|
|
||||||||
J Thorac Cardiovasc Surg 2008;136:1496-1502
© 2008 The American Association for Thoracic Surgery
Evolving Technology |
a The Interventional Centre, Rikshospitalet University Hospital, Oslo, Norway
b Department of Cardiothoracic Surgery, Rikshospitalet University Hospital, Oslo, Norway
c Department of Cardiology, Rikshospitalet University Hospital, Oslo, Norway
d Vestfold University College, Tønsberg, Norway
e The Faculty of Medicine, University of Oslo, Oslo, Norway
Received for publication May 5, 2008; revisions received August 8, 2008; accepted for publication August 28, 2008. * Address for reprints: Per Steinar Halvorsen, MD, Rikshospitalet University Hospital, The Interventional Centre, Sognsvannsveien 20, N-0027 Oslo, Norway. (Email: per.steinar.halvorsen{at}rikshospitalet.no).
| Abstract |
|---|
|
|
|---|
Methods: Ten patients with significant left anterior descending coronary artery stenosis underwent off-pump coronary artery bypass grafting. A 3-axis accelerometer (11 x 14 x 5 mm) was sutured onto the left anterior descending coronary artery–perfused region of left ventricle. Twenty episodes of ischemia were studied, with 3-minute occlusion of left anterior descending coronary artery at start of surgery and 3-minute occlusion of left internal thoracic artery at end of surgery. Longitudinal, circumferential, and radial accelerations were continuously measured, with epicardial velocities calculated from the signals. During occlusion, accelerometer velocities were compared with anterior left ventricular longitudinal, circumferential, and radial strains obtained by echocardiography. Ischemia was defined by change in strain greater than 30%.
Results: Ischemia was observed echocardiographically during 7 of 10 left anterior descending coronary artery occlusions but not during left internal thoracic artery occlusion. During ischemia, there were no significant electrocardiographic or hemodynamic changes, whereas large and significant changes in accelerometer circumferential peak systolic (P < .01) and isovolumic (P < .01) velocities were observed. During 13 occlusions, no ischemia was demonstrated by strain, nor was any change demonstrated by the accelerometer. A strong correlation was found between circumferential strain and accelerometer circumferential peak systolic velocity during occlusion (r = –0.76, P < .001).
Conclusions: The epicardial accelerometer detects myocardial ischemia with great accuracy. This novel technique has potential to improve monitoring of myocardial ischemia during cardiac surgery.
| Introduction |
|---|
|
|
|---|
An accelerometer permits continuous real-time assessments of heart wall motion.8-11
Measurements are independent of operator skills, and a 3-axis accelerometer allows simultaneous assessment of wall motion in the longitudinal, circumferential, and radial directions. This study was designed to investigate whether an epicardial 3-axis accelerometer could detect myocardial ischemia. In patients scheduled for off-pump CABG, occlusions of the left anterior descending coronary artery (LAD) at start of surgery and of the left internal thoracic artery (LITA) at end of surgery were performed to induce myocardial ischemia. The 2D strain obtained by transesophageal echocardiography was used as the reference value for the assessment of ischemia.
| Methods |
|---|
|
|
|---|
The patients were medicated before the operation with 5 to 10 mg diazepam orally. Anesthesia was induced with 2 to 5 µg/kg fentanyl, thiopentone, 2 to 5 mg/kg, and 0.15 mg/kg cisatracurium (INN cisatracurium besilate). Repeated doses of fentanyl and sevoflurane (1.0% to 2.5%) were used to maintain anesthesia. Heparin (1.5 mg/kg) was given to achieve an activated clotting time longer than 250 seconds. After sternotomy and pericardial opening, a prototype 3-axis accelerometer was sutured on the LV anterior apical region (Figure 1 ). A snare was placed around LAD distally to the first diagonal branch. The chest and the pericardium were left open, with the patients in the supine position.
|
Echocardiography
A Vivid 7 scanner (GE Vingmed Ultrasound AS, Horten, Norway) with a transesophageal transducer was used. Two-dimensional gray-scale recordings were obtained in 2- and 4-chamber views and in apical short-axis view. Longitudinal, circumferential, and radial LV functions were assessed with the 2D strain technique.5-7
This method, which calculates myocardial deformation from ultrasound speckles in the gray-scale images, is semiautomatic and angle independent. EchoPAC (GE Vingmed Ultrasound) was used for off-line analysis.
Accelerometer
We used a prototype 3-axis capacitive accelerometer (outer dimensions 11.0 x 14.5 x 5.2 mm),12
with range ± 2G and cross-sensitivity between axes of 2%. The acceleration and ECG signals were recorded synchronously at a sampling rate of 250 to 500 Hz with an NI USB 6009 AD converter (National Instruments Corporation, Franklin Park, Ill) and LabVIEW software (National Instruments). All signals were stored on a personal computer and analyzed off-line with Matlab (The MathWorks, Inc, Natick, Mass).
Experimental Protocol
After opening of the pericardium before CABG, baseline recordings were obtained with stable hemodynamics. Before grafting on the LAD, a 3-minute occlusion of LAD was performed, followed by 5 minutes of reperfusion. ECG, blood pressures, and pulse-contour CO and LV apical accelerations were continuously recorded. Echocardiographic recordings were obtained at baseline, at the end of LAD occlusion, and after 5 minutes of reperfusion. After bypass grafting was completed, the LITA was manually occluded for 3 minutes. Before LITA occlusion, activated clotting time was ensured to be above 250s. The same measurements described here were performed during LITA occlusion and reperfusion. Transit-time flow measurements with a 3-mm probe (Medistim KirOp AS, Oslo, Norway) of the LITA graft were obtained before, during, and after LITA occlusion. No manipulation of heart or body positions occurred during the occlusions of the LAD and LITA. In addition, vasoactive medications and fluid infusions were kept unchanged.
The 2D-strain was used as the reference method for the detection of ischemia. Ischemia was defined as a relative change in systolic strain greater than 30% in either direction.13
This cutoff was based on the fact that measurements of physiologic variables often lack precision, with errors of 10% to 20% not uncommon.14,15
Similarly, a change in either direction of the accelerometer recordings in peak systolic or isovolumic relaxation (IVR) velocity greater than 30% was defined as ischemia.
Calculations
Measurements of the ECG ST-segment were performed 80 ms from the ECG J-point. Mean arterial blood pressure (MAP), pulse-contour CO, and the positive systolic time derivative of the femoral arterial blood pressure (dP/dt) were assessed from the PiCCO catheter recordings. Systole was defined from the start of R on ECG to aortic valve closure, defined by a sharp spike in the acceleration signal.16
Diastole was defined from aortic valve closure to the start of R on ECG. The IVR phase was defined from aortic valve closure to mitral valve opening, as assessed by the acceleration signal.16
Peak systolic strains in the longitudinal, circumferential, and radial directions were calculated as percentages of LV end-diastolic dimensions. The mean of three consecutive heart cycles was used for statistical calculations. Frame rate was 69 ± 12 frames/s.
The acceleration signals were integrated to obtain epicardial velocities.17
Peak early systolic velocity and velocity in the IVR phase were measured.18
The accelerometer x-axis measured longitudinal velocity, the y-axis measured circumferential velocity, and the z-axis measure radial velocity (Figure 1). The mean of three consecutive heart cycles was used for statistical analysis. To assess interobserver variability, all baselines and occlusions were analyzed by another investigator who was blinded to the results from the reference method.
Statistical Analysis
The number of patients included was based on a power analysis. During ischemia, a clinical significant change in accelerometer peak velocity was set to 5.0 cm/s (SD 5.0 cm/s). With
= .05, this yielded a power of 0.80 with 8 patients included. Parametric statistical methods were used, and data are presented as mean ± SD (SD of sample distribution). For repeated measurements, paired-sample t tests were applied. To adjust for multiple comparisons, Bonferroni corrections of P values were performed. For the correlation analysis, the Pearson correlation coefficient was calculated. Sensitivity and specificity were computed from standard table analysis, and the 95% confidence intervals (CIs) were constructed with the f-distribution. To assess interobserver variability, the Bland–Altman method was applied.19
All statistical analyses were performed with SPSS statistical software (version 13.0, SPSS Inc; Chicago, Ill).
| Results |
|---|
|
|
|---|
LAD and LITA occlusions
Twenty occlusions were performed (10 LAD occlusions and 10 LITA occlusions). Ischemia, defined as changes in strain greater than 30%, developed in 7 of 10 LAD occlusions. No ischemia was observed during three LAD occlusions. One of these patients had a 100% distal LAD occlusion with no retrograde blood flow, and another had extensive collaterals to the LV apical region from the right coronary artery. LITA occlusion was not associated with myocardial ischemia in any patient.
The ST segment of the standard ECG lead II did not change significantly from baseline in the 7 patients with ischemia according to strain during LAD occlusion. ST-segment depression greater than 0.1 mV was not found in any patient. In addition, no significant changes in pulse-contour CO, mean arterial pressure, heart rate central venous pressure, and dP/dt were observed (Table 1 ). In contrast, marked changes in longitudinal, circumferential, and radial systolic strains were seen during ischemia (P < .01; Table 1). These values returned to baseline after 5 minutes of reperfusion. Figure 2 shows representative circumferential velocity curves generated from the accelerometer at baseline and during ischemia and reperfusion.
|
|
Increases in accelerometer velocities in the IVR phase were observed in all directions during ischemia. A significant change, however, was only found in the circumferential direction (–6.5 ± 4.0 to 2.0 ± 7.3, P < .001; Table 1).
During 13 occlusions without ischemia according to strain data, no significant changes were observed in the ECG ST-segment, pulse-contour CO, MAP, heart rate, central venous pressure, dP/dt, or accelerometer velocities (Table 1).
Individual changes in circumferential strain and accelerometer circumferential peak systolic velocity for all LAD occlusions can be seen in Figure 3 . A strong negative correlation was observed between circumferential systolic strain and accelerometer circumferential peak systolic velocity during LAD and LITA occlusions (Figure 4 ). In the 7 occlusions where ischemia was demonstrated by strain measurements, the accelerometer also exhibited ischemia. Ischemia was not found on accelerometer recordings in the 13 occlusions where strain did not show ischemia. The accelerometer therefore demonstrated a sensitivity of 100% (95% CI 56%–100%), a specificity of 100% (95% CI 72%–100%), a positive predictive value of 100% (95% CI 56%–100%), and a negative predictive value of 100% (95% CI 72%–100%) in detecting ischemia.
|
|
| Discussion |
|---|
|
|
|---|
The main finding in this study was that the accelerometer showed highly significant circumferential velocities changes during myocardial ischemia, whereas no changes could be seen in ECG and hemodynamics. The accelerometer demonstrated high sensitivity and specificity in the detection of myocardial ischemia. These findings, together with the strong negative correlation observed between 2D strain and circumferential peak systolic velocity, prove that accelerometer recordings may be applied successfully in the monitoring of myocardial ischemia.
Echocardiographic 2D strain was used as reference method, because this technique has the confirmed ability to detect and quantify myocardial ischemia.5-7
In this study, the LAD and LITA were occluded during stable hemodynamic periods, because a large decrease in LV function most likely would be caused by myocardial ischemia. Measurements of physiologic variables may vary 10% to 20% in either direction during stable hemodynamic conditions,14,15
and a relative decrease in myocardial function greater than 30% was therefore made a cutoff for the definition of ischemia.
In contrast to echocardiographic methods, accelerometer recordings are operator independent and have the potential for continuous real-time monitoring of myocardial ischemia. Velocities were calculated from the acceleration signals for two reasons. First, mitral valve closure may generate large systolic accelerations that do not necessarily reflect myocardial contraction.11,20,21
Second, velocity is an established and frequently used parameter for measuring myocardial ischemia.18,22
There is, however, a distinction between echocardiographic tissue velocity imaging and the accelerometer in assessing myocardial velocity. The ultrasonographic technique measures the velocity in one direction within the wall, whereas with the accelerometer, epicardial velocities are assessed simultaneously in three directions. Even so, the epicardial velocity traces appeared similar to traces generated by the ultrasonographic technique.18,22
During ischemia, a typical ischemic velocity pattern was observed, with a decrease in early peak systolic velocity and an increase in IVR velocity. Opening and closure of the mitral and aortic valves are reflected as distinct spikes in the acceleration signal. By recording ECG and acceleration simultaneously, the systolic and diastolic phases of the heart cycle were made easier to identify than with the ultrasonic technique.
Significant changes were observed only in the circumferential direction and not in the other axes. This may be due to the effects of pericardiotomy and the resulting loss of pericardial constraint. Pericardiotomy causes abnormal longitudinal apex motion,23,24
whereas LV rotation is unaffected.7
Radial myocardial contraction is caused by shortening of longitudinally and circumferentially oriented fibers,25-27
and therefore abnormal longitudinal contraction may affect radial motion. In the LV apical region, however, myocardial contraction is dominated by circumferentially orientated fibers, and it is therefore likely that the largest effects may be seen in the epicardial circumferential velocities during ischemia. The capability of the accelerometer to detect minor subepicardial or endocardial ischemia is unknown. The subepicardial and endocardial fibers are oriented longitudinally, but as myocardial deformation during ejection demonstrates extensive transmural tethering,28
an accelerometer placed on the epicardial surface may possibly detect subendocardial ischemia. Further studies are needed to draw firm conclusions.
Intraoperative ischemia is a predictor of postoperative myocardial infarction.1,29
Nevertheless, postoperative infarction may frequently occur without previously detected intraoperative ischemia.1
This emphasizes the need for sustained, continuous, and sensitive detection of ischemia even in the postoperative period. Echocardiography is an excellent tool to assess intraoperative ischemia, but the method is cumbersome and not routinely applied for continuous monitoring after surgery. For this period, the accelerometer may offer an important advantage relative to commonly used monitoring modalities. The development of an algorithm for automatic analysis of the accelerometer signals and further miniaturization of the device will be necessary for clinical implementation. The accelerometer could be incorporated into temporary pacemaker leads, minimizing invasiveness while enabling continued detection of ischemia.
Collateral flow from other coronary arteries may explain the lack of ischemic changes during LAD occlusion in 3 patients. In the experiments with occlusion of the LITA graft, collateral flow and continued flow through the native LAD may explain the lack of ischemic changes. These observations indicate that the monitoring of velocities from an accelerometer, although able to detect ischemia, may not always be able to detect malfunction of grafts. This drawback, however is also seen with ECG and echocardiography. Accelerometers not only may be used to monitor ischemia in CABG surgery but may possibly be applied to all cardiac surgery procedures where ischemia can occur.
Limitations
Measurements were performed in stable periods and not during bypass grafting, where surgical manipulation to a great extent affect LV performance.
Assessment of ischemia by transesophageal echocardiography may be performed in multiple heart regions. This study cannot be used to draw conclusions as to whether more than one accelerometer would be required to monitor the different heart regions. Apex motion has been shown to be affected by ischemia from all coronary arteries.10
An accelerometer attached to the LV apical anterior region might therefore be sufficient for detecting ischemia in various heart regions.
Repeated LAD occlusions may induced myocardial preconditioning, thereby decreasing the accelerometer changes caused by LITA occlusion. We consider this effect to be minor. The accelerometer may have limited utility in patients whose myocardium is "stunned" or "hibernating" at the start of surgery, because such patients have reduced baseline contractility.30
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. S. Halvorsen, E. W. Remme, A. Espinoza, H. Skulstad, R. Lundblad, J. Bergsland, L. Hoff, K. Imenes, T. Edvardsen, O. J. Elle, et al. Automatic real-time detection of myocardial ischemia by epicardial accelerometer J. Thorac. Cardiovasc. Surg., April 1, 2010; 139(4): 1026 - 1032. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |