J Thorac Cardiovasc Surg 2008;136:1082-1083
© 2008 The American Association for Thoracic Surgery
Hypoxemia after an axial flow pump Jarvik-2000 implantation: Catheter induced
Joonhwa Hong, MD,
Soon J. Park, MD*,
Sunil V. Mankad, MD,
Frank Cetta, Jr., MD,
Norman E. Torres, MD,
Morgan L. Brown, MD
Mayo Clinic, Rochester, Minnesota
Received for publication October 8, 2007; accepted for publication December 2, 2007.
* Address for reprints: Soon J. Park, MD, Cardiovascular Surgery, Mayo Clinic, 200 First Street SW, Rochester, MN 55905. (Email: park.soon{at}mayo.edu).
Left ventricular assist devices (LVADs) are now being implanted frequently as a bridge therapy to transplant or a destination therapy.1-3
Device technology is also changing in that nonpulsatile axial flow pumps are being used as frequently as pulsatile pumps at many centers. LVADs are effective in restoring circulation and unloading the left heart. LVAD support could result in a potential problem of right to left shunt through the patent foramen ovale (PFO),4-8
and the practice of performing transesophageal echocardiography (TEE) looking for PFO at the time of surgery has been well established. We describe an unusual case of significant hypoxemia resulting from a defect in the interatrial septum (IAS) created iatrogenically at the time of an electrophysiologic procedure in a patient who subsequently underwent an axial flow pump implantation.
Clinical Summary
The patient was a 41-year-old man with advanced hypertrophic cardiomyopathy. He had a history of atrial fibrillation and had undergone a transcatheter-based atrioventricular nodal ablation procedure in the past. His condition continued to deteriorate, and he had to be listed for heart transplantation as status 1A supported on 2 inotrope. His echocardiogram demonstrated severe concentric left ventricular hypertrophy with a small ventricular cavity. An estimated left ejection fraction was 35%, and he had a small noncompliant ventricle. He was recommended to undergo LVAD implantation as a bridge therapy to heart transplantation. Given the patient's small left ventricular cavity, the Jarvik-2000 axial pump (Jarvik Heart Inc, New York, NY) was thought to be the best option for him.
Intraoperative TEE demonstrated a small hole in the IAS with a left to right shunt (Figure 1, A). We inspected the atrial septum to close this hole at the time of LVAD implantation. No PFO was noted. There were multiple small holes that were visible in the IAS, and the largest one that appeared to be communicating to left atrium was closed primarily. However, a persistent shunting between the 2 atria was noted on TEE after the patient was weaned from cardiopulmonary bypass support. The shunt was now in the direction of right to left (Figure 1, B) with effective decompression of the left heart on Jarvik axial flow pump support. Because we had failed to identify the hole at the time of initial exploration of the IAS, we decided to close the hole percutaneously via the transcatheter route. The pump speed was reduced to 8000 from 11,000 rpm to maintain as high arterial saturation as possible on 100% oxygen.

View larger version (41K):
[in this window]
[in a new window]
|
Figure 1. A, Intraoperative TEE showing a left to right shunt through the atrial septum before the LVAD implantation. B, Intraoperative TEE showing a right to left shunt through the atrial septum after LVAD implantation.
|
|
The patient was taken to the catheterization laboratory. Intracardiac echocardiography demonstrated a 3-mm hole in the lower part of the IAS with a significant right to left shunt (Figure 2, A). This defect was consistent with a puncture hole usually created in the lower IAS with a catheter from the inferior vena cava at the time of atrioventricular nodal ablation. The Amplatzer (AGA Medical Corp, Golden Valley, Minn) atrial septal defect closure device (6 mm) was inserted to close the shunt. A small residual right to left shunt remained through the device itself (Figure 2, B). Arterial saturation improved from 77% to a modest 90%. Nonetheless, we brought the patient to the intensive care unit expecting further improvement in the arterial saturation over time. The small residual shunt within the Amplatzer device was expected to close soon by forming clots over the mesh. The patient also had significant intrapulmonary shunt based on sampled pulmonary venous blood PO
2 of 114 mm Hg on 100% oxygen before atrial septal defect closure. Arterial saturation increased up to 96% within 1 hour. The patient's condition continued to improve. His FIO
2 requirement decreased to 0.4 by the third day, and he was extubated on the sixth day after LVAD insertion. A follow-up echocardiography revealed no residual shunt through the IAS, and the patient remains on the waiting list for heart transplantation.

View larger version (41K):
[in this window]
[in a new window]
|
Figure 2. A, Intracardiac echocardiogram showing a right to left shunt through the atrial septum. B, Intracardiac echocardiogram showing a tiny residual shunt after atrial septal defect closure using the Amplatzer device (AGA Medical Corp, Golden Valley, Minn). RA, Right atrium; LA, left atrium.
|
|
Discussion
There are several case reports of hypoxemia resulting from right to left shunt through the IAS after LVAD implantation.4-8
Reported cases of shunting usually involve PFO after implantation of pulsatile LVADs. At most centers, performing a TEE study looking for intracardiac shunt through a PFO has been a matter of routine practice.4,5
Our patient developed significant hypoxemia from the right to left shunt through a 3-mm hole created iatrogenically at the time of the electrophysiology procedure. An axial flow pump may create a more significant right to left shunt than a pulsatile pump because it unloads the heart continuously and more effectively depending on the setting of the motor speed. Even a small hole can result in clinically significant hypoxemia with an axial flow pump.
Conclusions
All intracardiac shunts should be closed at the time of surgery for LVAD implantation. However, if hypoxemia from an untreated intracardiac shunt develops in a patient, measures such as reducing airway pressure, inotropic support of right heart function, and agents to reduce pulmonary vascular resistance can be helpful as a temporizing measure.5,8
LVAD flow should be reduced to elevate left atrial pressure while maintaining adequate tissue perfusion. Most importantly, the shunt needs to be closed promptly. A transcatheter-based percutaneous closure of an intracardiac shunt can be an effective option.
References
- Rose EA, Gelijns AC, Moskowitz AJ, et al. Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) Study Group Long-term mechanical left ventricular assistance for end-stage heart failure. N Engl J Med 2001;345:1435-1443.[Medline]
- Park S, Tector A, Piccioni W, et al. Left ventricular assist devices as destination therapy: a new look at survival. J Thorac Cardiovasc Surg 2005;129:9-17.[Abstract/Free Full Text]
- Lietz K, Long JW, Kfoury AG, et al. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection. Circulation 2007;116:497-505.[Abstract/Free Full Text]
- Baker JE, Stratmann G, Hoopes C, et al. Profound hypoxemia resulting from shunting across an inadvertent atrial septal tear after left ventricular assist device placement. Anesth Analg 2004;98:937-940.[Abstract/Free Full Text]
- Kilger E, Strom C, Frey L, et al. Intermittent atrial level right to left shunt with temporary hypoxemia in a patient during support with a left ventricular assist device. Acta Anaesthesiol Scand 2000;44:125-127.[Medline]
- Baldwin RT, Duncan JM, Frazier OH, et al. Patent foramen ovale: a cause of hypoxemia in patients on left ventricular support. Ann Thorac Surg 1991;52:865-867.[Abstract/Free Full Text]
- Nguyen DQ, Das GS, Grubbs BC, et al. Transcatheter closure of patent foramen ovale for hypoxemia during left ventricular assist device support. J Heart Lung Transplant 1999;18:1021-1023.[Medline]
- Srinivas CV, Collins N, Borger MA, Horlick E, Murphy PM. Hypoxemia complicating LVAD insertion: novel application of the Amplatzer PFO occlusion device. J Card Surg 2007;22:156-158.[Medline]