J Thorac Cardiovasc Surg 2008;136:419-423
© 2008 The American Association for Thoracic Surgery
Evolving Technology |
a Department of Cardiovascular Surgery, Institut del Tòrax, Hospital Clínic, University of Barcelona, Barcelona, Spain
b Department of Pathology, Hospital Clínic, University of Barcelona, Barcelona, Spain
Received for publication December 13, 2007; revisions received February 12, 2008; accepted for publication March 13, 2008. * Address for reprints: Manuel Castellá, MD, PhD, Department of Cardiovascular Surgery, Hospital Clínic, C/Villarroel 170—08036 Barcelona, Spain. (Email: mcaste{at}clinic.ub.es).
| Abstract |
|---|
|
|
|---|
Methods: Nine explanted fresh human hearts were studied, two of them with left coronary dominance. Two types of bipolar radiofrequency clamps were positioned to reach the mitral and tricuspid annuli, and relationships within the atrioventricular junction were analyzed, including coronary sinus and coronary arteries.
Results: In all hearts studied, the coronary arteries and veins within the adipose tissue of the right or left atrioventricular groove lay in the atrial side, 3 to 18 mm away from the mitral or tricuspid annuli. When the bipolar radiofrequency clamp was closed toward the mitral annulus, the coronary sinus was always included between the jaws, and in left coronary–dominant hearts, the circumflex artery was also included. Nevertheless, the clamp never reached the annulus owing to the increase in thickness of the adipose tissue around the groove and the ventricular mass, leaving 5 to 10 mm of atrial myocardium free from the radiofrequency electrodes. In the right atrium, clamp placement toward the tricuspid annulus excluding the right coronary left 8 to 18 mm of atrial muscle free from the bipolar electrodes.
Conclusions: Bipolar radiofrequency clamps are not sufficient to complete a Cox maze IV procedure. Moreover, they may compromise coronary arteries in patients with left coronary dominance. Lines to the atrioventricular annuli need to be completed with the cut-and-sew technique or with alternative monopolar energy devices.
| Introduction |
|---|
|
|
|---|
Despite the efficacy of the Cox maze III operation, the surgical complexity of the procedure has limited its use and few long series have been reported. Recent introduction of alternative energy devices that can create thermal lesions in a relatively short time and avoid long atrial sutures has impulsed this operation again, with similar conduction block lines as described by Dr James Cox, termed now as the Cox maze IV procedure (
Figure 1).
|
A recent report has suggested inferior efficacy of the Cox maze operation when performed with bipolar radiofrequency than with the cut-and-sew technique.4
In this study, bipolar ablation was combined with cryolesions to complete the lines toward the mitral and the tricuspid valve annuli, and lack of transmurality arose as a possible factor determining results.
The possibility of reproducing the pattern of lesions of the cut-and-sew maze III with total bipolar radiofrequency may mimic the efficacy of Cox's operation. Our study tried to clarify whether application of bipolar radiofrequency is feasible in the most controversial lines, connecting the mitral and the tricuspid valve annuli.
| Material and Methods |
|---|
|
|
|---|
|
We also studied a necropsy heart of a male patient 72 years of age with a history of aortic stenosis and long-standing persistent AF (left atrial diameter 51 mm). Preoperative coronary arteriography showed right dominance. The patient died 10 hours after an aortic valve replacement and Cox maze IV procedure by bipolar radiofrequency (AtriCure), following the lesion pattern of Figure 1, A, B, and C as we routinely perform for persistent and long-standing persistent AF. The cause of death was not related to the ablation but to sudden bleeding from a very calcified aortotomy after a hypertensive peak at the sixth hour after the operation. After the family signed authorization, a complete necropsy study was performed 8 hours after death. Hematoxylin and eosin staining and complement C9 immunostaining were performed on the ablation lines of the left and right atria. C9 staining has been attributed as one of the earliest and most sensitive markers of necrosis, being able to detect myocardial necrosis 4 to 6 hours after infarction.5
| Results |
|---|
|
|
|---|
Both Medtronic and AtriCure bipolar clamps were able to be placed in the direction of the mitral annulus. In all cases, what limited how far the clamp could be positioned was the unfeasibility of clamping the ventricular myocardium and the AV adipose tissue that protrudes in the epicardial side. At all times, the maximum amount of atrial tissue clamped was limited by the tip of the epicardial jaw reaching the ventricular myocardial mass. However, the clamp never reached the mitral annulus (
Figure 3), leaving from 5 to 10 mm of atrial muscle out of reach of the bipolar electrodes. This did not limit the device to include the coronary sinus within its jaws in all hearts. Most important, in both hearts in which a left coronary dominance was detected, the circumflex artery was within the bipolar jaws of the clamp.
|
|
|
| Discussion |
|---|
|
|
|---|
To really reproduce short- and long-term results of the cut-and-sew Cox maze operation with alternative energy application, one needs to assure transmural conduction block along the complete set of lines. Bipolar radiofrequency has demonstrated transmurality. However, our study including the two systems with more references in medical literature shows that it is not possible to complete the ablation lines to the mitral or tricuspid valve annuli with actual bipolar radiofrequency technology. Completing the lines from the pulmonary veins to the AV fibrous tissue still needs to be performed by cut-and-sew technique or by any means of monopolar thermal lesion, either warm or cold, compromising transmurality. At this moment, the combination of bipolar radiofrequency to perform the ablation lines and cryolesions to finish the lines up to the mitral and tricuspid annuli is probably the technique with best known results in Cox maze IV procedures. Nevertheless, this technique has shown less efficacy than the cut-and-sew Cox maze III.4
Newly introduced monopolar or transpolar radiofrequency devices still need to prove tissue transmurality and clinical efficacy. Not being able to complete the mitral line may even be detrimental, inasmuch as incomplete lines with narrow paths favor left atrial flutter.8
Electrophysiologic studies in patients with atrial flutter after modified radiofrequency maze procedures have shown incomplete surgical lesions at the mitral and cavotricuspid isthmus.9
It is important to remark that atrial flutter is most of the time more symptomatic and worse tolerated than long-standing persistent AF, and if the narrow atrial tissue not successfully ablated lies beneath the Dacron annulus of a newly implanted valve prosthesis, the possibilities that postoperative flutter may be controlled in the electrophysiology laboratory are not precisely high.
Our observations in this study are founded in basic anatomy that we believe needs to be emphasized with the increase of procedures based on new technologies. The left atrial wall folds down internally over the crest of the left ventricle for a variable distance. On the epicardial side, the adipose tissue in the AV groove covers basically atrial wall, and the structures that are embraced by this fat, the coronary arteries and sinus, lie in the epicardial side of both the left and right atria. This anatomic fact was already noted in the original descriptions by Cox of his technique, which included the use of a No. 15 blade knife into the epicardial fat through the overlapping atrial muscle to reach the annulus. This fat pad around the AV junction facilitates the progression in thickness from a thin atrial wall to a thick ventricular wall, most evident in the left side.
When applying the bipolar radiofrequency clamps in the left atrium to reach the mitral annulus, we detected two potential problems. First, the change in thickness from the atrium to the ventricle by the epicardial adipose tissue in the AV junction impeded closure of the clamp reaching the mitral annulus, leaving 5 to 10 mm of atrial muscle out of reach of the thermal injury. This was confirmed in our postmortem ablation study, in which we could confirm normal myocardial tissue between the ablation injury and the mitral annulus. Second, even if the mitral annulus could not be reached, the coronary sinus and the circumflex artery in the left coronary–dominant hearts were included within the clamp, inasmuch as both structures lie in the atrial side of the AV junction.
In the right side, the slim thickness of both the atrial and free wall of the right ventricle allow closure of the clamp to reach the tricuspid annulus. However, then it always includes the midsection of the right coronary artery, inasmuch as this artery also lies in the atrial side of the AV junction. If this artery is to be avoided, the clamp is always unable to reach the tricuspid annulus.
This study was performed in freshly dead human hearts, and lesions were not actually performed. Therefore, we cannot conclude whether application of bipolar radiofrequency energy during a variable time between 5 to 15 seconds at around 60°C would irreversibly affect coronary circulation. Several studies report coronary damage by endocardially applied unipolar radiofrequency ablation, including temporal spasm by neurogenic or chemotactic (peptidergic) mechanisms.10
Therefore, and as intuitively advanced, patients with left coronary dominance may not be suitable candidates for bipolar radiofrequency or any means of epicardially applied thermal lesions to perform the mitral line. This can also be true in patients with important posterolateral branches arising from the right coronary artery. Careful revision of the preoperative coronary arteriogram is mandatory before deciding the ablation pattern in each patient.
Bipolar radiofrequency ablation toward the mitral annulus certainly produces a circumferential thermal ablation around the coronary sinus. To our knowledge, there are no reports of coronary sinus perforation or bleeding as a result of the application of bipolar clamps. Whether the thermal lesion creates a certain degree of stenosis is yet to be defined, as it has been the case in the pulmonary veins when percutaneous monopolar radiofrequency is endocardially applied with the circumferential Lasso catheter. Nevertheless, the multiple connections of the cardiac venous system with the ventricular cavities may limit the importance of possible coronary sinus stenosis.
In conclusion, current bipolar radiofrequency ablation technology is not able to complete a Cox maze IV procedure. A unipolar energy source, or ultimately the cut-and-sew technique, is needed to perform ablation lines reaching the mitral or the tricuspid valve annuli. At this time, only the cut-and-sew technique ensures total transmurality along the lesion pattern of the Cox maze surgical procedure.
| Acknowledgments |
|---|
Disclosures and Freedom of Investigation. The bipolar clamps tested in the study were previously used in the clinical setting. Hearts were studied after homograft harvesting in our homograft bank (Transplant Services Foundation, Hospital Clínic, Barcelona, Spain) except one heart which was studied 18 hours after aortic valve replacement and radiofrequency Cox maze IV ablation. The study had no external funding. We had full control of the design of the study, methods used, outcome parameters, analysis of data, and production of the written report.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. C. Maroto, M. Carnero, J. A. Silva, J. Cobiella, N. Perez-Castellano, F. Reguillo, J. Perez-Villacastin, and J. E. Rodriguez Early recurrence is a predictor of late failure in surgical ablation of atrial fibrillation Interact CardioVasc Thorac Surg, May 1, 2011; 12(5): 681 - 686. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Minakata, T. Yunoki, E. Yoshikawa, M. Katsu, T. Oda, and K. Ujino Predictors of success of the modified maze procedure using radiofrequency device Asian Cardiovascular and Thoracic Annals, February 1, 2011; 19(1): 33 - 38. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Wu, S. Zhang, A. Dong, Z. He, S. Chen, and R. Chen Long-Term Outcomes of Maze Procedure Plus Valve Replacement in Treating Rheumatic Valve Disease Resulting in Atrial Fibrillation Ann. Thorac. Surg., June 1, 2010; 89(6): 1942 - 1949. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Castella Reply to the Editor. J. Thorac. Cardiovasc. Surg., June 1, 2009; 137(6): 1574 - 1575. [Full Text] [PDF] |
||||
![]() |
S. Benussi, D. Corradi, and O. Alfieri Clamping the mitral and tricuspid annuli with bipolar devices. J. Thorac. Cardiovasc. Surg., June 1, 2009; 137(6): 1573 - 1574. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||