J Thorac Cardiovasc Surg 2008;136:280-282
© 2008 The American Association for Thoracic Surgery
Expert Commentary |
The Cleveland Clinic Foundation, Cleveland, Ohio
Received for publication May 27, 2008; revisions received May 27, 2008; accepted for publication May 29, 2008. * Address for reprints: A. Marc Gillinov, MD, The Cleveland Clinic Foundation, Department of Thoracic/Cardiovascular Surgery, Desk F25, 9500 Euclid Avenue, Cleveland, OH 44195. (Email: gillinom{at}ccf.org).
| Introduction |
|---|
|
|
|---|
| Key Recommendations |
|---|
|
|
|---|
Cardiac Resynchronization Therapy
The need for optimization of medical therapy before CRT implantation is emphasized. CRT (with or without an ICD) is indicated (class IA recommendation) for patients who are in New York Heart Association (NYHA) class III or ambulatory class IV with optimal recommended medical therapy and who have a left ventricular ejection fraction (LVEF) less than or equal to 35%, a QRS duration greater than or equal to 0.12 seconds, and sinus rhythm. Similar patients with atrial fibrillation may also benefit from CRT. In patients meeting the aforementioned criteria, CRT has been associated with improved LV systolic function, favorable cardiac remodeling, and reduction in functional mitral regurgitation, heart failure symptoms, hospitalizations, and death. However, no large clinical trials have demonstrated the benefits of CRT for patients without prolongation of the QRS interval (>0.12 seconds). CRT is not indicated for asymptomatic patients with reduced LVEF in the absence of other indications for pacing.
Implantable Cardioverter Defibrillators
ICDs are recommended in specific circumstances for primary and secondary prevention of sudden cardiac death; secondary prevention refers to the prevention of sudden cardiac death in patients who have survived a previous cardiac arrest or sustained ventricular tachycardia. In such patients, ICD therapy improves survival. ICD recommendations apply only to patients receiving optimal medical therapy with a reasonable expectation of survival with good functional capacity for more than 1 year. Of particular relevance to surgeons, ICD therapy is indicated in the following groups of patients:
Surgical Considerations
The guidelines recognize that surgeons, with appropriate training, may participate in device-based therapies of cardiac rhythm abnormalities and heart failure. Surgeons may be involved with the transvenous or epicardial placement of leads and with the insertion of related generators and hardware. Surgeons most frequently engage in the placement of epicardial pacing leads.
Surgically placed epicardial pacing leads are indicated in selected instances when standard transvenous lead placement is not feasible or contraindicated. Examples of such circumstances include 1) the inability or failure to place an adequate LV lead in patients requiring biventricular pacing, 2) indications for permanent pacing in certain pediatric patients and in pediatric or adult patients with tricuspid valve prostheses or recurrent or prolonged bacteremia, and 3) congenital or acquired venous anomalies that preclude transvenous access to the heart.
Biventricular Pacing (Cardiac Resynchronization Therapy)
The reported success rate of coronary venous lead implantation for CRT ranges from 81% to 99%. Causes of failed percutaneous lead placement may be anatomic (superior vena cava or coronary sinus obstruction or inadequate coronary venous anatomy) or technical (failure to cannulate the coronary sinus, coronary sinus dissection, inadequately high pacing thresholds with intermittent capture, diaphragmatic pacing because of proximity of the phrenic nerve to the target coronary sinus branch, or lead dislodgement). When coronary sinus lead implantation fails, several nonrandomized studies have demonstrated that surgical LV lead placement is almost always successful. The key advantage of surgical lead placement is access to the entire posterior and lateral walls of the left ventricle, which enables choice of the best pacing site. The combination of echocardiography with tissue Doppler imaging and electrophysiologic measurements may facilitate the choice of a transthoracically directed LV epicardial pacing site. Placement of 2 epicardial leads may be considered to provide backup capability if one lead should fail or become dislodged. Steroid-eluting epicardial leads are generally preferable to screw-on leads.
The choice of surgical approach for LV epicardial leads seems to influence hospital morbidity. Surgical approaches for the placement of LV epicardial leads include left thoracotomy, left thoracoscopy, and robotically assisted port-based placement. Thoracotomy in frail patients with heart failure has been associated with bleeding, stroke, hypotension, and arrhythmias. In contrast, thoracoscopic and robotic approaches have been reported to be associated with minimal morbidity and are preferred. These less-invasive procedures generally require 60 to 90 minutes of operative time and a mean hospital stay of 4 to 5 days. However, not all patients are candidates for minimally invasive or robotic procedures. Patients who have undergone a previous thoracotomy or sternotomy may have limited pericardial/epicardial access.
In certain instances, it may be advisable to place an LV epicardial lead at the time of concomitant cardiac surgery. In patients who are currently or in the future may be candidates for CRT who require coronary artery bypass grafting or mitral valve surgery and who have medically refractory, symptomatic heart failure, ischemic or dilated cardiomyopathy, prolonged QRS interval, LV end-diastolic diameter greater than or equal to 55 mm, and ejection fraction less than or equal to 35%, the surgeon may elect to place an LV epicardial lead. The lead is tunneled to a prepectoral pocket for intraoperative or postoperative attachment to an appropriate pacing generator. This approach is probably not indicated for the patient who is expected to have substantial improvement in LVEF after cardiac surgery (eg, a patient with extensive viable myocardium undergoing revascularization). There are limited data documenting outcomes of this "preemptive" strategy.
Pediatric Patients
Epicardial leads may be necessary in some pediatric patients. The most common indications for permanent pacemaker implantation in the pediatric population are sinus node dysfunction or atrioventricular block after surgery for congenital heart disease or congenital atrioventricular block. In most instances, such pacing systems can be placed using standard transvenous techniques. However, epicardial leads may be needed in children as a result of their small size, the presence of congenital heart defects with a right-to-left shunt, or an inability to pace the chamber desired because of anatomic barriers (eg, after a Fontan procedure). In such instances, steroid-eluting leads provide excellent durability.
Bacteremia
Epicardial leads are suggested in some pediatric or adult patients who need pacing and who have recurrent or prolonged bacteremia. For a single episode of device-related bacteremia, extraction of all hardware followed by reimplantation by the transvenous route at a later date is appropriate.
Tricuspid Valve Surgery
Implantation of permanent epicardial pacing leads is indicated in the pacemaker-dependent patient undergoing tricuspid valve replacement. A mechanical tricuspid valve represents an absolute contraindication to placement of transvenous right ventricular leads, because such leads will cross the valve and interfere with valve function. This scenario occurs most commonly in the patient with tricuspid valve endocarditis and a transvenous pacemaker. At surgery, all hardware should be removed. If the tricuspid valve is reparable, standard transvenous pacing leads can be placed postoperatively. However, if tricuspid valve replacement is necessary, epicardial ventricular leads should be implanted at the time of surgery.
Conclusions
With appropriate training, surgeons can place both endovascular and epicardial leads and corresponding generators. The newly issued American College of Cardiology/American Heart Association/Heart Rhythm Society 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities1
provide detailed, well-referenced recommendations to guide patient selection and device type.
| References |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |