J Thorac Cardiovasc Surg 2007;134:1471-1476
© 2007 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
a Department of Diagnostic Radiology, Medical Physics, University Hospital Freiburg, Germany
b Department of Cardiovascular Surgery, University Hospital Freiburg, Germany.
Received for publication January 11, 2007; revisions received April 13, 2007; accepted for publication May 2, 2007. * Address for reprints: Michael Markl, PhD, University Hospital Freiburg, Department of Diagnostic Radiology, Medical Physics, Hugstetter Strasse 55, 79106 Freiburg, Germany. (Email: michael.markl{at}uniklinik-freiburg.de).
| Abstract |
|---|
|
|
|---|
Methods: A commercially available ventricular assist device was integrated into a magnetic resonance–compatible flow circuit. Flow-sensitive 3-dimensional magnetic resonance imaging was performed to measure time-resolved 3-directional flow velocities in the entire device. Advanced computer-aided 3-dimensional flow visualization methods were used to derive a comprehensive picture of flow dynamics within the ventricular assist device system.
Results: On the basis of the ventricular assist device model system, magnetic resonance imaging, and flow visualization, the first 4-dimensional functional magnetic resonance imaging analysis of flow characteristics inside an operating clinical routine ventricular assist device chamber system is reported. Detailed visualization of flow patterns and local changes in flow characteristics were successfully performed and revealed locally accelerated, vortical, and helical flow regions inside the geometry of the device.
Conclusions: Complex flow patterns such as vortex formation and locally accelerated flow demonstrate the potential of the presented method to further deepen the understanding of complex and regionally different flow characteristics inside ventricular assist devices.
| Introduction |
|---|
|
|
|---|
An ideal approach to investigate blood flow in a VAD would require direct assessment of flow characteristics during the in vivo setting. A potential solution may be offered by two- or three-dimensional (2D or 3D) time-resolved flow-sensitive magnetic resonance imaging (MRI) methods, which have been widely applied for the qualitative and quantitative assessment of in vivo blood flow.12-14
More recently, true 3D time-resolved MRI with 3-directional flow encoding (flow sensitive 4D-MRI) has been introduced.15,16
Such techniques permit the detailed analysis of dynamic blood flow characteristics within an acquired 3D data volume covering the entire thoracic aorta,17-19
the heart,20
or the peripheral arteries.21
However, implanted VADs are often not MR compatible or the patient setup does not allow MR acquisitions, and in vivo flow studies thus cannot be performed directly. For the understanding of flow characteristics within VADs to be enhanced, it is therefore of great interest to provide a realistic in vitro VAD model system.22
The experimental simulation and analysis of 3D flow patterns in such a system may have the potential to deepen the understanding of the development of flow patterns and evaluate functional consequences such as thrombus development.
To our knowledge, no detailed 3D flow analysis of VAD flow characteristics has been presented to date, and the actual flow patterns inside VADs in pulsatile flow mode are generally unknown. The overall aim of this study was to test the feasibility of an MR-compatible VAD model system for the assessment of regional 3D flow characteristics within the device. Thus, it was the purpose of this study to perform flow-sensitive 4D MRI measurements in combination with detailed 3D visualization for the assessment of 3D flow dynamics inside a VAD model system. Specifically, the inflow waveform at the inlet of the vascular model was adjusted to match the blood flow characteristics typically used in vivo by using an identical control and pump unit.
As a long-term objective, such model systems could also be used to optimize the design of VADs and to test different inflow conditions and their effects on the blood flow characteristics and VAD performance.
| Materials and Methods |
|---|
|
|
|---|
The VAD (chamber size = 60 mL) consisted of a pneumatically driven blood pump that may serve as support of the left or right ventricle as well as biventricular assistance for both ventricles. The pulsatile volume flow created by the control and pump unit was adjusted by two opposite permeable bulbs at the inlet or outlet side with inserted 3-leaflet valves similar to the natural aortic valve. The pressure wave was transmitted to the pump chamber by a double-layered membrane transferring the supplied periodic pressure changes into pulsatile blood flow through the VAD.
The pump and control unit consisted of a clinical routine device (MEDOS VAD-Driving Unit; MEDOS Medizintechnik AG) used for in vivo VAD operation, which was modified by use of long tubes (approximately 8 m) to connect the pump unit outside the MR room to the VAD circuit inside the MR scanner. Periodic pressure waveforms were transferred to the VAD circuit to generate pulsatile flow (systolic pressure = 180 mm Hg, diastolic pressure = –20 mm Hg, systolic fraction = 35% of flow cycle) within the VAD model system.
A reservoir with approximately 5 L of additional fluid was integrated into the flow circuit to simulate in vivo pressure conditions.
The fluid (water) was doped with a gadolinium-based contrast agent (Magnevist; gadopentetate dimeglumine, Schering, Germany) to maximize signal-to-noise ratio. A schematic illustration of the experimental setup is provided in Figure 1.
|
Data Analysis—3D Flow Visualization
After noise masking and eddy current correction,24
the measured velocity data were loaded into a commercially available software package (EnSight; CEI, Apex, NC) for 3D visualization of the flow characteristics inside the VAD system. The software permitted interactive 3D data navigation, definition of cut planes transecting the 3D data volume at any user-selected angle, and 3D visualization of the measured flow velocities as vector graphs and 3D streamlines.25
Vector graphs were generated on predefined 2D cut planes and illustrated the direction and magnitude of the measured flow velocities for different frames within the flow cycle.
The 3D streamlines represent lines locally parallel to the 3-directional flow velocity field for a given time-frame within the pulsatile flow cycle. All tracings were color-coded according to the measured local low velocities and originated from a set of equidistant grid points on an emitter plane, which was interactively positioned within the 3D imaging volume.
| Results |
|---|
|
|
|---|
|
The results for all 14 analyzed time frames within the flow cycle are shown in Figure 3. Similar to Figure 2, the magnitude data additionally reflect the state of the VAD chamber during the flow cycle. Higher signal intensities in the VAD chamber in frames 6 to 10 are consistent with minimal pressure and expanded VAD chamber volume.
|
A more detailed visualization of 3D flow patterns for two selected time frames is shown in Figure 4. During filling of the VAD chamber (Figure 4, A), high and accelerated flow at the inlet can clearly be appreciated. Moreover, the existence of a considerable flow vortex inside the expanded VAD chamber is clearly visible (white arrows).
|
Similar flow patterns were detected by vector graph visualization as depicted in Figure 5 for a time frame corresponding to the filling phase in Figure 4, A. Considerable vortex formation in the central VAD lumen (yellow arrows) as well as accelerated flow at the inlet (white arrow) are clearly visible.
|
| Discussion |
|---|
|
|
|---|
Flow-sensitive 4D MRI may thus have the potential to enhance the understanding of flow characteristics inside VAD systems with different size, geometry, and inflow conditions (eg, nonpulsatile versus pulsatile flow). Moreover, the detailed visualization flow patterns and/or local changes in flow velocities may help to identify vortical, helical flow regions inside the VAD geometry that are likely to develop thrombi or other functional deficits that may impair VAD performance.
It has to be noted, however, that thrombosis as a result of vortical or helical flow was not directly observed in this study. In our evaluation of the observed flow patterns inside the VAD, we hypothesize that such flow features may lead to thrombosis as described in the literature for similar in vivo flow patterns.26-30
Future studies should therefore include a more detailed analysis of parameters directly related to thrombogenesis, such as changes in wall shear forces that can be derived from the measured flow velocity vector fields.27,30
In addition, VAD model systems and experimental MR flow simulations in a controlled environment offer the opportunity to modify location, size, and shape of the VAD and optimize its performance with respect to generation of target flow characteristics. Moreover, MR flow analysis coupled with realistic VAD model systems has the potential to permit the detailed analysis of the influence of size and shape of different VAD geometries on local flow dynamics, for example, change in flow vortex formation as a function of VAD chamber size and shape. Larger studies may thus provide new information for the optimized design of VAD geometry and/or inflow conditions.
Limitations of this study include the use of water for all flow experiments. A more realistic experimental setup would require the use of human plasma or blood in combination with appropriate temperature control. Future studies using blood mimicking fluids or even real plasma or blood are thus warranted to analyze the influence of viscosity and density on flow dynamics inside VAD systems.
In addition, the inflow boundary characteristics may have been altered compared with in vivo conditions owing to the long tubes that were necessary to deliver the periodic pressure changes to the VAD system. Further studies are thus planned to understand the waveform variation of the propagating wave to permit a more precise generation of realistic inflow conditions.
The VAD used for this study demonstrated metal artifacts, for example, signal loss, in regions at the inlet and outlet.
The existence of such metal components limits the application of sensitive MRI for the assessment of flow characteristics inside all regions of the VAD. Therefore, no results could be obtained concerning the flow characteristics in the "sinus of Valsalva–type" recesses of the inflow and outflow cannulas that are originally designed to resemble nature in that respect. Clinically, these sinuses are frequently areas of thrombus formations (especially on the right side), leading eventually to the exchange of the pump chamber.
Nevertheless, metal components of the investigated VAD only partially obstructed the device, and flow patterns were successfully analyzed in large sections of the VAD, such as the pump chamber as well as the inlet and outlet. The metal components, which are used to support the artificial valve systems in the VAD, consist of simple rings mounted at the outer layer of the VAD at the level of the valve base. These can be removed and replaced by nonmetallic materials without any additional changes or modifications of the VAD chamber, valves, and flow channels. It is thus expected that the modified nonmetallic VAD will provide flow patterns identical to those of the original device. Future studies will therefore be performed with completely MR-compatible model systems that will provide insights into 3D flow characteristics within all regions of the VAD including flow patterns through and near the valve systems.
Once such optimized model systems are established, incremental changes in VAD design (eg, chamber geometry or valve design) and/or boundary conditions (eg, inflow pressures) can be used to systematically evaluate and optimize VAD operation by minimizing the extent of recirculation zones and vertical flow fields.
Although the MRI flow characterization reported in this article has not been compared with other flow analysis methods, additional measurement techniques or computational methods exist that have been used to assess flow within model systems.30-35
In addition to optical measurements (particle trace velocimetry)30,31
or ultrasound using gas bubbles,32
many studies using computational fluid dynamics33-35
have been performed. Analysis was typically based on model geometries or realistic vascular computer representations derived from angiography data. However, to our knowledge, no detailed analysis of 3D flow dynamics using full VAD coverage has been reported on the basis of these modalities. Specifically, no computational fluid dynamics study regarding flow patterns inside a VAD using realistic boundary conditions (ie, inflow waveforms and elasticity of VAD chamber) has been reported to date. Owing to the complexity of modeling realistic pump chamber performance using computer modeling, MRI may provide an advantage over computational fluid dynamics because the real VAD in combination with pump and control unit can be directly integrated into the 3D flow analysis process.
The methods used in this pilot study permitted the detailed evaluation of the 3D evolution of local flow patterns inside a routine clinical VAD. Results illustrate that flow-sensitive 4D MRI could successfully be used to analyze local and global flow dynamics in a realistic VAD model system.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. Benk, R. Lorenz, F. Beyersdorf, J. Bock, R. Klemm, J. G. Korvink, and M. Markl Three-dimensional flow characteristics in ventricular assist devices: Impact of valve design and operating conditions J. Thorac. Cardiovasc. Surg., November 1, 2011; 142(5): 1019 - 1026. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. Carlhall and A. Bolger Passing Strange: Flow in the Failing Ventricle Circ Heart Fail, March 1, 2010; 3(2): 326 - 331. [Full Text] [PDF] |
||||
![]() |
C. Heilmann, U. Geisen, C. Benk, M. Berchtold-Herz, G. Trummer, C. Schlensak, B. Zieger, and F. Beyersdorf Haemolysis in patients with ventricular assist devices: major differences between systems Eur J Cardiothorac Surg, September 1, 2009; 36(3): 580 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Geisen, C. Heilmann, F. Beyersdorf, C. Benk, M. Berchtold-Herz, C. Schlensak, U. Budde, and B. Zieger Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease Eur J Cardiothorac Surg, April 1, 2008; 33(4): 679 - 684. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||