J Thorac Cardiovasc Surg 2007;133:689-695
© 2007 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
a Department of Cardiothoracic Surgery, Medical University of Vienna, Wahringer Guertel, Vienna, Austria
b Department of Biomedical Engineering, Medical University of Vienna, Wahringer Guertel, Vienna, Austria
c Department of Cardiovascular Anesthesiology, Medical University of Vienna, Wahringer Guertel, Vienna, Austria.
Received for publication January 21, 2006; revisions received August 3, 2006; accepted for publication August 10, 2006. * Address for reprints: Daniel Zimfer, MD, Department of Cardiothoracic Surgery, Medical University Vienna, Waehringer Guertel, A-1090 Vienna, Austria (Email: daniel.zimpfer{at}meduniwien.ac.at).
| Abstract |
|---|
|
|
|---|
Methods: Thirty-five consecutive cardiac transplant candidates (age 56 ± 6 years, 88.5% were men) with fixed pulmonary hypertension (5.1 ± 2.6 Wood units) resistant to medical treatment received a left ventricular assist device as a bridge to transplantation. Three left ventricular assist device systems were used (pulsatile blood flow: Novacor [World Heart Inc, Oakland, Calif] n = 8; continuous blood flow: MicroMed DeBakey [MicroMed Technology Inc, Houston, Tex] n = 24, DuraHeart [Terumo Heart Inc, Ann Arbor, Mich] n = 3). Right-sided heart catheter data were obtained before left ventricular assist device implantation at 3-day and 6-week follow-ups. Clinical data and complications were recorded.
Results: Before left ventricular assist device implantation, the pulmonary vascular resistance was 5.1 ± 2.8 Wood units. Values were comparable in patients receiving pulsatile (5.1 ± 3.4 Wood units) or continuous blood flow left ventricular assist devices (5.1 ± 2.7 Wood units, P = .976). Left ventricular assist device implantation decreased pulmonary vascular resistance at 3-day (2.9 ± 1.3 Wood units, P < .0001) and 6-week (2.0 ± 0.8 Wood units, P < .0001) follow-ups compared with before implantation. This effect was independent of the type of left ventricular assist device system used (3-day follow-up: pulsatile flow: 3.2 ± 1.3 Wood units vs continuous flow: 2.7 ± 1.2 Wood units; P = .310 and 6-week follow-up: pulsatile flow: 1.9 ± 0.9 Wood units vs continuous flow: 2.1 ± 0.8 Wood units; P = .905). Twenty-four patients had successful bridges to transplantation (69%, mean time on left ventricular assist device 210 ± 83 days), and 11 patients died before transplantation (31%, mean time on left ventricular assist device 67 ± 30 days). The 1-year survival after transplantation was 95%.
Conclusion: Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates and allow patients to overcome a contraindication for cardiac transplantation. Therefore, left ventricular assist devices should be considered in all cardiac transplant candidates with fixed pulmonary hypertension.
| Introduction |
|---|
|
|
|---|
Pulmonary hypertension (PH) is a risk factor for early and late death after cardiac transplantation. The reason for this is the unacceptably high risk of acute donor right-sided heart failure immediately after implantation.3-7
There is consensus that the risk of death after cardiac transplantation is increased in patients with pulmonary vascular resistance (PVR) greater than 2.5 Wood units (WU), if PVR cannot be decreased by pharmacologic interventions (fixed PH).8,9
Various pharmacologic agents have been evaluated for their ability to decrease PVR in these patients.10
Pharmacologic agents tested include sodium nitroprusside, inhaled nitric oxide, phosphodiesterase inhibitors, urapidil, prostaglandins (PGE1 and PGI1), and levosimendan.10-14
However, even by combining these pharmacologic agents, PH cannot be significantly decreased in many of these patients. No safe and efficient treatment can be offered to those patients with end-stage heart failure and fixed severe PH today. Left ventricular assist devices (LVADs) have been reported to reduce PH in patients with terminal heart failure. We are the first to study this topic in a prospective fashion and to compare pulsatile and continuous flow LVADS.15-17
In this prospective study we report on our experience with the treatment strategy of improving fixed PH in cardiac transplant candidates by means of LVAD implantation and subsequent orthotopic cardiac transplantation.
| Materials and Methods |
|---|
|
|
|---|
The duration of support, survival, incidence of neurologic complications, device malfunctions, and infections were monitored. Infections were defined as the presence of a positive blood culture along with a leukocytosis.
