J Thorac Cardiovasc Surg 2007;134:657-662
© 2007 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
a Department of Thoracic and Cardiovascular Surgery of Johann Wolfgang Goethe University Hospital
b Frankfurt/Main, Germany; and Cardiovascular Engineering Group, Helmholtz Institute Aachen, Germany
Received for publication November 16, 2006; revisions received February 18, 2007; accepted for publication February 23, 2007. * Address for reprints: Farhad Bakhtiary, MD, Department of Thoracic and Cardiovascular Surgery, Johann Wolfgang Geothe University Hospital, Theodor-Stern-Kai 7, 60596 Frankfurt/Main, Germany (Email: farhad{at}bakhtiary.de).
| Abstract |
|---|
|
|
|---|
Methods: Edwards Perimount Magna (Edwards Lifesciences, Irvine, Calif) (n = 5) and Medtronic Mosaic Ultra (Medtronic Inc, Minneapolis, Minn) (n = 5) heart valves (23 mm) were investigated in an artificial circulation system (70 beats/min, cardiac output 5 L/min). Leaflet kinematics were visualized with a high-speed camera (3000 frames/sec). Valves were then exposed to a calcium-phosphate solution at a constant pulse rate of 300 beats/min for a total of 6 weeks. Repeated testing was performed after 1, 2, 3, 4, and 6 weeks of calcification. The calcification process might not be similar to in vivo performance.
Results: Initially, the Perimount Magna valves demonstrated lower pressure gradients compared with the Mosaic Ultra valves (9.7 ± 0.36 mm Hg vs 14.0 ± 1.16 mm Hg), but they showed higher closing volume and leakage flow. Total energy loss was equivalent after 1 week of calcification. Perimount Magna valves calcified significantly faster and more severely, leading to an increase in gradients and closure volume. Leaflet kinematics showed progressively longer opening and closing times for the pericardial valves (closing time Perimount Magna 135 ± 11 msec vs Mosaic Ultra 85 ± 9 msec after 6 weeks).
Conclusions: On the basis of visual inspection, despite the new ThermaFix (Edwards Lifesciences) tissue treatment, the Perimount Magna pericardial valves calcified in vitro faster and more severely than did the Mosaic Ultra porcine valves, which demonstrated a more constant performance throughout the calcification process. Leaflet kinematics showed a progressive prolongation of opening and closing times for pericardial valves, leading to higher closing volume.
| Introduction |
|---|
|
|
|---|
In vitro studies have some inherent limitations, such as nonphysiologic contraction of the artificial LV, with the consequence of an abnormal flow pattern through the aortic valve. In addition, only fresh biologic valve prostheses were investigated for hemodynamic performance in vitro, whereas changes in leaflet motion caused by progressive degeneration will contribute to the increased mortality after aortic valve replacement compared with the background population.8,9
The aim of the present study was to achieve a more complete view of biologic valve function. We investigated the hemodynamic performance and durability of 2 biologic valves (1 porcine and 1 pericardial prosthesis), especially focusing on leaflet kinematics and its consequences for valve opening and closure. Fresh and progressively degenerated valves were tested in an in vitro setup. The 2 selected valves represented the most recent supraannular modifications of well-known aortic valve substitutes aiming in maximization of EOA.
| Materials and Methods |
|---|
|
|
|---|
Five porcine (Mosaic Ultra, Medtronic Inc, Minneapolis, Minn) and 5 pericardial (Carpentier Edwards Perimount Magna, Edwards Lifesciences, Irvine, Calif) aortic valve prostheses, 23 mm in size, were tested in a previously described artificial circulation system (Figure 1).
10
|
Perimount Magna
This latest modification of the extensively studied Perimount valve is characterized by glutaraldehyde fixation, the new ThermaFix (Edwards Lifesciences) anticalcification treatment, and a true supraannular design to maximize EOA.
In Vitro Tests
At a constant hemodynamic level (cardiac output of 5 L/min, 70 beats/min), standard in vitro testing was performed (mean systolic pressure difference, EOA, closure volume, leakage flow), followed by calculation of systolic, diastolic, and total energy loss. Energy loss was defined as the time integral of the product of the mean systolic gradient and aortic flow during a defined interval such as systolic duration (systolic energy loss), closing interval (closing energy loss), or leakage duration (leakage energy loss). Energy loss values are expressed in the percentage of ventricular work load.
Leaflet kinematics were recorded using a high-speed camera (300 frames/sec). Three heart cycles were taped for further evaluation. The movies were evaluated for beginning and end of leaflet motion for valve opening (opening time) and beginning and end of leaflet motion for valve closure (closure time).
Valve Calcification
After the initial test run, all prostheses were incorporated into a pulse duplicator running at 300 beats/min. The valves were irrigated with a calcium-phosphate solution to imitate calcific valve degeneration. This in vitro method of calcification using calcium-phosphate at a physiologic pH and temperature has been used and investigated previously, demonstrating results comparable to in vivo valve degeneration.11,12
The higher pulse rate accelerates the calcification process as additional mechanical stress is applied to the leaflets.
During the first 4 weeks, valves were taken out of the pulse duplicator weekly to undergo photography, hemodynamic testing, and leaflet kinematics recording with the high-speed camera. The valves were subjected to an additional 2 weeks of calcification and then tested as noted above. The valves were then investigated for calcium uptake within the leaflet tissue.
Statistical Evaluation
For the statistical analysis of baseline characteristics and outcomes for the 2 valves, the Student t test and Fisher exact test were used. The results are reported as the mean ± standard deviation in text and Table 1.
|
| Results |
|---|
|
|
|---|
Hemodynamic Performance
Hemodynamic testing of the fresh and progressively degenerated valve prostheses revealed the following results.
Transvalvular gradients
Systolic pressure difference demonstrated significantly lower values for fresh Magna valves (mean 9.7 ± 0.36 mm Hg) compared with the fresh Ultra prostheses (mean 14.0 ± 1.16 mm Hg) (P < .05). This difference narrowed during the test period of 6 weeks and was not statistically significant at the 2-, 3-, and 6-week measurements (Figure 2).
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Most of the commercially available biologic valves consist either of porcine aortic leaflets or bovine pericardium. No randomized studies of patient outcome with large patient numbers comparing these 2 designs have been published, but a tendency toward lower transvalvular gradients for the pericardial valves has been reported.13,14
However, some groups showed that long-term survival seemed to favor the porcine prostheses.16,17
Factors other than pressure gradients and EOA may also contribute significantly to the clinical outcome after aortic valve replacement.
The present study aimed at gaining a detailed view of biologic valve performance with the special focus on valve closure and changes of leaflet kinematics with progressive calcification. The 2 selected valves were labeled 23 mm; however, studies comparing Perimount Magna with Medtronic Mosaic valves demonstrated smaller metric diameters for the Mosaic valve, leading to the ability to implant larger valve sizes in 28.4% for the porcine valve compared with only 8.3% for the pericardial valve.18,19
However, the authors decided to compare identically labeled valve sizes to avoid misunderstanding.
Our study confirmed the excellent systolic performance of the fresh Magna valve, but this advantage disappeared to a certain extent so that the difference was no longer significant after 2 weeks of in vitro calcification. This observation was explained by the fast initiation of the calcification process for the pericardial valves despite the new ThermaFix anticalcification treatment. The observed increase of systolic pressure differences was also observed in vivo for standard Perimount valves by Banbury and colleagues.16
EOA also favored the Magna prostheses. The large valve opening could be observed in the high-speed movies, with the pericardial leaflets moving completely out of the area of forward flow. However, this extensive leaflet motion during valve opening led to the delayed beginning of valve closure and significantly longer closure times. As a consequence, higher closure volumes were seen for the Magna prostheses, leading to an equivalent total energy loss. Valve incompetence was reported in vivo with transition from Grade 0 or Trace to Grade 1+, then to Grade 1+ or 2+, and finally to Grade 3+ or 4+ (14).
Mosaic valves showed higher transvalvular gradients and smaller EOAs, especially for fresh valves. The high-speed movies demonstrated a less-extensive leaflet motion compared with the Magna valves, so the systolic energy loss can be explained by the tissue and stent material within the forward flow area. On the other hand, this more physiologic valve opening led to fast and competent valve closure so closure volumes matched the results of native human aortic valves.18
The absence of regurgitation was also reported clinically for standard Mosaic valves.15
The porcine tissue treated with alpha-aminooleic acid resisted the in vitro classification method, so the hemodynamic results remained almost stable throughout the experiment. No valve malfunction could be observed.
On comparison of the 2 valve designs, one might summarize that the Magna pericardial valve was designed for maximal opening and thus matched the current opinion that systolic performance determines valve function and patient outcome.20-22
The price paid for maximizing the leaflet movement away from the coaptation zone is delayed and slow valve closure. If we consider valve closure as the primary function of a valve, the Magna valve failed to mimic physiology. It is not surprising that the use of native porcine valve leaflets in the Mosaic Ultra model led to leaflet motion closer to physiology. The need for stent mounting of the leaflets resulted in some restriction of forward flow leading to higher gradients (maximal difference of 4 mm Hg), but the leaflet motion allowed earlier and faster closure.
The main goal in aortic valve replacement is good quality of life for our patients and a long-term survival comparable to that of the background population. This can be achieved by a valve that imitates a healthy native human valve and avoids lesions that impact ventricular function (stenosis and regurgitation). First, the primary function has to be matched: an early, fast, and competent closure. Here, the porcine Mosaic Ultra valve demonstrated lower regurgitant flows compared with the pericardial prosthesis. Second, the valve opening has to be considered with favorable results for the pericardial Magna prosthesis. The investigation of leaflet kinematics using a high-speed camera delivered the explanation for the systolic and diastolic hemodynamic results. In regard to the susceptibility to calcification, the pericardial tissue seemed to show a tendency toward faster and more severe calcification regarding photographic documentation.
| Limitations |
|---|
|
|
|---|
The valves were removed from the calcifying solution for hemodynamic testing each week, so the calcification process might have been interrupted and thus affected the outcome. However, this was identical for all valves tested. A follow-up study with calcification for 4 weeks in a row is planned. Further investigation is needed to demonstrate the clinical relevance of our findings.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. von Knobelsdorff-Brenkenhoff, M. A. Dieringer, A. Greiser, and J. Schulz-Menger In vitro assessment of heart valve bioprostheses by cardiovascular magnetic resonance: four-dimensional mapping of flow patterns and orifice area planimetry Eur J Cardiothorac Surg, September 1, 2011; 40(3): 736 - 742. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Ruzicka, I. Hettich, A. Hutter, S. Bleiziffer, C. C. Badiu, R. Bauernschmitt, R. Lange, and W. B. Eichinger The Complete Supraannular Concept: In Vivo Hemodynamics of Bovine and Porcine Aortic Bioprostheses Circulation, September 15, 2009; 120(11_suppl_1): S139 - S145. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Bakhtiary, O. Dzemali, U. Steinseiffer, C. Schmitz, B. Glasmacher, A. Moritz, and P. Kleine Hydrodynamic comparison of biological prostheses during progressive valve calcification in a simulated exercise situation. An in vitro study Eur J Cardiothorac Surg, November 1, 2008; 34(5): 960 - 963. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bottio, V. Tarzia, G. Rizzoli, and G. Gerosa The changing spectrum of bioprostheses hydrodynamic performance: considerations on in-vitro tests Interact CardioVasc Thorac Surg, October 1, 2008; 7(5): 750 - 754. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. El Oakley, P. Kleine, and D. S. Bach Choice of Prosthetic Heart Valve in Today's Practice Circulation, January 15, 2008; 117(2): 253 - 256. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||