J Thorac Cardiovasc Surg 2008;136:1318-1323
© 2008 The American Association for Thoracic Surgery
Evolving Technology |
a Division of Cardiothoracic Surgery, University of Minnesota, Minneapolis, Minn
b Department of Surgery and Division of Cardiology, University of Minnesota, Minneapolis, Minn
Received for publication July 31, 2007; revisions received November 14, 2007; accepted for publication December 18, 2007. * Address for reprints: Ranjit John, MD, Division of Cardiothoracic Surgery, University of Minnesota, Minneapolis, MN 55455. (Email: johnx008{at}umn.edu).
| Abstract |
|---|
|
|
|---|
Methods: At our institution, 45 patients (mean age, 57.24 ± 14.2 years) underwent implantation of the HeartMate II; 30 underwent bridge-to-transplantation therapy, 7 underwent destination therapy, and 8 underwent left ventricular assist device exchange for a failed XVE left ventricular assist device. Total duration of HeartMate II support was 352.13 patient-months (mean duration, 7.2 ± 5.2 months). All 45 patients were treated postoperatively with warfarin and aspirin. We recorded use of these 2 medications and monthly international normalized ratios. Prospectively, we also monitored patients for any clinical thromboembolic events and for pump thrombus.
Results: Of our 45 study patients, 41 had a mean international normalized ratio of less than 2.0; of those 41 patients, 21 had a mean international normalized ratio of less than 1.6. Because of recurrent gastrointestinal bleeding episodes, 7 patients discontinued warfarin for a total duration of 39.1 patient-months. During the entire period of HeartMate II support, we noted 1 thromboembolic event. In addition, another patient had a suspected left ventricular assist device pump thrombus that resolved with a high-intensity heparin anticoagulation protocol (international normalized ratio, 1.3).
Conclusions: Our preliminary single-center analysis suggests that the HeartMate II is associated with an extremely low thromboembolic risk and with less stringent requirements for anticoagulation. Selected patients at high risk for bleeding can be safely followed with either no or extremely low anticoagulation requirements for prolonged periods.
| Introduction |
|---|
|
|
|---|
The new HeartMate II LVAD, which incorporates continuous-flow rotary pump technology, represents the next generation of devices. Its focus is enhanced durability and quality of life for extended periods of circulatory support. A principal advantage of these new pumps is their small size, thereby extending therapy to underserved patient populations, including women and even some children; in addition, the HeartMate II is associated with minimal noise and greater patient comfort.8,9
The simpler design of continuous-flow, rotary pump technology promises greater long-term mechanical reliability with a single moving part, the internal rotor (Figures 1 and 2)
. The HeartMate II LVAD is one seventh the size and one fourth the weight of the previous HeartMate XVE. We report our initial experience with changes in anticoagulation and in the risk of thromboembolism with the HeartMate II in 45 patients with end-stage heart failure at the University of Minnesota Medical Center.
|
|
| Materials and Methods |
|---|
|
|
|---|
Patients
From October 1, 2005, through May 31, 2007, a total of 45 patients (mean age, 57.24 ± 14.2 years) underwent HeartMate II implantation at the University of Minnesota Medical Center (30 underwent bridge-to-transplantation therapy, 7 underwent destination therapy, and 8 underwent LVAD exchange for a failed XVE model). Total duration of LVAD support was 352.1 patient-months (mean duration, 7.8 ± 5.38 months). All 45 patients were treated postoperatively with warfarin and aspirin. We recorded use of these 2 medications and monthly international normalized ratios (INRs). We calculated the mean INR from the mean of all the INRs recorded at monthly intervals for all patients receiving HeartMate II devices. Prospectively, we also monitored patients for any of 3 categories of clinical thromboembolic events and for device thrombosis, as defined below.
Neurologic event
Such events included any new, temporary, or permanent focal or global neurologic deficits, such as transient ischemic attack (TIA), metabolic encephalopathy, and seizure.
Stroke
A stroke is a neurologic deficit lasting more than 24 hours or lasting 24 hours or less if a brain-imaging study shows new infarction. A TIA is a neurologic deficit lasting 24 hours or less if an imaging study shows no evidence of new infarction.
Peripheral thromboembolic event
These events do not involve the central nervous system. They include a thrombus or thromboembolism in the pulmonary or systemic circulation confirmed by means of standard clinical and laboratory testing, operative findings, or autopsy findings or that requires empiric intervention.
Device thrombosis
This event entails any obstructive thrombus in the device or its conduits that is associated with clinical symptoms of impaired pump performance (eg, level 0, none; level 1, <10% obstruction [minimal]; level 2, 11% to 50% obstruction [moderate]; level 3, >50% obstruction [severe]; and level 4, 100% obstruction [extreme].
Anticoagulation
For our study, we agreed on this 6-step initial anticoagulation regimen as a guideline: (1) initiation of an intravenous infusion of unfractionated heparin 12 to 24 hours after HeartMate II implantation or at the point that thoracostomy tube drainage is less than 50 mL/h; (2) titration of the heparin infusion to a partial thromboplastin time (PTT) of 45 to 50 seconds for 24 hours after implantation; (3) titration of the heparin infusion to a PTT goal of 50 to 60 seconds after 24 hours; (4) titration of the heparin infusion to a PTT goal of 55 to 65 seconds after an additional 24 hours; (5) initiation of antiplatelet therapy on postoperative day 2 to 3 with 81 mg of aspirin daily; and (6) initiation of anticoagulation with warfarin on postoperative day 3 to 5 and after removal of thoracostomy tubes, titrating the dose to an INR of 2 to 3 and discontinuing heparin after obtaining a therapeutic INR.
However, as a result of significant gastrointestinal bleeding in 4 patients and delayed pericardial tamponade in 1 patient (among the first 14 patients), we decided to stop using postoperative heparin in all subsequent patients. Furthermore, we decided to initiate anticoagulation with warfarin (starting from day 3), titrating the dose to an INR of 1.5 to 2.0.
Data Collection
After obtaining written informed patient consent and enrollment into the study, we collected baseline and follow-up data, including patient characteristics, results of blood chemistry analyses, hematologic findings, neurologic status, and concomitant medications. After patients were discharged from the hospital to home, they returned to our institution for follow-up, device review, and general status assessment weekly for the first 4 weeks and then monthly until the study end point. We recorded readmissions to the hospital and patient adverse events (including suspected device malfunctions) throughout the study as they occurred using standardized definitions.
Statistical Analysis
All INRs recorded for each patient for each month were averaged to obtain 1 mean value for each patient. Values are reported as means ± standard deviations.
| Results |
|---|
|
|
|---|
The overall mean duration of HeartMate II support was 7.2 ± 5.2 months, the duration in the bridge-to-transplantation group was 6.9 ± 4.9 months, the duration in the destination therapy group was 12.1 ± 7.4 months, and the duration in the XVE exchange group was 7.8 ± 3.8 months. The total duration of support for all patients was 352.13 patient-months.
Anticoagulation
Of our 45 study patients, 41 had a mean INR of less than 2.0; of those 41 patients, 21 had a mean INR of less than 1.6 (Figure 3
). The mean INR for all 45 patients during follow-up is shown in Figure 4
. Because of recurrent gastrointestinal bleeding episodes, 7 patients did not receive warfarin for a total duration of 39 patient-months. Of those 7 patients, 1 has been off warfarin for a total of 12 months; 2 patients have been off warfarin for 6 months. The mean INR of patients receiving warfarin (after excluding those patients not receiving warfarin) was 1.7 ± 0.2.
|
|
Hemorrhagic Complications
Of the first 14 patients with the HeartMate II pump in our series who received heparin followed by warfarin, 4 had gastrointestinal bleeding, and 1 had delayed pericardial tamponade (5/14 [35.7%]). Of the remaining 31 patients who received only warfarin and no heparin, 4 had gastrointestinal bleeding, and none had pericardial tamponade (4/31 [12.9%]). Note that this analysis excluded patients who underwent mediastinal exploration for bleeding on the day of surgical intervention because that was unrelated to postoperative anticoagulation therapy.
Device Thrombus
Of our 45 study patients, 1 had a suspected pump thrombus that resolved with a high-intensity heparin anticoagulation protocol (INR, 1.3 at the time of presentation). This patient had gone to an outside hospital with HeartMate II flows in the mid-3 L/min range (in contrast to this patient's usual flows of >4 L/min) and with increased HeartMate II power. After 12 hours of high-intensity anticoagulation heparin therapy (PTT goal, 60–80 seconds), the patient's hemodynamic parameters returned to normal. Heparin therapy was continued for 48 hours. Warfarin therapy was continued to achieve an INR goal of 1.5 to 2.0. No thrombolytic therapy was used. As of the end of this study, this patient is doing well on the heart transplant waiting list without any further similar episodes 6 months after this suspected pump thrombus.
Thus, our overall incidence of device thrombus was 2 (4.4%) of 45, including 1 explanted pump that received a thrombus score of 3, as reported in the next section. Note that none of the patients who discontinued warfarin experienced any episodes of pump thrombus during HeartMate II support.
Explanted Pump Analysis
For the 14 patients in our study group whose HeartMate II was explanted (at the time of their heart transplantation), we analyzed their explanted pumps for thrombus deposition. All pumps except 1 received a thrombus score of 0 (none) or 1 (minimal). Figure 5
shows a typical photograph of an explanted pump with no evidence of thrombus deposition at the forward ruby bearing from a patient supported successfully by a HeartMate II for 164 days. The 1 explanted pump that received a high score of 3 (>50% obstruction with thrombus) had been in a patient who died after prolonged sepsis. He had several days of low pump flow preceding his death, which might have been associated with the development of pump thrombus.
|
| Discussion |
|---|
|
|
|---|
The development of novel materials used for implant operations and the increasing use of implanted devices has made it evident that no material is biologically inert. Commonly used biomaterials, including so-called inert compounds, such as titanium, polytetrafluoroethylene, and acrylics, can trigger an array of iatrogenic effects, including inflammation, fibrosis, coagulation, and infection. In the case of LVADs, in which the biomaterial is in direct contact with the blood circulation, significant changes in systemic immunologic and thrombostatic functions have been well documented. Like most other implanted devices, LVADs activate the coagulation system, resulting in device-related thrombus.11,12
The standard strategy to reduce the risk of thromboembolism has been systemic anticoagulation. Despite this strategy, the risk of thromboembolism continues to be as high as 30%. Furthermore, the risk of bleeding after LVAD implantation is exacerbated with anticoagulation.
The HeartMate II is a continuous-flow rotary LVAD composed of a blood pump, a percutaneous lead, and an external power source and system driver. The inlet and outlet cannulas include woven polyester grafts (CR Bard, Haverhill, Pa) that require preclotting. The pump motor and associated blood tube have smooth titanium surfaces; in an effort to duplicate the excellent biocompatibility of the original pulsatile HeartMate XVE, the inlet and outlet elbows and the intraventricular cannula are textured with titanium microsphere coatings.8,9
The rationale for using textured materials for the original HeartMate XVE was that they would absorb and entrap elements from the patient's blood to form a stable, densely adherent biologic lining on the inside of the device.13-15
The resultant tissue lining, rather than the underlying biomaterials, would then form the long-term blood-contacting interface, thus obviating the need for systemic anticoagulation. The HeartMate XVE actually encouraged the adherence of a thin layer of fibrin-based neointima to its rough interior surfaces, yet the risk of embolization was reduced by the dense adherence of this acquired biologic lining to its surface. The physiologic consequences of this biologic lining, however, were many and varied because of the nature and characteristics of the cells adhering to the HeartMate XVE surface. Spanier and colleagues11,16
described a phenomenon of a "compensated coagulopathy" underlying the apparent autoanticoagulation in textured-surface HeartMate XVE recipients, attributing this finding to procoagulant stimuli elicited from the LVAD cell-surface environment. Some investigators have suggested that such activation of anticoagulation was due largely to the continuous contact of blood with the foreign LVAD surface. However, others showed that specific cells that progressively adhere to the textured LVAD surface and become activated can also contribute to the coagulopathy.14,16
In addition to the thromboembolic risks inherent with all LVADs, including the HeartMate II, the left ventricle in patients with end-stage heart failure is also prone to the development of thrombus in areas of pre-existing scar tissue, of previous myocardial infarction, and of previous ventricular aneurysm; such patients can even have a pre-existing thrombus. Thus they might already have an increased risk of thromboembolism, even before LVAD implantation. Moreover, after LVAD implantation, the degree of blood stasis in the patient's heart is determined, in part, by the opening of the native aortic valve. At our institution, we usually adjust the fixed-rate speed of the HeartMate II to maximize left ventricular decompression and to improve cardiac output, simultaneously allowing for at least a 1:3 ratio of aortic valve opening. We believe this adjustment has a favorable effect on reducing the incidence of thromboembolism associated with continuous-flow pumps, such as the HeartMate II.
In patients undergoing cardiopulmonary bypass, an association between a higher rate of microembolic signals during cardiac surgery and postsurgical neurologic dysfunction has been described.17,18
The clinical relevance of detecting microemboli is relatively unclear; this phenomenon is mostly unrelated to the development of clinically significant adverse neurologic events in patients receiving LVADs (including HeartMate II).19-21
Whether routine surveillance with transcranial Doppler monitoring confers a significant clinical benefit remains unanswered.22
Our study was limited by its relatively small number of single-center patients. Also, using the mean INR to represent the degree of anticoagulation might not truly capture the level of anticoagulation accurately. It is certainly possible that between patients' monthly clinic visits, their levels of anticoagulation might be higher or lower than what is shown when the INR is being measured.
In conclusion, the new continuous-flow HeartMate II LVAD has benefited from duplicating several features of the original HeartMate XVE, especially in that both confer a low thromboembolic risk. The explanation for the improved thromboembolic rate most probably has more to do with the influence of high pump flows, well-executed computational fluid dynamics modeling, and the continuous-flow nature of the device than with the textured surfaces. The average flow in all 45 of our patients was consistently greater than 4 L/min. Such high flow is associated with reduced stasis and with better washing effect and consequently with a favorable effect on the risk of thrombus formation in both the device pump and the patient's heart. With increasing experience, we have also successfully reduced the intensity of anticoagulation (avoiding heparin and decreasing the therapeutic INR, which formerly was 2.0–3.0 and now is 1.5–2.0), without any increased risk of thromboembolic events. In addition, we currently treat patients with low-dose warfarin therapy (2 to 3 mg/d) to achieve a therapeutic INR of 1.5 to 2.0 by the end of the first week. The favorable low thrombogenicity and thromboembolic risk associated with the HeartMate II makes it well suited as destination therapy. With the increasing number of adverse reactions and the increasing intolerance to anticoagulation in the destination therapy population, our experience with avoiding anticoagulation (other than aspirin) in a select, albeit small, number of patients receiving HeartMate II LVADs further favors its use. Moreover, with the increasing number of LVAD recipients who later undergo noncardiac surgery (eg, hernia repairs and orthopedic procedures), the ability to discontinue anticoagulation without a fear of thromboembolic complications is paramount. Further studies are needed that seek to identify the molecular and biochemical markers of thromboembolism and bleeding to better understand thromboembolic and bleeding risks in LVAD recipients.