|
|
||||||||
J Thorac Cardiovasc Surg 2008;135:1081-1086
© 2008 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
a Department of Cardiovascular Surgery, University Hospital Schleswig-Holstein, Campus Kiel, Kiel, Germany
b Institute for Microtechnology IMT, University of Braunschweig, Braunschweig, Germany
c Institute of Anatomy, Christian-Albrechts-University of Kiel, Kiel, Germany
d Institute of Biomedical Optics, University of Luebeck, Luebeck, Germany
Received for publication May 28, 2007; revisions received August 28, 2007; accepted for publication November 26, 2007. * Address for reprints: Georg Lutter, MD, PhD, Department of Cardiovascular Surgery, University Hospital Schleswig-Holstein, Campus Kiel, Arnold-Heller-Strasse 7, D-24105 Kiel, Germany. (Email: lutter{at}kielheart.uni-kiel.de).
| Abstract |
|---|
|
|
|---|
Material and Methods: After anterolateral minithoracotomy, the aortic valve isolation chamber system was inserted into the descending aorta and pushed forward transluminally into the aortic position to generate a separate operation space between the subvalvular and the proximal ascending aortic area. After deployment and sealing of the chamber, stable function with a continuous chamber lavage of 1.58 L/min saline solution was established (8/10 cases). The endoscopically guided laser fiber was delivered via the right carotid artery. After fixation of a leaflet by a forceps catheter, the native leaflets were resected each by a thulium:YAG laser with 20-W power rating. Macropathology and micropathology of surrounding anatomic structures were analyzed.
Results: The duration of transluminal positioning and deployment of the aortic valve isolation chamber took 7.3 ± 5.8 minutes. Fluoroscopy confirmed sealed chambers. The resection was completed in all leaflets and took, on average, 6.0 ± 3.5 minutes per leaflet. The aortic wall was moderately injured in 4 of 10 cases and the aortic annulus in two cases with one aortic wall perforation. The surrounding tissue, the coronary ostia, the mitral valve, and the left ventricular outflow tract remained unaffected.
Conclusion: This study demonstrates the feasibility of endovascular resection of human aortic valves in situ. This is a subsequent step toward complete percutaneous replacement (resection and implantation) of human aortic valves.
| Introduction |
|---|
|
|
|---|
|
Fifteen years ago, the pioneers of percutaneous valve replacement published their experimental studies and showed the feasibility of this technology for the implantation site.1,2
Five years ago, the cardiologist Cribier and his group performed the first human percutaneous aortic valve implantation.3
This was the onset of the upcoming percutaneous technique for interventional aortic valve implantation as a preliminary clinical method. Currently, three transluminal techniques are in clinical testing: the percutaneous approach from the antegrade,3
retrograde,4
and transapical sites of the aortic valve.5,6
For a true replacement of the diseased valve to be performed percutaneously, the native aortic valve must first be percutaneously resected before the new valve is implanted.
All current experimental and clinical studies of percutaneous aortic valve implantation reported this profound problem: mild to severe paravalvular leakage.3,4,7
In addition, occlusion of coronary ostia,8,9
embolization of debris,10
migration of the valved stent,9,11
mild increase of mitral insufficiency,4
and a too small effective aortic valve area12
have been observed.
As has been pointed out since 2002, percutaneous resection of the diseased and calcified valve will be necessary if a safe and morphology-oriented consecutive implantation is to be performed in patients who require a curative approach. This method will come closer to the high conventional surgical standard.12
After a variety of resection methods had been tested, endoscope-guided resection of aortic valves could be successfully undertaken in 2005 with a high-pressure water-stream scalpel in an in vitro model. An aortic valve isolation chamber (AVIC; with support of STI Deutschland GmbH, Henstedt-Ulzburg, Germany) has been additionally developed to avoid the escape of any debris during valve excision, consequently preventing central or peripheral embolization.13
The aim of this study was to analyze the laser-assisted endovascular resection of sclerotic human aortic valves in situ in 10 human cadavers using a thulium:YAG laser with a 20-W power rating.
| Material and Methods |
|---|
|
|
|---|
|
|
The pressure of AVIC balloons was monitored online to prevent unnoticed pressure decrease caused by leakage with a subsequent unstable chamber function. A permanent lavage of 1.58 L/min was established by a rotary pump (Stockert Instruments GmbH, Munich, Germany) to ensure a clear view during the resection process. Additionally, the debris was washed out.
To facilitate insertion of AVIC and prevent blood leakages during the resection, we developed a special port system for temporal implantation into the distal descending aorta during the procedure (Figure 1, C).
Aortic Valve Resection Tool
The mild-to-severely sclerotic aortic valves were resected by a diode-pumped continuous-wave thulium:YAG laser (ITL 2000, Lisa Laser Products OHG, Katlenburg-Lindau, Germany) emitting at a wavelength of 2.01 µm with 20-W power rating. Laser power was transmitted via a commercially available 365-µm quartz fiber (core Ø, CeramOptec GmbH, Bonn, Germany). The fiber was controlled by a flexible endoscope (Richard Wolf GmbH, Knittlingen, Germany) with an outer diameter of 7.5F and a length of 600 mm. The front part of the endoscope (40 mm) was operated by manual control.
The endoscope was inserted via an additional port system (8F) through the right carotid artery.
Human Preparations
Before death, these subjects gave written consent to the Institute of Anatomy of the Christian-Albrechts University in Kiel to donate their bodies for research (n = 10; age, 73 ± 10 years; gender, 5 female, 5 male; weight, 66.4 ± 10.2 kg; height, 169.7 ± 7.6 cm). Our study was approved by the Ethics Committee of the University of Kiel from November 24, 2004 (D 434/04). This study was realized in agreement with the basic principles of ethics (Basic Constitutional Law of the Federal Republic of Germany,
1).
AVIC Deployment
For AVIC installation, two approaches were required: a thoracic aortic access (descending aorta) and a right common carotid artery access. Therefore, a left-sided lateral minithoracotomy was performed (ICR [intercostal space] 4/5). The descending aorta was prepared with two purse-string sutures. To optimize AVIC installation, we implanted the port system into the descending aorta. Then, the AVIC was positioned via a guide wire into the aortic valve position (Ø 0.035 inch; Terumo Europe Interventional Systems, Guyancourt Cedex, France). The subvalvular balloons were positioned below the aortic valve and the supravalvular balloon was placed between the brachiocephalic artery and the left common carotid artery (Figure 2, A). After inflation of the subvalvular and supravalvular balloons under fluoroscopic and endoscopic control, the resection chamber system was filled with saline solution.
AVIC Testing
Three pressures were separately monitored online: the subvalvular balloon pressure, the supravalvular balloon pressure, and the resection chamber pressure. In case of balloon or chamber leakage, the experiment would have been stopped and the chamber integrity re-established. After the AVIC had been filled with water, the inflow and outflow of the chamber was controlled to obtain a stable chamber lavage (flow of 1.58 L/min). In addition, the in-chamber pressure was measured to adjust the chamber perfusion.
Endovascular Aortic Valve Resection
The experimental setup is shown in Figure 2, A. The endoscope was controlled by a cardiac surgeon. Three assistants handled the forceps catheter, endoscopic visualization, chamber perfusion, and pressures of the AVIC (balloons and chamber).
The endoscope with the in-lying laser fiber was placed in the AVIC via the right common carotid artery. The forceps catheter was inserted via the AVIC system and the visualization was also realized via the AVIC system. The aortic leaflets, all moderately sclerosed, were gripped and held in place (one by one) with the forceps catheter (Ø 1.3 mm, Richard Wolf). The resection was performed with a continuous-wave laser source (wavelength at 2.01 µm) with a distal power level at 20 W transmitted via quarz fiber (365-µm core diameter). The resection process was video controlled (with a 30-inch television) and digitally recorded. The leaflets were excised 2 to 3 mm from the insertion of the annulus (Figure 2, B and C). Fragments of the resected leaflets were removed through the AVIC.
Analyzed Parameters
The procedure time (minithoracotomy and dissection of the aorta and the right carotid artery), deployment time for the port, and installation time for the AVIC were recorded (in minutes). The handling of the port, the AVIC, and the auxiliary tools (guidance, deployment, fixation, visualization, and removal) was analyzed. The sealing of the AVIC was verified with balloon- and chamber-pressure monitoring (10/10) and fluoroscopy (5/10). Visualization was accomplished endoscopically.
The duration of the laser resection required for one leaflet (minutes), size of the resected valve area (cm2), size of the excised fragments (cm2), resistance of the AVIC balloons, and any lesions observed in the surrounding tissue within 40 mm of the annulus were protocoled: macropathology and micropathology were analyzed (hematoxylin–eosin staining). The laser handling (resection accuracy, fiber flexibility, and usability of the laser source) were analyzed. The use of the laser-guiding endoscope and the forceps catheter (guidance, flexibility, usability, and operating accuracy) were analyzed. All data are expressed as mean ± SD.
| Results |
|---|
|
|
|---|
AVIC Testing
AVIC testing showed the following pressures on average: constant chamber pressure in 7 of 10 cases, subvalvular balloon pressure 182.8 ± 52.3 mm Hg, and supravalvular pressure 179 ± 82.2 mm Hg. The sealing efficiency was analyzed.
Resection Process
The resection process took a mean of 6.0 ± 3.6 minutes per cusp. Macropathology and micropathology demonstrated superficial lesions and one complete perforation of the aortic wall (Figure 2, D). The mitral valve, left ventricular outflow tract, and endomyocardium remained unaffected.
Debris Analysis
The leaflets were resected and removed in toto. On average, the excised leaflet size was 1.4 ± 0.5 cm2.
Evaluation of AVIC and the Resection Tools
Visualization was adequate to display the structures of interest during the entire experiment. The endoscope allowed acceptable guidance of the laser fiber during resection. Operating the forceps catheter was difficult owing to lack of steerability. The forceps possesses only open-and-close function and no direction control. Although the AVIC was pretested successfully in numerous preliminary tests, a few construction problems occurred: dismantling of bonding surfaces (3/10), twisting of tubes (1/10), and leakage of subvalvular balloons (3/10).
The supravalvular balloon remained stable in all cases (10/10). One of the two subvalvular balloons ruptured in two cases.
| Discussion |
|---|
|
|
|---|
AVIC
An AVIC must be used to avoid the embolization resulting from calcified particles. This catheter-based system was easily deployed and the chamber could be sealed in all cases. The ongoing enhancement of this system is moving toward minimization and more flexibility for in vivo deployment.
Resection
The results demonstrate that endoluminal resection with percutaneously inserted instruments is feasible. Except for the perforation of the aortic wall seen in one case, the macroscopic and microscopic analysis showed only mild damage. However, efforts must be undertaken to improve the tractability and the preciseness of the resection tools and to shorten the required resection time. In the case of highly calcified aortic valves, the implemented continuous laser was not effective. However, it has already been demonstrated that these human calcified valves can be cut by a high-pressure water-stream scalpel.13
The development of a newly designed, more flexible, smaller water scalpel is ongoing as is the enhancement of a suitable laser source.
Although research on the optimal valve substitute is still ongoing, endovascular techniques have recently been proposed as alternative therapies for both aortic and mitral disease. In contrast to robotics, the percutaneous valve techniques are still simpler than surgery. The opportunity to have an aortic valve replaced (resected and implanted) without sedation by a vascular approach is undoubtedly appealing to the patient.
The latest data showed very good procedural and short-term results for pulmonary valve implantation14
and satisfactory results for percutaneous aortic valve implantation.6
Nevertheless, despite the successes with percutaneous aortic valve implantation, optimism should be tempered. The aortic valve implantation series presented by Webb and coworkers4
evidenced some major drawbacks, such as paravalvular leakage (in 13/14 cases), which is caused by the persistence of empty space between the percutaneously implanted valve and the native valve owing to calcifications. This was observed in the majority of patients. Moreover, an increase in mitral insufficiency (from pre-procedural +2 to +3 of 4 grades after the procedure) has been observed on average in the overall patient cohort. In addition, a coronary flow obstruction caused by a bulky calcified leaflet that was displaced by the valved stent occurred in 1 patient and atheroembolism of calcic debris during the necessary valvuloplasty and the positioning of the device happened in another 2 patients (1 stroke).4
In fact, some intraprocedural complications, such as ejection of the valved stent into the ascending aorta at the time of balloon inflation, have been reported a few times by two groups.4,15
Hemodynamic collapse after balloon predilation also occurred.15
One could argue that this is all due to the stainless steel stent, which is often used for valved stent implantation. The option of using a valved nitinol stent has resulted in even more cardiovascular events and a higher comorbidity.16
Finally, several technical difficulties, that is, inability to cross the stenotic valve or unsuccessful prosthesis deployment, have been described for both the antegrade and retrograde approaches,3,4,17
so that transapical access has been discussed.18
Even if the device-related complications and the morbidity of this implantation procedure in these high-risk patients is reduced over time (Webb's group4
already has fewer problems than the first data reported by Cribier and associates3
), the fundamental problems will not be easily overcome: the necessary balloon valvuloplasty (with its own high mortality and especially morbidity rate) and the foreseen short-term durability of the implanted pericardial valves (dysfunction after high pressurized compression and re-expansion) are huge problems.
In contrast to the percutaneous aortic approach, transapical implantation seems to be more favorable. This direct approach allows better valved stent deployment and has a lower complication rate, especially in patients with generalized atherosclerosis. The short-term results are well accepted clinically.5,6
Therefore, evaluation of this approach is ongoing and remains encouraging.
In 2007, the device was optimized by numerous improvements, and this has also significantly improved the patient's outcome. New stents have been developed, from balloon-expandable19
to self-expanding stents,20
and new applicator systems have been designed for an antegrade,21
retrograde,4
or transapical approach.5
In addition, the preinterventional screening of the patients became more important owing to the fact that calcified peripheral and central vessels and highly calcified aortic valves dramatically increase the risk of periprocedural and postprocedural complications, especially mortality.22-24
To treat patients with moderately to highly calcified aortic valves percutaneously and at the same time adhere as much as possible to the surgical gold standard, the valve must be resected before the new valve can be implanted.11
However, it appears from the present experience that is important to leave an anchoring area (annulus plus a small rim of calcified leaflet) for the fixation of the new valved stent.
In a recent in vitro study, Flecher and associates8
demonstrated that coronary flow was significantly reduced after implantation of a valved stent owing to coronary ostial irritation by the left-in-place native leaflets. After removal of these leaflets, the flow reduction was no longer present. Earlier in vivo studies showed diametrically different results: no interference concerning coronary ostia and the native valve appeared after valve implantation25
versus complete ebbing of myocardial perfusion owing to stent position.26
Limitations
Owing to the experimental appendage, the presented system is not yet ready to be used in vivo, but the results will pave the way for future enhancement. For instance, in the next study, the coronary arteries will be blocked by cardioplegia catheters to prevent any debris from entering the ostia.
The improvement of the resection chamber technology is ongoing, especially the sealing method for a beating heart modus.
The described approach via the distal descending aorta and the carotid artery must be combined into a single approach, which will be reached after minimization of the overall resection catheter.
Currently, on-pump aortic valve replacement is carried out with very low mortality and acceptable morbidity in the vast majority of patients.24
Very few cases are rejected owing to too high a risk from the presence of significant comorbidity. Hence, this procedure remains the gold standard for the treatment of aortic valve stenosis and/or regurgitation.
Future Prospects
Extensive research is still required to achieve clinical off-pump aortic valve resection.27
In this study, we have shown the feasibility of endovascular aortic valve resection in a human preparation. The next step will be to resect the aortic valve under cardiopulmonary bypass support in an arrested heart with cardioplegic myocardial arrest. The end goal is to perform the resection with blood flowing through a central lumen of the resection chamber in a beating heart, without cardiopulmonary bypass.
Currently, novel microsystemic technology (actuator-based) for improved guidance of the resection tool is under construction. In our future experiments, only one ablation device containing all instruments will be applied for the transapical access and later for the percutaneous approach.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. Astarci, D. Glineur, G. Elkhoury, and B. Raucent A novel device for endovascular native aortic valve resection for transapical transcatheter aortic valve implantation Interact CardioVasc Thorac Surg, January 9, 2012; (2012) ivr125v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Bombien, L. Lozonschi, J. Cremer, and G. Lutter Transcatheter valve replacement: Resection and valved stent implantation in a beating heart J. Thorac. Cardiovasc. Surg., August 1, 2010; 140(2): 477 - 479. [Full Text] [PDF] |
||||
![]() |
S. Salizzoni, P. Bajona, K. J. Zehr, W. D. Anderson, S. Vandenberghe, and G. Speziali Transapical off-pump removal of the native aortic valve: a proof-of-concept animal study. J. Thorac. Cardiovasc. Surg., August 1, 2009; 138(2): 468 - 473. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Bombien, M. Appel, T. Attmann, G.-R. Klaws, M. Schunke, C. Hass, J. Cremer, and G. Lutter Percutaneous aortic valve replacement: gross anatomy and histological findings after transapical and endoluminal resection of human aortic valves in situ, Eur J Cardiothorac Surg, July 1, 2009; 36(1): 112 - 117. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The Year in Valvular Heart Disease J. Am. Coll. Cardiol., May 19, 2009; 53(20): 1894 - 1908. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |