|
|
||||||||
J Thorac Cardiovasc Surg 2003;125:1451-1460
© 2003 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
From the Departments of Surgical and Perioperative Science,a and Integrative Medical Biology,b Section for Anatomy and Department of Medical Biosciences, Section for Pathology, Umeå University Hospital, Umeå, Sweden.
This work was supported by Swedish Society for Medical Research and funds of the Medical Faculty, Umeå University Hospital, Swedish Medical Research Council (12X-11204), Swedish Heart and Lung Foundation, and the Heart Foundation of North Sweden.
Received for publication May 10, 2002. Revisions requested July 22, 2002; revisions received Aug 27, 2002. Accepted for publication Sept 11, 2002. Address for reprints: Patrik Boivie, Cardiothoracic Division, Department of Surgical and Perioperative Science, Umeå University Hospital, S-901 85 Umeå, Sweden (E-mail: patrik_boivie{at}hotmail.com).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Brain damage is a significant part of morbidity and mortality after cardiac surgery. The scenario of brain damage is complex and has two extremes, major stroke versus diffuse cognitive defects.
1 Atherosclerosis of the ascending aorta and aortic crossclamping (ACC) has been identified as major risk factors for perioperative stroke during cardiac surgery.
2,3 Diffuse symptoms of brain damage occurs in the postoperative period and are often discussed in relation to the use of extracorporeal circulation, such as from fat microembolism of retrieved mediastinal blood,
4,5 or in terms of immunological mechanisms.
6 In addition, cerebral injury may be secondary to hypoperfusion and anesthetic management.
7
Most types of cardiac surgery involve manipulation of the aorta by clamping.
8 A number of techniques have been used to reduce the risk of stroke during cardiac operation,
8,9 including echo visualization of the aorta.
10 In coronary artery bypass grafting (CABG) on the nonbeating heart, two different methods are employed to perform distal anastomoses, either cardioplegic arrest or the older method of fibrillating heart. The purpose of this study was to investigate the mechanism of injury caused by ACC. The focus was on repeated clamping, such as with the fibrillating heart method.
An experimental model was designed using human cadaveric aorta. The spectrum of dislodged material following ACC was evaluated using computerized image processing, both of macroscopic and microscopic particulate matter. The macro- versus microscopic approach was done with reference to possible mechanisms behind stroke and diffuse brain damage, respectively.
| Materials and methods |
|---|
|
|
|---|
|
Retrograde perfusion was implemented (Figure 1). A catheter (8/10.6 32, Polystan A/S, Värlöse, Denmark) was positioned in the descending aorta and fixed with ligatures so that the perfusion tip was localized at the subclavian artery. The cannula was connected to a constant pressure device consisting of a 10-L infusion bag positioned at a hydrostatic level corresponding to 60 mm Hg pressure. The perfusion medium consisted of 9 g/L NaCl. The aorta was flushed with about 400 mL perfusion medium to remove debris, followed by a 50-mL sample of the perfusate representing the baseline prior to ACC (denoted as ACC 0). A standard 70-mm aortic crossclamp (Pilling Co, Fort Washington, Pa) was applied to the ascending aorta at a location that resembles a standard surgical procedure. This was typically about 2 cm proximal to the brachiocephalic trunk. To precisely repeat the crossclamping at the same location, the crossclamp jaws were fixed to the aortic adventitia with stitches. To gain the same occluding force of repeated clamping, the clamp was closed to half of its full range in all experiments.
|
Sample processing
Each washout was centrifuged at 1500g for 10 minutes at 22°C. The supernatant was carefully aspirated, leaving the deposit in the test tube. Distilled water (50 mL) was added to the sample to lyse remaining erythrocytes and for washing purpose. The sample was recentrifuged and the supernatant was again aspirated. The collected material was fixed for 10 minutes at 22°C, by adding 5 mL of 10% formalin in phosphate buffer. The fixation was stopped by adding distilled water, followed by washing and recentrifugation. The material was stained for 10 minutes at 22°C with 20 µL of cresyl violet for basophilic material,
11 followed by 2 washing and centrifugation cycles. The sample was collected in a plastic Pasteur pipette and deposited on an uncoated microscopic slide. The droplet was spread out and left to dry at room temperature. When a sample contained high amounts of particulate matter, it was spread out on 2 separate microscopic slides. This was done to avoid severely condensed material that would interfere with the image analysis. With duplicate slides, which occurred in 20% of all samples, the particulate matter from both slides was added together during image analysis. A typical deposit is shown in Figure 2.
|
For microscopic evaluation the samples were viewed in an inverted microscope (Olympus CK40-F200, Olympus Optical Company Ltd, Tokyo, Japan) equipped with a CDPlan x10-FPL objective lens, and recorded using a black and white camera (C5405-01 Hamamatsu Photonics, Hamamatsu City, Japan). In this configuration 2 microscopic views were sampled at random from each slide. The microscopic data therefore represent a qualitative measure of the deposit.
The average volume of the washout samples was 35.0 ± 0.9 mL (mean ± SEM). Due to technical circumstances a few washouts deviated in volume with a range from 15.0 to 51.5 mL. Due to this variation the amount of deposits was corrected for the sample volume and expressed as particles per milliliter. For the microscopic evaluation, the data were expressed per milliliter perfusate and per microscopic view.
Image processing and geometric measurements
The images were processed by computerized technique to measure particle count, area, and shape factor. The image handling included: (1) contrast enhancement to full gray scale, (2) manual editing for technical artefacts, (3) 1-step pixel dilation followed by pixel erosion, and (4) geometric measurement from preset gray threshold levels.
On the macroscopic level the particulate matter was dense in contrast and evaluated by a single threshold setting for calcified particles. On the microscopic level, at which both calcified dense and cellular semitransparent particulate matter were detected, the image analysis was performed using 2 windows of gray scale attenuation, dense and soft spectrum, respectively. All threshold settings were constant throughout the study. The results were transferred to an Excel spreadsheet.
Statistical analysis
The statistical evaluation contained ordinal, discrete, and numerical data. The geometric data (eg, area and shape factor measurements) showed normality. Mean values ± SEM are given throughout. However, because of the different nature of data only nonparametric statistics were used. For within-group differences Wilcoxon signed rank test was applied. Between-groups differences were evaluated by Kruskal-Wallis test and post hoc comparison by Mann-Whitney U test. Correlation matrices were tested using Spearman's rank coefficient. Slope values refer to linear regression versus clamp opening. A P value above .05 was considered nonsignificant.
| Results |
|---|
|
|
|---|
Method evaluation
We developed and tested a new method of studying the injuries caused by repeated ACC. The method gave reproducible results by which the accumulated deposits varied in number and characteristics with repeated ACC. There was a significant correlation between number of deposits and the magnitude of aortic wall calcification (P = .014). There was no correlation between subject age and number of deposits.
Macroscopic particle evaluation (dense spectrum)
The first-time clamping procedure resulted in a marked increase in observed calcified particles versus the baseline recording (Figure 3, a). With repeated ACC the number of deposits decreased. ACC 1 to 5 produced significantly larger amounts of particles compared to ACC 6 to 10 (Table 2, column A). This finding was also confirmed by a negative slope that was significantly different from zero (Table 2
). There were no significant differences in the observed average area or shape factor of the calcified particles between ACC 1 to 5 compared with ACC 6 to 10 (Figures 3
, b and c, Table 2
). When the embolic particle load was calculated, by multiplying the area with particle numbers, there was again a significant difference between ACC 1 to 5 versus ACC 6 to 10 (Figure 3
, d, Table 2
). This variability produced a negative slope significantly different from zero (Table 2
). The average diameter of recorded particles of all measurements was 0.63 ± 0.03 mm, with a spread from 0.22 to 4.74 mm (0.22 mm represents the technical resolution for the macroscopic measurements).
|
|
|
|
Because of the different recording methods between macroscopic and microscopic methods it was not meaningful to compare particle numbers between these groups. However, the two microscopic evaluation methods were possible to compare. Of particular interest was the embolic load from the product of particle-x-area. There was a significantly larger amount of embolic output of cellular material, comparing columns C versus B (Table 2
). From a technical image-analysis perspective this was expected as the cellular measuring window also contained calcified material. If the calcified material was subtracted from the data, leaving only cellular debris, it was found that the perfusate contained 169 ± 41 (n x µm2/mL and microscopic view) of cellular embolic material. This was twice the amount of calcified particles, 85.7 ± 43.1 (n x µm2/mL and microscopic view) (Table 2
, columns B and C). However, this apparent difference was not statistically significant (P = .070).
Variability between subjects
The data showed great variability between the cases. Although 9 of the 10 subjects had atherosclerotic changes in their aorta, only half of the cases produced major bursts of dislodged output (Figure 6). This was evaluated by calculation of linear regression. In 9 subjects the slope was negative, although in 4 of these only minor negativity was found (Figure 6
). The patient without visual aortic calcifications belonged to the group that had a slope close to zero.
|
| Discussion |
|---|
|
|
|---|
Before the era of cardioplegic arrest, the method of choice for CABG was by transient fibrillation of the heart.
17 With this technique the aorta was clamped repeatedly, for each anastomosis. Even though this method may be considered old-fashioned, it is still in use at several clinics around the world.
18 A parallel scenario is the manipulation from side-biting aortic clamps. In this study we simulated the surgical methods of fibrillating heart or other forms of repeated clamping. The crossclamp was applied repeatedly and in the same location on the aorta, and the dislodged material was collected and geometrically analyzed. The hypothesis was that the magnitude of dislodged particles would diminish with repeated clamping. This assumption was here confirmed, although the baseline was reached after several clamp maneuvers.
Aortic particles that cause stroke are thought to be geometrically large, such as detached calcified plaques. These particles are best analyzed on the macroscopic level. We found an average particle area of macroscopically detected material of about 0.2 mm2. The average diameter was 0.6 mm, with maximum of 4.7 mm. Moreover, it was found that the macroscopic material was noncircular with a shape factor of about 0.6 (circular shape equals 1 and linear shape 0). These data can be compared with known diameters of embolic material in occluded cerebral arteries (Willis circle) of about 5 mm.
19 Our data corroborate histological observations from ongoing studies using intra-aortic filters, in which the diameter of captured particles ranged from 0.1 to 6.0 mm.
13
Although the macroscopic material seems to be the most interesting with respect to stroke, cardiac surgery is also known to cause diffuse defects in neuropsychology and cognition.
1 This may in part reflect small-artery brain emboli.
20 It has been speculated that pericardial suction blood contains liquid fat that occludes brain capillaries' blood flow function.
4,5 However, this study suggests an additional source of microscopic emboli from ACC. We investigated microscopic particles from two perspectives, both calcified and cellular material. The distinction between these particles was created by gray-scale attenuation at image analysis. Dense opaque material was calcified microparticles. Cellular material was measured using a softer threshold spectrum to expose semitransparent material. Red blood cells that contaminated the samples were lysed by water prior to the analysis, and the measured material was noncalcified cellular debris from the aortic wall. The average area of these particles was approximately 100 to 200 µm2, with measured diameters of about 10 to 15 µm for calcified and cellular material, respectively. The particles were more circular in shape than the large-size macroscopic deposits. The capillary diameter is known to be about 5 µm,
21 which suggests embolic potential.
We found that the magnitude of dislodged particles differed substantially between subjects. In this study only 5 of the 10 subjects produced large amounts of debris. There was a positive correlation between the amount of particles and the visual appearance of aortic atherosclerosis. Nine of the 10 cases had visible atherosclerosis, but in only half of these were severe amounts of particles dislodged at clamping.
Our study was limited by the use of cadaveric aorta. It is reasonable to expect that some postmortem changes may affect the results in different ways. Furthermore, the conditions at autopsy may not truly represent the surgical conditions in the operating room. The limited number of observations must also be considered. Our model employed retrograde washout, which is a simplification although it resembles the surgical situation at aortic declamping. However, the data appeared consistent, and image analysis revealed significant trends with repeated ACC, a fact that suggests a useful model in surgical science. The clinical impact of our findings can only be speculated upon as a question that relates to the complexity of flow dynamics and vascular anatomy.
In conclusion, we designed an experimental model with cadaveric aorta to analyze injury caused by crossclamping. Our main focus was on surgical methods using repeated ACC, such as with the fibrillating heart technique. Repeated crossclamping reduced the amount of dislodged particulate matter. However, the output of material was profound during the initial series of clamp maneuvers, a finding that questions repeated ACC in surgery.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
Related Article
This article has been cited by other articles:
![]() |
P. Boivie, M. Hansson, and K. G. Engstrom Intraluminal aortic manipulation by means of intra-aortic filter, cannulation, and external clamp maneuvers evaluated versus dislodged embolic material J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 283 - 289. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. F. Aranki, M. Nathan, P. Shekar, G. Couper, R. Rizzo, and L. H. Cohn Hypothermic Circulatory Arrest Enables Aortic Valve Replacement in Patients With Unclampable Aorta Ann. Thorac. Surg., November 1, 2005; 80(5): 1679 - 1687. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Boivie, C. Edstrom, and K. G. Engstrom Side differences in cerebrovascular accidents after cardiac surgery: A statistical analysis of neurologic symptoms and possible implications for anatomic mechanisms of aortic particle embolization J. Thorac. Cardiovasc. Surg., March 1, 2005; 129(3): 591 - 598. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Engstrom Contaminating fat in pericardial suction blood: a clinical, technical and scientific challenge Perfusion, January 1, 2004; 19(1_suppl): S21 - S31. [Abstract] [PDF] |
||||
![]() |
J. W. Hammon Aortic nightmares: Can we sleep better? J. Thorac. Cardiovasc. Surg., June 1, 2003; 125(6): 1200 - 1201. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |