J Thorac Cardiovasc Surg 2007;133:1147-1153
© 2007 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
a Department of Surgery, New York Presbyterian Hospital, College of Physicians & Surgeons, Columbia University, New York, NY
b Department of Medicine, New York Presbyterian Hospital, College of Physicians & Surgeons, Columbia University, New York, NY
c Department of Pathology, New York Presbyterian Hospital, College of Physicians & Surgeons, Columbia University, New York, NY.
Received for publication June 14, 2006; revisions received October 7, 2006; accepted for publication October 25, 2006. * Address for reprints: Mario C. Deng, MD, FACC, FESC, Director of Cardiac Transplantation Research, Department of Medicine, Columbia University College of Physicians & Surgeons, New York, NY 10032. (Email: md785{at}columbia.edu).
| Abstract |
|---|
|
|
|---|
Methods: Polyurethane patch material was placed in the wall of the mouse descending aorta. Mice were then infected with Staphylococcus aureus 1 or 14 days after implantation. In vitro adhesion studies were conducted with polyurethane membranes coated with endothelial cells and membranes coated with fibrinogen.
Results: Mice were susceptible to infection in both dose- and time-dependent fashions. The patch material was significantly more susceptible to infection at day 1 than day 14. Immunohistologic and morphologic studies demonstrated that the CD31+ cells deposited on the membrane surface phenotypically appeared to be endothelial cells. In vitro adhesion studies of polyurethane membranes coated with endothelial cells showed them to be less susceptible to S aureus binding than were membranes coated with fibrinogen.
Conclusion: Textured polyurethane membranes are less susceptible to infection as cellular deposition occurs. The time frame within which these membranes become populated with cellular material is consistent with the time-dependent clinical incidence of infection. Cellular coating of polyurethane may provide a strategy for reducing the risk of infection.
| Introduction |
|---|
|
|
|---|
Staphylococcus aureus is among the most common of the pathogens that are associated with LVAD-related infections.3-6
We therefore elected to use S aureus to investigate the pathogenesis of LVAD-related endovascular infections, focusing on the initial colonization step in which bacteria adhere to the neointimal lining that forms on the surface of the textured HeartMate (Thoratec Corporation Pleasanton, Calif) LVAD polyurethane membrane. These membranes have already been shown to reduce the incidence of thromboembolic complications associated with LVAD implantation.7
To mimic the clinical situation, we developed an in vivo murine aortic polyurethane patch model.8
In this model, the blood-contacting surface of the membrane was coated with cellular material in a time-dependent manner. With this model, we aimed to establish an animal model of LVAD-associated infection and to determine the effect of cellular population on the susceptibility of the polyurethane patch to S aureus infection.
| Materials and Methods |
|---|
|
|
|---|
Animals
Male C57BL/6J mice from 3 to 4 months old (Jackson Laboratories, Bar Harbor, Mass) were maintained for at least a week on a light-dark (12 hour/12 hour) cycle at 25°C before the experiment. All procedures were performed according to protocols approved by the Columbia University Institutional Animal Care and Use Committee.
Patch Implantation Procedure
The polyurethane patch was implanted in the mouse abdominal aorta as described previously.8
Briefly, the mice were anesthetized by intraperitoneal injection of ketamine and xylazine. An intraperitoneal injection of heparin (300 IU/kg) was given before the procedure. The abdomen was opened through a midline incision, and the abdominal aorta and the inferior vena cava were exposed and carefully freed. The segment of the abdominal aorta between the renal arteries and the aortic bifurcation was temporally occluded with microvascular clamps (Roboz Surgical Instrument Co, Inc, Gaithersburg, Md), and the aorta was cut open longitudinally. A sterile, polyurethane membrane patch (HeartMate) was sewn onto the aorta with 10-0 nylon running suture around the margin of the patch under 25x magnification (Leitz Wild microscope M651; Wild Surgical Microscopes Co, Heerbrugg, Switzerland). After the vascular clamp was removed, pulsatile flow was confirmed at the distal part of the aorta. The operative field was irrigated with saline solution, and the skin incision was closed. Buprenorphine was given subcutaneously for postoperative analgesia as necessary. The animal was given 1 mL normal saline solution subcutaneously and put into a 36°C incubator for recovery. The entire procedure took approximately 45 minutes.
In Vivo Infection Model
S aureus (Wb strain), an endocarditis isolate, was used in this study.9,10
Susceptibility to infection was determined in a series of dose escalation studies. One day after patch implantation, the mice were lightly anesthetized and received a tail vein injection concentration of 106, 107, or 108 colony-forming units (CFU) of S aureus (0.2 mL). A blood sample was obtained to confirm bacteremia 1 hour after inoculation. After infection, the mice were kept in different cages to avoid cross infection. The mice were killed 3 days after infection. Liver, kidneys, spleen, and the polyurethane patch were excised for bacterial assay. The tissue was washed in phosphate-buffered saline solution, weighed, and homogenized in phosphate-buffered saline solution; aliquots of the homogenate were then plated onto mannitol salt agar plates (Becton Dickinson Microbiology Systems, Sparks, Md). Each plate was incubated for 24 hours at 37°C, and the number of colonies was determined. Results were expressed as the CFU per gram of tissue.
Susceptibility to infection was also assessed across time. These studies were performed to test the hypothesis that time-dependent changes in polyurethane topology alter the susceptibility to infection. The mice were infected either 1 day or 14 days after patch implantation. After infection, the mice were processed in the same manner as outlined for the dose escalation studies. Because dose escalation and time dependencies were our variables of interest in the context of this study, the time point at which the animals were killed (3 days) was kept constant.
Histologic Studies
After they had been humanely killed, mice infected with S aureus 1 day after polyurethane patch implantation were perfused with phosphate-buffered saline solution through a cardiac puncture in the left ventricle at physiologic pressure with an infusion pump. The aorta, including the patch, was excised from the body, embedded in OCT freezing compound (Sakura Finetek USA, Inc, Torrance, Calif), frozen, and stored at –80°C. Transverse sections (8 µm) from the middle of the polyurethane patch were stained with hematoxylin and eosin or Gram stain.
In Vitro Assay Measuring Effect of Endothelial Cell Coating of Polyurethane Membranes on S Aureus Adhesion
An in vitro assay was developed to assess further the role of endothelialization in reducing bacterial adhesion to the polyurethane membrane. Endothelial cells (Cambrex Corporation, East Rutherford, NJ) were grown to confluence in endothelial cell growth medium.9
Adhesion of S aureus to fibrinogen- versus endothelial cell–coated membranes was then compared with a modification of a previously described adhesion assay.9,11
LVAD membrane (diameter 0.5 cm) was immobilized on the bottom of a 96-well tissue culture dish (Becton, Dickinson and Company, Franklin Lakes, NJ). Fibrinogen (0.5 mg/mL; Sigma, St Louis, Mo) was added to the membrane-containing wells and incubated overnight at 37°C. The next day wells were washed, seeded with endothelial cells, and incubated in endothelial cell growth media at 37°C in 5% carbon dioxide for an additional 2 days.
S aureus (Wb strain) was grown overnight in Todd Hewitt broth (Becton Dickinson). The next morning, an aliquot from the overnight sample was diluted and reincubated for 3 hours to bring the suspension to logarithmic growth phase. The sample was then centrifuged and resuspended in 2 mL M199 Earle Salts (Invitrogen Corporation, Carlsbad, Calif), and the suspension was adjusted to an optical density at 600 nm of 0.01.
The membrane-containing microtiter wells were washed once with Medium 199 (Invitrogen). The staphylococcal suspension was vortex mixed, and 100 µL was added to the membranes coated with either endothelial cells or fibrinogen. The microtiter plate was incubated at 37°C for 1 hour. The bacterial suspensions were then aspirated and the wells washed with Medium 199 at 37°C for 5 minutes at 80 rpm. The bacteria and cells were lifted with trypsin–ethylenediaminetetraacetic acid (Invitrogen). The number of endothelial cells per well was determined with a hemocytometer. Aliquots (100 µL) of the trypsin-digested samples were plated onto heart infusion Agar (Becton Dickinson) and incubated overnight at 37°C. The colonies grown in the plates were counted, and the results were expressed as the number of bacteria per well. Each assay was performed with 12 wells per condition and was repeated four times.
Statistical Analysis
Data were compared with an unpaired Student t test.
| Results |
|---|
|
|
|---|
Dose-dependence assay
Five of 7 polyurethane patches obtained from the mice infected with 106 CFU were sterile. All samples obtained from animals inoculated at 107 CFU were infected (106 CFU vs 107 CFU 1.1 ± 1.9 vs 6.8 ± 1.2, P < .001; Figure 1, A). The kidney and liver homogenates were all infected, regardless of the concentration of the inoculation. One spleen sample in the 106 CFU group was sterile. The organ bacterial density was higher in groups infected with the higher bacterial inoculum (106 CFU vs 107 CFU): kidneys, 4.6 ± 2.1 versus 6.5 ± 0.6 (P < .05; Figure 2, A); liver, 3.4 ± 0.7 versus 6.2 ± 0.8 (P < .001; Figure 2, B); and spleen, 2.9 ± 1.2 versus 5.1 ± 1.6 (P < .002; Figure 2, C).
|
|
All tissue samples obtained from mice infected POD 1 (n = 7) showed infection, whereas only 2 of 7 samples from mice infected POD 14 (n = 12) were infected. The bacterial densities in the patches from mice infected POD 1 were significantly higher than in those from mice infected POD 14 (POD 1 vs POD 14 6.8 ± 1.2 vs 1.77 ± 2.7, P < .002; Figure 1, B).
There was no difference in the bacterial density of homogenates from the kidneys and livers between the two groups. The number of bacteria in the spleen was lower in the POD 14 group: kidneys, POD 1 vs POD 14, 6.5 ± 0.6 versus 5.6 ± 0.5 (P = .23; Figure 2, D); liver, POD 1 versus POD 14, 6.2 ± 0.8 versus 5.2 ± 2.0 (P = .14; Figure 2, E); and spleen, POD 1 versus POD 14, 5.1 ± 1.6 versus 4.0 ± 0.9 (P < .05; Figure 2, F).
Histologic Studies
One day after implantation, the polyurethane membrane was covered with a homogeneous fibrin layer (Figure 3, A, and Figure 4, A). Some of the fibers remained uncoated and in contact with the circulation. Polymorphonuclear leukocytes and mononuclear cells were attached to the surface as well as embedded within the fibrin (Figure 3, A). Sections obtained from an animal infected with 107 CFU 1 day after implantation and killed at 3 days showed infiltration of the aortic wall with polymorphonuclear leukocytes in addition to the findings in the noninfected animals (Figure 3, B). Gram-positive cocci were also observed in this tissue. Some of the bacteria were attached directly to the fibers of the polyurethane material, whereas some were seen embedded in the fibrin layer (Figure 3, C). At day 14, the blood-contacting surface of the polyurethane membrane was totally covered with CD31+ cells; the morphology, location, and CD31 staining all suggest that these were endothelial cells (Figure 4, B and C).
|
|
|
| Discussion |
|---|
|
|
|---|
Textured surface polyurethane material was originally developed for clinical use to minimize the thromboembolic events that were frequently seen in patients who received devices with a smooth blood-contacting surface. In comparison, devices with a textured blood-contacting surface (eg, the HeartMate device) had a decreased rate of thromboembolic events, presumably as a result of the formation of a neointima.7
Because the polyurethane membranes were less susceptible to infection on day 14, when the patch was coated with CD31+ cells, we hypothesize from results in this study that the neointima was also the basis for the protection against S aureus infection. The results from the in vitro assay provide further support for this hypothesis.
The reduced susceptibility to infection with time in our animal model is also consistent with the results of clinical studies of LVAD infections. These studies demonstrate that most bacterial infections occur within 2 months of device implantation.3,5,12
Although there are several potential reasons for this, the absence of a cellular lining to the polyurethane membrane is one likely contributing factor. In a related in vitro study, Arrecubieta and colleagues23
demonstrated that LVAD polyurethane membranes explanted from patients after 6 months were significantly less susceptible to S aureus adhesion than LVADs explanted earlier than 6 months.
The nature of the cellular lining of LVAD membranes evolves with time but has been found to include pluripotent CD34+ hematopoietic cells, endothelial cells, macrophages, fibroblasts, monocytes, and smooth muscle cells.13-16
These cells eventually establish a protective barrier, reducing exposure to the fibrinogen and fibrin deposited on the membrane surface shortly after implantation. As a result, the susceptibility to infection is reduced.
The cells lining the patch were CD31+ and had the morphologic appearance of endothelial cells. The identity of these cells as endothelial cells was not, however, confirmed by another immunohistochemical marker. The contribution of endothelial cells to the onset of endovascular infections is dichotomous. On the one hand, endothelial cells provide a continuous, protective barrier preventing access of bacteria to underlying tissue. In this study, the aortic polyurethane patch lined with CD31+ cells appeared less susceptible to infection than patch material coated with matrix molecules such as fibrinogen. Previous studies, however, have also demonstrated that S aureus is capable of direct adhesion to endothelial cells.11,17,18
This binding appears to primarily occur through the S aureus surface proteins fibronectin-binding proteins A and B. These proteins are part of a family of structurally related proteins designated microbial surface components recognizing adhesive matrix molecules.
19
These molecules are capable of adhering to matrix and cellular surfaces though a number of different receptor-ligand interactions.17,20
Our hypothesis is that fibrinogen, a serum component known to be deposited on prosthetic surfaces shortly after implantation, is more susceptible to S aureus adhesion than is a cellular layer. Reduction of exposure to fibrinogen by cellular deposition reduces S aureus adhesion. Several proteins mediate adhesion to fibrinogen, including the fibronectin-binding proteins and clumping factors A and B. The additional variable of different parameters of flow has to date received limited attention; however, it does appear to affect the adhesion interaction.21
In conclusion, in our polyurethane aortic patch model of infection, the blood-contacting surface is covered with CD31+ cells with time. In addition to preventing thromboembolism, this textured membrane therefore achieves an element of protection from bacterial infection by rapid cell coating. This provides a potential strategy for future devices in reducing the incidence of bacterial infections, which constitute a major complication of LVAD use.22,24