Devices, Implant Procedure, and Anticoagulation
Three different systems of LVADs were used in the present study. The technical details and implantation procedures of the MicroMed DeBakey (MicroMed Technology Inc, Houston, Tex), DuraHeart (Terumo Heart Inc, Ann Arbor, Mich) (continuous blood flow), and Novacor LVADs (World Heart Inc, Oakland, Calif) (pulsatile blood flow) have been described.18-20
Anticoagulation protocol was identical in all patients. During extracorporal circulation and implantation of the pump, patients received intravenous heparin (300 U/kg body weight), and the heartlung machine was primed with 1,000,000 IU aprotinin. After discontinuation of extracorporal circulation, heparin was reversed with an appropriate dose of protamine. Intravenous heparin was instituted 6 hours after surgery to achieve activated partial thromboplastin target times of 50 to 60 seconds. Platelet anti-aggregation therapy with 150 mg per day of aspirin and 225 mg per day of dipyridamole was started after removal of all chest drains. Administration of heparin was stopped when anticoagulation with coumarin reached target levels of 2.5 to 3.5 international normalized ratio.
Measurement of Pulmonary Hypertension
Right-sided heart catheterization was performed according to the guidelines published by the American College of Cardiology/American Heart Association using a SwanGanz thermodilution catheter. Right-sided heart catheterization was performed before LVAD implantation and 3 days and 6 weeks after LVAD implantation.21
The PH variables assessed included PVR (Wood units), systolic pulmonary artery pressure (millimeters of mercury), mean pulmonary artery pressure (millimeters of mercury), pulmonary capillary wedge pressure (millimeters of mercury), and cardiac output (liters per minute). Cardiac output was measured with the Fick method.
Testing for Reversibility of Pulmonary Hypertension
PH was defined as PVR greater than 3.5 WU. Reversibility of PH was assessed by nitroglycerin, prostaglandin (PGI2), nitric oxide, and levosimendan (only available in the last 10 patients). Nitroglycerin was applied intravenously at increasing doses of 2 to 6 mg per hour with dose increments every 10 minutes. PGI2 was given intravenously at increasing doses of 10 to 200 ng/kg/min with dose increments every 5 minutes. Nitric oxide was administered in doses of 40, 60, and 80 ppm through a tight-fitting facemask. In the last 10 patients, levosimendan was additionally used for testing for reversibility of PH. If PH was not reversible with this treatment, it was considered as fixed. All of the mentioned substances were tested in all patients (Figure 1).
|
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Pulmonary vascular hypertension is a common complication of severe long-standing heart failure. Approximately 72% of patients with terminal heart failure who are eligible for cardiac transplantation have PH.22
The pathophysiology behind PH in patients with terminal heart failure is the result of a multifactorial process mainly relating to left ventricular failure.23
Left atrial hypertension resulting from left ventricular failure translates into increased postcapillary pressure in the pulmonary circulation.23
This enhances pulmonary endothelial dysfunction with decreased availability of nitric oxide and prostacyclin and increased production of thromboxane A2 and endothelin-1.23
The activity of serine elastase is up-regulated in the subendothelium causing glycoprotein deposition and smooth muscle cell hypertrophy and hyperplasia.24
Intermittent hypoxia stimulates vasoconstriction in the pulmonary vascular tree. Changes in the expression of von Willebrand factor result in the development of platelet fibrin microthrombi.25
In addition, this process is associated with smooth muscle cell hypertrophy and hyperplasia.23
Depending on the duration of these processes, remodeling of the pulmonary vascular tree takes place.
Fixed PH is considered to be present when elevated PVR cannot be significantly decreased (>20%) by pharmacologic interventions.6,26
Although there is no international consensus, most transplant centers will not offer cardiac transplantation in patients with PVR greater than 3 to 4 WU for the following reasons. In the case of cardiac transplantation, the relatively thin-walled right ventricle of the organ donor does not function adequately in the presence of abnormally elevated PVR. This carries the high risk of distension of the donor right ventricle, resulting in right ventricular failure. In this case, neither high doses of inotropes nor mechanical support of the right ventricle results in a satisfying outcome.26-29
Posttransplant survival in patients with fixed PH is significantly worse compared with patients with normal PVR in both short and long-term follow-ups.3-7,29
In the present study, LVADs decreased fixed PH during a 6-week period of support. PH in terminal heart failure, at least in part, is the consequence of elevated left atrial filling pressures resulting from impaired left ventricular systolic function. Therefore, it might be speculated that LVADs reverse this process by continuously unloading the left ventricle. It remains to be studied whether LVAD support results in a reverse-remodeling of the pulmonary vascular tree. Pulsatile and continuous blood flow LVADs equally decreased PH during a 6-week period of support. It was previously emphasized that left ventricular unloading is more efficiently done by continuous blood flow devices.30
This was based on the theoretic consideration that continuous blood flow devices unload the left ventricle during the whole cardiac circle. On the basis of this finding, the decision for the type of LVAD in cardiac transplant candidates with fixed PH should depend only on the patients demands. Especially in large patients, the flow provided by continuous flow LVADs may be too low to increase exercise tolerance and enable recovery of the patient.31
Compared with the high risk of donor heart failure in patients with fixed PH and orthotopic cardiac transplantation, the combined approach of LVAD support and subsequent cardiac transplantation seems promising. Successful bridging rates ranging from 65% to 70% have been reported in patients without PH.32
We successfully bridged 69% of patients to safe cardiac transplantation. Posttransplant survivals at 30-day (95%) and 12-month (95%) follow-ups were comparable to those reported in patients without PH at the time of cardiac transplantation.33
Because of these results, all cardiac transplant candidates with a PVR greater than 3.5 WU (after testing for reversibility) receive an LVAD before cardiac transplantation at our center. Alternative approaches to orthotopic cardiac transplantation, such as heterotopic cardiac transplantation and right ventricle-sparing transplant techniques, have been recommended in the past in patients with severe PH.34,35
In these select cases, the donor heart acts as a biological assist device to the native left ventricle or both ventricles. Survivals reported at 12-month follow-ups range between 83% and 59%.34,35
The major limitations of heterotopic transplantation are the availability of a suitable donor, technical difficulties during implantation, and, in particular, late interactions of donor and recipient heart.36
Survivals with these alternative approaches are significantly lower compared with LVAD implantation and subsequent orthotopic cardiac transplantation. The presented algorithm significantly reduces the incidence of right-sided heart failure after cardiac transplantation to less than 1% and is associated with good posttransplant survival.
Nevertheless, patient morbidity while on LVAD support still remains a major concern, because 51% of all patients exhibited a severe adverse event. Most common were neurologic events, which also significantly contributed to patient mortality on LVAD support. Other complications included infections and bleeding. No device-related complications were observed. In our series the incidence of adverse events compares with that reported by others.37,38
There was no difference in adverse events between pulsatile and continuous flow LVADs. Continuous research, improvement of devices, and careful patient selection are crucial to reduce LVAD-associated complications.
| Conclusions |
|---|
|
|
|---|
LVADs decrease fixed PH in cardiac transplant candidates and allow them to overcome fixed PH as a contraindication for cardiac transplantation. Therefore, LVADs should be considered in all cardiac transplant candidates with fixed PH.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. R. Banner, R. S. Bonser, A. L. Clark, S. Clark, P. J. Cowburn, R. S. Gardner, P. R. Kalra, T. McDonagh, C. A. Rogers, L. Swan, et al. UK guidelines for referral and assessment of adults for heart transplantation Heart, September 15, 2011; 97(18): 1520 - 1527. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Beyersdorf, C. Schlensak, M. Berchtold-Herz, and G. Trummer Regression of "fixed" pulmonary vascular resistance in heart transplant candidates after unloading with ventricular assist devices J. Thorac. Cardiovasc. Surg., October 1, 2010; 140(4): 747 - 749. [Full Text] [PDF] |
||||
![]() |
R. John, K. Liao, F. Kamdar, P. Eckman, A. Boyle, and M. Colvin-Adams Effects on pre- and posttransplant pulmonary hemodynamics in patients with continuous-flow left ventricular assist devices J. Thorac. Cardiovasc. Surg., August 1, 2010; 140(2): 447 - 452. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Melenovsky, H. Al-Hiti, L. Kazdova, A. Jabor, P. Syrovatka, I. Malek, J. Kettner, and J. Kautzner Transpulmonary B-Type Natriuretic Peptide Uptake and Cyclic Guanosine Monophosphate Release in Heart Failure and Pulmonary Hypertension: The Effects of Sildenafil J. Am. Coll. Cardiol., August 11, 2009; 54(7): 595 - 600. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Sandner, D. Zimpfer, P. Zrunek, A. Rajek, H. Schima, D. Dunkler, M. Grimm, E. Wolner, and G. M. Wieselthaler Renal Function and Outcome After Continuous Flow Left Ventricular Assist Device Implantation Ann. Thorac. Surg., April 1, 2009; 87(4): 1072 - 1078. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Viklicky, I. Netuka, M. Urbanova, M. Adamec, J. Maly, L. Voska, and E. Pokorna Kidney transplantation from a brain-dead heart transplant candidate treated with biventricular assist device: 12-month follow-up Clinical Kidney Journal, February 1, 2008; 1(1): 34 - 35. [Full Text] [PDF] |
||||
![]() |
M. E. Stone Current Status of Mechanical Circulatory Assistance Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2007; 11(3): 185 - 204. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |