JTCS KCI
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Dirk Seifert
Reiner Koerfer
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bairaktaris, A.
Right arrow Articles by Koerfer, R.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Bairaktaris, A.
Right arrow Articles by Koerfer, R.

J Thorac Cardiovasc Surg 2008;136:1076-1077
© 2008 The American Association for Thoracic Surgery


Brief Communication

Pitfalls in catheter-based interventions to treat paravalvular leaks

Andreas Bairaktaris, MD, Nikolaus A. Haas, MD, Dirk Seifert, MD, Rainer Schaeffler, MD, Heinrich Koertke, MD, Soren Schenk, MD*, Reiner Koerfer, MD

Department of Cardiovascular Surgery, Heart Center NRW, Bad Oeynhausen, Germany

Received for publication August 14, 2007; accepted for publication August 23, 2007.

* Address for reprints: Soren Schenk, MD, Cardiovascular Surgery, Heart Center NRW, Georgstrasse 11, 32545 Bad Oeynhausen, Germany. (Email: sschenk{at}arcor.de).

Paravalvular leaks of prosthetic heart valves can be observed at a rate as great as 1.4% leaks/patient-year.1Go Surgical correction can be performed with low mortality in most instances of clinically relevant regurgitation or hemolysis.2Go More recently, interventional cardiologic techniques are of growing interest. This report depicts 3 patients who required surgical treatment for interventional complications. Current challenges and indications for this evolving technology are discussed.

Clinical Summaries

PATIENT 1. A 73-year-old man was referred for grade III mitral regurgitation caused by a persistent paravalvular leak. The patient had undergone implantation of a bioprosthesis in the mitral position at 53 years of age. This prosthesis was replaced when the patient was 66 years old with a mechanical prosthesis because of valve degeneration. A relevant paravalvular leak was diagnosed 6 years after the last valve exchange and prompted 2 attempts at interventional closure with the Amplatzer PDA Occluder Device (AGA Medical Corporation, Golden Valley, Minn). The patient had increasing hemolysis, and his clinical status progressively declined within the subsequent 2 years. Fluoroscopy revealed mitral insufficiency caused by restriction of leaflet motion by an occluder device. Intraoperative findings confirmed the imaging studies (Figure 1 ). A third valve replacement and the subsequent postoperative course were uneventful.


Figure 1
View larger version (147K):
[in this window]
[in a new window]

 
Figure 1. Intraoperative findings of obstructed leaflet of prosthetic valve in mitral position (MP) in patient 1. Two occluder devices are placed into paravalvular leak (arrows). One occluder protrudes into valve hinges and arrests leaflet (arrowhead).

 
PATIENT 2. The second patient was a 21-year-old woman with 3 previous operations for truncus arteriosus communis type 1. Banding of the pulmonary artery at the age of 1 year was followed by an extracardiac conduit and ventricular septal defect closure 4 years later. Replacements of pulmonary and aortic valves with mechanical prostheses were necessary when the patient was 13 years old. Paravalvular leakage of the aortic valve was diagnosed at the age of 20 years and then treated by intervention with 2 Amplatzer PDA Occluder Devices (Figure 2 ). Although her aortic insufficiency was reduced to grade I, significant hemolysis developed. Six months later, the patient was referred for surgery for valve replacement. Intraoperatively, the 2 devices were found to be correctly placed to occlude 1 paravalvular leak, whereas another leak was present opposite to the first and not treated. This fourth-time major cardiac surgery to replace the aortic valve was uneventful, as was the subsequent postoperative course.


Figure 2
View larger version (125K):
[in this window]
[in a new window]

 
Figure 2. Interventional closure of paravalvular leak in patient 2. Two occluder devices are correctly placed (arrows). AP, Aortic prosthesis.

 
PATIENT 3. The third patient was a 24-year-old woman who had previously undergone 3 cardiac operations. A commissurotomy when she was 5 years old for aortic stenosis was followed by aortic valve replacements at 13 and 14 years of age, the latter for prosthesis dysfunction. A paravalvular leak diagnosed when the patient was 24 years old prompted interventional closure. Unfortunately, severe hemolysis developed immediately after the intervention. Fourth-time cardiac surgery for reoperative aortic valve replacement was performed 1 week later. As with our previous 2 patients, surgery in this challenging case and the subsequent postoperative course both went well.

Discussion

Interventional closure of paravalvular leaks may evolve into an attractive treatment option for patients with multiple previous operations. Initial case reports have confirmed the feasibility of this approach for selected patients but have also revealed a variety of challenges.3Go For instance, residual leaks were observed in 10 of 11 sites of device deployment in a series of 12 patients.4Go Hemolysis may persist, even if the regurgitation is reduced (see patients 2 and 3). Thus the event-free survival (freedom from reoperation, reintervention, and significant residual shunt) was as low as 48% in a series of 24 patients.5Go Common reasons for failure of this demanding intervention are poor imaging and unsuitable anatomy, such as crescent-shaped or multiple leaks and severe calcification. Additional limitations result from inadequate device configuration, because the occluders in use were originally designed for atrial and ventricular septal defects or patent ducts. Consequently, surgery still remains the criterion standard of treatment for paravalvular leaks while the technology of interventional closure is being advanced.

Our program comprised 1842 heart valve surgeries in 2006, including 26 patients (0.14%) operated on for paravalvular leaks. Even in this large series, the 3 cases presented here are unusual because of the multiple previous surgeries. These cases were technically challenging and of high risk, and avoiding fourth-time cardiac surgery for reoperative valve exchange by use of catheter-based intervention would have been preferable. We therefore advocate a multidisciplinary approach that includes interventional cardiologists and surgeons. Depending on level of hemolysis, severity of regurgitation, and anatomy of the paravalvular leak, the risk of surgery must be weighed against the likelihood of success of intervention. Both treatment options are clearly explained to the patient so that an informed decision can be made. Surgical backup in case of interventional complications is provided for every intervention as well. Future studies, perhaps with more suitable device technology, may address a hybrid approach, so that outcomes can be improved even for most complex cases of paravalvular leakage.

References

  1. Akins CW. Results with mechanical cardiac valvular prostheses. Ann Thorac Surg 1995;60:1836-1844.[Abstract/Free Full Text]
  2. Akins CW, Bitondo JM, Hilgenberg AD, Vlahakes GJ, Madsen JC, MacGillivray TE. Early and late results of the surgical correction of cardiac prosthetic paravalvular leaks. J Heart Valve Dis 2005;14:792-800.[Medline]
  3. Pate GE, Al Zubaidi A, Chandavimol M, Thompson CR, Munt BI, Webb JG. Percutaneous closure of prosthetic paravalvular leaks: case series and review. Catheter Cardiovasc Interv 2006;68:528-533.[Medline]
  4. Shapira Y, Hirsch R, Kornowski R, Hasdai D, Assali A, Vaturi M, et al. Percutaneous closure of perivalvular leaks with Amplatzer occluders: feasibility, safety, and shortterm results. J Heart Valve Dis 2007;16:305-313.[Medline]
  5. Hein R, Skowasch M, Wunderlich N, Wilson N, Sievert H. Catheter treatment of paravalvular leaks using Amplatzer devices. Am J Cardiol 2006;98(Suppl S):231M.




This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Dirk Seifert
Reiner Koerfer
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bairaktaris, A.
Right arrow Articles by Koerfer, R.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Bairaktaris, A.
Right arrow Articles by Koerfer, R.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS