J Thorac Cardiovasc Surg 2008;136:943-947
© 2008 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
Department of Cardiothoracic and Vascular Surgery, G. B. Pant Hospital, Maulana Azad Medical College, New Delhi, India
Received for publication December 30, 2007; revisions received April 28, 2008; accepted for publication June 7, 2008. * Address for reprints: Sumit Narang, MS, Department of Cardiothoracic and Vascular Surgery, G. B. Pant Hospital, Maulana Azad Medical College, New Delhi, India 110002. (Email: sumit_narang{at}yahoo.com).
| Abstract |
|---|
|
|
|---|
Methods: Sixty-two patients were recruited for the study. Our protocol was to consider all patients older than 55 years for bioprostheses, and also younger patients were implanted with a bioprosthesis if they wanted to avoid anticoagulation. Patients selected for bioprostheses were randomly assigned to receive stentless (group A) and stented (group B) bioprostheses, depending on the treating unit. Patients in groups A and B were further divided into subgroups I and II based on left ventricular ejection fractions of 50% or greater and less than 50%, respectively.
Results: At 18 ± 3 months postoperatively, the effective orifice area was greater in group A versus group B. Left ventricular ejection fraction, left ventricular mass index, functional class, and mean gradient were similar in patients of subgroup I (left ventricular ejection fraction >50%) from both groups. However, there was a significant difference between all except mean gradient in patients of subgroup II (left ventricular ejection fraction <50%) from both groups. Also, in the patient subgroup implanted with valves of less than 19 mm (group A, 4 patients; group B, 3 patients), there was a significant difference in left ventricular mass index and mean gradient.
Conclusion: In patients with left ventricular impairment or a small aortic annulus, stentless bioprostheses might allow for greater improvement in left ventricular function postoperatively.
| Introduction |
|---|
|
|
|---|
Their design is postulated to impart characteristics that more closely emulate the normal physiology of the aortic valve and root. It has been further suggested that more favorable hemodynamic performance might lead to enhanced resolution of left ventricular hypertrophy (LVH),2
a known potent cause of premature mortality.3,4
Modern stented valves have been shown to have excellent hemodynamic characteristics.1,5
Previous randomized controlled trials have used a variety of stented and stentless valves, several of which are currently outmoded.6,7
Our aim was to conduct a randomized controlled trial to compare widely accepted state-of–the-art stentless and stented prostheses. Clinical outcomes, hemodynamic performance, and postoperative left ventricular mass (LVM) regression were the principal outcomes assessed.
| Materials and Methods |
|---|
|
|
|---|
All patients older than 55 years were considered for a bioprosthesis. Patients younger than 55 years were implanted with a bioprosthesis if they wanted to avoid anticoagulation. All patients with aortic aneurysms, severely asymmetric aortic roots, annuloaortic ectasia, active aortic valve infection, renal failure requiring dialysis, and requirement for additional cardiac procedures were excluded from the study.
All patients were implanted with valves of 19 mm or larger. The Freestyle porcine tissue valve (Medtronic, Inc) has been used as a stentless bioprosthetic valve, and the Carpentier–Edwards Perimount Standard aortic valve (Edwards Life Sciences Corp) has been used as a stented bioprosthesis. In patients with an aortic annulus of less than 19 mm, the aortic root enlargement technique (Nicks) was used to implant a 19-mm valve. Patients with very small aortic roots requiring complex root enlargement procedures were excluded. We have accepted the 19-mm valve size because of the smaller average body surface area in our patients.
Stented valves were implanted with the continuous or interrupted suture technique, and stentless valves were implanted with the inclusion cylinder technique.
The inclusion cylinder implantation technique involves a proximal suture line (continuous or interrupted) like the subcoronary technique in a circular plane coursing below the commissures. Generous openings are made by excising the sinuses facing the right and left main coronary ostia. These are then tacked around the ostia, much as a root inclusion Bentall procedure, with continuous 5–0 polypropylene sutures. The aortotomy is then closed in standard fashion after trimming the device top down as necessary to incorporate the closure, making sure that the complete circle of the sinotubular junction is left intact.
Postoperatively, all patients received 75 mg of aspirin as prophylaxis against thromboembolism. Patients were formally anticoagulated only if there were specific indications, such as atrial fibrillation. Clinical and echocardiographic parameters were recorded preoperatively, and functional status and echocardiographic parameters were reevaluated. Mean follow-up was 18 ± 3 months for both groups. Echocardiographic parameters recorded were effective orifice area (EOA), LVEF, left ventricular end-diastolic diameter (LVEDD), mean gradient, and left ventricular mass index (LVMI). Functional status was recorded in terms of New York Heart Association functional class.
The protocol of the study was approved by the institutional review board before the start of the study. Informed consent was obtained from each patient before inclusion in the study.
The data have been expressed as means and standard deviations. The Student t test has been used for statistical analysis.
| Results |
|---|
|
|
|---|
|
50%) of both groups (Table 2
). However, there was a significant difference between all these parameters except mean gradient in patients of subgroup II (LVEF <50%) from both groups (LVEF, 56% vs 45%; LVMI, 115 vs 140 g/m2; and mean gradient, 8 vs 11 mm Hg), with more improvement in functional status of patients of group A (subgroup II) compared with those of group B (subgroup II; Table 3
). In the patient subgroup with a small aortic root who were implanted with a 19-mm valve (group A, 4 patients; group B, 3 patients), although the number of patients was much less, there was a significant difference in LVM, EOA, and mean gradient between this subgroup of patients of groups A and B (LVMI, 117 vs 138 g/m2; mean gradient, 11 vs 25 mm Hg; and EOA, 1.57 vs 1.34 cm2, respectively), with better functional status in patients of group A (Table 4
).
|
|
|
| Discussion |
|---|
|
|
|---|
Our inability to document any hemodynamic advantages or superior LVM regression after stentless AVR parallels the findings of other recent randomized trials.8-10
Newer-generation stented valves have a lower profile and more narrow sewing rings, which produce less impedance to transvalvular flow than older designs. Excellent hemodynamic performance of stented prostheses in the aortic position has been documented by a number of investigators.1,5
It has been inferred that the use of a stentless bioprosthesis might be particularly advantageous in the setting of preoperative ventricular impairment. This is based on the findings of a few investigators who suggest that stentless valves might allow for greater improvement in left ventricular function postoperatively.11-14
It has been demonstrated that early after AVR, stentless prostheses allow for a greater and more immediate reduction in left ventricular systolic wall stress. This observation has been correlated with improved ventricular function in the early postoperative period.11-14
This difference wanes after 12 months postoperatively, except in patients with left ventricular impairment. Our randomized study seems to substantiate this observation with a difference in LVMI regression, and improvement in functional status between groups A and B at a mean follow-up of 18 ± 3 months was statistically significant in patients with left ventricular impairment preoperatively. In patients with reduced ventricular function, this finding might have important implications for long-term outcome. Also, we found that preoperative LVEDD was an independent factor for improvement in left ventricular function and LVM regression in patients with an LVEF of less than 50%. All patients with AR and an LVEDD of greater than 6.5 cm had less improvement in both these parameters postoperatively compared with those who had an LVEDD of less than 6.5 cm preoperatively. Patients with aortic stenosis and an LVEDD of greater than 5.5 cm had less improvement in both these parameters postoperatively compared with those who had an LVEDD of less than 5.5 cm preoperatively. This finding is similar to that shown by Tafreshi and colleagues,15
who demonstrated end-diastolic dimensions to be an important predictor for persistent postoperative left ventricular dysfunction.
In the present study we have chosen to implant bioprosthetic valves in patients older than 55 years. The age group in our study is younger than that for bioprosthetic valve implantation seen in other studies (ie, >65 years age). We have lowered the age limit for implanting a bioprosthetic valve in our study because of the lower average life expectancy in our country.
Our observations suggest that the routine use of stentless bioprostheses over stented valves in all patients on the presumption that they confer hemodynamic advantages is unlikely to be justified. However, in specific patient groups undergoing AVR, stentless valves might provide important benefits. We noted a trend toward improved hemodynamic performance of stentless valves in patients with left ventricular impairment preoperatively and those with smaller aortic annuli. In patients with an annular diameter of less than or equal to 19 mm who were implanted with a 19-mm valve, the mean systolic gradient was significantly lower in stentless valve recipients, with significantly better functional status at 18 ± months postoperatively.14
However, this patient group comprised only a small proportion of our entire study cohort. In our study patients receiving stentless valves had a significantly greater EOA compared with those with stented valves 18 ± months after AVR. The absence of significant AR in both groups at a mean follow-up of 18 ± months shows durability of both these valve types; however, long-term follow-up is required to substantiate this finding. Also, this study suggests that patients with huge left ventricular dimensions (LVEDD >6.5 cm in severe AR and >5.5 cm in severe aortic stenosis) with an LVEF of less than 50% had less LVM regression and functional improvement after AVR. Hence these patients should undergo operations early before deterioration in LVEF.
In conclusion, both stented and stentless bioprostheses are associated with excellent clinical and hemodynamic outcomes at a mean follow-up of 18 ± 3 months. Comparable hemodynamics and LVM regression can be achieved with a newer-generation stented bioprosthesis. In patients with ventricular impairment and those with a small annulus, stentless bioprostheses might allow for greater improvement in left ventricular function and greater improvement in functional class postoperatively. Hence despite increased operative time for implantation of stentless valves, this is the subgroup of patients who were benefited with implantation of a stentless valve. With studies showing long-term durability of stentless valves13
and improved outcomes of redo operations, stentless valves can be considered as a first option in younger patients with left ventricular impairment or with a small aortic annulus.
We also conclude that patients with huge left ventricular dimensions should be operated on early before deterioration of left ventricular function to achieve better postoperative results.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. A. Funder Current status on stentless aortic bioprosthesis: a clinical and experimental perspective Eur J Cardiothorac Surg, December 21, 2011; (2011) ezr141v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Kalavrouziotis, J. Rodes-Cabau, R. Bagur, D. Doyle, R. De Larochelliere, P. Pibarot, and E. Dumont Transcatheter Aortic Valve Implantation in Patients With Severe Aortic Stenosis and Small Aortic Annulus J. Am. Coll. Cardiol., August 30, 2011; 58(10): 1016 - 1024. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Murtuza, J. R. Pepper, C. Jones, P. Nihoyannopoulos, A. Darzi, and T. Athanasiou Does stentless aortic valve implantation increase perioperative risk? A critical appraisal of the literature and risk of bias analysis Eur J Cardiothorac Surg, May 1, 2011; 39(5): 643 - 652. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||