J Thorac Cardiovasc Surg 2008;135:436-437
© 2008 The American Association for Thoracic Surgery
Brief Communication |
a Nebraska Heart Institute, Lincoln, Neb
b Cardiac Thoracic and Endovascular Therapies, Peoria, Ill
Received for publication July 31, 2007; accepted for publication September 18, 2007. * Address for reprints: James H. Wudel, MD, Nebraska Heart Institute, Lincoln, NE 68506. (Email: jwudel{at}neheart.com).
One of the ongoing challenges associated with the size of devices used in thoracic aortic endografting is arterial access. Intraoperative conduits are frequently required,1
or patients are excluded from endovascular repair because their anatomy precludes delivery of the device.2
Artery size, artery tortuosity, and previous aortic stent grafting all may complicate the retrograde endovascular navigation required to reach the target site safely. As an alternative, we describe the technique of placing a prosthetic vascular conduit on the right axillary artery to facilitate antegrade deployment of a thoracic endograft in a patient with an enlarging descending thoracic aneurysm, multiple comorbidities, and a previously placed infrarenal endograft.
A 79-year-old woman was seen with an enlarging descending thoracic aortic aneurysm measuring 7.5 cm. She had a history of hypertension, obesity, and an abdominal aortic bifurcated endograft (Gore Excluder; W. L. Gore & Associates, Inc, Flagstaff, Ariz). Preoperative aortography showed a type C arch without extracranial great vessel disease. Multiplanar computed tomography showed a suitably placed abdominal endograft within tortuous aorta and iliac arteries (
Figure 1). Her anatomy was considered suboptimal for retrograde deployment because of the iliac vessel size and tortuosity and because of the previous abdominal endograft. Antegrade deployment through a right axillary conduit was therefore performed.
|
Gore TAG grafting was then performed antegrade from just distal to the subclavian artery to just proximal to the celiac artery with a total of three overlapping endoprostheses (
Figure 2). Completion arteriography showed no endovascular leakage. The vascular graft was amputated and oversewn near its base. The patient awakened without any neurologic deficit and was discharged 5 days later. Two months after the operation, she continues to do well, with postprocedure computed tomography showing no clinical abnormalities.
|
The relatively large diameter of the thoracic aorta and the technical limits of catheter-mounted endografts make vascular access a critical part of the preoperative evaluation in stent graft candidates. Small iliac vessel size and tortuosity complicate the manipulation of a large sheath retrograde to the thoracic landing zone. In a recent series of 126 patients undergoing thoracic stent graft evaluation, inadequate vascular access accounted for nearly a third of the anatomic exclusions.2
In the first pivotal trial of an approved device in the United States, 15% of patients required an intraoperative access conduit.1
In addition, navigating a large sheath retrograde has the potential to disrupt a previously placed abdominal stent graft. All these factors contributed to the evaluation of an alternate access site in this case.
Access sites for deployment of thoracic endografts through either common carotid artery have been reported.3-5
Potential advantages of using the right axillary artery with a conduit include (1) direct avoidance of repeated manipulation of the carotid artery, with its potential for vessel injury, atheroembolism, and cerebral hypoperfusion, (2) ability of the extracorporeal graft to cover most of endograft sheath length and allow a lower intravascular device profile, and (3) provision by the right axillary femoral wire of improved trackability for an antegrade delivery through disadvantaged iliac arteries. Suitable arch anatomy must be confirmed preoperatively for this approach. In addition, this approach may not be suitable for other thoracic device models without modification.
Thoracic endografting continues to evolve and is being applied for increasingly complex cases. Accessory access sites and adjunctive maneuvers in these cases are vital. The appreciation that brachiocephalic vessels may be used for access will facilitate the use of these devices in the treatment of diseases involving not only the descending thoracic aorta but also the ascending aorta, root, and aortic valve.