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. E. A. Felix, J. R. Martina, J. H. Kirkels, C. Klopping, H. Nathoe, E. Sukkel, N. Hulstein, F. Z. Ramjankhan, P. A. F. M. Doevendans, J. R. Lahpor, et al. Continuous-flow left ventricular assist device support in patients with advanced heart failure: points of interest for the daily management Eur J Heart Fail, February 3, 2012; (2012) hfs012v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Murakawa, T. Murayama, J. Nakajima, and M. Ono Lung lobectomy in a patient with an implantable left ventricular assist device Interact CardioVasc Thorac Surg, December 1, 2011; 13(6): 676 - 678. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Suarez, C. B. Patel, G. M. Felker, R. Becker, A. F. Hernandez, and J. G. Rogers Mechanisms of Bleeding and Approach to Patients With Axial-Flow Left Ventricular Assist Devices Circ Heart Fail, November 1, 2011; 4(6): 779 - 784. [Full Text] [PDF] |
||||
![]() |
R. John, F. Kamdar, P. Eckman, M. Colvin-Adams, A. Boyle, S. Shumway, L. Joyce, and K. Liao Lessons Learned From Experience With Over 100 Consecutive HeartMate II Left Ventricular Assist Devices Ann. Thorac. Surg., November 1, 2011; 92(5): 1593 - 1600. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Topilsky, T. Hasin, J. K. Oh, D. D. Borgeson, B. A. Boilson, J. A. Schirger, A. L. Clavell, R. P. Frantz, R. Tsutsui, M. Liu, et al. Echocardiographic Variables After Left Ventricular Assist Device Implantation Associated With Adverse Outcome Circ Cardiovasc Imaging, November 1, 2011; 4(6): 648 - 661. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. John, Y. Naka, N. G. Smedira, R. Starling, U. Jorde, P. Eckman, D. J. Farrar, and F. D. Pagani Continuous Flow Left Ventricular Assist Device Outcomes in Commercial Use Compared With the Prior Clinical Trial Ann. Thorac. Surg., October 1, 2011; 92(4): 1406 - 1413. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jennings, J. McDonnell, and J. Schillig Assessment of long-term anticoagulation in patients with a continuous-flow left-ventricular assist device: A pilot study J. Thorac. Cardiovasc. Surg., July 1, 2011; 142(1): e1 - e2. [Full Text] [PDF] |
||||
![]() |
A. Tiede, J. H. Rand, U. Budde, A. Ganser, and A. B. Federici How I treat the acquired von Willebrand syndrome Blood, June 23, 2011; 117(25): 6777 - 6785. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Schaffer, G. J. Arnaoutakis, J. G. Allen, E. S. Weiss, N. D. Patel, S. D. Russell, A. S. Shah, and J. V. Conte Bleeding Complications and Blood Product Utilization With Left Ventricular Assist Device Implantation Ann. Thorac. Surg., March 1, 2011; 91(3): 740 - 749. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. R. Ensor, C. A. Paciullo, W. D. Cahoon Jr, and P. E. Nolan Jr Pharmacotherapy for Mechanical Circulatory Support: A Comprehensive Review Ann. Pharmacother., January 1, 2011; 45(1): 60 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Jennings, R. M. Chambers, and J. M. Schillig The Pharmacotherapy of the HeartMate II, a Continuous Flow Left Ventricular Assist Device, in Patients with Advanced Heart Failure: Integration of Disease, Device, and Drug Ann. Pharmacother., October 1, 2010; 44(10): 1647 - 1650. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Bhamidipati, G. Ailawadi, J. Bergin, and J. A. Kern Early thrombus in a HeartMate II left ventricular assist device: A potential cause of hemolysis and diagnostic dilemma J. Thorac. Cardiovasc. Surg., July 1, 2010; 140(1): e7 - e8. [Full Text] [PDF] |
||||
![]() |
S. E. Sandner, D. Zimpfer, P. Zrunek, A. Rajek, H. Schima, D. Dunkler, M. Grimm, E. Wolner, and G. M. Wieselthaler Renal Function and Outcome After Continuous Flow Left Ventricular Assist Device Implantation Ann. Thorac. Surg., April 1, 2009; 87(4): 1072 - 1078. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |