JTCS KCI
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Charles A. Yankah
Miralem Pasic
Michele Musci
Henryk Siniawski
Roland Hetzer
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yankah, C. A.
Right arrow Articles by Hetzer, R.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Yankah, C. A.
Right arrow Articles by Hetzer, R.
Related Collections
Right arrow Valve disease

J Thorac Cardiovasc Surg 2008;136:688-696
© 2008 The American Association for Thoracic Surgery


Surgery for Acquired Cardiovascular Disease

Aortic valve replacement with the Mitroflow pericardial bioprosthesis: Durability results up to 21 years

Charles A. Yankah, MD, PhD*, Miralem Pasic, MD, PhD, Michele Musci, MD, Julia Stein, MSc, Christine Detschades, SRN, Henryk Siniawski, MD, PhD, Roland Hetzer, MD, PhD

Deutsches Herzzentrum Berlin, Berlin, Germany

Received for publication November 12, 2007; revisions received March 28, 2008; accepted for publication May 13, 2008.

* Address for reprints: A. Charles Yankah, MD, PhD, Professor of Surgery, Humboldt, Charité Medical University, German Heart Institute Berlin, Augustenburger Platz 1, D-13353 Berlin, Germany. (Email: yankah{at}dhzb.de).


    Abstract
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 
Objective: The study aim was to analyze the performance profile of a large series of Mitroflow pericardial valves (Sorin Group Canada Inc. Mitroflow Division) in the very long term.

Methods: Data from 1513 patients with isolated aortic valve replacement who received pericardial bioprostheses between 1986 and 2007 were analyzed. Cumulative duration of follow-up was 6164 patient-years with a maximum duration of 21 years. Actuarial rates of valve-related events were calculated by the Kaplan–Meier method and the Cox multivariate analysis to identify independent determinants of outcome.

Results: Hospital mortality for elective surgery was 2.5%. Late death was 40.6%. Reoperation was required in 86 (5.7%) patients and was valve related in 83: structural valve deterioration in 64 (4.2%) patients, prosthetic valve endocarditis in 17 patients (1.1%), valve thrombosis in 1, and periprosthetic leak in 1. Rates of 20-year actuarial freedom from valve-related morbidity were as follows: structural valve deterioration 84.8% (actual 96.6%) in patients 70 years of age or older; thromboembolism 94.1%; and prosthetic valve endocarditis 96.8%. Twenty-year actual risk of reoperation for structural valve deterioration was 11.4% in all patients and 3.4%, in patients 70 years or age or older. Advanced age, renal insufficiency, pulmonary disease, and low body mass index were independent risk factors for late outcome (P < .001).

Conclusions: After 2 decades of follow-up, the Mitroflow pericardial aortic valve continues to be a valve of choice with a predictable low rate of valve-related events, particularly for patients over the age of 65 to 70 years and others with comorbidities.



Abbreviations and Acronyms NYHA = New York Heart Association; CI = confidence interval; CL = confidence limits; OR = odds ratio; PVE = prosthetic valve endocarditis; SVD = structural valve deterioration; TE = thromboembolism



    Introduction
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 
Although the pericardial bioprosthesis was once unpopular for clinical use because of design-related early structural deterioration,1-3Go it has demonstrated its potential for midterm and long-term durability beyond 15 years.4-11Go It is now apparent that durability and performance resulting from improved design of second-generation pericardial tissue valves are even better when the valves are implanted in a selected group of patients.6-11Go The Mitroflow pericardial bioprosthesis (Sorin Group Canada Inc. Mitroflow Division) has demonstrated excellent hemodynamics, comparable with stentless porcine valves and superior to similar stented bioprostheses,8,12-20Go especially in small sizes. The very long-term durability is not yet established. For this reason, we evaluated clinical data from 1513 patients who received single Mitroflow pericardial aortic valve replacement from March 1986 to determine the long-term clinical durability and threshold age for implantation.


    Patients and Methods
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 
Over a period of 21 years between March 1986 and December 2007, 1760 Mitroflow pericardial bioprostheses were implanted in 1722 consecutive patients in the aortic, mitral, tricuspid, and pulmonary positions as single, double, or triple procedures. A total of 1513 patients received single aortic valve replacements. There were 1071 (70.8%) women and 442 (29.2%) men. Preoperative characteristics of the study population are shown in Table 1 .


View this table:
[in this window]
[in a new window]

 
Table 1 Characteristics of patients
 
The results of 17-year durability on 1037 patients in the series have been reported8Go; the follow-up is now extended to 21 years. Four hundred seventy-six additional patients who were operated on in the interim have been followed up to 2.69 years for 482.06 patient-years (mean: 1.017 ± 0.35 SEM) with the previous series.

Surgical Technique
Standard surgical procedures were performed as described previously.8Go The major aspects of implantation are as follows: Meticulous decalcification is used to avoid abrasion and early alteration to the pericardial valve by residual calcific tissue. The decalcified annulus allows proper matching and secures good coaptation of leaflets.

A sizing obturator is used to measure the native annulus for proper selection of the pericardial valve size, using clinical judgment of the space in the supra-annular area of the aortic sinuses. This will avoid oversizing, which creates valve deformation. In situations in which subaortic obstructive septal hypertrophy is present, incision or preferably resection (myectomy) should be performed. Enlargement of a small aortic annulus is rare, but if it is necessary the Nicks procedure can be performed to avoid squeezing and deformation of the valve.

Despite the valve's low-profile characteristics, proper orientation of the anatomic units of the pericardial prosthesis should be observed.

The valve should be inserted in a supra-annular position to avoid hemodynamic stress, and the valve commissures should correspond to native commissures while the valve sinuses face the coronary ostia to achieve unobstructed coronary ostia. Finally, very tall knots that would expose valve tissue to abrasion, early alteration, and perforation should be avoided.

Data Collection and Postoperative Follow-up
The study, as well as the use of patients' data for research purposes and publication, was approved by the institutional review board and informed consent was obtained from the follow-up patients.

Data acquisition
Patients were examined at our institution or have been contacted by telephone interview and mailed questionnaire. Further patient data were obtained from hospital records, family physicians, and cardiologists. Patients with unknown addresses could be tracked through the district or state registry of residents or birth and deaths.

Total follow-up was 6163.5 patient-years (mean: 4.07 ± 0.12 years, SEM). Eleven (0.7%) patients were lost to follow-up (99.3% complete). A total of 1297 (85.7%) of the hospital survivors underwent routine echocardiographic studies at 3 and 9 months after operation and thereafter annually. Transthoracic Doppler echocardiography was performed at different institutions in a uniform manner and the evaluations were comparable. If a patient had undergone more than one echocardiographic or clinical evaluation, the result of the most recent investigation was reported.

We evaluated 189 measurements of mean gradients in 121 patients. Measurement of mean gradients and time intervals between serial echocardiographic studies were not uniform in some patients. Therefore, we did not apply the mixed model longitudinal regression analysis as described by Banbury and associates.4Go We are presenting scatterplots of mean gradients of valve sizes 19 to 25 mm and their median values to demonstrate the evolution of transvalvular mean gradients across time.

Structural and nonstructural valve deterioration of Mitroflow bioprostheses, diagnosed preoperatively by echocardiographic studies, was confirmed at the time of explantation.

At the time of the last follow-up study, 150 patients (29%) were in New York Heart Association (NYHA) class I, 232 (45%) in class II, 120 (23%) in class III, and 14 (3%) patients could not be classified because of advanced age.

Ninety-seven (7.6%) patients were receiving anticoagulation treatment with warfarin sodium (Coumadin) and 260 (20.5%) with antiplatelet drugs. Three hundred fifteen (24.8%) patients 65 years of age or older were receiving anticoagulants as compared with 42 (3.3%) patients younger than 65 years.

NYHA functional classification of the patients was inconsistent and unreliable to reproduce owing to a high proportion of patients at advanced age (mean age at operation 73 years). Many follow-up patients discontinued or irregularly took their drugs partly because of the risk and fear of bleeding and stroke. Therefore, a statistical approach to estimate a time-related proportion of patients in NYHA functional class and receiving anticoagulants did not yield meaningful, reproducible clinical information.

Statistics
The probability of survival, freedom from valve-related morbidity events, death, and reoperation were estimated by Kaplan–Meier actuarial analysis and by linearized occurrence rates. Linearized occurence rates of events and confidence limits were calculated according to Poisson distribution.

Actual competing risk analysis (cumulative incidence) was performed.

Tabular data are summarized by the mean and given with standard error for continuous variables and by percentages for categorical variables. Differences in actuarial freedom between groups of patients are determined by the log–rank test. Differences in prognostic variables between two groups were evaluated by t tests for continuous variables and the {chi}2 or Fisher exact test for categorical variables. Predictors of events during follow-up were identified by the Cox proportional hazards regression.

All variables listed in Table 1 were investigated for association with hospital death, overall death, and valve-related complications by univariate and multivariate analysis. Standard statistical formulas have been used for the analysis of data, as described previously.8Go


    Results
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 
Patient Survival
Hospital survival
Thirty-day hospital survival was 97.5% for elective patients and 90.2% for patients in NYHA functional class III–IV (odds ratio [OR], 3.9; confidence interval [CI] (95% confidence limits [CL]) 1.31–12.07; P value .015).

The causes of early death were cardiac related in 96 (61%) patients and noncardiac and nonvalve related in 61 (39%) patients. Significant predictive factors for early mortality were emergency operation (OR, 1.53; CI [95% CL] 1.33–1.75; P value < .001) and preoperative renal failure (OR, 2.76; CI [95% CL] 0.98–7.76; P value .054).

Long-term survival
A total of 615 patients died. Twenty-five (4.1%) deaths were valve related. Actuarial freedom from valve-related death at 20 years was 82.9% ± 4.0%.

The actuarial survival including operative death in our cohort at 10, 15, and 20 years was 31.9% ± 1.8%, 12.7% ± 1.4%, and 6.1% ± 1.5%, respectively (Figure 1 ; OR, 1.82; CI [95% CL] 1.55–2.18; P value < .001). Univariate and multivariate analysis showed that NYHA III–IV (OR, 1.37; CI [95% CL] 1.18–1.59; P value < .001), renal insufficiency (OR, 1.74; CI [95% CL] 1.40–2.17; P value < .001), chronic obstructive pulmonary disease (OR, 1.54; CI [95% CL] 1.22–1.95; P value .001), body mass index less than 20 kg/m2 (OR, 1.46; CI [95% CL] 1.13–1.88; P value .001), and advanced age (OR, 1.82; CI, 1.56–2.16; P value < .001) were predictive factors for late death.


Figure 1
View larger version (15K):
[in this window]
[in a new window]

 
Figure 1. Overall cumulative survival of all 1513 patients after aortic valve replacement and of a German population as background (data from Federal Bureau for Population Statistics, Wiesbaden, Germany).

 
On the other hand, small valve size (19 mm: OR, 1.09; CI, 0.86–1.38; P value 0.458; 21 mm: OR, 1.68; CI, 0.65–4.38: 1.38; P value .286) and gender (OR, 1.11; CI, 0.96–1.29; P value .158) were not independent risk factors.

Of 435 follow-up patients, after 1 year 136 were in NYHA class I, 202 in class II, 87 in class III, and 10 in class IV; after 5 years, of 45 patients 6 were in class I, 19 in class II, 18 in class III, and 2 in class IV; after 10 years, of 36 patients 8 were in class I, 11 in class II, 15 in class III, and 2 in class IV; after 15 years, of 6 patients 1 was in class I, another 1 in class II, 3 were in class III, and 1 in class IV. Anticoagulation was administered in all patients for 6 weeks after discharge. After 1 year 138 of 995 follow-up patients were receiving anticoagulants, after 5 years 116 of 493, after 10 years 52 of 156, and after 15 years 51 of 58 follow-up patients were receiving anticoagulants.

Valve-related Complications
Reoperation
Reoperation was required in 86 patients and was valve related in 83. Causes were as follows: structural valve deterioration (SVD) in 64 (4.2%) cases, valve thrombosis in 1 case (0.07%), periprosthetic leak in 1 case (0.07%), and prosthetic valve endocarditis (PVE) in 17 (1.1%) cases.

The 20-year actual risk of reoperation for all causes was 15.5% with a linearized rate of 1.4%/pt-y (CI, 1.05–1.62).

SVD
Sixty-four (4.2%) patients had primary tissue failure resulting from leaflet calcification or fibrosis causing stenosis in 8, incompetence in 30 (including 1 leaflet tear in Mitroflow early experience), or mixed lesion in 26 patients. Detailed analysis is shown in Table 2 . Overall 20-year freedom from SVD was 62.3 + 5.02%. Actuarial freedom from reoperation for SVD at 20 years in patients aged 65 years or older and 70 years or older was 71.8% ± 6.0% (actual, 92.6 ± 4.6%) and 84.8 ± 0.7% (actual, 96.6% ± 0.8%), respectively (Figures 2–4 and Table 3 Go Go). The 20-year actual risk (cumulative incidence) for SVD was 11.4%. The linearized rate of SVD for the entire patient cohort was 0.9%/pt-y. It was 0.76%/pt-y and 0.34%/pt-y for patients aged 65 years or older and 70 years or older, respectively. Younger patients, below the age of 65 years, had a linearized rate for SVD of 2.1%/pt-y (P = .004). Detailed linearized rates of events and confidence limits at different age groups are shown in Table 3 .


View this table:
[in this window]
[in a new window]

 
Table 2 Summary of causes for reoperation and explantation of 86 Mitroflow valves
 

Figure 2
View larger version (10K):
[in this window]
[in a new window]

 
Figure 2. Actuarial freedom from reoperation for structural valve deterioration (SVD) in age groups (50–59, 60–69, and ≥70 years) after pericardial aortic valve replacement.

 

View this table:
[in this window]
[in a new window]

 
Table 3 Summary of linearized rate of events by age group: Patients at risk and confidence limit
 

Figure 3
View larger version (15K):
[in this window]
[in a new window]

 
Figure 3. Actual (cumulative incidence) freedom from reoperation for structural valve deterioration (SVD) by age group (50–59, 60–69 and ≥70 years) after pericardial aortic valve replacement.

 

Figure 4
View larger version (15K):
[in this window]
[in a new window]

 
Figure 4. Actuarial and actual (cumulative incidence) freedom from reoperation for structural valve deterioration (SVD) in age group ≥65 years after pericardial aortic valve replacement.

 
The leaflet disruption observed in 1 patient occurred at the base of the commissural site facing the noncoronary aortic annulus (linearized rate, 0.016%/pt-y; CI, 0.0004–0.08).

Time-related transvalvular mean gradients and valve size
Repeated measurements of transvalvular mean gradients and valve sizes across time are shown in Table 4 and Figure 5 . Postoperative transvalvular mean gradient with valve size 19 mm was greater than that with 21 mm and 23 mm; however, all were decreased at 1 year.


View this table:
[in this window]
[in a new window]

 
Table 4 Postoperative echocardiographic data and labeled prosthesis size in 121 patients
 

Figure 5
View larger version (16K):
[in this window]
[in a new window]

 
Figure 5. Time-related transvalvular mean gradients in relation to labeled valve sizes. Each symbol represents a labeled valve size. Solid lines represent time-related estimates of median values for each prosthesis.

 
In patients aged 65 years or more, mean transvalvular gradient at 5 years for valve sizes 19 to 25 mm was 12 to 6 mm Hg. This increased to 18 to 8 mm Hg after 10 years and then remained stable at 15 and 20 years for valve sizes 21 and 23 mm, respectively (Figure 5).

In contrast, patients less than 65 years showed a significant increase in transvalvular mean gradient to 40 to 19 mm Hg or more after 10 years for valve sizes 21 and 23 mm (P =.004), leading to reoperation (Figure 2).

Nonstructural valve dysfunction
There was 1 (0.07%) case of nonstructural valve dysfunction: periprosthetic leak (0.07%). An additional 3 dysfunctions were not valve related: 2 technically related (0.13%) and 1 intraoperative mismatch (0.07%) (Table 2).

Thromboembolism
Forty-nine (3.2%) patients had episodes of thromboembolsm (TE) representing an actuarial freedom of 94.1% at 20 years (Figure 6 ). The linearized rate was 0.79%/pt-y.(CI, 0.60–1.01). TE occurred in 6 (0.4%) patients under 65 years of age (linearized rate, 0.5%/pt-y; CI, 0.24–1.06) and in 43 (2.8%) patients 65 years of age or older (linearized rate, 0.93%/pt-y; CI, 0.66–1.18). In the 49 patients with stroke, 4 neurologic events were permanent (linearized rate, 0.06%/pt-yr) and 9 (linearized rate, 0.09%/pt-yr; CI, 0.07–0.25) were fatal. TE was recurrent in nine patients. One patient experienced a valve thrombosis (linearized rate: 0.02%/pt-y).


Figure 6
View larger version (13K):
[in this window]
[in a new window]

 
Figure 6. Actuarial freedom from thromboembolism (TE) after pericardial aortic valve replacement.

 
Bleeding
There were 4 nonfatal cases of hemorrhagic events under anticoagulation therapy (linearized rate, 0.065%/pt-y; CI, 0.02–0.15) The indication for anticoagulation was atrial fibrillation or cardiologic decision. None of the patients experienced a recurrent bleeding episode.

PVE
Seventeen (1.1%) patients with a sterile aortic root at primary operation had a prosthetic infection. The organisms associated with late infections were staphylococcal and streptococcal species. Among the 202 patients with healed endocarditis and 40 with active endocarditis, there was 1 episode of reinfection in each group. Significant paravalvular leaks resulting from infection were observed in 5 patients (linearized rate, 0.08%/pt-y; CI, 0.03–0.17) and a leaflet tear in 1 of the patients.

The actuarial freedom from endocarditis was 96.8% ± 0.9% at 20 years and the linearized rate for endocarditis was 0.28%/pt-y (CI, 0.14–0.44, Figure 7 ). There were constant very low hazard events of PVE.


Figure 7
View larger version (8K):
[in this window]
[in a new window]

 
Figure 7. Hazard function of prosthetic valve endocarditis after pericardial aortic valve replacement. Constant hazard rate of events.

 

    Discussion
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 
Our current series of 1513 Mitroflow stented low-profile pericardial bioprostheses represents one of the largest single institutional experience with the longest follow-up (21 years).

For xenograft tissue valves, the porcine bioprosthesis is taken as the standard against which results can be compared. The potential for a very long-term durability of stented pericardial and stented porcine tissue valves for over 20 years has been demonstrated and reported recently.8,21,22Go

Statistical comparison between actuarial data reported from various institutions is very difficult, but literature review can help to gain a general impression.5,8,17,18,20,23-25Go Although the porcine and pericardial valves are collagen structures treated with glutaraldehyde, their shape, technique of mounting, mode of function, and fluidodynamic characteristics are dissimilar and therefore their long-term performance and mode of failure may be different.

Twenty-one–year durability of the Mitroflow pericardial bioprosthesis attests to the quality of the pericardial tissue and the design/shape of this bioprosthesis. In comparison with porcine valves, the Mitroflow pericardial valve is less obstructive especially in the 19- and 21-mm sizes and even competes well with porcine stentless valves.8,14-20Go Sustained long-term durability over 21 years has been demonstrated in patients older than 70 years with 84.8% (actual, 96.6%) free of SVD.

The 72% 21-year actuarial (actual, 92.6%) freedom from SVD in patients over 65 years with Mitroflow pericardial tissue valves compares favorably with Hancock II (Medtronic, Inc, Minneapolis, Minn) porcine valves (actuarial freedom from SVD of 73%.) in patients in similar age groups.21Go Other reports also confirmed that the pericardial valve is as durable as porcine and allograft valves at ages above 65 years.24Go

Several authors have reported leaflet tears and early SVD even in elderly patients, which was a rare event in our series.1-3,5,21,26Go The etiology of early SVD associated with leaflet tears in a subgroup of elderly patients is not clearly understood.

In our series, leaflet tears or disruptions occurred in 2 patients. Leaflet tear as a primary tissue failure occurred in 1 patient, which represents a linearized rate of 0.02%/pt-y (CI, 0.0004–0.08), whereas the other leaflet tear was associated with PVE. This indicates that if there is a history of endocarditis, as we encountered in our patient, the tear is secondary to the infection and the complication has to be classified as endocarditis and excluded from the SVD group, as reported in the guidelines of Edmunds and associates27Go and by other authors.21,29,30Go Acceleration of structural deterioration could occur after a successful treatment of endocarditis.7,28Go It is therefore important to consider the history of previous antibiotic treatment and exclude endocarditis as an underlying cause before classifying a tear as a primary tissue failure.

The analysis of our 21-year results with Mitroflow pericardial bioprostheses in the aortic position has shown it to be safe even at withdrawal of anticoagulation or antiplatelet treatment in 92% and 80% of the patients, respectively. The risk of late TE complications impact with an incidence of 3.8% and a linearized rate of 0.79%/pt-y (CI, 0.60–1.01).

The major indication for long-term anticoagulation was atrial fibrillation. Valve thrombosis and lethal TE were encountered in 1 (0.02%/pt-y) and another 9 (0.7% pt-y) patients in this series, respectively, and the potential hazard of anticoagulant-induced hemorrhage was significantly low (n = 4, 0.3%) with a linearized rate of 0.06%/pt-y (CI, 0.02–0.15).

The incidence of 3.4% TE complications in patients older than 65 years of age could be explained by the fact that the average patient age in the series was over 73 years and the risk of cerebral accidents developing from unrecognized cerebrovascular lesions without cardiac surgery was already high.26Go The disparity in TE rates between our series and recently published data with porcine valves is evident.5,18,21-23,25Go The lower propensity for TE in our cohort with the Mitroflow pericardial valve relates to the quality of the tissue and the design of the prosthesis. Although the exact etiology of early TE is not widely established, our experience has indicated that many patients could benefit from anticoagulant treatment for the initial 6 weeks after bioprosthetic aortic valve replacement. In the presence of one or more factors associated with an increased risk of TE (atrial fibrillation, enlarged left atrium with or without thrombus, recurrent atrial fibrillation after ablation), such patients with tissue valves should receive lifelong anticoagulant therapy.

Concomitant coronary artery bypass graft operation was not found to be a risk factor for operative death. The low patient survival of 6% at 21 years with a median probability of survival time of 6.8 years is attributed to the average patient age of 73 years at implantation in our series, and it is in line with the results obtained from meta-analysis of life expectancy, which was 10.7 years for a 65-year-old patient after aortic valve replacement.29Go

Furthermore, after 15 years the Kaplan–Meier estimates for patient survival are not as reliable as the initial estimates because of the limited number of patients at risk beyond 15 years. This translates in the mean follow-up per patient, which is 4.07 years per patient in the present study and 4.85 years per patient in the 2003 study.8Go

Actuarial freedom from valve-related death in our series was 94% at 10 years and 83% at 20 years. The result is in line with data published by others.4,17,19Go Although autopsies were not performed in some patients who died with known causes, they were performed in cases of unexplained or sudden death with some exceptions (religious or cultural). Figures on SVD were obtained at reoperation or autopsy.27Go The available autopsy results did not show any acute SVD or leaflet rupture as a cause of death. The long-term patient mortality was caused much more commonly by patient-related factors than by the presence of the Mitroflow pericardial valve prosthesis, as described in the paragraph on long-term survival of this article. Valve-related death of 4.1% (25/615) was therefore low in our series and was overestimated as sudden or unexplained patient death and was the most common cause of valve-related death. There was no direct correlation between valve size and early or late death in patients with 19- to 21-mm and 23- to 25-mm Mitroflow bioprostheses (P = .458 and P = .286). Further analysis to correlate with NYHA functional class revealed even better results for smaller valve sizes 19 to 21 mm (P = .003, P = .034).

Inasmuch as the physical capacity of elderly patients may diminish over time, we did not use change in NYHA class as a function for SVD or patient–prosthesis mismatch in these patients without echocardiographic evidence. The analysis has ruled out hemodynamically relevant patient–prosthesis mismatch and testifies to the excellent hemodynamic performance of small sizes of Mitroflow bioprostheses (mean transvalvular gradient [mm Hg]/effective orifice area [cm2] of 9.4 mm Hg/1.4 cm2, 7.1 mm Hg/1.6 cm2, and 4.7 mm Hg/1.85 cm2 for valve sizes 19 mm, 21 mm, and 23 mm, respectively) as compared with other bioprostheses.4,8,12-20Go

Repeated measurements of transvalvular mean gradients in relation to Mitroflow pericardial valve size across time in patients aged 65 years or older showed that mean transvalvular gradient at 5 years for valve sizes 19 to 23 mm was 12 to 5 mm Hg. This increased to 18 to 8 mm Hg after 10 years and then remained stable at 15 and 20 years for valve sizes 21 and 23 mm, respectively.

In contrast, patients younger than 65 years showed an increase in transvalvular mean gradient to 40 to 19 mm Hg after 10 years for valve sizes 21 and 23 mm, which led to reoperation.

Patient–prosthesis mismatch was rare in our series, however, 2 patients with 19-mm and 21-mm valves in the supra-annular position were identified with a classic prosthesis–patient mismatch (effective orifice area, 0.8 cm2/m2), which could have been avoided by myectomy and the Nicks procedure for enlargement of the noncoronary sinus, respectively.

Frater,17Go Banbury,4Go and their associates reported mean gradients of 22 to 15 mm Hg and 22 to 12 mm Hg for Carpentier–Edwards Perimount pericardial valve (Edwards Lifesciences, Irvine, Calif) sizes 19 to 25 mm at 12 and 17 years in patients at a mean age of 65 years, respectively. The gradients reported were somewhat greater than those we found in our cohort at 15 and 20 years. SVD observed in the first 5 years (n = 7; linearized rate, 0.4%/pt-y; CI, 0.18–0.87) in our cohort was confounded by sequelae of radiotherapy for cancer, implantation technical reasons including a rare patient–prosthesis mismatch, and subclinical endocarditis, whereas SVD beyond the 5-year period in the first decade was associated with age below 65 years.1,5,7,20,21Go

PVE was a rare event that occurred in 17 (1.1%) patients, and the rate of recurrent infection was 0.8% (n = 2). The actuarial freedom from PVE was 96.8% ± 0.0.9% and the linearized rate was 0.28%/pt-y (CI, 0.14–0.44). The hazard rate for PVE was constantly very low.20Go

Our results suggest that the freedom from valve-related SVD, as well as the freedom from reoperation, and the TE rate are all very competitive with results published for other bioprostheses.17,18,21-25Go The freedom from SVD and reoperation are age related, as is the case in other reports of series with bioprostheses.17,18,21,24Go These data support our proposals that the Mitroflow prosthesis is the valve of choice in patients over the age of 65 to 70 years and a logical and safe choice for patients younger than 65 years of age with comorbidities.


    Conclusions
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 
The Mitroflow pericardial bioprosthesis exhibits very low pressure gradients combined with easy implantability. After 2 decades of follow-up, the Mitroflow pericardial aortic valve continues to be a valve of choice with a predictable low rate of valve-related events, particularly for patients over the age of 65 to 70 years and others with comorbidities.


    Acknowledgments
 
We are grateful to Anne M. Gale, ELS, for editorial assistance, Astrid Benhennour for bibliographic support, and Carla Weber for providing the graphics.


    Footnotes
 
Charles Yankah reports consulting and lecture Fees From Sorin.


    References
 Top
 Abstract
 Introduction
 Patients and Methods
 Results
 Discussion
 Conclusions
 References
 

  1. Roselli EE, Smedira NG, Blackstone EH. Failure modes of the Carpentier–Edwards pericardial bioprosthesis in the aortic position. J Heart Valve Dis 2006;15:421-428.[Medline]
  2. Scully H, Goldman B, Fulop J, Butany J, Tong C, Azuma J, et al. Five-year follow-up of Hancock pericardial valves: management of premature failure. J Card Surg 1988;3(3 Suppl):397-403.[Medline]
  3. Leandri J, Bertrand P, Mazzucotelli JP, Loisance D. Mode of failure of the Mitroflow pericardial valve. J Heart Valve Dis 1992;1:225-231.[Medline]
  4. Banbury MK, Cosgrove 3rd DM, Thomas JD, Blackstone EH, Rajeswaran J, Okies JE, et al. Hemodynamic stability during 17 years of the Carpentier–Edwards aortic pericardial bioprosthesis. Ann Thorac Surg 2002;73:1460-1465.[Abstract/Free Full Text]
  5. Minami K, Zittermann A, Schulte-Eistrup S, Koertke H, Körfer R. Mitroflow synergy prostheses for aortic valve replacement: 19 years experience with 1,516 patients. Ann Thorac Surg 2005;80:1699-1705.[Abstract/Free Full Text]
  6. Pomar JL, Jamieson WR, Pelletier LC, Gerein AN, Castella M, Brownlee RT. Mitroflow pericardial bioprosthesis: clinical performance to ten years. Ann Thorac Surg 1995;60(2 Suppl):S305-S310.[Medline]
  7. Yankah CA, Weng Y, Meyer R, Siniawski H, Hetzer R. Twenty-two–year durability of Ionescu–Shiley pericardial aortic bioprosthesis implanted in a 49-year-old woman: a valuable insight into the performance of current pericardial bioprostheses. J Thorac Cardiovasc Surg 2006;132:427-428.[Free Full Text]
  8. Yankah CA, Schubel J, Buz S, Siniawski H, Hetzer R. Seventeen-year clinical results of 1,037 Mitroflow pericardial heart valve prostheses in the aortic position. J Heart Valve Dis 2005;14:172-180.[Medline]
  9. Moggio RA, Pooley RW, Sarabu MR, Christiana J, Ho AW, Reed GE. Experience with the Mitroflow aortic bioprosthesis. J Thorac Cardiovasc Surg 1994;108:215-220.[Abstract/Free Full Text]
  10. Sjögren J, Gudbjartsson T, Thulin LI. Long-term outcome of the MitroFlow pericardial bioprosthesis in the elderly after aortic valve replacement. J Heart Valve Dis 2006;15:197-202.[Medline]
  11. Thulin LI, Thilen UJ, Kymle KA. Mitroflow pericardial bioprosthesis in the aortic position. Low incidence of structural valve deterioration in elderly patients during an 11-year follow-up. Scand Cardiovasc J 2000;34:192-196.[Medline]
  12. Cosgrove DM, Lytle BW, Williams GW. Hemodynamic performance of the Carpentier–Edwards pericardial valve in the aortic position in vivo. Circulation 1985;72(3 Pt 2):II146-II152.[Medline]
  13. Garcia-Bengochea J, Sierra J, Gonzalez-Juanatey JR, Rubio J, Vega M, Fernandez AL, et al. Left ventricular mass regression after aortic valve replacement with the new Mitroflow 12A pericardial bioprosthesis. J Heart Valve Dis 2006;15(3):446-452.[Medline]
  14. David TE, Puschmann R, Ivanov J, Bos J, Armstrong S, Feindel CM, et al. Aortic valve replacement with stentless and stented porcine valves: a case–match study. J Thorac Cardiovasc Surg 1998;116:236-241.[Abstract/Free Full Text]
  15. Sintek CF, Fletcher AD, Khonsari S. Small aortic root in the elderly: use of stentless bioprosthesis. J Heart Valve Dis 1996;5(Suppl 3):S308-S313.[Medline]
  16. Westaby S, Jin XY, Katsumata T, Arifi A, Braidley P. Valve replacement with a stentless bioprosthesis: versatility of the porcine aortic root. J Thorac Cardiovasc Surg 1998;116:477-484.[Abstract/Free Full Text]
  17. Frater RW, Furlong P, Cosgrove DM, Okies JE, Colburn LQ, Katz AS, et al. Long-term durability and patient functional status of the Carpentier–Edwards Perimount pericardial bioprosthesis in the aortic position. J Heart Valve Dis 1998;7:48-53.[Medline]
  18. Jamieson WR, David TE, Feindel CM, Miyagishima RT, Germann E. Performance of the Carpentier–Edwards SAV and Hancock-II porcine bioprostheses in aortic valve replacement. J Heart Valve Dis 2002;11:424-430.[Medline]
  19. Minami K, Schereika S, Kortke H, Gleichmann U, Koerfer R. Long term follow-up of Mitroflow pericardial valve prostheses in the small aortic annulus. J Cardiovasc Surg (Torino) 1993;34:189-193.[Medline]
  20. Rao V, Jamieson WR, Ivanov J, Armstrong S, David TE. Prosthesis–patient mismatch affects survival after aortic valve replacement. Circulation 2000;102(19 Suppl. 3):III5-III9.[Medline]
  21. Borger MA, Ivanov J, Armstrong S, Christie-Hrybinsky D, Feindel CM, David TE. Twenty-year results of the Hancock II bioprosthesis. J Heart Valve Dis 2006;15:49-55discussion 55-6.[Medline]
  22. Khan SS, Trento A, DeRobertis M, Kass RM, Sandhu M, Czer LS, et al. Twenty-year comparison of tissue and mechanical valve replacement. J Thorac Cardiovasc Surg 2001;122:257-269.[Abstract/Free Full Text]
  23. Legarra JJ, Llorens R, Catalan M, Segura I, Trenor AM, de Buruaga JS, et al. Eighteen-year follow up after Hancock II bioprosthesis insertion. J Heart Valve Dis 1999;8:16-24.[Medline]
  24. Smedira NG, Blackstone EH, Roselli EE, Laffey CC, Cosgrove DM. Are allografts the biologic valve of choice for aortic valve replacement in nonelderly patients? Comparison of explantation for structural valve deterioration of allograft and pericardial prostheses. J Thorac Cardiovasc Surg 2006;131:558-564e4.[Abstract/Free Full Text]
  25. Hammermeister K, Sethi GK, Henderson WG, Grover FL, Oprian C, Rahimtoola SH. Outcomes 15 years after valve replacement with a mechanical versus a bioprosthetic valve: final report of the Veterans Affairs randomized trial. J Am Coll Cardiol 2000;36:1152-1158.[Abstract/Free Full Text]
  26. Butchart EG, Moreno de la Santa P, Rooney SJ, Lewis PA. Arterial risk factors and ischemic cerebrovascular events after aortic valve replacement. J Heart Valve Dis 1995;4:1-8.[Medline]
  27. Edmunds Jr. LH, Clark RE, Cohn LH, Grunkemeier GL, Miller DC, Weisel RD. Guidelines for reporting morbidity and mortality after cardiac valvular operations. J Thorac Cardiovasc Surg 1996;112:708-711.[Free Full Text]
  28. Nollert G, Miksch J, Kreuzer E, Reichart B. Risk factors for atherosclerosis and the degeneration of pericardial valves after aortic valve replacement. J Thorac Cardiovasc Surg 2003;126:965-968.[Abstract/Free Full Text]
  29. Puvimanasinghe JP, Takkenberg JJ, Edwards MB, Eijkemans MJ, Steyerberg EW, Van Herwerden LA, et al. Comparison of outcomes after aortic valve replacement with a mechanical valve or a bioprosthesis using microsimulation. Heart 2004;90:1172-1178.[Abstract/Free Full Text]
  30. Blackstone EH, Kirklin JW. Death and other time-related events after valve replacement. Circulation 1985;72:753-767.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
Eur J Cardiothorac SurgHome page
S. Martens, J. Sadowski, F. S. Eckstein, K. Bartus, B. Kapelak, H.-H. Sievers, C. Schlensak, and T. Carrel
Clinical experience with the ATS 3f Enable(R) Sutureless Bioprosthesis
Eur J Cardiothorac Surg, September 1, 2011; 40(3): 749 - 755.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
W. R. E. Jamieson, C. T. P. Lewis, M. P. Sakwa, D. A. Cooley, V. R. Kshettry, K. W. Jones, T. E. David, J. A. Sullivan, G. J. Fradet, and D. S. Bach
St Jude Medical Epic porcine bioprosthesis: Results of the regulatory evaluation
J. Thorac. Cardiovasc. Surg., June 1, 2011; 141(6): 1449 - 1454.e2.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
P. Myken and O. Bech-Hansen
Reply to the editor.
J. Thorac. Cardiovasc. Surg., February 1, 2011; 141(2): 596 - 597.
[Full Text] [PDF]


Home page
Eur J Cardiothorac SurgHome page
The ISTHMUS Investigators
The Italian study on the Mitroflow postoperative results (ISTHMUS): a 20-year, multicentre evaluation of Mitroflow pericardial bioprosthesis
Eur J Cardiothorac Surg, January 1, 2011; 39(1): 18 - 26.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
C. A. Yankah
Mitroflow pericardial aortic bioprosthesis in patients younger than 60 years
J. Thorac. Cardiovasc. Surg., November 1, 2010; 140(5): e83 - e84.
[Full Text] [PDF]


Home page
Eur J Cardiothorac SurgHome page
W. R. E. Jamieson
Reply to Quiroga et al.
Eur J Cardiothorac Surg, November 1, 2010; 38(5): 649 - 650.
[Full Text] [PDF]


Home page
Eur J Cardiothorac SurgHome page
C. Valfre, P. Ius, G. Minniti, L. Salvador, T. Bottio, F. Cesari, G. Rizzoli, and G. Gerosa
The fate of Hancock II porcine valve recipients 25 years after implant
Eur J Cardiothorac Surg, August 1, 2010; 38(2): 141 - 146.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
J. Conte, N. Weissman, J. A. Dearani, J. Bavaria, D. Heimansohn, W. Dembitsky, and D. Doyle
A North American, Prospective, Multicenter Assessment of the Mitroflow Aortic Pericardial Prosthesis
Ann. Thorac. Surg., July 1, 2010; 90(1): 144 - 152.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
S. H. Rahimtoola
Choice of Prosthetic Heart Valve in Adults: An Update
J. Am. Coll. Cardiol., June 1, 2010; 55(22): 2413 - 2426.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
R. S. McClure, N. Narayanasamy, E. Wiegerinck, S. Lipsitz, A. Maloney, J. G. Byrne, S. F. Aranki, G. S. Couper, and L. H. Cohn
Late Outcomes for Aortic Valve Replacement With the Carpentier-Edwards Pericardial Bioprosthesis: Up to 17-Year Follow-Up in 1,000 Patients
Ann. Thorac. Surg., May 1, 2010; 89(5): 1410 - 1416.
[Abstract] [Full Text] [PDF]


Home page
Eur J Cardiothorac SurgHome page
W. R. E. Jamieson, R. Koerfer, C. A. Yankah, A. Zittermann, R. I. Hayden, H. Ling, R. Hetzer, and W. B. Dolman
Mitroflow aortic pericardial bioprosthesis -- clinical performance,
Eur J Cardiothorac Surg, November 1, 2009; 36(5): 818 - 824.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Charles A. Yankah
Miralem Pasic
Michele Musci
Henryk Siniawski
Roland Hetzer
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yankah, C. A.
Right arrow Articles by Hetzer, R.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Yankah, C. A.
Right arrow Articles by Hetzer, R.
Related Collections
Right arrow Valve disease


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS