.6, P = .002).
CONCLUSION: Intraoperative spinal cord ischemia was detectable with near-infrared spectroscopy in pigs weighing as much as 70 kg. The potential utility of this technique in patients undergoing thoracic aortic surgery warrants investigation.
| Introduction |
|---|
|
|
|---|
Current spinal cord monitoring techniques rely on somatosensory-evoked potentials (SSEPs) or motor-evoked potentials (MEPs). Monitoring SSEPs has several disadvantages, including slow response time (caused by delays between the onset of ischemia and the disappearance of potentials) and poor overall sensitivity and specificity.2-4
Although MEP monitoring has been successfully used to detect spinal cord ischemia, guide surgical strategy, and prevent postoperative neurologic deficits, it has limitations that have prevented it from being widely adopted.1,5-7
Transcutaneous near-infrared spectroscopy (NIRS), which exploits the unique near-infrared absorption profiles of hemoglobin, oxyhemoglobin, and cytochrome aa3, is currently widely used for cerebral oximetry during cardiovascular surgery.8-12
This technique assesses the oxyhemoglobin fraction within a focal area of underlying tissue by measuring the differential absorption of two wavelengths of near-infrared light (730 and 810 nm) that reflect deoxyhemoglobin and total hemoglobin concentration. The purpose of this pilot study was to assess the feasibility of using NIRS to detect spinal cord ischemia during intercostal artery ligation in the pig.
| Materials and Methods |
|---|
|
|
|---|
Anesthetic Management
Four domestic swine (weighing 21, 37, 48, and 70 kg) were premedicated with intramuscular atropine sulfate (0.5 mg/kg), acepromazine maleate (0.1 mg/kg), and ketamine hydrochloride (20 mg/kg). Isoflurane (0.5%-3.0%) was given by mask for induction. Crystalloid fluid was infused throughout the procedure, and boluses of hetastarch were given when needed. Monitors included a pulse oximeter placed on the ear, electrocardiographic leads, and a rectal temperature probe. The animals were orally intubated with a cuffed endotracheal tube through direct laryngoscopy and connected to a volume ventilator that delivered 100% oxygen at a tidal volume of 10 mL/kg. General anesthesia was maintained with inhaled isoflurane (0.5%-3.0%) and pancuronium bromide (0.1 mg/kg). A warming blanket was placed underneath the pigs to maintain normothermia. A carotid artery catheter was used for blood pressure monitoring and arterial blood gas sampling.
Operative Procedure
The dorsal area was shaved and cleaned, and 5100SAF SomaSensors (Somanetics Corporation, Troy, Mich) were placed in the midline over the upper (T6-T7) and lower (T9-T11) thoracic vertebrae. These sensors were connected to an INVOS 5100 Cerebral Oximeter (Somanetics Corporation). A pediatric spinal drainage catheter was inserted into the subarachnoid space through either a laminectomy or direct puncture between the third and fourth lumbar vertebrae.
A left thoracoabdominal incision was made through the sixth intercostal space. The diaphragm was divided, and the entire thoracoabdominal aorta was exposed. Regional spinal oxygen saturation (SrO 2) was monitored continuously by both upper and lower sensors. Raw optical data from the sensors were stored in a computer at 4-second intervals. After baseline SrO 2 levels were recorded, segmental intercostal and lumbar arteries from T6 through L1 were sequentially occluded at approximately 10-minute intervals. Each artery was initially occluded with a bulldog clamp; after approximately 10 minutes of clamping, SrO 2 was recorded, and the artery was ligated with metallic clips and divided.
After all segmental arteries were ligated, 1 mL indocyanine green dye (2.5 mg/mL; Akorn, Inc, Buffalo Grove, Ill) was injected into the subarachnoid space through the spinal catheter. This dye absorbs near-infrared light in a band centered at 805 nm. The catheter was flushed twice with 1 mL saline solution to distribute the dye evenly within the space surrounding the spinal cord. Optical density (the log of the ratio of measured intensity to incident intensity) at 810 nm was recorded with the oximeter to determine changes in light absorption. After ligation of the L1 segmental arteries in each of the 3 largest pigs, the animals were briefly awakened and examined for hind limb paralysis. After this examination, the animals were reanesthetized and humanely killed with intravenous potassium chloride.
Histopathology
The entire spinal column was removed from each of the 3 largest pigs and placed in formalin after the upper and lower segments monitored by the sensors were marked. Identification of the arteria radicularis magna (Adamkiewicz artery) was not attempted. The cords were sectioned, and representative portions of both regions were stained with either hematoxylin and eosin or luxol fast blue dye. A pathologist (F.J.C.) who was blind to the origin of each section (upper or lower cord) examined the sections and quantified ischemic change by counting the number of normal neurons and the number of ischemic neurons per section.
Statistical Analysis
The statistical analyses were performed with SPSS version 12.0 for Windows (SPSS Inc, Chicago, Ill). The following intraoperative variables were compared: mean upper- and lower-cord SrO
2 values at baseline and after ligation of the segmental spinal arteries T6 through L1; absolute percentage SrO
2 decline from baseline after each vessel was ligated; and mean heart rate, temperature, and mean arterial pressure. The upper and lower cords were compared in terms of the number of ischemic neurons and the ratio of ischemic to normal neurons. Continuous variables are reported as mean ± SD and were analyzed with the Student t test for between-group differences. We used analysis of variance for comparisons among three or more groups. The Bonferroni correction was used for multiple comparisons.
| Results |
|---|
|
|
|---|
|
|
Histopathology
The regional difference in the degree of ischemia was confirmed by microscopic examination of the lateral corticospinal tracts. Lower-cord sections exhibited more pronounced ischemic changes than did upper-cord sections, including increased vacuolization, retraction of neurons, and loss of nucleoli (Figure 3). Lower-cord sections also had more ischemic neurons (mean 21.3 ± 13.6 per section) than did upper-cord sections (mean 8.8 ± 9.4 per section, P = .002). These numbers corresponded to a significantly lower ratio of ischemic to normal neurons in the upper cord (0.17) than in the lower cord (0.50, P = .005), indicating that the regional differences in spinal ischemia indicated by the NIRS monitor paralleled the regional histopathologic differences.
|
| Discussion |
|---|
|
|
|---|
Attempts to monitor spinal cord perfusion indirectly during aortic aneurysm surgery began with the monitoring of SSEPs, measurable electrical patterns that occur within the spinal cord when a peripheral sensory nerve (usually the posterior tibial nerve) is stimulated.15
Although SSEP monitoring showed promise in canine studies,16
a clinical trial by Crawford and associates2
showed that the measurement of SSEPs during aortic surgery did not reduce the incidence of postoperative neurologic deficits.17
The poor sensitivity and specificity of SSEP monitoring is partially attributable to the fact that the sensory and motor pathways of the spinal cord are anatomically separate and have different blood supplies. The amplitude of SSEPs only indicates the function of sensory tracts, which are located in the dorsal portion of the spinal cord and are not supplied by the anterior spinal artery, where blood flow is reduced during aortic clamping. As a result, SSEP monitoring provides only delayed information about spinal cord ischemia.18
In contrast, MEPs reflect the functional integrity of the motor pathways in the anterior (ventral) cord, especially the more vulnerable motor neurons in the anterior gray matter. Because the anterior cord is supplied by the anterior spinal artery, MEP monitoring is an anatomically and physiologically sound means of detecting spinal ischemia during aortic clamping. Jacobs and colleagues1,5-7
have used MEP monitoring as part of their spinal cord protection strategy and have achieved excellent results. In a pilot study involving 52 consecutive patients with Crawford extent I or II TAAAs, Jacobs and colleagues1
recorded MEPs in all patients and reported that spinal cord ischemia was detected as early as 2 minutes after intercostal artery ligation. During distal aortic perfusion, 14 patients showed a rapid decrease in MEP amplitude to less than 25% of baseline, indicating spinal cord ischemia; this was corrected by increasing distal aortic pressure. Additionally, in 33 patients, aggressive reattachment of intercostal arteries returned MEP amplitudes to baseline levels, and no early or late paraplegia occurred. Jacobs and Mess5
recently reported preventing neurologic deficits in 98% of patients undergoing TAAA repair by using MEP monitoring.
MEP monitoring is not, however, without limitations.19,20
For example, the neuromuscular blocking agents that are part of the usual anesthetic regimen interfere with MEP monitoring. For this reason, MEP monitoring requires alterations in standard anesthetic management to prevent complete neuromuscular blockage. Also, MEP monitoring depends on signal averaging across a span of seconds, which limits its ability to detect vascular compromise before neurons have been irreversibly injured. Additional limitations of this technique include the need for monitoring by a neurophysiologist during the operation and its inability to be used in conscious postoperative patients. For these reasons, although MEP monitoring has been shown to reduce the risk of paraplegia, the technique is not widely used.
Other methods of monitoring spinal cord ischemia currently being studied include intrathecal monitoring of cerebrospinal fluid oxygen tension21,22
and transesophageal NIRS monitoring of spinal cord Sro2.23
An ideal spinal cord monitoring method would (1) be noninvasive, (2) be simple enough to use that no additional personnel are required, (3) be highly sensitive and specific to changes in the anterior spinal cord, (4) allow standard anesthesia delivery, (5) exhibit no delay in ischemia detection, and (6) be usable in conscious postoperative patients. Transcutaneous NIRS has the potential to satisfy these criteria.
Our pilot study was designed to assess the potential usefulness of transcutaneous NIRS as a noninvasive monitor that provides continuous, real-time information about spinal cord oxygenation. NIRS uses transcutaneous sensors to measure the focal oxyhemoglobin fraction in underlying tissue. Currently, this technique is used to monitor cerebral oxygenation during a wide variety of cardiovascular operations.4,9,10,12,24
In addition to being entirely noninvasive, the NIRS monitor is simple to use, requires no additional technical personnel, and does not require modifications to anesthetic management. In our experiment, we showed that transcutaneous NIRS can detect intraoperative spinal cord ischemia in a porcine model.
An initial concern was that the sensors would detect oxygenation levels in the surrounding tissues, such as muscle and subcutaneous tissue, but not in the underlying spinal cord. Given the anatomy of the blood supply shared between the spinal cord and these surrounding tissues, however, NIRS data on regional oxygenation may adequately reflect spinal cord perfusion, albeit indirectly. To address this issue specifically, we injected indocyanine green dye into the subarachnoid space and found an immediate change in light absorption (Figure 2), suggesting that at least some of the photons were reaching the spinal cord. Additionally, Macnab and coworkers25
found that, in a porcine model of spinal cord hypoxia-ischemia, NIRS detected ischemic changes in the spinal cord whether the sensors were placed directly on the spinal cord, the spinal lamina, or the spinous processes. Further experiments correlating direct measurements of spinal oxygenation with transcutaneous NIRS readings would help clarify this issue.
Real-time information about spinal cord perfusion during TAAA repair would allow the surgical team to intervene when ischemia occurs. Intervention strategies used during SrO 2 monitoring could be based on those outlined in Jacobs and colleagues' report1 regarding MEP monitoring during TAAA surgery. For example, when ischemia is detected, the mean arterial pressure could be immediately increased. Distal aortic flow and pressure could also be increased when left heart bypass is being used. Additionally, temporary occlusion catheters could be placed in segmental arteries with significant back-bleeding, thereby reducing steal from the anterior spinal artery. Finally, decline in SrO 2 could prompt the reattachment of specific segmental arteries within the isolated aortic segment, and aortic endarterectomy could be performed when segmental arteries are not readily apparent.
Because of its potential value in guiding these types of interventions, NIRS monitoring of spinal cord oxygenation during thoracic aortic surgery is worthy of clinical investigation. Expanding the application of NIRS to postoperative spinal cord monitoring as a means of preventing delayed-onset paraplegia also merits study.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |