J Thorac Cardiovasc Surg 2007;133:485-493
© 2007 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
Gaseous and solid cerebral microembolization during proximal aortic anastomoses in off-pump coronary surgery: The effect of an aortic side-biting clamp and two clampless devices
Lorenzo Guerrieri Wolf, MD,
Yasir Abu-Omar, MRCS,
Bikram P. Choudhary, MRCS,
David Pigott, FRCA*,
David P. Taggart, MD (Hons), PhD, FRCS
Department of Cardiothoracic Surgery, John Radcliffe Hospital, Oxford, United Kingdom.
* Address for reprints: Professor David P. Taggart, MD, PhD, FRCS, Professor of Cardiovascular Surgery, University of Oxford, Department of Cardiothoracic Surgery, John Radcliffe Hospital, Headley Way, Headington, Oxford OX3 9DU, United Kingdom (Email: david.taggart{at}orh.nhs.uk).
 |
Abstract
|
|---|
OBJECTIVES: Intraoperative cerebral microembolism is a cause of cerebral dysfunction after cardiac surgery, and particulate microemboli are the most damaging. Using a new-generation transcranial Doppler ultrasound, we compared the number and nature of microemboli in patients undergoing off-pump coronary artery bypass grafting during performance of proximal anastomoses with three techniques: an aortic side-biting clamp and two clampless devices (the Enclose II device [Novare Surgical Systems, Inc, Cupertino, Calif] and the Heartstring II device [Guidant Corporation, Santa Clara, Calif]) developed to obviate the need for an aortic side-biting clamp, thereby reducing the number of cerebral microemboli.
METHODS: Bilateral continuous monitoring of the middle cerebral arteries was performed with a multirange, multifrequency transcranial Doppler device that both automatically rejects artifacts online and discriminates between solid and gaseous microemboli. Recordings were continuously undertaken during performance of 66 proximal aortic anastomoses in 42 patients. Thirty-five anastomoses were performed with an aortic side-biting clamp, 20 with the Enclose device, and 11 the Hearstring device.
RESULTS: Most microemboli occurred during application/insertion and removal of each device from the ascending aorta. The median number (interquartile range) of total microemboli was 11 (6-32) during side clamping, 11 (6-15) with the Enclose device, 40 (31-48) with the Heartstring device (P < .01). The proportion of solid microemboli was significantly higher in the side-clamp group (23%) compared with 6% and 1% in the Enclose and Heartstring groups, respectively (P < .01).
CONCLUSIONS: Avoidance of aortic side clamping results in a significant reduction in the proportion of solid microemboli detected with transcranial Doppler. As solid microemboli are probably the most damaging, use of the Enclose and Heartstring devices may represent an important strategy for minimizing cerebral injury during proximal aortic anastomoses.
Abbreviations and Acronyms CABG = coronary artery bypass grafting; IQR = interquartile range; OPCAB = off-pump coronary artery bypass grafting; TCD = transcranial Doppler
 |
Introduction
|
|---|
Cerebral injury remains the most significant and disabling complication of coronary artery bypass grafting (CABG).1
It occurs in two forms2
: type I injury, which includes stroke, transient ischemic attack, and coma (incidence 3%-6%), and type II injury, which is more subtle and includes impairment of cognitive function. The latter affects up to half of all patients soon after surgery and persists in a quarter at 6 months3
and in up to 42% at 5 years.4
Although macroembolization of atheromatous debris during manipulation of the diseased aorta is the main cause of stroke,5-10
intraoperative cerebral microembolization is considered to be the most important cause of postoperative neurocognitive decline.11-14
Indeed, we15
recently added support to this hypothesis by demonstrating a correlation between the number of intraoperative cerebral microemboli and subsequent impairment of verbal memory defined by functional magnetic resonance imaging.
Using a new generation of advanced transcranial Doppler (TCD) ultrasound (Embo-Dop; DWL Elektronische Systeme, GmbH, Singen, Germany), which uses multifrequency and multirange recordings both to reject artifacts online and to discriminate between solid and gaseous microemboli, we previously confirmed that cardiopulmonary bypass and aortic manipulation (cannulation, decannulation, and application and removal of an aortic crossclamp or side-biting clamp) are the main causes of cerebral microemboli during CABG.16
While off-pump composite arterial grafts permit a true "no touch" aortic technique and may be the most effective way to reduce the macroembolic/microembolic load to the brain,1,17-23
one or more proximal anastomoses to the aorta are often necessary. This has driven the development of novel anastomotic devices, to enable the surgeon to perform proximal anastomoses without an aortic side-biting clamp in a fully pressurized aorta. Although several groups have reported their clinical experience with these methods,24-31
to the best of our knowledge their effects on the relative proportions of gaseous and solid microemboli have not been reported.
In the current study we compared the number and nature of cerebral microemboli in patients undergoing off-pump coronary artery bypass grafting (OPCAB) during construction of proximal aortic anastomoses using three different techniques: a conventional aortic side-biting clamp and two clampless devices, the Enclose II proximal seal system (Novare Surgical Systems, Inc, Cupertino, Calif; Figure 1) and the Heartstring II proximal seal system (Guidant Corporation, Santa Clara, Calif; Figure 2). We were particularly interested in the nature of the microemboli (solid vs gaseous) and their generation in temporal relation to application and removal of the devices.

View larger version (44K):
[in this window]
[in a new window]
|
Figure 1. The Enclose II proximal seal system (Novare). Foreground, The device with uppermost extraluminal jaw with a nonexpandable wire form and intraluminal jaw with expandable membrane below; behind upper and lower knob and vented plug. Background, Actuator tool supplied with the kit used to open and close expandable membrane and control vertical movement of upper jaw. (Adapted from Ethicon Web site: www.novaresurgical.com).
|
|
 |
Patients and Methods
|
|---|
A total of 42 patients undergoing OPCAB were prospectively recruited for the study. All patients gave informed consent, and full ethical approval was obtained from the local research ethics committee (Oxford Research Ethics Committee number C01.258). A total of 66 proximal anastomoses were performed, 35 with an aortic side-biting clamp (group S), 20 with the Enclose device (group E), and 11 with the Heartstring device (group H).
All proximal aortic anastomoses were performed in palpably32
normal aortas. Patients with symptomatic carotid disease, peripheral vascular disease, preoperative neurologic history, or atrial fibrillation were excluded. Although epiaortic scanning would also have been very useful, this is not routinely available in our center.
The operating surgeon made no choice in selection of the anastomotic device. A limited number of 11 Enclose and 11 Heartstring devices were randomly selected by two nonoperating researchers and not by the surgeon. The researchers were not involved in the perioperative evaluation or management of the study patients but were responsible for TCD measurements. Therefore, we believe that selection bias, although not completely eliminated, was effectively decreased in this study.
Surgical Technique
All operations were performed by one surgeon (D.P.T.). After midline sternotomy, one or two internal thoracic arteries were harvested in a skeletonized fashion. The radial artery, the long saphenous vein graft, or both were simultaneously harvested. Full anticoagulation was achieved with heparin. The distal anastomoses were performed first, with the aid of a suction-assisted stabilizer (Octopus; Medtronic, Inc, Minneapolis, Minn; or Guidant, Guidant Corporation) and visualization was enhanced with a surgical blower-mister (Medtronic ClearView; Medtronic, Inc).
After completion of the distal anastomoses, the proximal radial artery or saphenous vein graft anastomoses were performed on a disease-free segment of the aorta as assessed by gentle palpation. An anterior segment of normal aorta with a diameter of at least 1 cm, for use of the Heartstring device, and of at least 2.5 cm for the Enclose device or side-clamp was required.
Side-biting clamp
After removal of aortic adventitia, a palpably normal segment of ascending aorta was grasped with Duvall forceps and a side-biting clamp applied. The area of enclosed aorta was from 1 to 2 cm2. A small (around 5 mm) linear aortotomy was performed, and the conduit was anastomosed with continuous 6-0 polypropylene (Prolene; Ethicon, Inc, Somerville, NJ). The side-biting clamp was removed after completion of the anastomosis.
Enclose device
The device was used as previously described.33
A 14-gauge needle was introduced into the ascending aorta through a 4-0 Prolene purse-string suture. The lower jaw of the Enclose device, containing an expandable membrane (Figure 1), was then inserted and hemostasis was secured by tightening the purse-string suture. The Enclose device was then advanced toward an area of palpably normal aorta to perform the anastomosis. The nonexpandable wire-form upper jaw was lowered toward the aortic wall (with an actuator tool furnished with the kit), the membrane of the lower jaw was opened (still with the actuator tool), and an aortotomy was made in the middle of the aortic wall within the upper jaw, using an aortic punch; hemostasis was secured by the device membrane seal, which was fixed to the inner wall of the aorta (Figure 3). The anastomosis was then constructed manually in a bloodless field with continuous 6-0 Prolene polypropylene. After completion, the device was released and, if necessary, was repositioned and deployed again by rotation, without the need for complete removal, allowing a second or third anastomosis to be constructed.

View larger version (153K):
[in this window]
[in a new window]
|
Figure 3. The Enclose II proximal seal system in use. A 1 cm2 area of aortic wall lies between the tightened upper and lower jaws of the device; the proximal anastomosis will be performed manually in the middle of this bloodless area without using a side-biting clamp. (Adapted from Ethicon Web site: www.novaresurgical.com).
|
|
Heartstring device
In a palpably normal segment of aorta, a proprietary aortic cutter, supplied with the Heartstring device (Figure 2), was used to create a circular aortotomy. The coiled Heartstring device was then delivered through the aortotomy, temporarily occluded by a finger, and deployed to create a hemostatic seal against the inner wall of the aorta. The anastomosis was then performed manually with continuous 6-0 Prolene polypropylene (Figure 4). In contrast to the other two techniques, a blower-mister device was used (as recommended by the manufacturer34
) to enhance visualization during construction of the anastomosis. After completion of the anastomosis, the device was uncoiled and withdrawn before final tightening of the suture line.
TCD Monitoring
Bilateral continuous TCD monitoring of the middle cerebral arteries was performed, as we16
have described in detail previously. This is a noninvasive technique with no known complications. All recordings were performed throughout the procedure using a multirange, multifrequency TCD (Embo-Dop), which rejects artifacts online and automatically discriminates between solid and gaseous microemboli with a relatively high sensitivity and specificity.16,35,36
Russell and Brucher37
recently reported that results are most reliable for embolus differentiation when, as in the current study, the Doppler signal enhancement, that is, embolus-to-blood ratio, is greater than 28 dB/ms simultaneously in 2.0-MHz and 2.5 MHz channels. Experience from "in vitro" studies35,36
suggests that the Embo-Dop TCD can detect with the most reliable results microemboli ranging between 2 to 3 µm and 40 µm in case of gaseous microemboli and between 80 µm and 450 µm in case of solid microemboli. Gaseous emboli larger than 40 µm or solid emboli larger than 450 µm passing through the sample volume are associated with overloading of the detection mechanism. Very small emboli (gaseous <2-3 µm or solid <80 µm) may cause such a small increase in Doppler energy that they do not exceed the detection threshold of the instrumentation.
Statistical Analysis
Patient characteristics are presented as mean ± standard deviation. The number of microemboli is presented as median and interquartile ranges (IQRs). Statistical comparison, for normally distributed data, was perfomed with the Student t test for continuous variables and the
2 test or Fisher exact test for categorical variables. Inasmuch as the data for the number of microemboli were not normally distributed, the nonparametric Kruskal-Wallis test was used to compare the differences among the three groups.
 |
Results
|
|---|
Perioperative patient characteristics were similar in all groups and are summarized in Table 1. Long saphenous vein and radial artery conduits were used to perform 24 (70%) and 11 (30%) proximal aortic anastomoses, respectively, in group S, 16 (80%) and 4 (20%) proximal anastomoses in group E, and 9 (82%) and 2 (18%) in group H. The number of proximal anastomoses performed per patient ranged between 1 and 3 with a median of 1.6 (66/42). Mean time for performance of the proximal anastomosis with a side-biting clamp was 6 ± 1.4 minutes, with an Enclose device, 10.4 ± 2.1 minutes, and with a Heartstring device, 10.6 ± 2.1 minutes (P < .05).
 |
Microemboli
|
|---|
Only around 5% of all high-intensity transient signals were microemboli, with the remaining 95% being rejected as artifacts. The median number (IQR) of total microemboli was 11 (6-32) during side clamping, 11 (6-15) with the Enclose device, and 40 (31-48) with the Heartstring device (P < .01), as shown in Figure 5. The median numbers (IQRs) of gaseous and solid microemboli are shown in Figures 6A and 6B, respectively. The proportion of solid microemboli was significantly higher in group S (23%) than in group E (6%) and group H (1%) (P < .01) (Table 2).

View larger version (10K):
[in this window]
[in a new window]
|
Figure 5. Median number (IQR) of total microemboli detected in the three groups. The line inside the box plots represent the median number of microemboli. The upper and lower edges of each box are respectively the first and third quartile values. The top and bottom end of each whisker represent the range of microemboli. Circles correspond to the outliers (any value that lies between 1.5 and 3 box heights from the upper or lower edge of the box). Asterisks correspond to the extreme values (any value that lies more than 3 box heights from the upper or lower edge of the box). S, Side-biting clamp; E, Enclose; H, Heartstring.
|
|

View larger version (9K):
[in this window]
[in a new window]
|
Figure 6. A, Median number (IQR) of gaseous microemboli detected in the three groups. The median number (IQR) of gaseous microemboli was significantly higher when the Heartstring device was used compared with side-biting clamp or Enclose device. See Figure 5 for right interpretation of box plots and whiskers. B, Median number (IQR) of solid microemboli detected in the three groups. The use of a side-biting clamp resulted in a significantly higher number of solid microemboli compared with the use of the two new clampless proximal seal devices. See Figure 5 for right interpretation of box plots and whiskers.
|
|
View this table:
[in this window]
[in a new window]
|
TABLE 2. Median number (IQR) of total microemboli and proportion of gaseous and solid microemboli detected in the three groups
|
|
In each group the median number (IQR) of total microemboli was significantly higher during application/insertion and removal of the device compared with that occurring during performance of the anastomosis, as shown in Figure 7. However, significantly more microemboli occurred during insertion and removal of the Heartstring device than during the same maneuvers with the other two techniques. The median number of total microemboli during construction of the anastomosis using the Heartstring device was higher than in the other two groups but did not reach statistical significance (P = .239). The median number of gaseous and solid microemboli during the three different stages (application of the device, construction of anastomosis, and removal of the device) is shown in Figure 8. Application and removal of the side-biting clamp caused a significantly higher number of solid microemboli compared with either group E or H for the same maneuvers.

View larger version (29K):
[in this window]
[in a new window]
|
Figure 7. Median number (IQR) of total microemboli detected in the three groups during three different stages (applicat, application/insertion of the devices to the aortic wall; anastomo, manual construction of proximal anastomosis; removal, removal of the devices after completion of the anastomosis). See Figure 5 for right interpretation of box plots and whiskers.
|
|

View larger version (15K):
[in this window]
[in a new window]
|
Figure 8. Median number of solid and gaseous microemboli detected in the three groups during the three different stages. S, Side-biting clamp; E, Enclose; H, Heartstring.
|
|
 |
Discussion
|
|---|
To our knowledge, this is the first study quantifying the number of gaseous and solid cerebral microemboli, and their temporal generation, during proximal aortic anastomoses using a conventional side-biting clamp and two novel clampless proximal anastomotic devices in OPCAB. Our study demonstrated two findings: an increase in the total number of microemboli with the Heartstring device compared with the side-biting clamp or Enclose device but a far lower proportion of solid microemboli with the clampless devices (1% in group H, 6% in group E, and 23% in group S; Table 2). This finding is important because solid microemboli are known to be the most clinically damaging and therefore use of the side-biting clamp is likely to have the greatest detrimental effect on neurocognitive function. Furthermore, considering that all our patients had palpably normal aortas, more marked changes may have been observed in patients with more severe atherosclerotic disease. Moreover, the number of microemboli detected with TCD ultrasonography only represents a very small fraction of the total systemic microembolic load. Therefore, apparently small numbers of detected microemboli cannot really be considered insignificant because small differences should in reality be magnified several times. At last, the relationship between microembolic load and cerebral injury may not be a direct relationship.
In our study, most of the microemboli occurred during the application and removal of all three devices. Barbut and associates9
studied the temporal generation of microemboli in on-pump CABG during different maneuvers. They found that crossclamp and side-biting clamp removal were responsible for 58% of the total cerebral microembolic load (34% and 24%, respectively) with a far lower number of microemboli during clamp application. Inasmuch as the largest number of microemboli occurred in patients with the most severe aortic atherosclerosis, the authors concluded that the emboli detected at clamp removal may be due to dislodgment of fragments of atheromatous plaque from the aortic wall and not air bubbles or platelet clumps in the bypass equipment. In our study in the side-biting clamp group, 39% and 46% of the total microemboli were generated during clamp application and removal, respectively, and a significant proportion of these (Figure 8) were solid. These findings, detected in off-pump patients with a palpably normal aorta, add to the important conclusions from Barbut and associates study.
We detected considerably more gaseous microemboli during application and removal of the Heartstring device than with the other two techniques. Inasmuch as several reports38-40
have demonstrated direct deleterious effects of gaseous microembolization on the cerebral microvasculature, such findings may represent a significant disadvantage in using the Heartstring device. Between performing and covering the fully pressurized aortotomy with a finger and delivering the coiled seal system into the lumen, the formation of gaseous microemboli may be inevitable due to turbulence. A similar but more trivial process occurs on removal of the device. In an animal model, Nollert and colleagues41
speculated that most gaseous microemboli with the Heartstring device were attributable to the use of a blower-mister when constructing the anastomosis. We have, however, clearly demonstrated that the vast majority of gaseous microemboli occur when inserting and removing the Heartstring device rather than during construction of the anastomosis and therefore cannot be attributed solely to the use of the blower-mister.
Microembolization during proximal aortic anastomosis with clampless devices has been assessed with TCD ultrasound33,41,42
and intra-aortic filters29
by different authors. However, none of them used this new-generation multifrequency TCD or evaluated the temporal release of microemboli in terms of application of devices, construction of anastomosis, or removal of devices. The limitations inherent with traditional TCD ultrasound with its simple intensity threshold renders it unable to discriminate the nature of such microemboli. In addition, as we have previously reported (and confirmed in this current study), up to 95% of emboli may be artifacts, and this therefore has a considerable impact on the findings and conclusions of previous studies. Furthermore, the findings in some of those previous articles41,42
have been complicated by having a mixture of on-pump and off-pump patients.
Notwithstanding these limitations, Akpinar and coworkers33
reported a lower microembolic load with the Enclose II device than with the use of a side-biting clamp. Compared with a side-biting clamp, the sutureless Symmetry proximal anastomotic device (St Jude Medical, Inc, St Paul, Minn) has produced conflicting results, with Scarborough and colleagues42
reporting a reduction in microemboli and Martens and coworkers,29
using an intra-aortic filter, reporting no reduction in embolization.
With the Heartstring device and side-biting clamp, each proximal anastomosis requires repeat application and removal of the device. However, with the Enclose device it is possible to perform more than one proximal anastomosis without removing and reinserting the device; it is sufficient to manipulate the closed device to an adjacent area on the aorta. Inasmuch as the majority of microemboli occur during insertion and removal of these devices, it would be intuitive to assume that a single insertion, of the Enclose device, may result in a significant reduction in cerebral microembolic load with repeated anastomoses. However, we are concerned about the potential for dislodgment of atheromatous material from a diseased aorta when attempting to reposition the Enclose device for construction of a subsequent anastomoses.
 |
Potential Limitations
|
|---|
Although lack of a rigorous randomization is a potential limitation of the present study, we believe that this is unlikely to have significantly influenced the results because two independent observers, with no direct responsibility for clinical care of the patient, randomly chose which anastomotic device was to be used. As shown in Table 1, there were no systematic differences in baseline characteristics of the patients.
Inasmuch as this is a small study, a larger sample size would have increased the statistical power of the study, especially regarding the relative proportions of gaseous and solid microemboli. Nevertheless, significant differences were still obvious between the different devices and provide a basis for future randomized controlled trials in selected patient groups (eg, high-risk patients and those with extensive atherosclerosis).
Epiaortic ultrasound is a very sensitive method for demonstrating aortic atherosclerosis but is not routinely available in our center. All the patients in our study had a palpably normal aortic wall, and the risk of embolization and stroke would be significantly greater in those patients with diseased and calcified aortic walls, particularly elderly5,6,43,44
and diabetic patients.8,45
None of the patients had a perioperative stroke, and we did not use formal neuropsychologic assessment because the primary objective was simply to compare the degree of cerebral microembolization in the different groups. However, a strong correlation between intraoperative microembolization and postoperative neuropsychologic deficits has already been well documented,11-15
and inasmuch as particulate microemboli are believed to be most damaging, it is intuitive that any technique able to minimize solid cerebral microembolization will reduce the risk of even subclinical injury.
The authors of the original "in vivo" validation of the Embo-Dop system reported excellent results both in automatic rejection of artifacts and in discrimination between solid and gaseous microemboli,35,36
reducing bias in the interpretation of high-intensity transient signals and minimizing interobserver variability. Even so, showers of emboli entering the sample volume of the insonating probes may be wrongly identified as artifacts, underestimating the true embolic burden. As showers of microemboli are less likely with an off-pump "no touch" aortic technique, the phenomenon of "undercounting" showers of microemboli may underestimate the benefits of this technique.
A recent study by Markus and Punter46
has shown that the Embo-Dop system has a sensitivity of 50% and a specificity of 96% for detecting solid micremboli. Therefore, possibly as many as 50% of all solid microemboli may be misclassified as gaseous by the machine. If it is subsequently confirmed that an even greater proportion of microemboli are in fact solid rather than gaseous, it will only serve to emphasize the importance of using measures that may reduce the number of solid microemboli such as OPCAB and the avoidance of any aortic manipulation and especially side-biting clamps.
 |
Conclusions
|
|---|
The majority of microemboli occur during active aortic manipulation with application and removal of a side-biting clamp or insertion and removal of clampless aortic anastomotic devices. The use of such devices results in a significant reduction in the proportion of solid microemboli and may represent an important strategy for minimizing cerebral injury during proximal anastomoses, particularly in the highest risk groups.
 |
References
|
|---|
- Taggart DP, Westaby S. Neurological and cognitive disorder after CABG. Curr Opin Cardiol 2001;16:271-276.[Medline]
- Roach GW, Kanchuger M, Mangano CM, Newman M, Nussmeier N, Wolman R, et al. Adverse cerebral outcomes after coronary bypass surgery. N Engl J Med 1996;335:1857-1863.[Medline]
- van Dijk D, Keizer AM, Diephuis JC, Durand C, Vos LJ, Hijman R. Neurocognitive dysfunction after coronary artery bypass surgery: a systematic review. J Thorac Cardiovasc Surg 2000;120:632-639.[Abstract/Free Full Text]
- Newman MF, Kirchner JL, Phillips-Bute B, Gaver V, Grocott H, Jones RH, et al. Longitudinal assessment of neurocognitive function after coronary artery bypass surgery. N Engl J Med 2001;344:395-402.[Medline]
- Blauth CI, Cosgrove DM, Webb BW, Ratliff NB, Boylan M, Piedmonte MR, et al. Atheroembolism from the ascending aorta: an emerging problem in cardiac surgery. J Thorac Cardiovasc Surg 1992;103:1104-1111; discussion 1111-2.[Abstract]
- Bar-El Y, Goor DA. Clamping of the atherosclerotic ascending aorta during coronary artery bypass operations: its cost in stroke. J Thorac Cardiovasc Surg 1992;104:469-474.[Abstract]
- Abu-Omar Y, Taggart DP. Off-pump coronary artery bypass grafting. Lancet 2002;360:327-329.[Medline]
- Lynn GM, Stefanko K, Reed 3rd JF, Gee W, Nicholas G. Risk factors for stroke after coronary artery bypass. J Thorac Cardiovasc Surg 1992;104:1518-1523.[Abstract]
- Barbut D, Hinton RB, Statrowski TP, Hartman GS, Bruefach M, Williams-Russo P, et al. Cerebral emboli detected during bypass surgery are associated with clamp removal. Stroke 1994;25:2398-2402.[Abstract/Free Full Text]
- Wareing TH, Davila-Roman VG, Barzilai B, Murphy SF, Kouchoukos NT. Management of the severely atherosclerotic ascending aorta during cardiac operations. J Thorac Cardiovasc Surg 1992;103:453-462.[Abstract]
- Stump DA, Tegeler CH, Rogers AT, Coker LH, Newman SP, Wallenhaupt SL, et al. Neuropsychological deficits are associated with the number of emboli detected during cardiac surgery. Stroke 1993;24:509-516.
- Pugsley W, Klinger L, Paschalis C, Treasure T, Harrison M, Newman S. The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning. Stroke 1994;25:1393-1399.[Abstract/Free Full Text]
- Stump DA, Rogers AT, Hammon JW, Newman SP. Cerebral emboli and cognitive outcome after cardiac surgery. J Cardiothorac Vasc Anesth 1996;10:113-119.[Medline]
- Sylivris S, Levi C, Matalanis G, Rosalion A, Buxton BF, Mitchell A, et al. Pattern and significance of cerebral microemboli during coronary artery bypass grafting. Ann Thorac Surg 1998;66:1674-1678.[Abstract/Free Full Text]
- Abu-Omar Y, Cifelli A, Matthews PM, Taggart DP. The role of microembolisation in cerebral injury as defined by functional magnetic resonance imaging. Eur J Cardiothorac Surg 2004;26:586-591.[Abstract/Free Full Text]
- Abu-Omar Y, Balacumaraswami L, Pigott D, Matthews PM, Taggart DP. Solid and gaseous microembolization during off-pump, on-pump, and open cardiac surgery procedures. J Thorac Cardiovasc Surg 2004;127:1759-1765.[Abstract/Free Full Text]
- Lev-Ran O, Braunstein R, Sharony R, Kramer A, Paz Y, Mohr R, et al. No-touch aorta off-pump coronary surgery: the effect on stroke. J Thorac Cardiovasc Surg 2005;129:307-313.[Abstract/Free Full Text]
- Kapetanakis EI, Stamou SC, Dullum MKC, Hill PC, Haile E, Boyce SW, et al. The impact of aortic manipulation on neurologic outcomes after coronary artery bypass surgery: a risk-adjusted study. Ann Thorac Surg 2004;78:1564-1571.[Abstract/Free Full Text]
- Hangler HB, Nagele G, Danzmayr M, Mueller L, Ruttmann F, Laufer G, et al. Modification of surgical technique for ascending aortic atherosclerosis: impact on stroke reduction in coronary artery bypass grafting. J Thorac Cardiovasc Surg 2003;126:391-400.[Abstract/Free Full Text]
- Patel NC, Pullan, MD, Fabri BM. Does off-pump total arterial revascularization without aortic manipulation influence neurological outcome? A study of 226 consecutive, unselected cases. Heart Surg Forum 2002;5:28-32.[Medline]
- Kim KB, Kang CH, Chang WI, Lim C, Kim JH, Ham BM, et al. Off-pump coronary artery bypass with complete avoidance of aortic manipulation. Ann Thorac Surg 2002;74(Suppl):S1377-S1382.[Abstract/Free Full Text]
- Leacche M, Carrier M, Bouchard T, Pellerin M, Perrault LP, Paga P, et al. Improving neurologic outcome in off-pump surgery: the "no touch" technique. Heart Surg Forum 2003;6:169-175.[Medline]
- Bonatti J, Nagele G, Hangler H, Danzmayr M, Mueller L, Rieger M, et al. Extraanatomical coronary artery bypass grafts on the beating heart for management of the severely atherosclerotic ascending aorta. Heart Surg Forum 2002;5(Suppl 4):S272-S281.[Medline]
- Calafiore AM, Bar-El Y, Vitolla G, Di Gianmarco G, Teodori G, Iaco AL, et al. Early clinical experience with a new sutureless anastomotic device for proximal anastomosis of the saphenous vein to the aorta. J Thorac Cardiovasc Surg 2001;121:854-858.[Abstract/Free Full Text]
- Antona C, Scrofani R, Lemma M, Vanelli P, Mangini A, Danna P, et al. Assessment of an aortosaphenous vein graft anastomotic device in coronary surgery: clinical experience and early angiographic results. Ann Thorac Surg 2002;74:2101-2105.[Abstract/Free Full Text]
- Medalion B, Meirson D, Hauptman E, Sasson L, Schachner A. Initial experience with the Heartstring proximal anastomotic system. J Thorac Cardiovasc Surg 2004;128:273-277.[Abstract/Free Full Text]
- Weber A, Reuthebuch O, Turina M. Guidant Heartstring: initial experience in OPCAB surgery. Heart Surg Forum 2005;8:E4-E8.[Medline]
- Vicol C, Oberhoffer M, Nollert G, Eifert S, Boekstegers P, Wintersperger B, et al. First clinical experience with the HEARTSTRING, a device for proximal anastomoses in coronary surgery. Ann Thorac Surg 2005;79:1732-1737; discussion 1737.[Abstract/Free Full Text]
- Martens S, Dietrich M, Herzog C, Doss M, Schneider G, Moritz A, et al. Automatic connector devices for proximal anastomoses do not decrease embolic debris compared with conventional anastomoses in CABG. Eur J Cardiothorac Surg 2004;25:993-1000.[Abstract/Free Full Text]
- Katariya K, Yassin S, Teherani HY, Lombardi P, Masroor S, Salerno TA. Initial experience with sutureless proximal anastomoses performed with a mechanical connector leading to clampless off-pump coronary artery bypass surgery. Ann Thorac Surg 2004;77:563-567; discussion 567-8.[Abstract/Free Full Text]
- Eckstein FS, Bonilla LF, Englberger L, Stauffer E, Berg TA, Schmidli J, et al. Minimizing aortic manipulation during OPCAB using the Symmetry aortic connector system for proximal vein graft anastomoses. Ann Thorac Surg 2001;72:S995-S998.[Abstract/Free Full Text]
- Mills NL, Everson CT. Atherosclerosis of the ascending aorta and coronary artery bypass: pathology, clinical correlates, and operative management. J Thorac Cardiovasc Surg 1991;102:546-553.[Abstract]
- Akpinar B, Guden M, Sagbas E, Sanisoglu I, Ergenoglu MU, Turkoglu C. Clinical experience with the Novare Enclose II manual proximal anastomotic device during off-pump coronary artery surgery. Eur J Cardiothorac Surg 2005;27:1070-1073.[Abstract/Free Full Text]
- Guidant Corporation. HEARTSTRING Proximal Seal System. Instructions for use. Santa Clara (CA). 2002.
- Brucher R, Russell D. Automatic online embolus detection and artifact rejection with the first multifrequency trancranial Doppler. Stroke 2002;33:1969-1974.[Abstract/Free Full Text]
- Russell D, Brucher R. Online automatic discrimination between solid and gaseous cerebral microemboli with the first multifrequency trancranial Doppler. Stroke 2002;33:1975-1980.[Abstract/Free Full Text]
- Russell D, Brucher R. Embolus detection and differentiation using multifrequency transcranial Doppler. Stroke 2005;36:706.[Free Full Text]
- Deklunder G, Roussel M, Lecroart J-L, Prat A, Gautier C. Microemboli in cerebral circulation and alteration of cognitive abilities in patients with mechanical prosthetic heart valves. Stroke 1998;29:1821-1826.[Abstract/Free Full Text]
- Steinberg GK, De La Paz R, Mitchell SM, Bell TE, Albers GW. MRI and cerebrospinal fluid enzymes as sensitive indicators of subclinical cerebral injury after open-heart valve replacement surgery. Am J Neuroradiol 1996;17:205-212.[Abstract]
- Hills BA, James PB. Microbubble damage to the blood-brain barrier: relevance to decompression sickness. Undersea Biomed Res 1991;18:111-116.[Medline]
- Nollert G, Oberhoffer M, Reichart B, Vicol C. Combination of the HEARTSTRING proximal seal system with a blower mister: a possible source of gas emboli. J Thorac Cardiovasc Surg 2003;126:1192-1194.[Free Full Text]
- Scarborough JE, White W, Derilus FE, Mathew JP, Newman MF, Landolfo KP, Neurological Outcome Research Group Combined use of off-pump techniques and a sutureless proximal aortic anastomotic device reduces cerebral microemboli generation during coronary artery bypass grafting. J Thorac Cardiovasc Surg 2003;126:1561-1567.[Abstract/Free Full Text]
- Ricci M, Karamanoukian HL, Abraham R, Von Fricken K, DAncona G, Choi S, et al. Stroke in octogenarians undergoing coronary artery surgery with and without cardiopulmonary bypass. Ann Thorac Surg 2000;69:1471-1475.[Abstract/Free Full Text]
- Demaria RG, Carrier M, Fortier S, Martineau R, Fortier A, Cartier R, et al. Reduced mortality and strokes with off-pump coronary artery bypass grafting surgery in octogenarians. Circulation 2002;106(12 Suppl 1):I5-I10.[Medline]
- Weinberger J, Biscarra V, Weisberg MK, Jacobson JH. Factors contributing to stroke in patients with atherosclerotic disease of the great vessels: the role of diabetes. Stroke 1983;14:709-712.[Abstract/Free Full Text]
- Markus HS, Punter M. Can transcranial Doppler discriminate between solid and gaseous microemboli? Assessment of a dual-frequency transducer system. Stroke 2005;36:1731-1734.[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
M. Y. Emmert, B. Seifert, M. Wilhelm, J. Grunenfelder, V. Falk, and S. P. Salzberg
Aortic no-touch technique makes the difference in off-pump coronary artery bypass grafting
J. Thorac. Cardiovasc. Surg.,
December 1, 2011;
142(6):
1499 - 1506.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Biancari and S. Yli-Pyky
Meta-analysis on the use of the Heartstring anastomotic device to prevent stroke in patients undergoing off-pump coronary artery bypass grafting
Eur J Cardiothorac Surg,
November 1, 2011;
40(5):
1236 - 1240.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. D. Puskas, M. E. Halkos, H. Balkhy, M. Caskey, M. Connolly, J. Crouch, A. Diegeler, J. Gummert, W. Harringer, V. Subramanian, et al.
Evaluation of the PAS-Port Proximal Anastomosis System in coronary artery bypass surgery (the EPIC trial)
J. Thorac. Cardiovasc. Surg.,
July 1, 2009;
138(1):
125 - 132.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. C. Groom, R. D. Quinn, P. Lennon, D. J. Donegan, J. H. Braxton, R. S. Kramer, P. W. Weldner, L. Russo, S. D. Blank, A. A. Christie, et al.
Detection and Elimination of Microemboli Related to Cardiopulmonary Bypass
Circ Cardiovasc Qual Outcomes,
May 1, 2009;
2(3):
191 - 198.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. F. Buxton, P. A. R. Hayward, A. E. Newcomb, S. Moten, S. Seevanayagam, and I. Gordon
Choice of conduits for coronary artery bypass grafting: craft or science?
Eur J Cardiothorac Surg,
April 1, 2009;
35(4):
658 - 670.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Dittrich and E. B. Ringelstein
Occurrence and Clinical Impact of Microembolic Signals During or After Cardiosurgical Procedures
Stroke,
February 1, 2008;
39(2):
503 - 511.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Lynch and J. Riley
Microemboli detection on extracorporeal bypass circuitsa
Perfusion,
January 1, 2008;
23(1):
23 - 32.
[Abstract]
[PDF]
|
 |
|