JTCS KCI
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Predescu, D.
Right arrow Articles by Lee, K.-J.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Predescu, D.
Right arrow Articles by Lee, K.-J.
Related Collections
Right arrow Congenital - acyanotic
Right arrow Congestive Heart Failure

J Thorac Cardiovasc Surg 2008;136:1223-1228
© 2008 The American Association for Thoracic Surgery


Surgery for Congenital Heart Disease

Complete heart block associated with device closure of perimembranous ventricular septal defects

Dragos Predescu, MD, Rajiv R. Chaturvedi, MD, PhD, Mark K. Friedberg, MD, Lee N. Benson, MD, Akira Ozawa, MD, Kyong-Jin Lee, MD*

Cardiac Diagnostic and Interventional Unit, the Labatt Family Heart Centre, the Hospital for Sick Children, the University of Toronto School of Medicine, Toronto, Ontario, Canada

Received for publication November 24, 2007; revisions received January 31, 2008; accepted for publication February 19, 2008.

* Address for reprints: Kyong-Jin Lee, MD, the Hospital for Sick Children, 555 University Ave, Toronto, Ontario, Canada M5G 1X8. (Email: kyong-jin.lee{at}sickkids.ca).


    Abstract
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
Objective: The development of the Amplatzer Membranous VSD Occluder (AGA Medical Corp, Plymouth, Minn) for closure of the perimembranous ventricular septal defect has ameliorated many of the technical difficulties of previous devices. Application of this new technology requires comparative evaluation with the current standard of surgical repair. We report our experience of complete heart block associated with device closure of a large perimembranous ventricular septal defect with unequivocal indications for intervention.

Methods: We performed a retrospective review of 20 patients between January 2003 and August 2005 who underwent perimembranous ventricular septal defect device closure, 18 with hemodynamically large shunts meeting the surgical criteria for intervention. The median age was 1.6 years (range, 0.5–16.2 years), and the median weight was 9.7 kg (range, 6.2–43 kg).

Results: Acute complete shunt occlusion was achieved in all patients. There were no acute procedural complications. The median follow-up time was 23.1 months (range, 1–37.8 months). Four (22%) had complete heart block at 17 days, 4.2 months, 8.8 months, and 37.5 months after implantation, respectively. No risk factors were identified for development of complete heart block, including age, weight, trisomy 21, preceding conduction abnormalities, perimembranous ventricular septal defect size related to body surface area or device size, and progressive device flattening.

Conclusions: Device closure of large perimembranous ventricular septal defects in infants and children with the Amplatzer Membranous VSD Occluder resulted in excellent closure rates but an unacceptably high rate of complete heart block.



Abbreviations and Acronyms CHB = complete heart block; PMVSD = perimembranous ventricular septal defect; PPM = permanent pacemaker; RBBB = right bundle branch block; VSD = ventricular septal defect



    Introduction
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
Perimembranous ventricular septal defects (PMVSDs) in infants and young children are the most common cause of a hemodynamically significant ventricular septal defect (VSD). Although device closure of certain VSDs (muscular, residual postsurgical, posttraumatic, and postinfarction defects) is highly effective, either by means of percutaneous catheterization or collaborative hybrid-type approaches,1-4Go initial attempts at transcatheter PMVSDs were largely unsuccessful.5,6Go

The development of the Amplatzer Membranous VSD Occluder (AGA Medical Corp, Plymouth, Minn) has led to numerous case series, and more than 2000 devices have been implanted worldwide.7-13Go Short- and medium-term follow-up indicates that many of the shortcomings of previous devices have been overcome, especially with respect to technical feasibility, shunt occlusion, and valvular dysfunction. However, there has been considerable uncertainty about the clinical importance of some of the occluded VSDs and also the long-term follow-up of these patients.10Go This new technology requires comparison with the standard practice of surgical repair.

After this preliminary experience, our institution embarked on an evaluation of device closure of the PMVSD with unequivocal indications for intervention in patients who would have otherwise undergone surgical closure. Reports of device-associated complete heart block (CHB) have emerged.13-17Go

We report on our experience of CHB associated with device closure of the hemodynamically significant PMVSD.


    Materials and Methods
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
Review of the cardiac database identified all patients who underwent PMVSD device closure with the Amplatzer Membranous VSD Occluder. Clinical records, angiograms, electrocardiograms, and echocardiograms were reviewed. The study was approved by the Research Ethics Board at the Hospital for Sick Children.

Patient Population
Patients were eligible for device closure if 1 or more of the following indications were present: pulmonary-to-systemic blood flow (Qp/Qs) ratio of greater than 1.5, heart failure not controlled with medications, evidence of left heart volume loading, and a history of previous endocarditis. Exclusion criteria included weight of less than 6 kg, inlet extension of the PMVSD, pulmonary vascular resistance index of greater than 7 WU/m2 that was unresponsive to oxygen, additional lesions requiring surgical intervention (right/left ventricular outflow tract obstruction and aortic regurgitation), and parental preference for surgical intervention. For the purpose of this report, we focus on patients with either a large shunt (Qp/Qs ratio of ≥2 or with pulmonary hypertension in the presence of a Qp/Qs ratio of <2). Pulmonary hypertension was defined as a mean pulmonary artery pressure of 20 mm Hg or greater based on the original natural history studies.18Go

VSD Sizing and Device Implantation
PMVSD anatomy and initial sizing were determined by means of transthoracic echocardiography. The parasternal short-axis and apical 4-chamber views were used to measure the short- and long-axis diameters, respectively, of the defect. Where the VSD was restrictive, size was based on the smallest jet width by means of color imaging. Baseline valvular regurgitation, aortic cusp prolapse, and the pressure gradient across the VSD were assessed.

All PMVSDs were closed with the Amplatzer Membranous VSD Occluder. Cardiac catheterization was performed after achievement of general anesthesia. Right and left heart hemodynamic assessments were performed. Device implantation followed the recommended protocol.7,8,11Go Briefly, the PMVSD was crossed with a wire from the left ventricular side, the wire was snared in the pulmonary artery and exteriorized to the femoral vein to form an arteriovenous circuit, and the device was implanted through a long sheath traversing the PMVSD from the right to the left ventricle. Transesophageal echocardiography was used to assist in assessment of defect size and device position and to exclude device distortion of surrounding cardiac structures. The device was generally selected to be 1 to 2 mm larger than the VSD, as assessed by the combination of echocardiography and left ventricular angiography. Given that the device is circular in shape and the PMVSD more elliptical, the larger of the short- and long-axis defect diameters was used. The Amplatzer Membranous VSD Occluder was available in 2-mm increments at the time of this study.

Follow-up
All patients were monitored in the hospital for 24 hours after device implantation, and chest radiography, electrocardiography, and echocardiography were performed before discharge. Outpatient follow-up was with an electrocardiogram and echocardiogram at 1 week, 1 month, 6 months, 1 year, and then yearly.

Follow-up echocardiograms were reviewed to assess residual shunt and valvular regurgitation. We found attempts to measure disk diameter by means of echocardiography unreliable and therefore documented our subjective impression of whether the device had flattened over time.

Statistical Analysis
Descriptive statistics are reported as medians and ranges. The numeric variables for the 18 patients with hemodynamically significant PMVSDs were split into CHB and non-CHB groups and compared with the Mann–Whitney test. The Fisher test was used to compare categorical variables between the groups. All data analysis was performed in R 2.5.1.19Go


    Results
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
Between January 2003 and August 2005, 20 patients (12 male patients) underwent PMVSD device implantation. Patient demographics and clinical and hemodynamic data are summarized in Go Table 1. The median age was 1.6 years (range, 0.5–16.2 years), and the median weight was 9.7 kg (range, 6.2–43 kg). Eighteen patients met unequivocal criteria for VSD intervention (Qp/Qs ratio of ≥2 [n = 10] or mean pulmonary artery pressure of ≥20 mm Hg with a Qp/Qs ratio of <2:1 [n = 8]). Patient 12 had left ventricular noncompaction with restrictive cardiomyopathy and high pulmonary artery pressures. Defects that did not meet the criteria of a large shunt were in patient 19, who was referred for closure after a documented episode of infective endocarditis, and patient 20, who was a 7-year-old who was asymptomatic, taking no medication, and who had a cardiac nuclear angiogram that was difficult to interpret and reported a Qp/Qs ratio of between 1.5:1 and 3:1. Echocardiographic analysis demonstrated a gradient of 103 mm Hg across the VSD, a left ventricular end-diastolic dimension z score of +2.7, and a left atrial diameter z score of +1.9. At catheterization, the Qp/Qs ratio was 1.8:1, and the mean pulmonary artery pressure was 15 mm Hg. There was a strong preference for closure by the referring cardiologist and the parents.


View this table:
[in this window]
[in a new window]

 
Table 1 Patient characteristics and hemodynamic and procedural data
 
During this time period, 25 patients weighing 6 kg or more underwent surgical intervention with a PMVSD as the primary indication for the operation. Those with the secondary diagnoses of atrial septal defect, patent ductus arteriosus, and patent foramen ovale were included. Of this group, 13 (52%) weighed 8 kg or more, 12 (48%) weighed 10 kg or more, and only 1 weighed 16 kg or more, reflecting our institutional criteria for PMVSD intervention being predominantly in small patients with large shunts. None of these patients have had CHB.

A device was deployed successfully in the standard fashion in all cases. The median fluoroscopic time was 27 minutes (range, 13–50 minutes). Acute complete VSD shunt occlusion was achieved in all patients. There were no major procedural complications. In 4 children (patients 7, 10, 11, and 12) the initial device prolapsed through the defect and was replaced with the next size up (2 mm larger). In 2 children the device size was 3 mm greater than the VSD size. In patient 12 the initial 12-mm device prolapsed through the 11-mm VSD, and a 14-mm device was successfully implanted. Patient 6 had a 9-mm VSD closed with a 12-mm device. This device looked bulky after delivery but could not be retrieved because of entanglement within the tricuspid valve. Initially, tricuspid valve regurgitation was moderate, but 13 months later, it was mild.

Follow-up time at the last clinical assessment or until heart transplantation or development of CHB was a median of 23.1 months (range, 1–37.8 months). One patient had only 1 month of follow-up (patient 13). There were no deaths or device embolization. Twelve patients had new or increased valvular regurgitation (60%), none of more than a mild degree at last follow-up. One child had self-limiting hemolysis that did not require blood transfusion (patient 1). The patient with restrictive cardiomyopathy had disease progression and underwent successful transplantation 2.2 years after device implantation.

New conduction abnormalities were observed after device implantation in 16 (80%) children, 7 of whom had pre-existing abnormalities. The new abnormalities were bifascicular block (n = 1), incomplete right bundle branch block (RBBB; n = 2), complete RBBB (n = 3), and first-degree heart block (n = 1). These were transient in 3 and permanent in 13 (65%) patients. Four (20%) patients had CHB presenting at 17 days, 4.2 months, 8.8 months, and 37.5 months, respectively, after implantation.

Patient 4 presented with 3 syncopal episodes, 2 of which were associated with seizures. High-grade second-degree heart block and periods of asystole up to 30 seconds were observed. Emergency treatment with an isoproterenol infusion and transvenous pacing was initiated. The parents declined a trial of corticosteroids, and the device was removed at the time of permanent pacemaker (PPM) implantation and sinus rhythm recovered 1 day later. In the other 3 patients, CHB was diagnosed during routine follow-up. Ventricular rates were 43, 49, and 50 beats/min. All patients had at least 1 Holter examination demonstrating continuous sinus rhythm before development of CHB. All patients underwent PPM implantation within 0 to 10 days of diagnosis. The device was left in situ in 3 patients with late-onset CHB, and all remain pacemaker dependent.

We sought to identify risk factors for CHB using a detailed review of the cases, including procedural details, echocardiograms, and angiograms. There was no discernible difference between the CHB and non-CHB groups (Go Table 2), irrespective of whether we examined only the 18 patients who met the hemodynamic criteria for intervention or the extended group of all 20 patients (data not shown). Electrocardiograms with or without Holter examination after implantation and before diagnosis of CHB were normal for 2 patients and in 1 patient indicated new RBBB. Patient 1 with trisomy 21 had baseline PR-interval prolongation and progressed to bifascicular block (24 hours), trifascicular block (2 weeks), and then CHB (4.2 months). In the other 3 patients, there was no warning of CHB. Neither of the 2 patients with a device 3 mm larger than the VSD size had CHB.


View this table:
[in this window]
[in a new window]

 
Table 2 Comparison between non-CHB and CHB groups (n = 18)
 
Immediately after implantation, the device disks did not appear fully flattened in 14 (78%) patients and progressively flattened in 7 (50%) patients. However, the 2 groups did not differ significantly with respect to initial or subsequent device shape (P = 1.00 and P = .6, respectively).

Concern that use of the largest diameter of the VSD might lead to device oversizing and CHB led the manufacturer (AGA Medical Corp) to recommend a new formula (square root of the product of the short- and long-axis diameters of the PMVSD) for device sizing based on an assumed elliptical shape. Retrospective application of this new formula in our series recommended smaller devices in 10 (71%) of the patients without CHB and in 3 (75%) of the patients with CHB. The ratio of the new device to the used device did not differ between the 2 groups (Table 2). All 4 devices that dislodged and required larger sizes were in the non-CHB group. In 2 of the 4 instances of device dislodgement, a device that was the same size or smaller than that recommended by this new formula was actually initially implanted but subsequently required a larger implant.


    Discussion
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
The development of the Amplatzer Membranous VSD Occluder and similar devices has ameliorated many of the technical difficulties and decreased distortion of the surrounding valve apparatus that occurred with previous devices. These advantages have lead to renewed interest in this technique. Initial experience from multiple institutions reported no mortality, high acute technical success (89% to 100%), and high occlusion rates (84% to 100%) in the absence of any major concerns.7-13Go Similar to others, we also achieved high technical success with 100% shunt occlusion and good acute outcomes. Longer-term follow-up in our institution unveiled a high frequency (22%) of CHB. Concurrently, sporadic published reports of acute and late CHB emerged with a reported incidence of 1% to 5%.14-17Go Based on these data, PMVSD device implantation was terminated at our institution. Most recently, Butera and colleagues13Go report longer-term follow-up in 104 patients at a median of 38.5 months. The incidence of CHB was 8.7%, and PPM was required in 6 (5.7%) patients, 2 in the early phase and 4 during late follow-up.

Conduction system injury from mechanical trauma/compression by the delivery system or device causing acute intraprocedural CHB has a reasonable probability of early resolution. Chronic inflammation or fibrosis is more likely to be responsible for the late-onset type.15,20Go Clearly, some cases of early CHB resolve spontaneously or with temporary pacing, corticosteroid and/or high-dose acetylsalicylic acid therapy, or device removal.10,16Go Recovery of sinus rhythm after PPM implantation without device removal has also been reported.7,14Go However, in early-onset CHB it is likely that the highest chance of sinus rhythm recovery is with device removal.13Go Of most concern is the late onset, lack of predictive factors, and potentially life-threatening presentation of CHB. Hemodynamic instability, including syncope, has been the presenting event in 10 reported cases, including our patient with seizure. Others were discovered on routine scheduled follow-up.13-17Go CHB occurred as late as 37 months after device implantation in our series.

Identifying Risk Factors for Device-associated CHB
PMVSD anatomy
The simple anatomic classification of a VSD in the perimembranous region belies the wide variability of its shape. The common denominator of such defects is the fibrous continuity between the leaflets of the tricuspid and aortic valves, which form part of the defect's margins, with the conduction tissue coursing posteroinferiorly.21Go This begs the question of whether one device shape fits all PMVSDs and more prudently whether such a device can avoid damaging the conduction tissue.

Because of concern that inlet extension of the VSD might predispose to CHB, these patients were specifically excluded in our series.16Go

Device sizing
VSD sizing is challenging because the complexity of PMVSD geometry cannot be captured by measuring 1 or 2 dimensions available by means of standard echocardiographic and angiographic imaging. Echocardiographic analysis can overestimate VSD diameter by 21%.22Go Efforts to minimize possible oversizing include new recommendations for device sizing, accounting for the elliptical shape of the PMVSD. Retrospective application of this new device size formula failed to identify oversizing as a risk factor. There are no reports of prospective application of this new method of device size selection. Device oversizing was not a risk factor for CHB in the series by Butera and colleagues.13Go

Progressive device flattening of an originally oversized device has been hypothesized as a mechanism for the development of CHB. However, this was not observed in the patients with CHB in other series14,15Go and was not significantly different in the patients with and without CHB in our series.

Patient characteristics
It has been suggested that smaller infants might be at higher risk, but in the reported cases of CHB, age ranged from 2.5 to 12 years, and weights ranged from 14 to 45 kg.13-16Go Weight in our series was lower than that of other series, with the median weight for the entire cohort being 9.7 kg, reflecting our criteria for VSD intervention. There was no significant difference in weight between the CHB and non-CHB groups. In fact, the smallest 2 patients, weighing 6.2 and 6.8 kg, both with large shunts and pulmonary hypertension, did not have CHB. The minimum weight at which device implantation can be considered remains to be determined.

PMVSD surgery
VSD management aims to avoid morbidity and prevent premature mortality from chronic heart failure and pulmonary hypertension.23,24Go Excellent outcomes are expected for the timely diagnosed PMVSD. Surgical intervention for the isolated PMVSD in experienced centers has achieved this with minimal mortality (0% at our institution since the year 2000, unpublished data) and a very low incidence of CHB (0.7% to 1%).25,26Go Tucker and associates26Go report from the Pediatric Cardiac Care Consortium database an incidence of PPM implantation in 0.8% of 1739 patients who weighed more than 8 kg and were primarily operated on to repair PMVSDs (ie, comparable patients eligible for PMVSD device closure). In the contemporary era of 1315 patients, the PPM rate was 0.4%, and hospital mortality was 0.2%.26Go Morbidity remains with requirement of cardiopulmonary bypass, sternotomy, and psychosocial effect and thus provides impetus for the search for alternative therapies.27,28Go Until there are data to demonstrate that the incidence of CHB can be decreased by changes in patient selection or device modification, the potential benefits of reducing surgical intervention–associated morbidity is not superseded by the risk of sudden, potentially life-threatening, late-onset CHB. With this rationale, in these patients with hemodynamically significant PMVSDs (ie, those with an unequivocal indication for intervention), we believe that surgical closure remains the treatment of choice.


    Limitations
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
This is a single-center experience with small patient numbers that is focused on the large PMVSDs that in our institution would otherwise have been referred for surgical intervention. Patient number was limited by the termination of this technique based on CHB incidence. It is not representative of the entire PMVSD device experience, which includes many older and larger patients and smaller defects. In the study by Butera and coleagues,13Go median weight was 26.5 kg, mean pulmonary artery pressure was 16 ± 5 mm Hg, and 37% of patients were older than 10 years. In this regard controversy exists regarding the contemporary criteria for VSD intervention because many of the patients in this series would not have met the criteria for VSD intervention in our institution.27-30Go Only one type of device was used, and it is conceivable that alternative designs might be more favorable.


    Conclusion
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 
We sought to close significant PMVSDs in infants and small children who would otherwise have required cardiac surgery. Device architecture and sizing to avoid the conduction tissue around the complex geometry of the PMVSD has significant limitations. Despite excellent results for closure, there was a high rate of CHB (22%). No risk factors for CHB were identified. The late and insidious nature of the onset of CHB is a critical caveat for device closure of these large defects. Long-term follow-up is essential in patients who have had these devices implanted.


    References
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Limitations
 Conclusion
 References
 

  1. Bacha EA, Cao QL, Galantowicz ME, Cheatham JP, Fleishman CE, Weinstein SW, et al. Multicenter experience with perventricular device closure of muscular ventricular septal defects. Pediatr Cardiol 2005;26:169-175.[Medline]
  2. Holzer R, Balzer D, Cao QL, Lock K, Hijazi ZM. Amplatzer Muscular Ventricular Septal Defect Investigators. Device closure of muscular ventricular septal defects using the Amplatzer muscular ventricular septal defect occluder: immediate and mid-term results of a U.S. registry. J Am Coll Cardiol 2004;43:1257-1263.[Abstract/Free Full Text]
  3. Holzer R, Balzer D, Amin Z, Ruiz CE, Feinstein J, Bass J, et al. Transcatheter closure of postinfarction ventricular septal defects using the new Amplatzer muscular VSD occluder: results of a U.S. registry. Catheter Cardiovasc Interv 2004;61:196-201.[Medline]
  4. Knauth AL, Lock JE, Perry SB, McElhinney DB, Gauvreau K, Landzberg MJ, et al. Transcatheter device closure of congenital and postoperative residual ventricular septal defects. Circulation 2004;110:501-507.[Abstract/Free Full Text]
  5. Rigby ML, Redington AN. Primary transcatheter umbrella closure of perimembranous ventricular septal defect. Br Heart J 1994;72:368-371.[Abstract/Free Full Text]
  6. Sideris EB, Walsh KP, Haddad JL, Chen CR, Ren SG, Kulkarni H. Occlusion of congenital ventricular septal defects by the buttoned device. "Buttoned device" Clinical Trials International Register. Heart 1997;77:276-279.[Abstract/Free Full Text]
  7. Hijazi ZM, Hakim F, Haweleh A, Madani A, Tarawna W, Hiari A, et al. Catheter closure of perimembranous ventricular septal defects using the new Amplatzer membranous VSD occluder: initial clinical experience. Catheter Cardiovasc Interv 2002;56:508-515.[Medline]
  8. Fu Y-C, Bass J, Amin Z, Radtke W, Cheatham JP, Hellenbrand WE, et al. Transcatheter closure of perimembranous ventricular septal defects using the new Amplatzer membranous VSD Occluder. J Am Coll Cardiol 2006;47:319-335.[Abstract/Free Full Text]
  9. Holzer R, de Giovanni J, Walsh KP, Tometzki A, Goh TH, Hakim F, et al. Transcatheter closure of perimembranous ventricular septal defects using the Amplatzer Membranous VSD Occluder: immediate and midterm results of an international registry. Catheter Cardiovasc Interv 2006;68:620-628.[Medline]
  10. Masura J, Gao W, Gavora P, Sun K, Zhou AQ, Jian S, et al. Percutaneous closure of perimembranous ventricular septal defects with the eccentric Amplatzer device: multicenter follow-up study. Pediatr Cardiol 2005;26:216-219.[Medline]
  11. Pedra CA, Pedra SR, Esteves CA, Pontes Jr. SC, Braga SL, Arrieta SR, et al. Percutaneous closure of perimembranous ventricular septal defects with the Amplatzer device: technical and morphological considerations. Catheter Cardiovasc Interv 2004;61:403-410.[Medline]
  12. Pinto RJ, Dalvi BV, Sharma S. Transcatheter closure of perimembranous ventricular septal defects using Amplatzer Asymmetric Ventricular Septal Defect Occluder: preliminary experience with 18–month follow up. Catheter Cardiovasc Interv 2006;68:145-152.[Medline]
  13. Butera G, Carminati M, Chessa M, Piazza L, Micheletti A, Negura DG, et al. Transcatheter closure of perimembranous ventricular septal defects—early and long-term results. J Am Coll Cardiol 2007;50:1189-1195.[Abstract/Free Full Text]
  14. Walsh MA, Bialkowski J, Szkutnik M, Pawelec-Wojtalik M, Bobkowski W, Walsh KP. Atrioventricular block after transcatheter closure of perimembranous ventricular septal defects. Heart 2006;92:1295-1297.[Abstract/Free Full Text]
  15. Butera G, Chessa M, Carminati M. Late complete atrioventricular block after percutaneous closure of a perimembranous ventricular septal defect. Catheter Cardiovasc Interv 2006;67:938-941.[Medline]
  16. Yip WC, Zimmerman F, Hijazi ZM. Heart block and empirical therapy after transcatheter closure of perimembranous ventricular septal defect. Catheter Cardiovasc Interv 2005;66:436-441.[Medline]
  17. Sullivan ID. Transcatheter closure of perimembranous ventricular septal defect: is the risk of heart block too high a price?. Heart 2007;93:284-286.[Abstract/Free Full Text]
  18. Weidman WH, Blount SG, DuShane JW, Gersony WM, Hayes CJ, Nadas AS. Clinical course in ventricular septal defect. Circulation 1977;56:I56-I69.[Medline]
  19. R Development Core Team R: a language and environment for statistical computing. Vienna (Austria): R Foundation for Statistical Computing; 2007Available at:http://www.R-project.org 2007.
  20. Kreutzer J, Ryan CA, Gauvreau K, Van Praagh R, Anderson JM, Jenkins KJ. Healing response to the Clamshell device for closure of intracardiac defects in humans. Catheter Cardiovasc Interv 2001;54:101-111.[Medline]
  21. McCarthy KP, Ho SY, Anderson RH. Categorisation of ventricular septal defects: review of the perimembranous morphology. Images Paediatr Cardiol 2000;3:24-40.
  22. Johnson TB, Fyfe DA, Thompson RP, Kline CH, Swindle MM, Anderson RH. Echocardiographic and anatomic correlation of ventricular septal defect morphology in newborn Yucatan pigs. Am Heart J 1993;125:1067-1072.[Medline]
  23. Campbell M. Natural history of ventricular septal defect. Br Heart J 1971;33:246-257.[Free Full Text]
  24. Weidman WH, Blount SG, DuShane JW, Gersony WM, Hayes CJ, Nadas AS. Clinical course in ventricular septal defect. Circulation 1977;56:I56-I69.[Medline]
  25. Andersen HØ, de Leval MR, Tsang VT, Elliott MJ, Anderson RH, Cook AC. Is complete heart block after surgical closure of ventricular septum defects still an issue?. Ann Thorac Surg 2006;82:948-957.[Abstract/Free Full Text]
  26. Tucker EM, Pyles LA, Bass JL, Moller JH. Permanent pacemaker for atrioventricular conduction block after operative repair of perimembranous ventricular septal defect. J Am Coll Cardiol 2007;50:1196-1200.[Abstract/Free Full Text]
  27. Neumayer U, Stone S, Somerville J. Small ventricular septal defects in adults. Eur Heart J 1998;19:1573-1583.[Abstract/Free Full Text]
  28. Backer CL, Winters RC, Zales VR, Takami H, Muster AJ, Benson Jr. DW, et al. Restrictive ventricular septal defect: how small is too small to close?. Ann Thorac Surg 1993;56:1014-1018.[Abstract/Free Full Text]
  29. Gabriel HM, Heger M, Innerhofer P, Zehetgruber M, Mundigler G, Wimmer M, et al. Long-term outcome of patients with ventricular septal defect considered not to require surgical closure during childhood. J Am Coll Cardiol 2002;39:1066-1071.[Abstract/Free Full Text]
  30. Waldman JD. Why not close a small ventricular septal defect?. Ann Thorac Surg 1993;56:1011-1012.[Free Full Text]



This article has been cited by other articles:


Home page
Eur J Cardiothorac SurgHome page
Q. Xing, Q. Wu, S. Pan, Y. Ren, and H. Wan
Transthoracic device closure of ventricular septal defects without cardiopulmonary bypass: experience in infants weighting less than 8 kg
Eur J Cardiothorac Surg, September 1, 2011; 40(3): 591 - 597.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
P. Oses, N. Hugues, N. Dahdah, S. J. Vobecky, J. Miro, M. Pellerin, and N. C. Poirier
Treatment of Isolated Ventricular Septal Defects in Children: Amplatzer Versus Surgical Closure
Ann. Thorac. Surg., November 1, 2010; 90(5): 1593 - 1598.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
H. Chen, J. Liu, W. Gao, and H. Hong
Late complete atrioventricular block and tricuspid regurgitation after percutaneous closure of a perimembranous ventricular septal defect
J. Thorac. Cardiovasc. Surg., September 1, 2010; 140(3): e60 - e61.
[Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
A. Lin, W. T. Mahle, P. A. Frias, P. S. Fischbach, B. E. Kogon, K. R. Kanter, and P. M. Kirshbom
Early and delayed atrioventricular conduction block after routine surgery for congenital heart disease
J. Thorac. Cardiovasc. Surg., July 1, 2010; 140(1): 158 - 160.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
Q. Xing, S. Pan, Q. An, Z. Zhang, J. Li, F. Li, Q. Wu, and Z. Zhuang
Minimally invasive perventricular device closure of perimembranous ventricular septal defect without cardiopulmonary bypass: Multicenter experience and mid-term follow-up
J. Thorac. Cardiovasc. Surg., June 1, 2010; 139(6): 1409 - 1415.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
D. B. McElhinney
Patent Foramen Ovale Closure: Let's Keep the Heart in Mind
Circulation, June 16, 2009; 119(23): 2967 - 2968.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Predescu, D.
Right arrow Articles by Lee, K.-J.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Predescu, D.
Right arrow Articles by Lee, K.-J.
Related Collections
Right arrow Congenital - acyanotic
Right arrow Congestive Heart Failure


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS