JTCS Medtronic Endurant
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Daniel Zimpfer
Martin Czerny
Michael Grimm
Ernst Wolner
Georg Wieselthaler
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Zimpfer, D.
Right arrow Articles by Wieselthaler, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zimpfer, D.
Right arrow Articles by Wieselthaler, G.
Related Collections
Right arrow Mechanical Circulatory Assistance
Right arrow Transplantation - heart

J Thorac Cardiovasc Surg 2007;133:689-695
© 2007 The American Association for Thoracic Surgery


Cardiopulmonary Support and Physiology

Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates

Daniel Zimpfer, MDa,*, Philipp Zrunek, MSb, Wilfried Roethy, MDa, Martin Czerny, MDa, Heinz Schima, PhDb, Leopold Huber, PhDb, Michael Grimm, MDa, Angela Rajek, MDc, Ernst Wolner, MD, PhDa, Georg Wieselthaler, MDa

a Department of Cardiothoracic Surgery, Medical University of Vienna, Wahringer Guertel, Vienna, Austria
b Department of Biomedical Engineering, Medical University of Vienna, Wahringer Guertel, Vienna, Austria
c Department of Cardiovascular Anesthesiology, Medical University of Vienna, Wahringer Guertel, Vienna, Austria.

Received for publication January 21, 2006; revisions received August 3, 2006; accepted for publication August 10, 2006.

* Address for reprints: Daniel Zimfer, MD, Department of Cardiothoracic Surgery, Medical University Vienna, Waehringer Guertel, A-1090 Vienna, Austria (Email: daniel.zimpfer{at}meduniwien.ac.at).


    Abstract
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Objective: Fixed pulmonary hypertension is a contraindication for cardiac transplantation because of the increased risk of donor heart failure. We sought to determine whether left ventricular assist devices improve fixed pulmonary hypertension in cardiac transplant candidates to enable safe cardiac transplantation.

Methods: Thirty-five consecutive cardiac transplant candidates (age 56 ± 6 years, 88.5% were men) with fixed pulmonary hypertension (5.1 ± 2.6 Wood units) resistant to medical treatment received a left ventricular assist device as a bridge to transplantation. Three left ventricular assist device systems were used (pulsatile blood flow: Novacor [World Heart Inc, Oakland, Calif] n = 8; continuous blood flow: MicroMed DeBakey [MicroMed Technology Inc, Houston, Tex] n = 24, DuraHeart [Terumo Heart Inc, Ann Arbor, Mich] n = 3). Right-sided heart catheter data were obtained before left ventricular assist device implantation at 3-day and 6-week follow-ups. Clinical data and complications were recorded.

Results: Before left ventricular assist device implantation, the pulmonary vascular resistance was 5.1 ± 2.8 Wood units. Values were comparable in patients receiving pulsatile (5.1 ± 3.4 Wood units) or continuous blood flow left ventricular assist devices (5.1 ± 2.7 Wood units, P = .976). Left ventricular assist device implantation decreased pulmonary vascular resistance at 3-day (2.9 ± 1.3 Wood units, P < .0001) and 6-week (2.0 ± 0.8 Wood units, P < .0001) follow-ups compared with before implantation. This effect was independent of the type of left ventricular assist device system used (3-day follow-up: pulsatile flow: 3.2 ± 1.3 Wood units vs continuous flow: 2.7 ± 1.2 Wood units; P = .310 and 6-week follow-up: pulsatile flow: 1.9 ± 0.9 Wood units vs continuous flow: 2.1 ± 0.8 Wood units; P = .905). Twenty-four patients had successful bridges to transplantation (69%, mean time on left ventricular assist device 210 ± 83 days), and 11 patients died before transplantation (31%, mean time on left ventricular assist device 67 ± 30 days). The 1-year survival after transplantation was 95%.

Conclusion: Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates and allow patients to overcome a contraindication for cardiac transplantation. Therefore, left ventricular assist devices should be considered in all cardiac transplant candidates with fixed pulmonary hypertension.



Abbreviations and Acronyms LVAD = left ventricular assist device; PG = prostaglandin; PH = pulmonary hypertension; PVR = pulmonary vascular resistance; WU = Wood units



    Introduction
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
The prevalence of congestive heart failure is continuously increasing in the Western world.1Go Despite continuous advances of medical therapy for congestive heart failure, cardiac transplantation is the most effective treatment for patients with end-stage heart failure.2Go

Pulmonary hypertension (PH) is a risk factor for early and late death after cardiac transplantation. The reason for this is the unacceptably high risk of acute donor right-sided heart failure immediately after implantation.3-7Go There is consensus that the risk of death after cardiac transplantation is increased in patients with pulmonary vascular resistance (PVR) greater than 2.5 Wood units (WU), if PVR cannot be decreased by pharmacologic interventions (fixed PH).8,9Go Various pharmacologic agents have been evaluated for their ability to decrease PVR in these patients.10Go Pharmacologic agents tested include sodium nitroprusside, inhaled nitric oxide, phosphodiesterase inhibitors, urapidil, prostaglandins (PGE1 and PGI1), and levosimendan.10-14Go However, even by combining these pharmacologic agents, PH cannot be significantly decreased in many of these patients. No safe and efficient treatment can be offered to those patients with end-stage heart failure and fixed severe PH today. Left ventricular assist devices (LVADs) have been reported to reduce PH in patients with terminal heart failure. We are the first to study this topic in a prospective fashion and to compare pulsatile and continuous flow LVADS.15-17Go

In this prospective study we report on our experience with the treatment strategy of improving fixed PH in cardiac transplant candidates by means of LVAD implantation and subsequent orthotopic cardiac transplantation.


    Materials and Methods
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Patients
During the study period, 242 cardiac transplant candidates were admitted to our hospital. Of those, 72 patients received an LVAD. Indication for LVAD implantation was terminal heart failure without fixed PH in 36 patients (50%), destination therapy in 1 patient (1%), and terminal heart failure with fixed PH (>3.5 WU) in 35 patients (49%). The 35 consecutive patients with terminal heart failure and fixed PH entered this prospective study. To qualify, all patients had to fulfill institutional inclusion criteria for terminal heart failure and had to have fixed PH unresponsive to maximum medical treatment. With the exception of fixed PH, patients had to be suitable for cardiac transplantation. The study was approved by the institutional review board, and all patients gave their written and informed consent before LVAD implantation and subsequent cardiac transplantation.

The duration of support, survival, incidence of neurologic complications, device malfunctions, and infections were monitored. Infections were defined as the presence of a positive blood culture along with a leukocytosis.

Devices, Implant Procedure, and Anticoagulation
Three different systems of LVADs were used in the present study. The technical details and implantation procedures of the MicroMed DeBakey (MicroMed Technology Inc, Houston, Tex), DuraHeart (Terumo Heart Inc, Ann Arbor, Mich) (continuous blood flow), and Novacor LVADs (World Heart Inc, Oakland, Calif) (pulsatile blood flow) have been described.18-20Go

Anticoagulation protocol was identical in all patients. During extracorporal circulation and implantation of the pump, patients received intravenous heparin (300 U/kg body weight), and the heart–lung machine was primed with 1,000,000 IU aprotinin. After discontinuation of extracorporal circulation, heparin was reversed with an appropriate dose of protamine. Intravenous heparin was instituted 6 hours after surgery to achieve activated partial thromboplastin target times of 50 to 60 seconds. Platelet anti-aggregation therapy with 150 mg per day of aspirin and 225 mg per day of dipyridamole was started after removal of all chest drains. Administration of heparin was stopped when anticoagulation with coumarin reached target levels of 2.5 to 3.5 international normalized ratio.

Measurement of Pulmonary Hypertension
Right-sided heart catheterization was performed according to the guidelines published by the American College of Cardiology/American Heart Association using a Swan–Ganz thermodilution catheter. Right-sided heart catheterization was performed before LVAD implantation and 3 days and 6 weeks after LVAD implantation.21Go The PH variables assessed included PVR (Wood units), systolic pulmonary artery pressure (millimeters of mercury), mean pulmonary artery pressure (millimeters of mercury), pulmonary capillary wedge pressure (millimeters of mercury), and cardiac output (liters per minute). Cardiac output was measured with the Fick method.

Testing for Reversibility of Pulmonary Hypertension
PH was defined as PVR greater than 3.5 WU. Reversibility of PH was assessed by nitroglycerin, prostaglandin (PGI2), nitric oxide, and levosimendan (only available in the last 10 patients). Nitroglycerin was applied intravenously at increasing doses of 2 to 6 mg per hour with dose increments every 10 minutes. PGI2 was given intravenously at increasing doses of 10 to 200 ng/kg/min with dose increments every 5 minutes. Nitric oxide was administered in doses of 40, 60, and 80 ppm through a tight-fitting facemask. In the last 10 patients, levosimendan was additionally used for testing for reversibility of PH. If PH was not reversible with this treatment, it was considered as fixed. All of the mentioned substances were tested in all patients (Figure 1).


Figure 1
View larger version (22K):
[in this window]
[in a new window]

 
Figure 1. Pharmacologic testing: decision-making tree. TX, Transplant; PVR, pulmonary vascular resistance; WU, Wood units; PG, prostaglandin; NO, nitric oxide; LVAD, left ventricular assist device.

 
Statistical Analysis
Data are presented as frequency distributions and percentages. Values of continuous variables are expressed as mean ± standard deviation. Continuous variables were compared using analysis of variance (Bonferroni). Categoric variables were compared by means of the chi-square or Fisher exact test as appropriate. Kaplan–Meier analysis was used to calculate long-term survival along with a log–rank P value when comparing groups. The study was analyzed with the Statistical Package for the Social Sciences 11.5 (SPSS Inc, Chicago, Ill).


    Results
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Demographics
The characteristics of LVAD recipients are shown in Table 1. The mean overall age was 56.6 years (88.5% were male). All patients were in New York Heart Association class IV and had fixed PH (PVR 5.1 ± 2.8 WU) before LVAD implantation. Idiopathic cardiomyopathy was the most common cause of heart failure present in 23 patients (65.5%), and ischemic cardiomyopathy was present in 12 patients (34.5%). Recipients of pulsatile and continuous flow LVAD were comparable with regard to patient characteristics (Table 1). Overall mechanical support time was 202 ± 169 days (range, 35-700 days).


View this table:
[in this window]
[in a new window]

 
TABLE 1 Patient characteristics and heart failure therapy
 
Survival
Twenty-four patients (69%) had successful bridges to transplantation (bridge to transplant time 210 ± 83 days), and 11 patients (31%) died during LVAD support (survival time 67 ± 30 days). One patient (5%) died after cardiac transplantation. We found no difference in bridge to transplant success between patients treated with pulsatile (62.5%) or continuous flow (71.1%) LVAD (P = .147). One-year survival after cardiac transplantation was 95%. The mean follow-up after cardiac transplantation was 25 ± 17 months. Overall survival is displayed in Figure 2.


Figure 2
View larger version (9K):
[in this window]
[in a new window]

 
Figure 2. Kaplan–Meier curve displaying overall survival after LVAD implantation.

 
Right-sided Heart Catheterization
Before LVAD implantation, PVR was 5.1 ± 2.8 WU. Values were comparable in patients receiving pulsatile (5.1 ± 3.4 WU) and continuous blood flow LVADs (5.1 ± 2.7 WU, P = .976). LVAD implantation decreased PVR at 3-day (2.9 ± 1.3 WU, P < .0001) and 6-week follow-ups (2.0 ± 0.8 WU, P < .0001) compared with before implantation. This effect was independent of the type of LVAD system used (3-day follow-up: pulsatile flow 3.2 ± 1.3 WU vs continuous flow 2.7 ± 1.2 WU, P = .310; 6-week follow-up: pulsatile flow 1.9 ± 0.9 WU vs continuous flow 2.1 ± 0.8 WU, P = .905). Detailed hemodynamic data are shown in Table 2.


View this table:
[in this window]
[in a new window]

 
TABLE 2 Data from right-sided heart catheterization
 
Complications and Causes of Death
The incidence and nature of complications and the causes of death are shown in Table 3. None of the patients required mechanical right-sided heart support after LVAD implantation. A total of 51.3% of patients experienced a complication while on LVAD support. Cerebrovascular events occurred in 31% of patients, and infections occurred in 25.7% of patients. We found no difference in the incidence and nature of complications between patients treated with pulsatile and continuous blood flow devices (Table 3). Cerebrovascular events were the most common cause of death present in 45.4% of patients, followed by multiorgan failure (36.3%) and infections (18.3%). We found no difference in the incidence and nature of complications between patients treated with pulsatile and continuous blood flow devices (Table 3).


View this table:
[in this window]
[in a new window]

 
TABLE 3 Complications and causes of death
 

    Discussion
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
Severe fixed PH is a contraindication for cardiac transplantation because of an increased risk of postoperative donor heart failure. LVADs decrease fixed PH in cardiac transplant candidates and enable them to undergo orthotopic cardiac transplantation.

Pulmonary vascular hypertension is a common complication of severe long-standing heart failure. Approximately 72% of patients with terminal heart failure who are eligible for cardiac transplantation have PH.22Go The pathophysiology behind PH in patients with terminal heart failure is the result of a multifactorial process mainly relating to left ventricular failure.23Go Left atrial hypertension resulting from left ventricular failure translates into increased postcapillary pressure in the pulmonary circulation.23Go This enhances pulmonary endothelial dysfunction with decreased availability of nitric oxide and prostacyclin and increased production of thromboxane A2 and endothelin-1.23Go The activity of serine elastase is up-regulated in the subendothelium causing glycoprotein deposition and smooth muscle cell hypertrophy and hyperplasia.24Go Intermittent hypoxia stimulates vasoconstriction in the pulmonary vascular tree. Changes in the expression of von Willebrand factor result in the development of platelet fibrin microthrombi.25Go In addition, this process is associated with smooth muscle cell hypertrophy and hyperplasia.23Go Depending on the duration of these processes, remodeling of the pulmonary vascular tree takes place.

Fixed PH is considered to be present when elevated PVR cannot be significantly decreased (>20%) by pharmacologic interventions.6,26Go Although there is no international consensus, most transplant centers will not offer cardiac transplantation in patients with PVR greater than 3 to 4 WU for the following reasons. In the case of cardiac transplantation, the relatively thin-walled right ventricle of the organ donor does not function adequately in the presence of abnormally elevated PVR. This carries the high risk of distension of the donor right ventricle, resulting in right ventricular failure. In this case, neither high doses of inotropes nor mechanical support of the right ventricle results in a satisfying outcome.26-29Go Posttransplant survival in patients with fixed PH is significantly worse compared with patients with normal PVR in both short and long-term follow-ups.3-7,29Go

In the present study, LVADs decreased fixed PH during a 6-week period of support. PH in terminal heart failure, at least in part, is the consequence of elevated left atrial filling pressures resulting from impaired left ventricular systolic function. Therefore, it might be speculated that LVADs reverse this process by continuously unloading the left ventricle. It remains to be studied whether LVAD support results in a reverse-remodeling of the pulmonary vascular tree. Pulsatile and continuous blood flow LVADs equally decreased PH during a 6-week period of support. It was previously emphasized that left ventricular unloading is more efficiently done by continuous blood flow devices.30Go This was based on the theoretic consideration that continuous blood flow devices unload the left ventricle during the whole cardiac circle. On the basis of this finding, the decision for the type of LVAD in cardiac transplant candidates with fixed PH should depend only on the patient’s demands. Especially in large patients, the flow provided by continuous flow LVADs may be too low to increase exercise tolerance and enable recovery of the patient.31Go

Compared with the high risk of donor heart failure in patients with fixed PH and orthotopic cardiac transplantation, the combined approach of LVAD support and subsequent cardiac transplantation seems promising. Successful bridging rates ranging from 65% to 70% have been reported in patients without PH.32Go We successfully bridged 69% of patients to safe cardiac transplantation. Posttransplant survivals at 30-day (95%) and 12-month (95%) follow-ups were comparable to those reported in patients without PH at the time of cardiac transplantation.33Go Because of these results, all cardiac transplant candidates with a PVR greater than 3.5 WU (after testing for reversibility) receive an LVAD before cardiac transplantation at our center. Alternative approaches to orthotopic cardiac transplantation, such as heterotopic cardiac transplantation and right ventricle-sparing transplant techniques, have been recommended in the past in patients with severe PH.34,35Go In these select cases, the donor heart acts as a biological assist device to the native left ventricle or both ventricles. Survivals reported at 12-month follow-ups range between 83% and 59%.34,35Go The major limitations of heterotopic transplantation are the availability of a suitable donor, technical difficulties during implantation, and, in particular, late interactions of donor and recipient heart.36Go Survivals with these alternative approaches are significantly lower compared with LVAD implantation and subsequent orthotopic cardiac transplantation. The presented algorithm significantly reduces the incidence of right-sided heart failure after cardiac transplantation to less than 1% and is associated with good posttransplant survival.

Nevertheless, patient morbidity while on LVAD support still remains a major concern, because 51% of all patients exhibited a severe adverse event. Most common were neurologic events, which also significantly contributed to patient mortality on LVAD support. Other complications included infections and bleeding. No device-related complications were observed. In our series the incidence of adverse events compares with that reported by others.37,38Go There was no difference in adverse events between pulsatile and continuous flow LVADs. Continuous research, improvement of devices, and careful patient selection are crucial to reduce LVAD-associated complications.


    Conclusions
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 
The primary limitation of the present study is the nonrandomized design. The reason for this design is because fixed PH is a contraindication for cardiac orthotopic transplantation. Therefore, comparing cardiac transplantation with and without prior LVAD implantation would be ethically questionable.

LVADs decrease fixed PH in cardiac transplant candidates and allow them to overcome fixed PH as a contraindication for cardiac transplantation. Therefore, LVADs should be considered in all cardiac transplant candidates with fixed PH.


    Acknowledgments
 
We thank Daniela Dunkler, MSc, from the Institute of Medical Biometry, Medical University of Vienna, for assistance with the statistical analysis.


    References
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Conclusions
 References
 

  1. Smith ER. Heart failure—are we making progress?. Can J Cardiol 2002;18:1124-1125.[Medline]
  2. Meyer B, Moertl D, Huelsmann M, Pacher R, Berger R. In the long term heart transplantation provides additional benefit in transplant candidates with uptitrated ACE-inhibitor- and ß-blocker-therapy. J Heart Lung Transplant 2005;24:S69.
  3. Chen JM, Levin HR, Michler RE, Prusmack CJ, Rose EA, Aaronson KD. Reevaluating the significance of pulmonary hypertension before cardiac transplantation: determination of optimal thresholds and quantification of the effect of reversibility on perioperative mortality. J Thorac Cardiovasc Surg 1997;114:627-634.[Abstract/Free Full Text]
  4. Costard-Jackle A, Fowler MB. Influence of preoperative pulmonary artery pressure on mortality after heart transplantation: testing of potential reversibility of pulmonary hypertension with nitroprusside is useful in defining a high risk group. J Am Coll Cardiol 1992;19:48-54.[Abstract]
  5. Espinoza C, Manito N, Castells E, Roca J, Rodriguez R, Octavio de Toledo MC, et al. Pretransplant risk factors of early mortality after orthotopic heart transplantation. Transplant Proc 1999;31:2507-2508.[Medline]
  6. Murali S, Uretsky BF, Armitage JM, Tokarczyk TR, Betschart AR, Kormos RL, et al. Utility of prostaglandin E1 in the pretransplantation evaluation of heart failure patients with significant pulmonary hypertension. J Heart Lung Transplant 1992;11:716-723.[Medline]
  7. Murali S, Kormos RL, Uretsky BF, Schechter D, Reddy PS, Denys BG, et al. Preoperative pulmonary hemodynamics and early mortality after orthotopic cardiac transplantation: the Pittsburgh experience. Am Heart J 1993;126:896-904.[Medline]
  8. Delgado JF, Gomez-Sanchez MA, Saenz de la Calzada C, Sanchez V, Escribano P, Hernandez-Afonso J, et al. Impact of mild pulmonary hypertension on mortality and pulmonary artery pressure profile after heart transplantation. J Heart Lung Transplant 2001;20:942-948.[Medline]
  9. Deng MC, Gradaus R, Hammel D, Weyand M, Gunther F, Kerber S, et al. Heart transplant candidates at high risk can be identified at the time of initial evaluation. Transpl Int 1996;9:38-45.[Medline]
  10. Balzer DT, Kort HW, Day RW, Corneli HM, Kovalchin JP, Cannon BC, et al. Inhaled Nitric Oxide as a Preoperative Test (INOP Test I): the INOP Test Study Group. Circulation 2002;106(12 Suppl 1):I76-I81.[Medline]
  11. Zink M, Gombotz H, Wasler A, Grasser B, Rehak P, Metzler H. Urapidil reduces elevated pulmonary vascular resistance in patients before heart transplantation. J Heart Lung Transplant 2002;21:347-353.[Medline]
  12. Givertz MM, Hare JM, Loh E, Gauthier DF, Colucci WS. Effect of bolus milrinone on hemodynamic variables and pulmonary vascular resistance in patients with severe left ventricular dysfunction: a rapid test for reversibility of pulmonary hypertension. J Am Coll Cardiol 1996;28:1775-1780.[Abstract]
  13. Radovancevic B, Vrtovec B, Thomas CD, Croitoru M, Myers TJ, Radovancevic R, et al. Nitric oxide versus prostaglandin E1 for reduction of pulmonary hypertension in heart transplant candidates. J Heart Lung Transplant 2005;24:690-695.[Medline]
  14. Schulze-Neick I, Luther YC, Ewert P, Lehmkuhl HB, Hetzer R, Lange PE. End-stage heart failure with pulmonary hypertension: levosimendan to evaluate for heart transplantation alone versus combined heart-lung transplantation. Transplantation 2004;78:1237-1238.[Medline]
  15. Martin J, Siegenthaler MP, Friesewinkel O, Fader T, van de Loo A, Trummer G, et al. Implantable left ventricular assist device for treatment of pulmonary hypertension in candidates for orthotopic heart transplantation-a preliminary study. Eur J Cardiothorac Surg 2004;25:971-977.[Abstract/Free Full Text]
  16. Salzberg SP, Lachat ML, von Harbou K, Zund G, Turina MI. Normalization of high pulmonary vascular resistance with LVAD support in heart transplantation candidates. Eur J Cardiothorac Surg 2005;27:222-225.[Abstract/Free Full Text]
  17. Smedira NG, Massad MG, Navia J, Vargo RL, Patel AN, Cook DJ, et al. Pulmonary hypertension is not a risk factor for RVAD use and death after left ventricular assist system support. ASAIO J 1996;42:M733-M735.[Medline]
  18. Wieselthaler GM, Schima H, Hiesmayr M, Pacher R, Laufer G, Noon GP, et al. First clinical experience with the DeBakey VAD continuous-axial-flow pump for bridge to transplantation. Circulation 2000;101:356-359.[Abstract/Free Full Text]
  19. Pennington DG, McBride LR, Swartz MT. Implantation technique for the Novacor left ventricular assist system. J Thorac Cardiovasc Surg 1994;108:604-608.[Abstract/Free Full Text]
  20. Takatani S, Matsuda H, Hanatani A, Nojiri C, Yamazaki K, Motomura T, et al. Mechanical circulatory support devices (MCSD) in Japan: current status and future directions. J Artif Organs 2005;8:13-27.[Medline]
  21. American College of Cardiology/American Heart Association Ad Hoc Task Force on Cardiac Catheterization ACC/AHA guidelines for cardiac catheterization and cardiac catheterization laboratories. Anonymous J Am Coll Cardiol 1991;18:1149-1182.[Medline]
  22. Butler J, Chomsky DB, Wilson JR. Pulmonary hypertension and exercise intolerance in patients with heart failure. J Am Coll Cardiol 1999;34:1802-1806.[Abstract/Free Full Text]
  23. Rabinovitch M. Pulmonary hypertension: pathophysiology as a basis for clinical decision making. J Heart Lung Transpl 1999;18:1041-1053.[Medline]
  24. Todorovich-Hunter L, Dodo H, Ye C, McCready L, Keeley FW, Rabionvitch M. Increased pulmonary artery elastolytic activity in adult rats with monocrotaline induced progressive hypertensive pulmonary vascular disease compared with infant rats with nonprogressive disease. Am Rev Respir Dis 1992;146:213-223.[Medline]
  25. Turner Gomes SO, Boudreau N, Rabinovitch M. Effect of ambient oxygen changes on platelet activating factor production by fetal ovine endothelial cells. Prostaglandines 1991;41:463-472.
  26. Taghavi S, Zuckermann A, Ankersmit J, Wieselthaler G, Rajek A, Laufer G, et al. Extracorporeal membrane oxygenation is superior to right ventricular assist device for acute right ventricular failure after heart transplantation. Ann Thorac Surg 2004;78:1644-1649.[Abstract/Free Full Text]
  27. Kirklin JK, Naftel DC, Kirklin JW, Blackstone EH, White-Williams C, Bourge RC. Pulmonary vascular resistance and the risk of heart transplantation. J Heart Transplant 1988;7:331-336.[Medline]
  28. Kormos RL, Thompson M, Hardesty RL, Griffith BP, Trento A, Uretsky BF, et al. Utility of perioperative right heart catheterization data as a predictor of survival after heart transplantation. J Heart Transplant 1986;5:391.
  29. Erickson KW, Costanzo-Nordin MR, O’Sullivan EJ, Johnson MR, Zucker MJ, Pifarre R, et al. Influence of preoperative transpulmonary gradient on late mortality after orthotopic heart transplantation. J Heart Transplant 1990;9:526-537.[Medline]
  30. Haddad H, Elabbassi W, Moustafa S, Davies R, Mesana T, Hendry P, et al. Left ventricular assist devices as bridge to heart transplantation in congestive heart failure with pulmonary hypertension. ASAIO J 2005;51:456-460.[Medline]
  31. Siegenthaler MP, Martin J, van de Loo A, Doenst T, Bothe W, Beyersdorf F. Implantation of the permanent Jarvik-2000 left ventricular assist device: a single-center experience. J Am Coll Cardiol 2002;39:1764-1772.[Abstract/Free Full Text]
  32. Oz MC, Argenziano M, Catanese KA, Gardocki MT, Goldstein DJ, Ashton RC, et al. Bridge experience with long-term implantable left ventricular assist devices. Are they an alternative to transplantation?Circulation. 1997;95:1844-1852.
  33. International Society for Heart & Lung Transplantation Registry 2005. Available at: www.ishlt.org. Accessed January 2006.
  34. Bleasdale RA, Banner NR, Anyanwu AC, Mitchell AG, Khaghani A, Yacoub MH. Determinants of outcome after heterotopic heart transplantation. J Heart Lung Transplant 2002;21:867-873.[Medline]
  35. Elefteriades JA, Lovoulos CJ, Tellides G, Goldstein LJ, Rocco EJ, Condos SG, et al. Right ventricle-sparing heart transplant: promising new technique for recipients with pulmonary hypertension. Ann Thorac Surg 2000;69:1858-1863.[Abstract/Free Full Text]
  36. Akasaka T, Lythall D, Cheng A, Yoshida K, Yoshikawa J, Mitchell A, et al. Continuous aortic regurgitation in severely dysfunctional native hearts after heterotopic cardiac transplantation. Am J Cardiol 1989;63:1483-1488.[Medline]
  37. Stevenson LW, Rose EA. Left ventricular assist devices: bridges to transplantation, recovery, and destination for whom?. Circulation 2003;108:3059-3063.[Free Full Text]
  38. Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, et al. Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) Study Group Long-term mechanical left ventricular assistance for end-stage heart failure. N Engl J Med 2001;345:1435-1443.[Medline]



This article has been cited by other articles:


Home page
HeartHome page
N. R. Banner, R. S. Bonser, A. L. Clark, S. Clark, P. J. Cowburn, R. S. Gardner, P. R. Kalra, T. McDonagh, C. A. Rogers, L. Swan, et al.
UK guidelines for referral and assessment of adults for heart transplantation
Heart, September 15, 2011; 97(18): 1520 - 1527.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
F. Beyersdorf, C. Schlensak, M. Berchtold-Herz, and G. Trummer
Regression of "fixed" pulmonary vascular resistance in heart transplant candidates after unloading with ventricular assist devices
J. Thorac. Cardiovasc. Surg., October 1, 2010; 140(4): 747 - 749.
[Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
R. John, K. Liao, F. Kamdar, P. Eckman, A. Boyle, and M. Colvin-Adams
Effects on pre- and posttransplant pulmonary hemodynamics in patients with continuous-flow left ventricular assist devices
J. Thorac. Cardiovasc. Surg., August 1, 2010; 140(2): 447 - 452.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
V. Melenovsky, H. Al-Hiti, L. Kazdova, A. Jabor, P. Syrovatka, I. Malek, J. Kettner, and J. Kautzner
Transpulmonary B-Type Natriuretic Peptide Uptake and Cyclic Guanosine Monophosphate Release in Heart Failure and Pulmonary Hypertension: The Effects of Sildenafil
J. Am. Coll. Cardiol., August 11, 2009; 54(7): 595 - 600.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
S. E. Sandner, D. Zimpfer, P. Zrunek, A. Rajek, H. Schima, D. Dunkler, M. Grimm, E. Wolner, and G. M. Wieselthaler
Renal Function and Outcome After Continuous Flow Left Ventricular Assist Device Implantation
Ann. Thorac. Surg., April 1, 2009; 87(4): 1072 - 1078.
[Abstract] [Full Text] [PDF]


Home page
Clin Kidney JHome page
O. Viklicky, I. Netuka, M. Urbanova, M. Adamec, J. Maly, L. Voska, and E. Pokorna
Kidney transplantation from a brain-dead heart transplant candidate treated with biventricular assist device: 12-month follow-up
Clinical Kidney Journal, February 1, 2008; 1(1): 34 - 35.
[Full Text] [PDF]


Home page
SEMIN CARDIOTHORAC VASC ANESTHHome page
M. E. Stone
Current Status of Mechanical Circulatory Assistance
Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2007; 11(3): 185 - 204.
[Abstract] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Daniel Zimpfer
Martin Czerny
Michael Grimm
Ernst Wolner
Georg Wieselthaler
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Zimpfer, D.
Right arrow Articles by Wieselthaler, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zimpfer, D.
Right arrow Articles by Wieselthaler, G.
Related Collections
Right arrow Mechanical Circulatory Assistance
Right arrow Transplantation - heart


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS